Raphaëlle Lautraite1, Lise Bernard1, Pascale Halle2, Philip Chennell1, Yoann Le Basle1, Justyna Kanold2,3, Valérie Sautou1.
Abstract
BACKGROUND: The treatment of relapsed or refractory leukemia remains a major problem. Among the new therapeutic approaches, the use of modified T lymphocytes, called chimeric antigen receptor T cells (CAR-T cells), seems promising. The first step of their preparation is leukapheresis, which involves the collection of mononuclear cells from the patient. This medical procedure requires numerous medical devices (MDs) made of plasticized polyvinylchloride (PVC). These compounds can leach out of the devices during contact with the patient's blood. The aim of our study was to evaluate the migration of the plasticizers contained in the MD during a simulated pre-CAR-T cell leukapheresis procedure, and to measure the patient's and their lymphocytes' exposure to them.Entities:
Keywords: CAR-T cells; di(2-ethylhexyl) phthalate; exposure; leukapheresis; medical devices; plasticizer
Year: 2022 PMID: 35202265 PMCID: PMC8875078 DOI: 10.3390/toxics10020079
Source DB: PubMed Journal: Toxics ISSN: 2305-6304
Characteristics of drugs and medical devices used in the study.
| Name | Supplier | Batch Number | Shelf-Life | Use |
|---|---|---|---|---|
| ACD-A | Fresenius Kabi | FC19H21L | 2021/01 | Anticoagulation |
| 0.9% sodium chloride | B Braun Medical | 15PDD680 | 03/22 | Priming and restitution |
| Medical devices | ||||
| Spectra Optia® IDL Set (N°10315) | Terumo BCT | 1911273130 | 01/10/2021 | Apheresis device |
Figure 1DEHP contents in the apheresis kit, and details of the sections for its determination. 1: Inlet line; 2: diversion bag; 3: inlet line manifold; 4: inlet saline line clamp (red); 5: anticoagulant (AC) check valve; 6: return line; 7: return line manifold; 8: return saline line clamp (blue); 9: AC line (orange); 10: saline line (green); 11: inlet line trap; 12: inlet pressure sensor diaphragm; 13: centrifuge pressure sensor diaphragm; 14: reservoir; 15: return pressure sensor diaphragm; 16: collect pressure sensor diaphragm; 17: centrifuge loop; 18: channel; 19: connector; 20: collection Bag; 21: collect line; 22: sample bulb assembly; 23: accessory line; 24: manifold to channel.
Figure 2The apheresis model used in the study.
Amounts of plasticizers in the different sections of the apheresis set, and in the anticoagulant bag.
| Location of Cut | Detail Cut | Average Content of Plasticizer (% i.e., g/100 g of PVC) (Mean ± SD) | Relative Standard Deviation (RSD) |
|---|---|---|---|
| Inlet line (1) |
| DEHP 1 = 36.89 ± 2.03 | 5.5% |
|
| DEHP 1 = 34.43 ± 2.81 | 8.2% | |
|
| NA 2 | NA 2 | |
| Inlet line manifold (3) | Orange | DEHP 1 = 39.45 ± 1.94 | 4.9% |
| White | DEHP 1 = 40.55 ± 3.61 | 8.9% | |
| Green | DEHP 1 = 49.55 ± 1.30 | 2.6% | |
| Return line (6) | DEHP 1 = 30.69 ± 1.73 | 5.6% | |
| Manifold to channel (24) | White 1 | DEHP 1 = 29.06 ± 0.53 | 1.8% |
| White 2 | DEHP 1 = 28.69 ± 1.12 | 3.9% | |
| Pink | DEHP 1 = 31.99 ± 1.56 | 4.9% | |
| Yellow | DEHP 1 = 27.45 ± 1.72 | 6.3% | |
| Return line manifold (7) | White | DEHP 1 = 30.56 ±1.66 | 5.4% |
| Green | DEHP 1 = 31.13 ± 0.48 | 1.5% | |
| Channel (18) | NA 3 | NA 3 | |
| Anticoagulant (AC) line (9) | DEHP 1 = 32.45 ± 1.26 | 3.9% | |
| Saline line (10) |
| DEHP 1 = 59.44 ± 5.69 | 9.6% |
|
| DEHP 1 = 34.86 ± 1.56 | 4.5% | |
| Collect line (21) | DEHP 1 = 24.99 ± 0.97 | 3.9% | |
| Collection bag (20) | NA 3 | NA 3 | |
| Manifold | DEHP 1 = 28.34 ± 2.16 | 7.6% | |
| ACD-A Bag | Spike port | DEHP 1 = 45.42 ± 20.17 | 44.4% |
| DEHP 1 = 28.29 ± 1.69 | 6.0% |
1 DEHP: bis(2-ethylhexyl) phthalate; 2 DEHA: bis(2-ethylhexyl) adipate; 3 NA: not applicable.
Concentrations of DEHP released during our apheresis model with water/ethanol simulant.
| Assay Number | CDEHP 1 in Each Sample (µg/mL) | Mean CDEHP 1 ± SD (µg/mL) | RSD | Calculated Dose (mg/kg/day) |
|---|---|---|---|---|
| 1 | 17.16 | 17.1 ± 0.70 | 4.1% | 1.2 |
| 17.82 | ||||
| 16.42 | ||||
| 2 | 19.86 | 19.9 ±0.09 | 0.4% | 1.4 |
| 20 | ||||
| 19.84 | ||||
| 3 | 22.28 | 22.3 ±0.52 | 2.3% | 1.6 |
| 21.8 | ||||
| 22.84 |
1 CDEHP: concentration of bis(2-ethylhexyl) phthalate in the simulant.