Development of a fluorescent probe hydrolysis-insulated isothermal PCR for rapid and sensitive on-site detection of African swine fever virus.

Tianli Zou1, Junhua Deng2, Xiangdong Li3, Shiyin Zhang4, Lingyan Chen2, Liying Hao2, Jinshan Zhuang2, Heng Wang5, Guihong Zhang5, Shengxiang Ge6, Kegong Tian7.   

Abstract

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35257964      PMCID: PMC9243616          DOI: 10.1016/j.virs.2022.03.002

Source DB:  PubMed          Journal:  Virol Sin        ISSN: 1995-820X            Impact factor:   6.947


× No keyword cloud information.
Dear Editor, African swine fever virus (ASFV), the sole member of the family Asfarviridae, is the causative agent of African swine fever (ASF), a viral disease that leads to high mortality in domestic pigs. Since firstly identified in Kenya in the 1920s, ASFV has been prevalent in Africa, Europe, and Russian Federation (Sanchez et al., 2019). Recently, ASFV was introduced to Asian countries including China, Mongolia, Vietnam, Cambodia, Laos and South Korea, which lead to huge economic losses to local pig industries (Li and Tian, 2018; Gaudreault and Richt, 2019). The first case of ASFV in China was reported in August 2018 (Zhou et al., 2018). Since then, hundreds of cases in most provinces were officially announced and more than 1 million pigs were slaughtered under the strict stamping-out policy (Miao et al., 2019). Currently, the World Organisation for Animal Health (OIE) recommends conventional and real-time PCRs as the commonly used assays to detect virus DNA in clinical samples (Luo et al., 2017). However, these molecular methods require expensive devices and well-trained personnel and can only be performed in the well-equipped laboratories. Therefore, a low-cost, sensitive and user-friendly method is urgently required for onsite detection of virus in the field. Insulated isothermal PCR (iiPCR) has been recently described to detect some veterinary viruses with the equivalent of sensitivity and specificity of real-time PCR (Lung et al., 2016; Ambagala et al., 2017; Zhang et al., 2019). Based on Rayleigh-Bénard convection PCR (Tsai et al., 2012), spontaneous fluid convection is triggered in a capillary tube heated within a thermally baffled device, where the denaturation, annealing, and extension of PCR are automatically completed passing sequentially through different temperature zones in a capillary tube. Fluorescence signals produced by probe hydrolysis during the reaction are collected and converted to Signal after/Signal before (S/N) ratios by a data processing module. Finally, the positive or negative results are displayed on the device screen, which is user-friendly enough to perform by untrained personnel. In this study, we developed a simple and rapid iiPCR for ASFV field surveillance targeting a highly conserved region of B646L gene, which only takes 25 minutes and the sensitivity is equivalent to that of real-time PCR in the laboratory. The sequences of forward and reverse primers were 5′-ATT TCA TTA ATG ACT CCT GG-3′ and 5′- TAC TAT CAG GCC CCC TCT-3',respectively. The sequence of probe was 5'-[FAM]-ATA AAC CAT GGT TTA AAG-3′-BHQ1. The primers and probe were synthetized by Shanghai BIOLIGO Biotechnology Co. Ltd, China. For iiPCR setup, 35 μL PCR reaction mix (Luoyang Putai Biotechnology Co., Ltd, China) including 0.15 μmol/L primers, 0.05 μmol/L probe was added to individual PCR tubes. Five microliter template (diluted plasmid or extracted nucleic acid) and 10 μL liquid paraffin wax were added before a briefly spun in a mini centrifuge. The PCR tubes were put into the reaction chamber in the iiPCR analyzer (Beijing Wantai Biological Pharmacy Enterprise Co., Ltd., China) and run on the preset program, which top and bottom temperatures were set as 58 ​°C and 95 ​°C, respectively (Fig. 1). A threshold was formulated as mean S/N ratio of non-template blank reactions +5 standard deviations (SD) as previously described (Tsai et al., 2012). The mean and SD of S/N ratios from 20 blank ASFV iiPCR reactions were 1.02 and 0.08, respectively, resulting in a threshold of 1.42. The results were calculated with defaulted S/N thresholds and displayed as “+” or “–” for positive and negative samples.
Fig. 1

Illustration of ASFV iiPCR device and PCR tube. The PCR tubes were put into the individual reaction chambers in the compact and portable iiPCR device. The reaction started when pushing the start button on the touch screen and the real-time process of reaction visualized in percentage appeared on the screen until the results shown with “+” for positive sample or “–” for negative sample. The iiPCR could be completed within 25 minutes.

Illustration of ASFV iiPCR device and PCR tube. The PCR tubes were put into the individual reaction chambers in the compact and portable iiPCR device. The reaction started when pushing the start button on the touch screen and the real-time process of reaction visualized in percentage appeared on the screen until the results shown with “+” for positive sample or “–” for negative sample. The iiPCR could be completed within 25 minutes. To determine the sensitivity of ASFV iiPCR, 10-fold serial dilutions of a plasmid DNA containing B646L gene were used and compared with OIE recommended real-time PCR (Chen et al., 2019). The purified plasmid was quantified by measuring OD260 with the spectrophotometer. The concentration was converted into copy numbers using the following formula: y (copies/μL) = (6.02 ​× ​1023) ​× ​(x (ng/μL) ​× ​10−9 DNA)/(DNA length ​× ​660) (Chen et al., 2019). As shown in Table 1, detection limits of iiPCR reached at 10−8-fold dilution with positive results of five repeats. By contrast, limits of detection of real-time PCR reached same dilution with four positive results and one suspected result of five repeats. The one suspected sample was proved to be positive when it was repeated as the manual described. The above results indicated that the established iiPCR had equivalent sensitivity as the OIE recommended real-time PCR.
Table 1

Comparison of detection limits between ASFV iiPCR and real-time PCR.

DilutionsCopy number/reactionii PCR result (S/N)Real-time PCR result (Ct)
10−62000+(4.75)+(4.11)+(4.32)+(4.08)+(4.29)+(30.79)+(29.93)+(31.22)+(30.07)+(31.21)
10−7200+(3.23)+(3.18)+(3.69)+(4.11)+(3.75)+(34.12)+(33.87)+(34.76)+(33.52)+(34.69)
10−820+(2.02)+(2.34)+(1.89)+(2.15)+(2.11)+(37.45)+(37.23)+(37.95)+(37.89)±(38.23)
10−92–(1.01)–(1.00)–(1.02)–(0.99)–(0.98)–(N/A)–(N/A)–(N/A)–(N/A)–(N/A)

Notes: “+” indicates positive results, “–” indicates negative results, “±” indicates suspected result. The cutoff Ct value for the real-time PCR was set as 38 according to the manual instruction.

Comparison of detection limits between ASFV iiPCR and real-time PCR. Notes: “+” indicates positive results, “–” indicates negative results, “±” indicates suspected result. The cutoff Ct value for the real-time PCR was set as 38 according to the manual instruction. To assess the specificity of ASFV iiPCR, nucleic acids were extracted from 200 different tissues of pigs including serum, nasal swabs, spleen, kidney, lung, lymph nodes, heart, liver, and brain, which had been collected before 2018 and confirmed ASFV negative using OIE recommended real-time PCR. The S/N values of above samples ranged from 0.96 to 1.07, indicating the good specificity of ASFV iiPCR. Nucleic acids were also extracted from other swine viruses including classical swine fever virus (CSFV) C strain, highly pathogenic porcine reproductive and respiratory syndrome virus (PRRSV) CH-1R, pseudorabies virus (PRV) Bartha-K61, porcine circovirus 2 (PCV2) HH3, porcine epidemic diarrhea virus (PEDV) HN1301, transmissible gastroenteritis virus (TGEV) HN1203, porcine rotavirus (PRoV) HN03 and porcine parvovirus (PPV) HN-2011. The RNAs of these RNA viruses were reverse transcribed into cDNA using PrimeScript™ RT reagent Kit (TaKaRa) before performed ASFV iiPCR. As shown by Table 2, positive signals were only generated from ASFV plasmid sample but not from other swine pathogens tested, which indicated the established ASFV iiPCR has no cross-reactivity with other swine viruses.
Table 2

Specificity analysis of ASFV iiPCR.

VirusaStrainASF iiPCR
ASFV Real-time PCR (Ct)
ResultsS/N
ASFVPlasmid (2000 copies)+4.1729.96
CSFVC strain0.97N/A
PRRSVCH-1R0.99N/A
PRVBartha-K610.98N/A
PCV2HH31.02N/A
PEDVHN13010.99N/A
TGEVHN12031.01N/A
PRoVHN030.96N/A
PPVHN-20111.01N/A

ASFV, African swine fever virus; CSFV, classical swine fever virus; PRRSV, porcine reproductive and respiratory syndrome virus; PRV, pseudorabies virus; PCV2, porcine circovirus 2; PEDV, porcine epidemic diarrhea virus; TGEV, transmissible gastroenteritis virus; PRoV, porcine rotavirus; PPV, porcine parvovirus.

Specificity analysis of ASFV iiPCR. ASFV, African swine fever virus; CSFV, classical swine fever virus; PRRSV, porcine reproductive and respiratory syndrome virus; PRV, pseudorabies virus; PCV2, porcine circovirus 2; PEDV, porcine epidemic diarrhea virus; TGEV, transmissible gastroenteritis virus; PRoV, porcine rotavirus; PPV, porcine parvovirus. Fifty-six DNA samples were isolated from clinical samples including whole blood, sera, heart, lung, spleen, brain, liver, lymph node, tonsil, kidney and nasal swabs of 56 suspected infected pigs. Nucleic acids were extracted using the Ambion®MagMAXTM Total Nucleic Acid Isolation Kit (Applied Biosystems) by ASFV Regional Reference Laboratory in Guangzhou, China. Surveillance protocol was approved by the ethics committee of reference laboratory and clinical samples were inactivated before genomic DNA extraction and detection. Performance of the established ASFV iiPCR for detecting above clinical samples was evaluated by comparing test results of a real-time PCR assay. Nucleic acid extracts of clinical samples were tested in parallel. Among the 56 tested clinical samples, 30 samples were tested positive by ASFV iiPCR and real-time PCR (Table 3). The iiPCR S/N ratio of ASFV positive samples ranged from 3.96 to 9.85 and the Ct values of real-time PCR were from 22.75 to 34.65. The above results revealed ASFV iiPCR has comparable sensitivity with an OIE recommended real-time PCR on testing clinical samples.
Table 3

Comparison of ASFV iiPCR with real-time PCR in clinical samples.

Real-time PCR
Total
PositiveNegative
iiPCRPositive30030
Negative02626
Total302656
Comparison of ASFV iiPCR with real-time PCR in clinical samples. To test if clinical blood samples could be directly used as the PCR template in ASFV iiPCR without nucleic acid extraction, 10 plasma samples (seven ASFV negative and three ASFV positive samples) were prepared by a briefly spun in a mini centrifuge. Consistent with previous results, three ASFV positive plasma samples had S/N values of 9.23, 9.34 and 8.85, respectively. The seven ASFV negative plasma samples had S/N values from 0.97 to 1.04. Next, the above three ASFV positive plasma samples were 10-fold serially diluted with ASFV negative plasma as the PCR templates. In parallel, the nucleic acids were extracted from 100 μL of plasma samples and eluted with same volume of TE buffer. The extracted nucleic acids were then serially 10-fold diluted with TB buffer. As shown in Table 4, the detection limits for both unextracted or extracted samples were 10−4 dilutions, which indicated the unextracted plasma samples could be directly used as template in ASFV iiPCR without compromising its sensitivity.
Table 4

Sensitivity comparison of ASFV iiPCR between unextracted and extracted plasma.

Dilution/Sample IDUnextracted positive plasma sample (S/N)
Extracted positive DNA sample (S/N)
123123
10−1+(8.77)+(8.61)+(8.13)+(9.01)+(8.84)+(8.75)
10−2+(6.33)+(6.25)+(5.69)+(6.78)+(6.31)+(6.14)
10−3+(3.21)+(3.42)+(2.87)+(3.97)+(3.72)+(3.03)
10−4+(1.53)+(2.20)+(1.64)+(2.14)+(1.98)+(1.77)
10−5–(1.00)–(0.99)–(0.99)–(1.01)–(0.98)–(1.02)
Sensitivity comparison of ASFV iiPCR between unextracted and extracted plasma. The early and accurate diagnosis of ASFV was prerequisite for ASFV control to implement stringent measures and stamping-out policy before the onsite of clinical symptoms which appear as early as 4 days post-infection (Zhao et al., 2019). Currently, the OIE-recommended real-time PCR is widely used to diagnose ASFV in China. However, the sophisticated thermocycler, complicated procedures and laboratory-dependent operations limited their applications in point-of-care testing. Moreover, it could take several days to collect and transport samples to ASFV reference laboratory for real-time PCR test. The delayed testing may endanger nearby pig herds or farms exposed to ASFV infection. Therefore, a rapid, highly sensitive and specific, and field-deployable assay is necessary for ASFV control. Relying on the fluorescent probe hydrolysis driven by natural liquid convection in a capillary tube, iiPCR was recently developed and applied to several swine viral pathogens including CSFV, Foot and mouth disease virus, and Senecavirus A (Lung et al., 2016; Ambagala et al., 2017; Zhang et al., 2019). Compare to the real-time PCR, iiPCR has several advantages. Firstly, ASFV iiPCR is simple and rapid to diagnose with the comparable sensitivity and specificity as real-time PCR does. Excluding the time needed to extract DNA, it only takes 25 minutes to read the results. Second, the compact iiPCR device is field-deployable for onsite disease surveillance and outbreak monitoring. It could be used in remote, resource-deprived areas to get rapid results at a relatively low cost. Third, the automated interpretation by iiPCR machine makes it easier to read the results for untrained users who work in veterinary clinics, pig farms, slaughtering houses, and diagnostic laboratories. Tran et al. had established an iiPCR assay for rapid and on-site detection of ASFV (Tran et al., 2021). In their study, the established ASFV iiPCR has a good sensitivity as real-time PCR by testing two ASFV positive samples (one blood and one tissue homogenate) with 10-fold dilution. To valid the feasibility of ASFV iiPCR in the field, more clinical samples should be included in their study. Therefore, 56 clinical samples were used in our study and the results proved that ASFV iiPCR could be applied in the field with high sensitivity and accuracy as real-time PCR (Table 3). Besides conventional and real-time PCR, the loop-mediated isothermal amplification (LAMP) and recombinase polymerase amplification (RPA) were developed for onsite ASFV diagnosis (James et al., 2010; Miao et al., 2019). These molecular methods need nucleic acid extraction from clinical samples, which is inconvenient to perform in the field. To overcome this bottleneck, plasma samples were directly used as templates in the established ASFV iiPCR in this study. The results showed that ASFV iiPCR had same detection limits by the positive results of 10−4 dilution for both unextracted and extracted samples (Table 4). Since the plasma and sera are the commonly used in the field for ASFV diagnosis, ASFV iiPCR could work as a practical onsite tool for virus detection using only 5 μL volume of samples.

Footnotes

This research was supported by Luoyang Major Science and Technology Project (1901029A). Prof. Xiangdong Li now works at Yangzhou University, and this work has partially supported by Jiangsu Agricultural Science and Technology Independent Innovation Fund Project [CX(21)2035] and Jiangsu Provincial Key R&D plan (BE2020398). This study does not contain any studies with human participants or animals performed by any of the authors. The authors declare that they have no conflict of interest.
  14 in total

1.  Emergence of African Swine Fever in China, 2018.

Authors:  Xintao Zhou; Nan Li; Yuzi Luo; Ye Liu; Faming Miao; Teng Chen; Shoufeng Zhang; Peili Cao; Xiangdong Li; Kegong Tian; Hua-Ji Qiu; Rongliang Hu
Journal:  Transbound Emerg Dis       Date:  2018-09-04       Impact factor: 5.005

2.  African swine fever in China.

Authors:  Xiangdong Li; Kegong Tian
Journal:  Vet Rec       Date:  2018-09-08       Impact factor: 2.695

Review 3.  Development of vaccines against African swine fever virus.

Authors:  Elena G Sánchez; Daniel Pérez-Núñez; Yolanda Revilla
Journal:  Virus Res       Date:  2019-03-25       Impact factor: 3.303

4.  Detection of African swine fever virus by loop-mediated isothermal amplification.

Authors:  Heather E James; K Ebert; R McGonigle; Scott M Reid; Neil Boonham; Jennifer A Tomlinson; Geoffrey H Hutchings; Mick Denyer; Chris A L Oura; Juliet P Dukes; Donald P King
Journal:  J Virol Methods       Date:  2009-12-04       Impact factor: 2.014

5.  Development and evaluation of a real-time RT-PCR and a field-deployable RT-insulated isothermal PCR for the detection of Seneca Valley virus.

Authors:  Jianqiang Zhang; Charles Nfon; Chuan-Fu Tsai; Chien-Hsien Lee; Lindsay Fredericks; Qi Chen; Avanti Sinha; Sarah Bade; Karen Harmon; Pablo Piñeyro; Phillip Gauger; Yun-Long Tsai; Hwa-Tang Thomas Wang; Pei-Yu Alison Lee
Journal:  BMC Vet Res       Date:  2019-05-24       Impact factor: 2.741

Review 6.  Subunit Vaccine Approaches for African Swine Fever Virus.

Authors:  Natasha N Gaudreault; Juergen A Richt
Journal:  Vaccines (Basel)       Date:  2019-06-25

7.  Insulated Isothermal Reverse Transcriptase PCR (iiRT-PCR) for Rapid and Sensitive Detection of Classical Swine Fever Virus.

Authors:  O Lung; J Pasick; M Fisher; C Buchanan; A Erickson; A Ambagala
Journal:  Transbound Emerg Dis       Date:  2015-01-27       Impact factor: 5.005

8.  Field-Deployable Reverse Transcription-Insulated Isothermal PCR (RT-iiPCR) Assay for Rapid and Sensitive Detection of Foot-and-Mouth Disease Virus.

Authors:  A Ambagala; M Fisher; M Goolia; C Nfon; T Furukawa-Stoffer; R Ortega Polo; O Lung
Journal:  Transbound Emerg Dis       Date:  2016-09-03       Impact factor: 5.005

9.  Development of TaqMan probe-based insulated isothermal PCR (iiPCR) for sensitive and specific on-site pathogen detection.

Authors:  Yun-Long Tsai; Hwa-Tang Thomas Wang; Hsiao-Fen Grace Chang; Chuan-Fu Tsai; Ching-Ko Lin; Ping-Hua Teng; Chen Su; Chien-Chung Jeng; Pei-Yu Lee
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

View more
  1 in total

1.  Development of a Duplex Insulated Isothermal PCR Assay for Rapid On-Site Detection and Differentiation of Genotypes 1 and 2 of African Swine Fever Virus.

Authors:  Ruilong Song; Penggang Liu; Yang Yang; Hu Suk Lee; Changhai Chen; Xiaodong Wu; Xiangdong Li
Journal:  Front Cell Infect Microbiol       Date:  2022-07-07       Impact factor: 6.073

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.