Literature DB >> 35245018

A Demonstration of Broadband Cavity-Enhanced Absorption Spectroscopy at Deep-Ultraviolet Wavelengths: Application to Sensitive Real-Time Detection of the Aromatic Pollutants Benzene, Toluene, and Xylene.

Meng Wang1, Ravi Varma2, Dean S Venables3, Wu Zhou1, Jun Chen1.   

Abstract

Benzene, toluene, and xylene (BTX) are serious air pollutants emitted by the chemical industry. Real-time monitoring of these air pollutants would be a valuable tool to regulate emissions of these compounds and reduce the harm they cause to human health. Here, we demonstrate the first detection of BTX using incoherent broadband cavity-enhanced absorption spectroscopy (IBBCEAS). The instrument was operated in the deep-ultraviolet spectral region between 252 and 286 nm, where aromatic compounds have intense π → π* absorption bands. The mirror reflectivity was calibrated by two methods and exceeded 99.63% at 266 nm. At an integration time of 60 s, the 1σ measurement sensitivities were estimated to be 7.2 ppbv for benzene, 21.9 ppbv for toluene, 10.2 ppbv for m-xylene, and 4.8 ppbv for p-xylene, respectively. The absorption cross sections of BTX were measured in this work with an uncertainty of 10.0% at a resolution of 0.74 nm. The absorption cross sections reported in this work were in good agreement with those from earlier studies after accounting for differences in spectral resolution. To demonstrate the ability of the instrument to quantify complex mixtures, the concentrations of m-xylene and p-xylene have been retrieved under five different mixing ratios. Instrumental improvements and measurements strategies for use in different applications are discussed.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 35245018      PMCID: PMC8928152          DOI: 10.1021/acs.analchem.1c04940

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


Introduction

Benzene, toluene, and xylene (BTX) are an important class of volatile organic compounds in the atmosphere that primarily enter the atmosphere from vehicles, biomass burning, solvent use, and the petrochemical industry.[1,2] Benzene is a known carcinogen and increases the risk of other illnesses. It is a notorious cause of bone marrow failure. The primary reaction pathways for monocyclic aromatics are reaction with hydroxyl radicals to form phenols or degradation to carbonyls, such as glyoxal.[3−5] Oxidation of BTX also produces secondary organic aerosols and ozone, thereby aggravating air pollution and exacerbating impacts on public health.[6,7] Analytical methods for quantifying BTX in the air usually include absorption traps and subsequent separation by gas chromatography with detection by flame ionization, photo-ionization, or mass chromatography.[8,9] Measurement time resolution is usually about 30–60 min per sample. In recent years in situ optical methods, such as long path differential optical absorption spectroscopy (LP-DOAS), have also been applied.[10] Unlike other methods, LP-DOAS uses a long physical path (typically 500–1000 m) and measures the path-integrated concentration over this distance. However, the correlation between the two types of methods is poor because of the different spatial scales; DOAS usually reports higher values than in situ methods.[11] Incoherent broadband cavity-enhanced absorption spectroscopy (IBBCEAS) has been applied to many gaseous molecules since it was first proposed by Fiedler et. al, mostly at visible wavelengths.[12] There are few applications in the UV range.[13−17] Optical cavity measurements in the UV range are challenging for several reasons: (1) CCD detectors have relatively low quantum efficiency in the UV, and light sources are less intense in this region. (2) Losses at optical cavity mirror surfaces by scattering and absorption are much higher than at longer wavelengths. (3) System sensitivity is limited by Rayleigh scattering at short wavelengths. (4) Increasing Mie scattering from particles, and strong ozone absorption is a challenge for atmospheric samples. As a result of the high inherent losses in the sample gas and the lower mirror reflectivity, the extremely long effective light pathlengths attained by optical cavity instruments at visible and near-infrared wavelengths are not feasible below 300 nm. Nevertheless, the ability to measure absorption and scattering at deep-UV wavelengths with high sensitivity would be valuable for fundamental research, monitoring of ambient air quality[10] and industrial emissions,[11] and other uses. To date, there have been few applications of optical cavity methods below 300 nm. Ityaksov and coworkers used cavity ring-down spectroscopy with a dye laser to measure Rayleigh scattering cross sections of N2, CH4, and SF6 between 198 and 270 nm.[18] Prior work with broadband cavity instruments includes a deep-UV CEAS instrument using a mercury atomic emission lamp (254 nm) to measure gaseous mercury and ozone with detection limits of 8.1 and 8.4 ppbv, respectively, in 10 s acquisition time,[19] an IBBCEAS system combining a laser-driven light source (LDLS) and LED to measure acetone between 272 and 292 nm with an LOD of 8 ppmv in 1 s acquisition time,[20] and a deep UV-LED system at 280 nm that measured ozone with an LOD below 0.1 ppm with an accuracy of 0.5%.[21] The aim of this work is to build on these prior studies and demonstrate that IBBCEAS is a useful approach for highly sensitive measurements of sample extinction at deep-UV wavelengths. We describe an LED IBBCEAS spectrometer operating in the 252–286 nm wavelength range, which achieves an LOD of 3.7 × 10–7 cm–1 with a 60 s acquisition time, a sensitivity high enough to be valuable for studying many compounds that absorb strongly at these wavelengths. We demonstrate that this instrument can be used to make sensitive, real-time measurements of aromatic compounds, specifically the important environmental pollutants BTX, and has potential applications in laboratory research, emissions monitoring, and other combined applications.[22]

Experimental Section

Instrument Setup

The IBBCEAS approach was first described by Fiedler et al.[12] and has been applied in many studies.[23,24] In this work, the deep-UV IBBCEAS instrument comprised three subsystems: the light source, optical cavity, and detection module (Figure a). The light source was a 255 nm LED (HSE255H-M807X, Hasunopto, China) with an FWHM of 11 nm, and the operation status was 5.3 V forward voltage and 0.17 A current. The LED was mounted on a thermostat and driven by a constant current source to minimize current and temperature fluctuations. The thermostat was stabilized to 15 ± 0.1 °C using a PID controller.
Figure 1

(a) A schematic of the DUV-IBBCEAS instrument for the measurement of benzene, toluene, and xylene. The optical portion of the instrument consists of a temperature-stabilized LED, collimating and focusing optics (lenses L1 and L2), a short-pass filter (F1), the optical cavity formed by plano-concave HR dielectric mirrors (M1 and M2), specialized fiber bundle, grating spectrometer, and a CCD detector. (b) Gas mixture and sampling system for generating different concentrations of benzene, toluene, and xylene. The sample gas is drawn through the cavity from the inlet using a diaphragm pump.

(a) A schematic of the DUV-IBBCEAS instrument for the measurement of benzene, toluene, and xylene. The optical portion of the instrument consists of a temperature-stabilized LED, collimating and focusing optics (lenses L1 and L2), a short-pass filter (F1), the optical cavity formed by plano-concave HR dielectric mirrors (M1 and M2), specialized fiber bundle, grating spectrometer, and a CCD detector. (b) Gas mixture and sampling system for generating different concentrations of benzene, toluene, and xylene. The sample gas is drawn through the cavity from the inlet using a diaphragm pump. The light emitted from the LED was collected and focused by a lens (f = 35 cm) into a high-finesse optical cavity formed by a pair of high-reflectivity (HR) mirrors (Layertec GmbH). The reflectivity of the HR mirror between 250 and 280 nm was specified by the manufacturer to be greater than 0.995. The cavity mirrors were separated by 96.75 ± 0.02 cm. Light transmitted through the cavity was focused by another lens (f = 75 cm), filtered by a short-pass filter (Semrock FF01-300/SP-25; transmission >70% between 255 and 290 nm), and coupled into a quartz fiber bundle. The fiber bundle was 1.5 m long and consisted of 20 × 200 μm core fibers (Figure a). The fibers were vertically arranged at the entrance slit of an imaging spectrograph to optimize light coupling and maximize the signal. The detection module was a grating spectrograph (Andor SR-303i-B, Oxford Instruments) with a CCD detector (Andor DH334T-18 U-E3, Oxford instruments) cooled to −25.0 °C to minimize the dark current. The detector’s Micro-Channel Plate (MCP) image intensifier can be varied to amplify the signal intensity. The spectrograph had a 1200 L mm–1 diffraction grating (300 nm blaze) and covered the range between 252 and 286 nm. The wavelength resolution was 0.74 nm, based on a fit to the 253.65 nm emission line of a commercial mercury lamp (Hg-1, Ocean Optics).

Sample Handling

Samples were introduced in a quartz flow tube (25 mm inner diameter) with a separation of 69.7 cm between the inlet and outlet (Figure b). The concentration of benzene, toluene, m-xylene, and p-xylene was generated by mixing a flow of N2 (>99.999% purity) with the output of a gas cylinder comprising 1.04 ± 0.02 ppmv for benzene, 1.02 ± 0.02 ppmv for toluene, 1.00 ± 0.02 ppmv for m-xylene, and 1.02 ± 0.02 ppmv for p-xylene (Dalian Special Gases). Flow rates of the N2 and aromatic gas mixture streams were controlled by mass flow controllers (MFC). The flow rates of BTX gases were controlled at 1000 sccm, and the N2 flow rate was controlled at 666, 250, and 0 sccm, respectively. Thus, the BTX gas was diluted to 600, 800, and 1000 ppbv at about 298 ± 1 K and 101 ± 1 kPa. One hand of the cavity was connected to a buffer bottle, from which sample was drawn into the cavity at a flow rate of 800 sccm (controlled by an MFC) by a pump. Excess gas in the buffer bottle was discharged to the fume hood. MFC were calibrated by a flow meter (Gilibrator-2, Sensidyne).

HR-Mirror Calibration

In IBBCEAS systems, the extinction coefficient εext(λ) of the gas sample in the cavity is related to the measured properties and system parameters through the following relation: Here, instrument parameters are the cavity length d, the mirror reflectivity R(λ), the measured reference and sample intensity spectra (I0(λ) and I(λ), respectively), and the extinction due to the Rayleigh scattering αRayl(λ). The mirror reflectivity R(λ) is essential for quantitative results and must be calibrated. In previous work, R(λ) has been determined using different methods based on a known extinction in the cavity. There were also many other different calibration approaches.[16,25−30] In this study, we used two approaches to determine the mirror reflectivity, R(λ). The first approach used a known gas absorption of SO2 to determine the reflectivity, RSO(λ). A 1005 ppm SO2 gas mixture (in N2) was further diluted in a stream of pure nitrogen to 600 and 1000 ppbv with a multigas calibrator (164i, Thermal Fisher). The reflectivity RSO(λ) can be determined from:where IN(λ) and ISO(λ) are the measured light intensity when the cavity was filled with nitrogen and a known number density (nSO) of SO2, respectively, and σSO(λ) is the literature absorption cross section of SO2[31] convolved to our spectrograph instrument function. This approach gave an uncertainty in RSO(λ) of about 4%, which is dominated by the uncertainty in the SO2 cross section. The second approach used the difference in Rayleigh scattering of two pure gases for another determination of the mirror reflectivity, RRayl(λ). Pure nitrogen (>99.999%) and pure helium (>99.999%) were used to determine RRayl(λ) according to eq :where IN(λ) and IHe(λ) represent the measured intensities when either nitrogen or helium, respectively, is in the cavity and σRayl,N(λ) and σRayl,He(λ) are the Rayleigh scattering cross sections of nitrogen and helium, respectively. The uncertainty of RRayl(λ) is about 5% and mainly determined by the uncertainty in the scattering cross section of N2 reported by Sneep and Ubachs.[32] The intensity of the cavity at 1 s integration time is shown in Figure a. A wide spectral range from 250 to 295 can be achieved. As shown in Figure b, the reflectivity of the mirror exceeded 99.63% at its center near 266 nm. The averaged reflectivity R(λ) was used in this instrument. Based on Gaussian error propagation, the uncertainty in R(λ) is 6.4%.
Figure 2

(a) The intensity of the cavity with 1800 gain at 1 s integration time. (b) Mirror reflectivities were calibrated with two methods: based on the difference in Rayleigh scattering of nitrogen and helium (green line) and based on a known SO2 absorption (red line) all with 1σ precision. The black line represents the calculated mean mirror reflectivity from both methods.

(a) The intensity of the cavity with 1800 gain at 1 s integration time. (b) Mirror reflectivities were calibrated with two methods: based on the difference in Rayleigh scattering of nitrogen and helium (green line) and based on a known SO2 absorption (red line) all with 1σ precision. The black line represents the calculated mean mirror reflectivity from both methods.

Results and Discussion

System Performance

The optimal integration time for an instrument is commonly determined by calculating the Allan deviation:[14,23,33−35]where t is the integration time, M is the number of time series, and α(t) is the absorption coefficient in each subset from i = 1 to i = M – 2. The Allan deviation σA(t) is the square root of σA2(t). A total of 20,000 reference spectra with 1 s exposure time were continuously measured over 5 h while the cavity was filled with pure nitrogen (>99.999%). System sensitivity increased with integration time up to 1000 s (Figure ). This result is not surprising because the system precision was fundamentally limited by low light levels and longer averaging implies that more photons are measured. The system gain setting also influenced instrument precision for a given measurement time, with a higher gain producing better precision for a given integration time. The maximum precision of the instrument was 2.3 × 10–7 cm–1 without the MCP image intensifier (Gain = 0); the precision improved to 1.1 × 10–7 cm–1 using the MCP image intensifier with Gain = 1800. The retrieval precision of the DUV-IBBCEAS system to BTX over a 60 s acquisition time was estimated by fitting the zero air spectrum to the BTX absorption cross sections reported by Fally et al.[36] A histogram analysis of 600 zero measurements is shown in Figure b. The 1σ measurement precision was 7.2 ppbv for benzene, 21.9 ppbv for toluene, 10.2 ppbv for m-xylene, and 4.8 ppbv for p-xylene, respectively.
Figure 3

(a) Allan deviation of the absorption coefficient at 266 nm. For an integration time of 60 s and at a gain setting of 1800, the precision of the instrument is 3.70 × 10–7 cm–1. (b) The histogram analysis of the concentration measurement of benzene, toluene, m-xylene, and p-xylene based on 600 measurements of 60 s. Concentrations were retrieved by fitting the measured spectrum with the absorption cross sections convolved from Fally et al.[36]

(a) Allan deviation of the absorption coefficient at 266 nm. For an integration time of 60 s and at a gain setting of 1800, the precision of the instrument is 3.70 × 10–7 cm–1. (b) The histogram analysis of the concentration measurement of benzene, toluene, m-xylene, and p-xylene based on 600 measurements of 60 s. Concentrations were retrieved by fitting the measured spectrum with the absorption cross sections convolved from Fally et al.[36]

Absorption Cross-Section Measurement

Table summarizes previous determinations of gas-phase absorption cross sections of benzene, toluene, m-xylene, and p-xylene at room temperature and at different pressures and resolutions.
Table 1

Instrumental Parameters of Literature Absorption Cross Sections

referencestemp. (K)pressure (mbar)wavelength (nm)resolution (nm)
Benzene
Trost et al. (1997)[38]293.51000230–2780.11
Etzkorn et al. (1999)[39]2981000235–2760.15
Fally et al. (2009)[36]2931.33239–2700.007
Olive (2015)[40]298N.A.200–2650.13
Dawes et al. (2017)[37]29810–9115–3300.1
Toluene
Etzkorn et al. (1999)[39]2981000237–2780.15
Koban et al. (2004)[41]2961000227–2731
Olive (2005)[40]298N.A.225–2850.08
Fally et al. (2009)[36]2933.05242–2780.007
m-Xylene
Bolovinos et al. (1982)[42]29833.3139–2800.25
Trost et al. (1997)[38]293.51000239–2780.11
Etzkorn et al. (1999)[39]2981000242–2830.15
Olive (2005)[40]298N.A.235–2850.05
Fally et al. (2009)[36]2934.12242–2860.007
p-Xylene
Bolovinos et al. (1982)[42]29833.3139–2800.25
Trost et al. (1997)[38]2931000239–2780.11
Etzkorn et al. (1999)[39]2981000242–2830.15
Olive (2005)[40]298N.A.235–2850.05
Fally et al. (2009)[36]2932.08242–2860.007

Benzene

The absorption coefficients were measured over a 30 min integration time. The absorption coefficients of benzene at 600, 800, and 1000 ppbv obtained in this work are shown in Figure S1(a). Figure a shows that the absorption coefficient at 253 nm is linearly correlated with concentration (R2 > 0.999). The intercept (6.88 × 10–8 cm–1) and its standard deviation (1.95 × 10–7 cm–1) are both small.
Figure 4

(a) Dependence of the benzene absorption coefficient at 252.96 nm on concentration. The correlation coefficient of the linear regression is R2 > 0.999. (b) Benzene absorption cross section from this work and recent literature spectrum.[36−40] Note that the y-axis uses a logarithmic scale.

(a) Dependence of the benzene absorption coefficient at 252.96 nm on concentration. The correlation coefficient of the linear regression is R2 > 0.999. (b) Benzene absorption cross section from this work and recent literature spectrum.[36−40] Note that the y-axis uses a logarithmic scale. The UV absorption of benzene is attributed to the S1(1B2u) ← S0(1A1g) electronic transition.[36] Narrow vibrational transitions dominate in this range, and spectral resolution strongly affects the measured spectrum. The absorption cross section of benzene measured in this work is compared against prior work in Figure b. Two major absorption peaks are apparent in our measurement window. The position of the absorption peaks reported in our study is consistent with the data of Fally et al.[36] and Dawes et al.,[37] whereas the spectrum reported by Olive[40] is shifted −0.2 nm compared to others. For highly structured spectral features, the instrument resolution has a large effect, and high-intensity structures are apparent in the high-resolution measurements of Fally et al.[36] These features are smoothed out at coarser resolution. Indeed, the resolution of our instrument is a factor of 4 or more lower than other measurements. The absorption cross section of recent literature convolved to this work’s resolution is shown in Figure S2. The absorption cross-sections reported in this work were the mean value obtained by averaging the absorption cross-sections of three concentrations (600, 800, and 1000 ppbv) and were in good agreement with cross sections obtained by linear regression analysis.

Toluene

The absorption coefficients of toluene at 600, 800, and 1000 ppbv obtained in this work are shown in Figure S1(b). As with benzene, the absorption (at 267 nm) was highly linear (R2 > 0.999) with toluene concentration (Figure a), and the intercept was small 4.65 × 10–8 cm–1 (standard deviation of 8.22 × 10–8 cm–1).
Figure 5

(a) Dependence of the toluene absorption coefficient on the different concentrations at 267 nm. The correlation coefficient of the linear regression is R2 ≥ 0.999. (b) Toluene absorption cross-section from this work and recent literature spectrum.[36,39−41] Note that the y-axis scale is linear.

(a) Dependence of the toluene absorption coefficient on the different concentrations at 267 nm. The correlation coefficient of the linear regression is R2 ≥ 0.999. (b) Toluene absorption cross-section from this work and recent literature spectrum.[36,39−41] Note that the y-axis scale is linear. Figure b shows the absorption cross section of toluene measured with the DUV-IBBCEAS system, together with the cross sections of Etzkorn et al.,[39] Koban et al.,[41] Olive,[40] and Fally et al.[36] The agreement between our cross section and those of Etzkorn et al.,[39] Koban et al.,[41] and Olive[40] is generally good, while the cross section reported by Fally et al.[36] is lower than other studies.

m-Xylene

The absorption coefficients of m-xylene at 600, 800, and 1000 ppbv obtained in this work are shown in Figure S1(c). The three absorption coefficients at 271 nm are compared in Figure a and also highly linear with concentration (R2 > 0.999) with a small y-intercept 5.78 × 10–8 cm–1 (standard deviation of 1.55 × 10–7 cm–1).
Figure 6

(a) Dependence of the m-xylene absorption coefficient on the different concentrations at 271 nm. The correlation coefficient of the linear regression is R2 > 0.999. (b) m-Xylene absorption cross section from this work and recent literature spectrum.[36,38−40,42]

(a) Dependence of the m-xylene absorption coefficient on the different concentrations at 271 nm. The correlation coefficient of the linear regression is R2 > 0.999. (b) m-Xylene absorption cross section from this work and recent literature spectrum.[36,38−40,42] The absorption cross section of m-xylene measured in this work is shown in Figure b, together with the spectra of Bolovinos et al.,[42] Trost et al.,[38] Etzkorn et al.,[39] Olive,[40] and Fally et al.[36] The agreement between the absorption cross section measured in this work, and those reported previously are good, with the notable exception of the cross section reported by Trost et al.,[38] which had a relatively large positive offset (see also Figure b for comparison with the work of Trost et al.).

p-Xylene

The absorption coefficients of p-xylene at 600, 800, and 1000 ppbv obtained in this work are shown in Figure S1(d), and the three absorption coefficients at 273 nm are compared in Figure a. The absorption was linear with concentration (R2 > 0.999) with a small y-intercept 1.82 × 10–7 cm–1 with a standard deviation of 4.17 × 10–7 cm–1.
Figure 7

(a) Dependence of the p-xylene absorption coefficient on the different concentrations at 272.79 nm. The correlation coefficient of the linear regression is R2 > 0.999. (b) p-Xylene absorption cross section from this work and recent literature spectrum.[36,38−40,42]

(a) Dependence of the p-xylene absorption coefficient on the different concentrations at 272.79 nm. The correlation coefficient of the linear regression is R2 > 0.999. (b) p-Xylene absorption cross section from this work and recent literature spectrum.[36,38−40,42] The absorption cross section of p-xylene measured in this work is shown in Figure b, together with the spectra of Bolovinos et al.,[42] Trost et al.,[38] Etzkorn et al.,[39] Olive,[40] and Fally et al.[36] The agreement between the spectrum in this work, and the spectra reported by literature are good over this region. The spectrum reported by Trost et al.[38] was marginally higher than other values.

Uncertainty of Cross Section

The uncertainty in our absorption cross-section measurements was estimated using standard uncertainty propagation. The uncertainty of extinction coefficient ε, can be calculated by eq :[16]where Δd, Δ(1 – R), and ΔI0 are the uncertainties in the cavity length, mirror reflectivity, and spectral intensity, respectively. The relative uncertainties in eq are 1% for d, 6.4% for 1 – R (according to the uncertainty of R), and 3.5% for intensity fluctuations at a 1 min integration time. The overall uncertainty in the extinction measurement is therefore around 9.8% and is dominated by the uncertainty in the mirror reflectivity. The uncertainty in the absorption cross section must also account for the uncertainty in the concentration of each species (ΔN): The uncertainties in the concentrations of gas cylinder are 2% for benzene, toluene, m-xylene, and p-xylene according to the manufacturer. The uncertainty in flowrate is 0.35% F.S. according to the manufacturer and calibrated by a flow meter (Gilibrator-2, Sensidyne). And the combined uncertainty in absorption cross sections is thus 10.0% for all four gases at a resolution of 0.74 nm.

Measurement of Xylene Isomers

The least-squares spectral fitting method was used to demonstrate that deep-UV spectral analysis could be used to quantify the concentrations of different xylene isomers at the same time. The optimized spectral fitting window was from 258 to 278 nm. A third-order polynomial was applied to fit the shape of the background spectrum, which was mainly produced by the spectral dependence of the mirror reflectivity and from sample scattering. The reported absorption cross sections of xylene in this work were used to retrieve the concentration of xylene isomers. As shown in Figure a, five sets of m-xylene/p-xylene mixtures diluted by standard gas cylinders were measured with an integration time of 60 s. During the experiment, the fractional flowrate of p-xylene was reduced in a stepwise manner (80, 60, 50, 40, and 20%); the fractional flow rate of m-xylene was increased sequentially from 20, 40, 50, 60, and 80%. The uncertainty of retrieved concentration was dominated by absorption cross section and reflectivity.
Figure 8

(a) Measurement result of five different concentrations of p-xylene (red line and circle) and m-xylene (black lines and squares) standard gases. The blue positive triangle and green inverted triangle represent the flow ratios of the p-xylene and m-xylene standard gas mixtures, respectively. (b) An example spectral fit to one spectrum measured 83 min from the start of the experiment. The retrieved spectral fits of m-xylene (c) and p-xylene (d). The red line represents the fitted spectrum, and the black line is the fitted result plus the residual with 1σ error bar.

(a) Measurement result of five different concentrations of p-xylene (red line and circle) and m-xylene (black lines and squares) standard gases. The blue positive triangle and green inverted triangle represent the flow ratios of the p-xylene and m-xylene standard gas mixtures, respectively. (b) An example spectral fit to one spectrum measured 83 min from the start of the experiment. The retrieved spectral fits of m-xylene (c) and p-xylene (d). The red line represents the fitted spectrum, and the black line is the fitted result plus the residual with 1σ error bar. Figure b,c,d shows an example of the spectral fitting of a spectrum of the m-xylene and p-xylene mixture measured 83 min from the start of the experiment. The retrieved mixing ratios of m-xylene and p-xylene were 493 ± 12 ppbv and 505 ± 5 ppbv, respectively. The corresponding fitting residual is in the range of ±7.6 × 10–7 cm–1.

Discussion

The instrument presented in this work represents an important advance in the application of broadband optical cavity techniques to deep-UV wavelengths. We applied our system to quantify the important anthropogenic environmental pollutants BTX. Nevertheless, we emphasize that the method has general application to other compounds that absorb in this region, as well as to particle extinction measurements. This work therefore has broad applicability to a range of applications and research fields, whether for air quality or emissions monitoring, pure or applied laboratory research, or other technological applications. In the following, we discuss considerations around the BTX measurements presented in this work, propose instrumental improvements and measurement strategies, and assess their use in different applications. The 1σ measurement precisions of the DUV-IBBCEAS system over a 60 s acquisition time (7.2 ppbv (1σ) for benzene, 21.9 ppbv (1σ) for toluene, 10.2 ppbv (1σ) for m-xylene and 4.8 ppbv (1σ) for p-xylene) are already sufficient to be relevant to a range of applications, including fence-line and continuous emissions monitoring of these species. It also provides a potentially useful tool for analytical chemistry in combination with pre-concentration or chromatographic approaches. These ideas will be explored in future work to monitor atmospheric BTX. Several strategies are possible to further improve the system sensitivity. Our spectrometer used a relatively low-resolution configuration to increase light throughput owing to the challenges of working with low light levels in the deep UV. Although there is a trade-off between resolution and optical throughput, higher resolution spectra would be advantageous for capturing highly structured spectral features (for benzene, in particular) with improved sensitivity and selectivity. Light source intensity is a major technical challenge in this spectral region. The advantages of deep-UV LEDs are that they are relatively inexpensive light sources, produce stable emission intensities with careful control of temperature and current and do not need extensive spectral filtering. Other light sources with deep-UV output could prove advantageous for some applications. LDLS or Xe flashlamps have relatively high-intensity DUV output and could be useful alternative sources but would require stringent filtering to remove unwanted wavelengths from the optical system. Deuterium lamps would be challenging to use at these wavelengths because they are not particularly bright. For applications where lower absorption sensitivities would suffice, light throughput could be raised by using lower reflectivity mirrors in the optical cavity, resulting in greater throughput and higher time resolution, at the cost of a shorter effective light path length and lower sensitivity. In addition to the technological challenges described above, absorption by ozone is also at its maximum in this spectral region and could prove challenging for working with ambient air samples. Appreciable absorption by O3 in a sample would reduce the overall instrument sensitivity. An O3 denuder in the inlet line could circumvent this issue, albeit at the cost of greater system complexity. Alternatively, sample preconcentration could be used to attain much higher sensitivities and avoid some interferences while retaining almost real-time measurement capability. We will explore the application of the DUV-IBBCEAS system to the measurement of ozone concentrations in future work. The influence of other atmospheric constituents on system performance should also be considered. The common atmospheric pollutant NO2 has an absorption cross section between 250 and 280 nm that is around two orders of magnitude lower than ozone and BTX species. Unless present in much higher concentrations than BTX, the influence of NO2 absorption on the retrieval of BTX concentration will be negligible. However, aerosol extinction, mainly from Mie scattering, can exceed 10–6 cm–1 in a polluted environment.[27,43] Such high extinctions will substantially reduce the effective path length of light in the cavity and hence lower the instrument sensitivity. In our instrument, for instance, the effective path length at 266 nm is ca. 189 m under minimal aerosol conditions but decreases to below 160 m when the aerosol extinction coefficient is 10–6 cm–1. To minimize the effect of ambient aerosols, a simple solution is to install a particle filter in the inlet stream. This work shows the potential of the IBBCEAS approach for measurements of absorbing gases in the deep UV. Many other volatile species absorb between 250 and 280 nm, notably carbonyl species and other small aromatics, SO2 (as used to calibrate our instrument), organic reaction intermediates and radicals (like the ethyl, benzyl, and phenyl radicals, the phenoxy radical, and the phenyl, methyl, and benzyl peroxy radicals), and a range of chlorine oxide species (ClO, ClOO, ClOOCl, Cl2O, Cl2O6). For research applications, spectroscopic methods would have exceptional sensitivity to these highly unstable species and the ability to measure them in situ would be valuable tools. It is also worth pointing out that the 200–230 nm spectral region is particularly important for the measurement of the key environmental gases NH3, SO2, isoprene, and NO, even if commercially available DUV LEDs cannot yet attain such short wavelengths, and mirror reflectivities are worse at these short wavelengths. Some active DOAS systems have already demonstrated high sensitivity in this region with light paths of 100 m,[44] which is useful for sampling particularly sticky gases like NH3.

Conclusions

This study presented a novel IBBCEAS system for the measurement of BTX. Combined with the 255 nm LED, optical cavity, and short-pass filter, a spectral range from 250 to 295 nm was achieved. The absorption cross sections of benzene, toluene, p- and m-xylene measured with the spectrometer were in good agreement with previous work by others. We demonstrated the application of the system to the simultaneous retrieval of mixtures of m-xylene and p-xylene. The extension of the approach to other species of scientific and technological interest, as well as ways to optimize the spectrometer for different purposes, was discussed.
  15 in total

1.  Cavity-enhanced absorption using an atomic line source: application to deep-UV measurements.

Authors:  Steven B Darby; Paul D Smith; Dean S Venables
Journal:  Analyst       Date:  2012-04-03       Impact factor: 4.616

2.  Near-ultraviolet absorption cross sections of nitrophenols and their potential influence on tropospheric oxidation capacity.

Authors:  Jun Chen; John C Wenger; Dean S Venables
Journal:  J Phys Chem A       Date:  2011-10-14       Impact factor: 2.781

3.  High sensitivity in situ monitoring of NO3 in an atmospheric simulation chamber using incoherent broadband cavity-enhanced absorption spectroscopy.

Authors:  Dean S Venables; Titus Gherman; Johannes Orphal; John C Wenger; Albert A Ruth
Journal:  Environ Sci Technol       Date:  2006-11-01       Impact factor: 9.028

4.  Incoherent broadband cavity-enhanced absorption spectroscopy in the near-ultraviolet: application to HONO and NO2.

Authors:  Titus Gherman; Dean S Venables; Stewart Vaughan; Johannes Orphal; Albert A Ruth
Journal:  Environ Sci Technol       Date:  2008-02-01       Impact factor: 9.028

5.  Long optical cavities for open-path monitoring of atmospheric trace gases and aerosol extinction.

Authors:  Ravi M Varma; Dean S Venables; Albert A Ruth; Uwe Heitmann; Eric Schlosser; Sophie Dixneuf
Journal:  Appl Opt       Date:  2009-02-01       Impact factor: 1.980

6.  Demonstration of a novel laser-driven light source for broadband spectroscopy between 170 nm and 2.1 μm.

Authors:  Meez Islam; Luca Ciaffoni; Gus Hancock; Grant A D Ritchie
Journal:  Analyst       Date:  2013-07-05       Impact factor: 4.616

7.  Insights into the Formation and Evolution of Individual Compounds in the Particulate Phase during Aromatic Photo-Oxidation.

Authors:  Kelly L Pereira; Jacqueline F Hamilton; Andrew R Rickard; William J Bloss; Mohammed S Alam; Marie Camredon; Martyn W Ward; Kevin P Wyche; Amalia Muñoz; Teresa Vera; Mónica Vázquez; Esther Borrás; Milagros Ródenas
Journal:  Environ Sci Technol       Date:  2015-10-28       Impact factor: 9.028

8.  Proton-Transfer-Reaction Mass Spectrometry: Applications in Atmospheric Sciences.

Authors:  Bin Yuan; Abigail R Koss; Carsten Warneke; Matthew Coggon; Kanako Sekimoto; Joost A de Gouw
Journal:  Chem Rev       Date:  2017-10-04       Impact factor: 60.622

9.  Vacuum ultraviolet photoabsorption spectroscopy of crystalline and amorphous benzene.

Authors:  Anita Dawes; Natalia Pascual; Søren V Hoffmann; Nykola C Jones; Nigel J Mason
Journal:  Phys Chem Chem Phys       Date:  2017-10-18       Impact factor: 3.676

10.  Smog chamber study of the effects of NOx and NH3 on the formation of secondary organic aerosols and optical properties from photo-oxidation of toluene.

Authors:  Xue Qi; Shuping Zhu; Chenzhang Zhu; Jing Hu; Shengrong Lou; Li Xu; Junguo Dong; Ping Cheng
Journal:  Sci Total Environ       Date:  2020-04-13       Impact factor: 7.963

View more
  1 in total

1.  Detection of Sulfur Dioxide by Broadband Cavity-Enhanced Absorption Spectroscopy (BBCEAS).

Authors:  Ryan Thalman; Nitish Bhardwaj; Callum E Flowerday; Jaron C Hansen
Journal:  Sensors (Basel)       Date:  2022-03-29       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.