Physiological pacing with a DF-1 single chamber defibrillator in a patient with permanent atrial fibrillation and heart block: A case report.

Mathieu Nasarre1, Nicolas Clementy1, Alexandre Bodin1, Dominique Babuty1, Arnaud Bisson1,2.   

Abstract

We report a case report of a 68-year-old man with chemotherapy-induced cardiomyopathy and uncontrolled permanent atrial fibrillation. Cardiac resynchronization therapy implantation and atrioventricular junction ablation were planned. DF-1 single chamber defibrillator was connected to lead's defibrillation and a lead destinated to left bundle branch area pacing. This system leads to reduce costs by one-third, improve battery longevity, and provide a more physiological pacing.
© 2022 Wiley Periodicals LLC.

Entities:  

Keywords:  DF-1 single chamber defibrillator; left bundle branch area pacing; resynchronization