| Literature DB >> 35198936 |
G B John Mancini1, Craig Kamimura1, Eunice Yeoh1, Arnold Ryomoto1, C David Mazer2.
Abstract
BACKGROUND: Coronary computed tomography angiography (CCTA) is used to assess plaque characteristics, remodelling, and progression and regression. Few papers address standard operating procedures that ensure achievement of high interobserver reproducibility. Moreover, assessment of coronary artery bypass grafts has not been reported.Entities:
Year: 2021 PMID: 35198936 PMCID: PMC8843959 DOI: 10.1016/j.cjco.2021.09.022
Source DB: PubMed Journal: CJC Open ISSN: 2589-790X
Figure 1Plaque analysis. This figure shows a composite of screen shot outputs of the plaque analysis software. The lumen is depicted in green, low-attenuation plaque in red, mixed plaque in blue, and calcified plaque in yellow. (A) Shows a representative cross-section with volumetric parameters shown below. (B) Shows the linear reconstruction of the proximal right coronary artery from which the indexes are calculated. (C) Shows the location of key reference points for the measurements and the Hounsfield unit distribution plots.
Figure 2Distribution of plaque components. The panel of 4 cross-section images illustrates the varying distribution of plaque component from 4 different coronary segments. The lumen is depicted in green, low-attenuation plaque in red, mixed plaque in blue, and calcified plaque in yellow.
Suggested standard operating procedures for training to optimize reproducibility.
| Vessel probe (This is Vitrea’s single-click tool to automatically define the centerline through the vessel lumen and perform automated detection of the inner lumen and vessel wall | After opening a CCTA case in the Vitrea’s Cardiac Analysis application, the program launches the “auto vessel probe” feature, which creates a centre line through the lumen of each coronary vessel and automatically segments and labels the main coronary arteries. A centre line serves as a reference point from which contours of the lumen and outer vessel wall are automatically defined. If the automatic vessel detection does not provide appropriate lumen and vessel contours, or when analyzing bypass grafts, manual vessel analysis and probing can be performed. With situations in which a significant lesion is present, selection of a vessel location distal to the lesion to initiate the automatic definition of the lumen centre line and associated contours is recommended. |
| Lumen and vessel contours | Review the lumen centerline and edit as necessary. Editing the centre line can improve the edge detection for the lumen and vessel contours. If the automatically provided contours remain suboptimal, reprobe by selecting another point along the vessel. Reprobing is preferred to editing poorly tracked contours, as editing is difficult and time consuming. Manual editing of vessel contours is difficult. The lack of visual cues to indicate the outer edge of the vessel make evaluating and repositioning the vessel contour extremely challenging, even with adjustment of the window and level display settings, as this has minimal effect on improving the ability to resolve the outer edges of the vessel wall. |
| Identifying segment landmarks | Evaluate vessel anatomy with MPR, CPR, and SPR displays with a slice thickness setting of 5 to10 mm to aid the identification of branches. Rotate the SPR view, and confirm the vessel segmentation landmark locations with the MPR and CPR views. (Note: A left button mouse selection on a location on the SPR display will automatically position the cross hairs on the corresponding location on the MPR views. Other software programs provide this ability as well.) |
| Start and end contours of the vessel segment (defining length) | The first and last cross-sections of a vessel segment are positioned to the segment landmarks by adjusting the segment boundary markers on CPR and SPR displays. Ensure that the lumen and vessel contours of the first and last cross- sections are not affected by the segment landmarks such as branches (see |
| Reference cross-section | Using the lumen diameter histogram, identify the cross-section(s) with maximum lumen diameter. Cross-sections with calcium and overt vessel wall thickening should not be selected for the reference. If there is more than 1 cross-section with the same maximum lumen diameter, the cross-section with the largest diameter and largest value for remodelling index can be used to determine the reference diameter. |
CCTA, coronary computed tomography angiography; CPR, curved planar reformatting; MPR, multiplanar reformatting; SPR, straightened planar reformatting.
Figure 3Effect of segment boundary selection. The series of sequential, cross-section images progress from left (1) to right (5) and illustrate an emerging branch at the 2 o’clock position, affecting the lumen and vessel contours. When identifying the last cross-section of a vessel segment, 1 and 2 would be included in the analysis segment but 3, 4, and 5 would not. In this example, the change in axis of the maximum lumen diameter (green line) aided in detecting the influence of the branch on the lumen contour.
Interobserver variability measurements for volumetric measures of vessels and plaques
| Parameter | Measurements | Coronary segments | Graft segments | All segments |
|---|---|---|---|---|
| Vessel volume (mm3) | Mean | 494 to 498 | 1813 to 1842 | 1022 to 1025 |
| Standard error of the mean | 53 to 55 | 247 to 248 | 146 to 146 | |
| Mean of differences | –2.0 to 3.2 | –12.6 to 16.2 | –5.7 to 8.4 | |
| Mean of absolute differences | 7 to 12 | 13 to 23 | 9 to 16 | |
| Pearson correlation | 0.998 to 0.999 | 1.000 to 1.000 | 1.000 to 1.000 | |
| Total plaque (mm3) | Mean | 281 to 282 | 933 to 941 | 542 to 545 |
| standard error of the mean | 35 to 36 | 120 to 122 | 73 to 73 | |
| Mean of differences | –0.8 to 1.1 | –3.2 to 4.7 | –1.7 to 1.7 | |
| Mean of absolute differences | 6 to 11 | 8 to 11 | 7 to 11 | |
| Pearson correlation | 0.996 to 0.999 | 1.000 to 1.000 | 0.999 to 1.000 | |
| Low-attenuation plaque (mm3) | Mean | 47 to 48 | 207 to 210 | 111 to 113 |
| Standard error of the mean | 5 to 6 | 28 to 28 | 17 to 17 | |
| Mean of differences | –0.4 to 0.5 | –1.2 to 2.3 | –0.7 to 1.2 | |
| Mean of absolute differences | 1 to 2 | 2 to 3 | 2 to 2 | |
| Pearson correlation | 0.996 to 0.997 | 1.000 to 1.000 | 1.000 to 1.000 | |
| Mixed plaque (mm3) | Mean | 195 to 196 | 731 to 736 | 409 to 411 |
| Standard error of the mean | 24 to 24 | 98 to 100 | 59 to 59 | |
| Mean of differences | –0.9 to 0.5 | –3.3 to 4.4 | –1.0 to 1.2 | |
| Mean of absolute differences | 5 to 9 | 8 to 10 | 6 to 10 | |
| Pearson correlation | 0.994 to 0.999 | 1.000 to 1.000 | 0.999 to 1.000 | |
| Calcified plaque (mm3) | Mean | 37 to 39 | 1 to 2 | 23 to 24 |
| Standard error of the mean | 10 to 12 | 1 to 1 | 7 to 7 | |
| Mean of differences | –0.9 to 0.8 | –0.3 to 0.3 | –0.5 to 0.4 | |
| Mean of absolute differences | 2 to 3 | 0 to 0 | 1 to 2 | |
| Pearson correlation | 0.995 to 0.999 | 0.996 to 0.998 | 0.996 to 0.999 |
Interobserver variability measurements for parameters derived from volumetric measurements of vessels and plaques
| Parameter | Measurements | Coronary segments | Graft segments | All segments |
|---|---|---|---|---|
| Total plaque % | Mean | 55 to 55 | 53 to 54 | 54 to 55 |
| Standard error of the mean | 2 to 3 | 1 to 2 | 2 to 2 | |
| Mean of differences | –0.5 to 0.3 | –0.1 to 0.2 | –0.4 to 0.3 | |
| Mean of absolute differences | 1 to 2 | 0 to 0 | 1 to 1 | |
| Pearson correlation | 0.978 to 0.993 | 0.997 to 0.999 | 0.986 to 0.995 | |
| Low-attenuation plaque % | Mean | 18 to 18 | 22 to 23 | 18 to 20 |
| Standard error of the mean | 1 to 1 | 1 to 1 | 1 to 1 | |
| Mean of differences | –0.1 to 0.2 | –0.1 to 0.1 | –0.1 to 0.2 | |
| Mean of absolute differences | 1 to 1 | 0 to 0 | 1 to 1 | |
| Pearson correlation | 0.947 to 0.985 | 0.996 to 0.999 | 0.969 to 0.991 | |
| Mixed plaque % | Mean | 71 to 72 | 77 to 77 | 71 to 74 |
| Standard error of the mean | 2 to 2 | 1 to 1 | 2 to 2 | |
| Mean of differences | –0.1 to 0.2 | –0.1 to 0.1 | –0.1 to 0.1 | |
| Mean of absolute differences | 1 to 1 | 0 to 0 | 0 to 1 | |
| Pearson correlation | 0.989 to 0.998 | 0.996 to 0.999 | 0.991 to 0.998 | |
| Calcified plaque % | Mean | 10 to 11 | 0 to 0 | 6 to 6 |
| Standard error of the mean | 2 to 2 | 0 to 0 | 2 to 2 | |
| Mean of differences | –0.1 to 0.2 | –0.03 to 0.03 | –0.1 to 0.1 | |
| Mean of absolute differences | 0 to 1 | 0 to 0 | 0 to 0 | |
| Pearson correlation | 0.997 to 0.999 | 0.996 to 0.999 | 0.997 to 0.999 |
Figure 4Interobserver variability for measures of plaque components. This illustration summarizes the key interobserver performance characteristics for the laboratory as a whole. On average, the set of test arterial segments and grafts displayed total plaque volume accounting for 55% of the total vessel volume. The lumen volume averaged 45%. With respect to the total plaque volume, mixed plaque was 74%, calcified plaque was 6%, and low-attenuation plaque (LAP) was 20%. The ranges of the key performance measurements are shown in tabular form.