Fernando Ocasio-Villa1, Luisa Morales-Torres2, Norma Velez-Medina1, Luis A Cubano1, Juan C Orengo2, Edu B Suarez Martinez1,3,4.
Abstract
Breast cancer is the leading cause of sex-specific female cancer deaths in the United States. Detection at earlier stages contributes to decreasing the mortality rate. The mammogram is the "Gold Standard" for breast cancer screening with an estimated sensitivity of 86.9% and a specificity of 88.9%. However, these values are negatively affected by the breast density considered a risk factor for developing breast cancer. Herein, we validate the novel LED-based medical device Pink Luminous Breast (PLB) by comparison with the mammogram using a double blinded approach. The PLB works by emitting a LED red light with a harmless spectrum of 640-800 nanometers. This allows the observation of abnormalities represented by dark or shadow areas. In this study, we evaluated the sensitivity and specificity of the PLB device as a screening tool for the early detection of breast abnormalities. Our results show that the PLB device has a high sensitivity (89.6%) and specificity (96.4%) for detecting breast abnormalities comparable to the adjusted mammogram values: 86.3 and 68.9%, respectively. The percentage of presence of breast density was 78.2% using PLB vs. 72.9% with the mammogram. Even with higher findings of breast density, the PLB is still capable of detecting 9.4% of calcifications compared to 6.2% in mammogram results and the reported findings for cysts, masses, or tumor-like abnormalities was higher using the PLB (6.5%) than the mammogram (5.6%). A 100% of the participants felt comfortable using the device without feeling pain or discomfort during the examination with 100% acceptability. The PLB positive validation shows its potential for routine breast screening at non-clinical settings. The PLB provides a rapid, non-invasive, portable, and easy-to-use tool for breast screening that can complement the home-based breast self-examination technique or the clinical breast examination. In addition, the PLB can be conveniently used for screening breasts with surgical implants. PLB provides an accessible and painless breast cancer screening tool. The PLB use is not intended to replace the mammogram for breast screening but rather to use it as an adjunct or complemental tool as part of more efficient earlier detection strategies contributing to decrease mortality rates.Entities:
Keywords: LED-based; breast; early-screening; imaging-based diagnoses; mammogram
Year: 2022 PMID: 35223883 PMCID: PMC8868042 DOI: 10.3389/fmed.2021.805182
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Figure 1Sample size of the recruited participants and resulting individual events. PLB, Pink Luminous Breast.
Figure 2Summary of pink luminous breast instructions.
Evaluation of PLB test performance with an imperfect reference standard: Mammogram.
|
|
|
| ||
|---|---|---|---|---|
|
|
|
|
| |
| Sensitivity (%) | 86.3 | 81.7–90.9 | 89.6 | 84.8–94.3 |
| Specificity (%) | 68.9 | 59.6–78.2 | 96.4 | 81.3–115.9 |
| Validity index (%) | 80.9 | 76.6–85.2 | 91.3 | 85.7–96.7 |
| Predictive value + (%) | 86.0 | 81.3–90.6 | 98.8 | 92.8–104.4 |
| Predictive value – (%) | 69.5 | 60.2–78.8 | 74.4 | 62.7–86.1 |
| Prevalence (%) | 68.8 | 63.8–73.9 | 76.2 | 69.6–82.8 |
| Youden index | 0.6 | 0.5–0.7 | 0.9 | 0.7–1.1 |
| Likelihood ratio + | 2.8 | 2.1–3.7 | 24.9 | −145.1–162.1 |
| Likelihood ratio – | 0.2 | 0.1–0.3 | 0.11 | 0.06–0.17 |
| Pre-test probability | 54.7 | |||
| Post-test probability | 98.2 | |||
| Kappa | 0.56 | |||
Imperfect reference test (mammogram: sensitivity 86.9%, specificity 88.9%).
Mammogram findings reports compared to PLB.
|
|
| |
|---|---|---|
|
|
|
|
| Dense breast | 248 (72.9) | 266 (78.2) |
| Cyst, mass or tumor-like | 19 (5.6) | 22 (6.5) |
| Calcifications | 21 (6.2) | 32 (9.4) |
PLB, Pink Luminous Breast.
Responses for acceptability of breast cancer screening tools and PLB (n = 169).
|
| |
|---|---|
| Age | |
| Mean (SD) | 53.3 (9.9) |
| Range | 40–77 |
| I feel comfortable taking screening tests? | |
| Yes | 164 (97.0) |
| Is the PLB device easy to use and to visualize my breasts? | |
| Yes | 169 (100.0) |
| The discomfort felt by the PLB device was: | |
| None | 169 (100.0) |
| Reasons to avoid a mammogram test | |
| I don't like it | 16 (9.5) |
| Afraid might hurt | 119 (70.4) |
| Anxiety | 34 (20.1) |
Represents one (1) missing value; PLB, Pink Luminous Breast.