Abstract
The pulsed-injection method for measuring the velocity of blood flow in intraarterial digital subtraction angiography is described. With this technique, contrast material is injected at a pulsing frequency as high as 15 Hz, so that two or more boluses can be imaged simultaneously. The velocity of flow is determined by measuring the spacing between the boluses and multiplying it by the pulsing frequency. Results of tests with phantoms correlate well with flow measurements obtained with a graduated cylinder for velocities ranging from 8 to 60 cm/sec. The potential of the method for time-dependent velocity measurement has been demonstrated with simulated pulsatile flows.Mesh:
Year: 1986 PMID: 3523598 DOI: 10.1148/radiology.160.2.3523598