| Literature DB >> 35195801 |
Yanwei Wang1, Simone Rink2, Antje J Baeumner2, Michael Seidel3.
Abstract
Gold nanoparticle-catalyzed chemiluminescence (CL) of luminol is an attractive alternative to strategies relying on enzymes, as their aggregation leads to significantly enhanced CL signals. Consequently, analytes disturbing such aggregation will lead to an easy-to-quantify weakening of the signal. Based on this concept, a homogeneous aptamer-based assay for the detection of sulfadimethoxine (SDM) has been developed as a microfluidic CL flow-injection platform. Here, the efficient mixing of gold nanoparticles, aptamers, and analyte in short channel distances is of utmost importance, and two-dimensional (2D) and three-dimensional (3D) mixer designs made via Xurography were investigated. In the end, since 2D designs could not provide sufficient mixing, a laminated 3D 5-layer microfluidic mixer was developed and optimized with respect to mixing capability and observation by the charge-coupled device (CCD) camera. Furthermore, the performance of standard luminol and its more hydrophilic derivative m-carboxy luminol was studied identifying the hydrophilic derivative to provide tenfold more signal enhancement and reliable results. Finally, the novel detection platform was used for the specific detection of SDM via its aptamer and yielded a stunning dynamic range over 5 orders of magnitude (0.01-1000 ng/ml) and a limit of detection of 4 pg/ml. This new detection concept not only outperforms other methods for SDM detection, but can be suggested as a new flow-injection strategy for aptamer-based rapid and cost-efficient analysis in environmental monitoring and food safety.Entities:
Keywords: 3D mixer; Aptamer; AuNPs; Chemiluminescence; Sulfadimethoxine; m-Carboxy luminol
Mesh:
Substances:
Year: 2022 PMID: 35195801 PMCID: PMC8866360 DOI: 10.1007/s00604-022-05216-6
Source DB: PubMed Journal: Mikrochim Acta ISSN: 0026-3672 Impact factor: 5.833
Comparison of different methods for SDM detection
| Method | Materials | LODs (ng/ml) | Linear ranges (ng/ml) | Advantages | Disadvantages | Reference |
|---|---|---|---|---|---|---|
| HPLC | Sodium dodecylbenzene sulfonate enhanced micro solid-phase extraction | 0.59 | 1–200 | High sensitive | Complex operation, expensive instrument | [ |
| Magnetic mixed hemimicelles solid-phase extraction | 0.033 | - | [ | |||
| Colorimetric | AuNPs, aptamer | 50 | 50–1000 | Rapid, visible | Insufficient sensitivity | [ |
| Graphene/nickel@palladium, aptamer | 0.7 | 1–500 | [ | |||
| AuNPs, aptamer | 10 | 10–106 | [ | |||
| Fluorescence | Magnetized upconversion nanoparticles,aptamer | 0.11 | 1–9 | High sensitive | Complex labeling operations, narrow linear range | [ |
| Cadmium telluride (CdTe) quantum dots | 2.24 | 25–300 | Rapid, low cost | Relatively low sensitivity | [ | |
| Coordination polymer nanobelt (CPNB), aptamer | 10 | 10–1000 | [ | |||
| Electrochemical | Plastic membrane electrode | 7.5 | - | Rapid | High cost and relatively low sensitivity | [ |
| Molecularly imprinted overoxidized polypyrrole | 2.17 × 104 | 4.65 × 104–1.15 × 107 | [ | |||
| Boron-doped diamond (BDD) | 2 | 10–1.2 × 105 | [ | |||
| Micro-flow injection CL | AuNPs, aptamer | 0.004 | 0.01–103 | High sensitive, low cost, simple operation | This work |
Fig. 1a Scheme of chemiluminescence (CL) measurement system: the reagents were pumped into mixer to generate CL signal which was recorded by a CCD camera (the observation area was marked by red dashed lines); b scheme of 2D mixer structure with 3 layers; c scheme of 3D mixer structure with 5 layers
Fig. 2Schematic illustration of homogeneous chemiluminescence detection of SDM using AuNPs probe
Fig. 3Images of chemiluminescence reaction in a 2D and c 3D mixer. b and d The mixing profile of 2D and 3D mixer which have the same structure as (a) and (c), respectively
Fig. 4a UV − Vis spectra of AuNPs in water, PBS, and various concentrations of aptamer in PBS; b UV − Vis spectra of AuNPs solution in 1 μM aptamer (in PBS) and various concentrations of sulfadimethoxine (in water)
Fig. 5The calibration plots of chemiluminescence intensity versus the logarithm of sulfadimethoxine concentration using AuNPs catalyzed chemiluminescence with m-carboxy luminol in a homogeneous aptamer-based assay in a flow-based 3D mixer chip, n = 3
Recovery results of water sample (n = 3)
| Spiked (ng/ml) | Detected (ng/ml) | Recovery (%) | RSD (%) |
|---|---|---|---|
| 50 | 48.42 ± 0.16 | 96.8 ± 0.31 | 0.32 |