Stuart et al.[1] method of estimating gas consumption in an Intensive Care Ventilator
| Gas consumption = (minute ventilation+bias flow) × [(FiO2-0.2)/0.8] + cycling requirement. |
| A more accurate approximation can be obtained using: |
| Gas consumption = [minute volume + (bias flow × [Ti/(Ti+Te)] × [(FiO2-0.2)/0.8] + cycling requirement. |
(Ti-Inspiratory time; Te-Expiratory time; FiO2-Inspiratory oxygen concentration)
Oxygen usage by Non-invasive ventilation (NIV)/Invasive ventilation in different ventilators
| VENTILATOR | MAKE | MODEL | NIV– Oxygenutilisation | Invasive ventilation -Oxygen utilisation |
|---|---|---|---|---|
| Mindray | Shenzhen Mindray Bio-Medical Electronics Co., Limited | Mindray SV 300 | 45 L/min | 14.5 L/min |
| Drager | Dragerwerk AG & Co. | Drager Savina | 47 L/min | 15.5 L/min |
| Altima | Medion Healthcare Private Limited | Altima | 33 L/min | 11.5 L/min |
| Maquet | Maquet Critical Care AB | SERVO s Base unit | 49 L/min | 10 L/min |
| Vela | Care fusion | Vela 16532-10 | 57.3 L/min | 15 L/min |
| Schiller | TECME S. A | Graphnet Ts | 49.6 L/min | 14 L/min |
| Orion G | Air Liquide Medical Systems Private Limited | Orion G | 45 L/min | 12 L/min |
| Bharat Electronics Ltd (BEL) | Bharat Electronics Limited | PM-CARES CV200 | 57.3 L/min | 15.5 L/min |
This table is based on flow rates under experimental conditions. Please note that flow rates may vary considerably depending on patient factors, oxygen supply factors and device/ventilator modes, most notably bias flow settings. Please consult manufacturer specifications when making any estimations as newer data may be available