| Literature DB >> 35254295 |
Mohammad Vakili1, Johan Bielecki1, Juraj Knoška2, Florian Otte1, Huijong Han1, Marco Kloos1, Robin Schubert1, Elisa Delmas1, Grant Mills1, Raphael de Wijn1, Romain Letrun1, Simon Dold1, Richard Bean1, Adam Round1, Yoonhee Kim1, Frederico A Lima1, Katerina Dörner1, Joana Valerio1, Michael Heymann3, Adrian P Mancuso1, Joachim Schulz1.
Abstract
The Sample Environment and Characterization (SEC) group of the European X-ray Free-Electron Laser (EuXFEL) develops sample delivery systems for the various scientific instruments, including systems for the injection of liquid samples that enable serial femtosecond X-ray crystallography (SFX) and single-particle imaging (SPI) experiments, among others. For rapid prototyping of various device types and materials, sub-micrometre precision 3D printers are used to address the specific experimental conditions of SFX and SPI by providing a large number of devices with reliable performance. This work presents the current pool of 3D printed liquid sample delivery devices, based on the two-photon polymerization (2PP) technique. These devices encompass gas dynamic virtual nozzles (GDVNs), mixing-GDVNs, high-viscosity extruders (HVEs) and electrospray conical capillary tips (CCTs) with highly reproducible geometric features that are suitable for time-resolved SFX and SPI experiments at XFEL facilities. Liquid sample injection setups and infrastructure on the Single Particles, Clusters, and Biomolecules and Serial Femtosecond Crystallography (SPB/SFX) instrument are described, this being the instrument which is designated for biological structure determination at the EuXFEL. open access.Entities:
Keywords: FEL physics; X-ray scattering; aerosols; crystallography; high-viscosity extrusion; instrumentation; liquid jets; microfluidics; rapid mixing; sample delivery; single-particle imaging
Mesh:
Year: 2022 PMID: 35254295 PMCID: PMC8900844 DOI: 10.1107/S1600577521013370
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616
Figure 1The high repetition rate X-ray pulse pattern at the EuXFEL. X-ray pulses arrive in 10 Hz trains at the sample and each train can provide up to 2700 ultrashort pulses.
Overview of X-ray parameters for various example settings on the SPB/SFX instrument
The pulse energies in settings 2–4 are comparable as the accelerator is usually running at 2.26 MHz and electron bunches are delivered interleaved to SASE1 and SASE3 (i.e. maximum 1.13 MHz delivered to each of SASE1 and SASE3). The number of X-ray pulses per 10 Hz train can be varied, leading to different intra-train repetition rates.
| Example values | ||||
|---|---|---|---|---|
| Parameter | 1 | 2 | 3 | 4 |
| Pulse duration (fs) |
| |||
| Photon energy (keV) | 3–16 | 9.3 | 9.3 | 9.3 |
| Pulse energy (mJ) | 4 | 1–2 | 1–2 | 1–2 |
| Beam size (µm) |
| |||
| Photon flux (photons pulse−1) | 1.0 × 1012 | |||
| Photon flux (photons s−1) | 2.7 × 1016 | 1.4 × 1016 | 6.8 × 1015 | 2.0 × 1015 |
| Brilliance [photons s−1 mm−2 mrad−2 (0.1% bandwidth)−1] | 5 × 1033 | |||
| No. of pulses per train | 2700 | 1350 | 352 | 202 |
| Time between two pulses (ns) | 221.6 | 443.3 | 886.5 | 1773.0 |
| Duration of pulse train (µs) | 598 | 598 | 312 | 358 |
| Intra-train rep. rate of pulses (MHz) | 4.51 | 2.26 | 1.13 | 0.564 |
| Time between two trains (ms) | 100 | |||
| Repetition rate of trains (Hz) | 10 | |||
| No. of pulses per second | 27000 | 13500 | 3520 | 2020 |
| Jet gap-to-gap spacing (µm) | ∼11 | ∼22 | 44 | ∼88 |
| Jet velocity (m s−1) | 49.6 | |||
Observed spacing between two gaps along the irradiated liquid jet (displacement between two pulses) in 1.13 MHz mode.
List of microchannel design parameters for the 2PP-3D printed liquid sample injection devices with an indication of required liquid flow rates (in µl min−1) for stable operation and compatible pulse operation modes (intra-train pulse repetition rates)
The denoted parameters for GDVNs describe the liquid orifice diameter D liquid, the gas orifice diameter D gas and the distance between these two orifices H liquid-gas. Mixing devices are further described by the diameter of the mixing channel D channel, the diameter of the main channel (sample inlet) D MC, the diameter of the side channel (reactant inlet) at the mixing initiation region D SC and the mixing channel length L channel. HVEs and CCTs for electrospray purposes are further described with their wall thickness at the tip z wall and the angle of the cone. This list and the referenced designs can be found online in our corresponding GitHub repository, https://github.com/flmiot/EuXFEL-designs.
| Device type | Design parameters |
| Pulse mode | Design name |
|---|---|---|---|---|
| GDVN |
| |||
| 30–30–30 | ≥5 | ≤2.3 MHz | JKMH_5 | |
| 50–60–60 | ≥8 | ≤2.3 MHz | JKMH_6 | |
| 75–60–75 | ≥8 | ≤1.1 MHz | MVED_B | |
| 100–90–100 | ≥10 | ≤0.5 MHz | MVED_C | |
| 150–100–150 | ≥75 | ≤0.14 MHz | MVED_D | |
| 180–145–180 | ≥100 | ≤0.14 MHz | MVED_E | |
| Micromixer |
| |||
| 100–75–180–300 | ≥10 | Depends on GDVN | MVED_Y | |
| 200–100–231.7–300 | ≥10 | MVED_Z | ||
| 200– (200/2)–(200/2)–950 | ≥10 | JKMH_10 | ||
| 100–(100/2)–(100/2)–950 | ≥10 | JKMH_10H | ||
| Mixing-GDVN |
| |||
| 100–75–180–2122.3–75–60–75 | ≥10 | ≤1.1 MHz | MVED_S | |
| 100–75–180–331.8–75–60–75 | ≥10 | ≤1.1 MHz | MVED_V | |
| DFFN | 100–75–96–445.6–75–70–70 | ≥5 + ≥15 (EtOH) | ≤1.1 MHz | JKMH_8 |
| HVE |
| |||
| 50–365–600–20–9 | ≥0.3 | 10 Hz | MV_K | |
| 75–365–600–20–9 | ≥0.3 | 10 Hz | MV_L | |
| 100–365–600–20–9 | ≥0.3 | 10 Hz | MV_T | |
| Mixing-HVE |
| |||
| 231.7–100–231.7–2570–100–345–600–20–9 | 0.3–3 | 10 Hz | MV_I | |
| 231.7–100–231.7–2570–75–345–600–32.5–9 | 0.3–3 | 10 Hz | MV_J | |
| CCT |
| |||
| 40–200–50–30 | 0.02–0.8 | ≤4.5 MHz | MV_W | |
| 10–200–10–20 | 0.02–0.8 | ≤4.5 MHz | MV_X | |
These designs were previously discussed elsewhere (Knoška et al., 2020 ▸).
Concentric cones.
Kenics.
Figure 2(a) An overview of the 2PP-3D printed mix-and-inject device (i.e. mixing-GDVN) consisting of a micromixer, a GDVN (type C) and a connective capillary in between. The capillary extension was chosen to be L 2 = 28 mm. (b) A microscopy image showing lysozyme crystal delivery on the SPB/SFX instrument (10× magnification, NA 0.28, pixel size ∼0.65 µm). Upstream from the depicted X-ray interaction region, the crystals enter from the main channel of the mixing device at 7 µl min−1 and are flow-focused by pure water entering from the side channel at 70 µl min−1. Downstream, the 11-fold diluted sample then enters the GDVN (type C, 100 µm liquid orifice). With a helium pressure of 550 psi (Q g = 34 mg min−1), a liquid jet of 7.5 µm in diameter delivers the crystals into the X-ray focus. With our prediction formula [Fig. 4(b)], the jet velocity was determined to be 30.5 m s−1. (c) A microscopy image of the lysozyme crystal dispersion from the utilized sample reservoir showing the near-monodisperse microcrystals. (d) A background-corrected detector image of lysozyme diffraction from the same 11-fold dilution collected on the AGIPD 1M detector. The detector (pixel size is 200 µm × 200 µm) consists of four movable quadrants, each quadrant consisting of four static independent modules. Each module is 26 mm × 103 mm (128 × 512 pixels) large and consists of 2 × 8 ASICs (application-specific integrated circuits). The magnified region (green rectangle) shows Bragg reflections within 2 × 2 ASICs.
Figure 3Side view images (10× magnification, NA 0.28, 1075 × 310 pixels2 detection area, pixel size ∼0.65 µm) taken inside the SPB/SFX sample chamber, each depicting a thin liquid jet (water) generated by a 75–60–75 (µm) GDVN using a liquid flow rate of Q l = 10 µl min−1 and an applied gas pressure of p He = 400 psi (Q g = 25 mg min−1). (a) The liquid jet in the absence of X-rays. (b) X-ray pulses arrive at f = 0.564 MHz (with 30 pulses per train) and create gaps in the liquid column (gap-to-gap spacing Δx = 104.7 µm). The X-ray interaction with the jet hence reveals a jet velocity of v jet = Δx × f pulse = 59.1 m s−1. (c) Two optical lasers (λ = 532 nm), each with a 5 ns pulse duration, illuminate the droplets 2 mm downstream of the jet region. The delay time between the two laser pulses is 119 ns and reveals a droplet displacement of Δx = 7.2 µm. Therefore, dual-pulse laser illumination reveals a droplet velocity of v droplet = Δx/Δt opt.pulse = 60.1 m s−1. The determined jet velocities imply a jet diameter of d jet = [4Q l/(πv)]1/2 ≃ 2 µm.
Figure 4(a, b) Plots of the experimentally determined jet velocities for the four different GDVN types as a function of applied liquid and gas flow rates. Circles represent data from the jet explosion method, while stars denote velocities from the droplet PIV method. In panel (b) the broken line describes the jet velocity prediction formula v jet = a + b × (Q g/D g 2)1/2/(Q l)1/4, where the constants have the numerical values a = 8.35 and b = 1022 if Q g is given in mg min−1, D g in µm, Q l in µl min−1 and v jet in m s−1. (c) The corresponding jet diameters are calculated via d jet = [4Q l/(πv)]1/2.
Figure 5(a) A photograph of an assembled micromixer (type Y with 100 µm ID) before connection to a GDVN. (b) A microscopic image showing a detailed view of the mixing initiation area. (c) Magnified regions of the mixing channel–capillary interface, showing the 3D hydrodynamic flow focusing of a central ink stream (main channel) where the diluting water (entering from the side channel) runs at various flow rates. The mixed species is seamlessly transferred from the 2PP-3D printed part into the capillary extension.
SFX data collection and refinement statistics
The entire peak search and index processing was performed identically across all data sets. The hit rates for dilution set 8 are unusually high. Under normal experimental conditions, each data set would be optimized independently. In this case, hit-finding is treated identically and the higher hit rates are probably due to the changes in jetting conditions and subsequent changes in detector background. More accurate values can be seen in the number of indexed crystals. The respective data quality statistics fall within typically acceptable values consistent with previous experiments.
| Dilution | |||||
|---|---|---|---|---|---|
| Parameter | 11 | 8 | 6 | 5 | 4 |
| Total number of frames | 3645696 | 3038080 | 1822848 | 608020 | 634684 |
| Total number of frames with hits | 206283 | 3037676 | 1155073 | 75838 | 36550 |
| Hit rate (%) | 5.66 | 99.99 | 63.37 | 12.47 | 5.76 |
| Number of indexed crystals | 80254 | 5 605 | 10 214 | 1 695 | 5 046 |
| Proportion of indexable hits (%) | 38.9 | 0.18 | 0.88 | 2.24 | 13.81 |
| Unit-cell lengths (Å) | 79.4, 79.5, 38.6 | ||||
| Unit-cell angles (°) | 90, 90, 90 | ||||
| Resolution (Å) | 32.1–2.0 (2.07–2.0) | ||||
|
| 6.89 (14.79) | 24.11 (53.17) | 20.24 (36.81) | 44.05 (93.68) | 28.88 (72.36) |
| CC1/2 (%) | 99.20 (95.06) | 91.12 (64.27) | 90.04 (71.85) | 74.66 (25.53) | 87.99 (52.61) |
| CC* (%) | 99.80 (98.73) | 97.65 (88.46) | 97.34 (91.44) | 92.46 (63.78) | 96.75 (83.04) |
| Signal-to-noise ratio | 13.24 (7.14) | 3.94 (1.96) | 4.98 (2.89) | 2.23 (1.03) | 3.07 (1.26) |
| Completeness (%) | 100 (100) | 100 (100) | 100 (100) | 99.76 (97.82) | 99.98 (99.81) |
Details of the respective mixing/jetting parameters can be found in the supporting information, Section S4.2.
Figure 6(a) Droplet diameters achieved as a function of applied liquid flow rate Q for different conductivities K. (b) In the dimensionless parametrization of the flow rate and droplet (Maißer et al., 2013 ▸), the droplet diameters from the different conductivities fall on the same curve. (c) A side-view microscopy image (0.92 µm pixel size) of the 2PP-3D printed CCT. The 40 µm ID of the electrospray tip exactly matches the ID of the utilized fused silica capillary and the 30° angle at the tip is seamlessly transferred to the angle of the formed liquid cone. The tip–liquid interface is indicated with the dotted red line.