Tereza Cristina Favre1, Lilian Christina Nóbrega Holsback Beck1, Fernando Schemelzer Moraes Bezerra2, Carlos Graeff-Teixeira3,4, Paulo Marcos Zech Coelho5, Martin Johannes Enk6, Naftale Katz5, Ricardo Riccio Oliveira7, Mitermayer Galvão Dos Reis7,8,9, Otávio Sarmento Pieri1.
Abstract
BACKGROUND: The World Health Organization recommends reliable point-of-care (POC) diagnostic testing to eliminate schistosomiasis. Lateral flow immunoassay that detects schistosome circulating cathodic antigen (CCA) in urine to establish prevalence thresholds for intervention in endemic areas is recommended. Stored urine may be useful if surveying at-risk populations is delayed or interrupted by unforeseen circumstances, such as the current COVID-19 pandemic. This study evaluated the manufacturer's claim that Schistosoma mansoni infection can be reliably diagnosed in urine samples stored at -20°C for one year.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35239906 PMCID: PMC8909435 DOI: 10.1590/0037-8682-0389-2021
Source DB: PubMed Journal: Rev Soc Bras Med Trop ISSN: 0037-8682 Impact factor: 1.581
Results of POC-ECO by infection status determined by a robust reference standard (KK+HTX) from freshly-collected urine and urine retested after being stored at -20°C for one year.
| Infection status (KK+HTX) | POC-ECO outcome | Freshly collected urine* | Urine stored at -20°C for one year** | ||
|---|---|---|---|---|---|
| POC-ECO | POC-ECO | POC-ECO/FLT | |||
| Egg-negative | Positive | N (%) | 3 (9.7) | 0 (0.0) | 11 (35.5) |
| 95% CI | 0.8% - 28.6% | - | 15.5% - 57.4% | ||
| Trace | N (%) | 25 (80.6) | 3 (9.7) | 17 (54.8) | |
| 95% CI | 56.6% - 93.5% | 1.1% - 27.5% | 31.1% - 74.9% | ||
| Negative | N (%) | 3 (9.7) | 28 (90,3) | 3 (9.7) | |
| 95% CI | 0.8% - 28.6% | 70.0% - 98.1% | 0.8% - 28.6% | ||
|
|
|
|
|
| |
| Egg-positive | Positive | N (%) | 9 (20.0) | 3 (6.7) | 16 (35.6) |
| 95% CI | 7.4% - 37.1% | 0.6% - 20.5% | 18.7% - 53.7% | ||
| Trace | N (%) | 23 (51.1) | 8 (17.8) | 21 (46.7) | |
| 95% CI | 31.9% - 68.5% | 6.0% - 34.6% | 28.0% - 64.4% | ||
| Negative | N (%) | 13 (28.9) | 34 (75.6) | 8 (17.8) | |
| 95% CI | 13.6% - 46.8% | 56.0% - 88.4% | 6.0% - 34.6% | ||
|
|
|
|
|
| |
| All | Positive | N (%) | 12 (15.8) | 3 (3.9) | 27 (35.5) |
| 95% CI | 6.9% - 27.9% | 0.3% - 12.5% | 22.4% - 49.4% | ||
| Trace | N (%) | 48 (63.2) | 11 (14.5) | 38 (50.0) | |
| 95% CI | 48.2% - 75.5% | 6.0% - 26.3% | 35.4% - 63.6% | ||
| Negative | N (%) | 16 (21.1) | 62 (81.6) | 11 (14.5) | |
| 95% CI | 10.7% - 33.9% | 67.8% - 90.5% | 6.0% - 26.3% | ||
|
|
|
|
|
| |
POC-ECO: point-of-care circulating cathodic antigen manufactured by Eco Diagnóstica Ltd; KK+HTX: combined Kato-Katz (one stool sample, two slides) and Helmintex® methods *tested with batch 180907091; **tested with batch 201907001; POC-ECO/FLT: filtration-concentration method. N: number of cases; CI: confidence interval.
FIGURE 1:Performance of Kato-Katz, POC-ECO (batch 180907091) with freshly-collected urine, and both POC-ECO and POC-ECO/FLT (batch 201907001) with urine storePerformance of Kato-Katz, POC-ECO (batch 180907091) with freshly collected urine, and both POC-ECO and POC-ECO/FLT (batch 201907001) with urine stored at -20°C for one year, taking as reference the joint outcome of Kato-Katz and Helmintex® (KK+HTX). Percentage (%) values for sensitivity, specificity, accuracy and 95% Confidence Intervals (CI) are depicted on vertical lines. POC-ECO and POC-ECO/FLT results were calculated considering traces as positive (t+) and as negative (t-).