The effect of positive end expiratory pressure on rebreathing and gas dilution in the Ayre's T-piece system--a laboratory study.

T L Dobbinson, E R Fawcett, D P Bolton.   

Abstract

The effect of positive end expiratory pressure on the function of the Ayre's T-Piece System was studied using a simple lung model. Positive end expiratory pressure did not effect end tial CO2 during "controlled breathing" but caused an increase during "spontaneous breathing" when fresh gas flow was less than 3 times the minute volume. Gas dilution did not occur under any of the test conditions.

Entities:  

Mesh:

Substances:

Year:  1978        PMID: 352189     DOI: 10.1177/0310057X7800600102

Source DB:  PubMed          Journal:  Anaesth Intensive Care        ISSN: 0310-057X            Impact factor:   1.669


  1 in total

Review 1.  Anaesthesia breathing circuits.

Authors:  J W McIntyre
Journal:  Can Anaesth Soc J       Date:  1986-01