Pollyana Ferreira da Silva1, Bruna Ferreira Gomes1, Carlos Manuel Silva Lobo1, Marcelo Carmo2, Christina Roth3, Luiz Alberto Colnago4. 1. Instituto de Química de São Carlos, Universidade de São Paulo, Av. Trabalhador São-carlense, 400, 13566-590, São Carlos, SP Brazil. 2. Forschungszentrum Jülich GmbH, Wilhelm-Johnen-Straße, 52428 Jülich, Germany. 3. Electrochemical Process Engineering, University of Bayreuth, Universitätsstraße 30, 95447 Bayreuth, Germany. 4. Embrapa Instrumentação, Rua XV de Novembro, 1452, 13560-970 São Carlos, SP Brazil.
Abstract
The in situ coupling between electrochemistry and spectrometric techniques can help in the identification and quantification of the compounds produced and consumed during electrochemical reactions. The combination of electrochemistry with nuclear magnetic resonance is quite attractive in this respect, but it has some challenges to be addressed, namely, the reduction in the quality of the NMR signal when the metallic electrodes are placed close to or in the detection region. Since NMR is not a passive technique, the convective effect of the magnetic force (magnetoelectrolysis), which acts by mixing the solution and increasing the mass transport, has to be considered. In seeking to solve the aforementioned problems, we developed a system of miniaturized electrodes inside a 5 mm NMR tube (outer diameter); the working and counter electrodes were prepared with a mixture of graphite powder and epoxy resin. To investigate the performance of the electrodes, the benzoquinone reduction to hydroquinone and the isopropanol oxidation to acetone were monitored. To monitor the alcohol oxidation reaction, the composite graphite-epoxy electrode (CGEE) surface was modified through platinization. The electrode was efficient for in situ monitoring of the aforementioned reactions, when positioned 1 mm above the detection region of the NMR spectrometer. The magnetoelectrolysis effect acts by stirring the solution and increases the reaction rate of the reduction of benzoquinone, because this reaction is limited by mass transport, while no effect on the reaction rate is observed for the isopropanol oxidation reaction.
The in situ coupling between electrochemistry and spectrometric techniques can help in the identification and quantification of the compounds produced and consumed during electrochemical reactions. The combination of electrochemistry with nuclear magnetic resonance is quite attractive in this respect, but it has some challenges to be addressed, namely, the reduction in the quality of the NMR signal when the metallic electrodes are placed close to or in the detection region. Since NMR is not a passive technique, the convective effect of the magnetic force (magnetoelectrolysis), which acts by mixing the solution and increasing the mass transport, has to be considered. In seeking to solve the aforementioned problems, we developed a system of miniaturized electrodes inside a 5 mm NMR tube (outer diameter); the working and counter electrodes were prepared with a mixture of graphite powder and epoxy resin. To investigate the performance of the electrodes, the benzoquinone reduction to hydroquinone and the isopropanol oxidation to acetone were monitored. To monitor the alcohol oxidation reaction, the composite graphite-epoxy electrode (CGEE) surface was modified through platinization. The electrode was efficient for in situ monitoring of the aforementioned reactions, when positioned 1 mm above the detection region of the NMR spectrometer. The magnetoelectrolysis effect acts by stirring the solution and increases the reaction rate of the reduction of benzoquinone, because this reaction is limited by mass transport, while no effect on the reaction rate is observed for the isopropanol oxidation reaction.
The coupling of electrochemistry with
nuclear magnetic resonance
(NMR) has been attracting the attention of the scientific community
in recent decades due to the possibility of detecting the formation
and consumption of species, in real time, which can be used to study
reaction mechanisms and kinetics. There are a few dozen models of
miniaturized cells that aim to minimize the effects resulting from
interference between the techniques, since the placement of electrodes
close to or in the NMR detection region can degrade the NMR signal
by introducing noise and increasing the width of the peaks.[1−19] Furthermore, when performing the electrochemistry–NMR coupling,
it is important to keep in mind the existence of the magnetoelectrolysis
effect, where the magnetic field force, F,a plays a major role, as it originates
from the interaction between the magnetic field, B, and
the ion flow density, j, as the charged species move
through the solution toward the electrode surface, as shown in eq .[19−27] This convection force, also termed the magnetohydrodynamic force,
stirs the solution with an intensity proportional to the intensity
of B; therefore, the greater the intensity of the magnetic
field, the greater the agitation caused in the system. For systems
that are limited by mass transport, this effect is very evident, as
it significantly increases the reaction rate.[19,28−33] By placing electrodes above the NMR detection region and by taking
advantage of the aforementioned effect, it is possible to greatly
reduce the delay between the consumption/production of species and
the detection of these events. This is because the magnetohydrodynamic
force, through its action, quickly homogenizes the concentration of
species in the solution.To address the current
limitations and challenges inherent to the
coupling of electrochemistry and NMR, a miniaturized system of composite
electrodes was developed. The electrodes were prepared by mixing a
graphite and expoxy (CaldoFix) resin. The choice in materials stems
from the inherent limitation of using metallic electrodes in the detection
region of the NMR spectrometer[6,19] as well as from the
fact that carbon-based electrodes can be used in a wide range of applications.[34−37] Furthermore, composite electrodes have been shown to have good mechanical
stability, are able to withstand a wide range of applied potentials,
and are compatible with nonaqueous solvents.[38−40] In addition,
the electrode’s surface can be renewed simply by polishing
the material, which shows the advantages of using such electrodes
and why their application in electrochemical systems has been investigated
over the years.[38−41]In this work we demonstrate the applicability of carbon–graphite
composite electrodes (CGEEs) as an alternative that aids in reducing
the challenges faced by the electrochemistry–NMR coupling.
Both working and counter electrodes were miniaturized CGEEs, and they
were used to analyze two electrochemical systems: the electro-reduction
of bezoquinone and the electro-oxidation of isopropanol (ISP). Naturally,
simultaneous observations of the systems were made using NMR; i.e.,
the experiments were performed in situ.
Results and Discussion
The use of metallic electrodes for electrochemistry–NMR
coupling causes several interference problems between the techniques,
such as the increase in noise in electrochemical measurements and
the loss of spectral quality in the NMR spectra.[6] Thus, a system of three miniaturized electrodes using mainly
graphite and epoxy resin as raw materials was developed in order to
minimize these issues. The applicability of this electrode system
was evaluated by monitoring the reactions of electro-reduction of
benzoquinone and electro-oxidation of the ISP in real time. In addition,
the magnetoelectrolysis effect was evaluated for both reactions. A
great advantage of this approach is the possibility of replacing graphite
with other carbon-based materials such as commercial catalysts with
different percentages of nanoparticles or even by monitoring different
electrochemical reactions (e.g., in fuel cells).
Benzoquinone Reduction
The study of the interference
in the NMR spectra due to the positioning of the electrodes in the
electrochemical cell relative to the detection region of the spectrometer
was done by considering two positions: 1 mm above the detection region
and 1 cm within it. For this analysis, the peaks of the TSPd4 standard,
the water suppression region, and the peaks corresponding to the analyte,
benzoquinone, and the product, hydroquinone, were monitored. Figure shows the spectra
obtained in both conditions. Due to the fact that WE and CE were made
from a composite of graphite and epoxy resin, it was expected that
the insertion of these electrodes in the detection region would not
significantly compromise the quality of the NMR spectrum. However,
upon analysis of Figure , it was possible to verify that the homogeneity of the magnetic
field was greatly compromised and, consequently, there was a broadening
of all the peaks of the spectrum: the definition of the peak was lost
(Figure A, bottom)
and the full width at half-maximum (fwhm) is much worse when the electrodes
are inserted in the detection region (9.62 Hz) compared to when the
electrodes are placed 1 mm above it (1.86 Hz), the suppression of
the solvent signal was severely impaired (Figure B) and the separation of the peaks referring
to benzoquinone and hydroquinone was lost (Figure C). This could be due to the fact that copper
wires were used to make the electric contact with the electrodes and
that this contact point was placed near the detection region. Another
explanation could be that while graphite is not as magnetically active
as some metals, it still has a much higher magnetic susceptibility
than water, which changes the magnetic field distribution in its vicinity.
With this in mind, the position chosen for electrodes was 1 mm above
the detection region, since the magnetic force promotes homogenization
of the sample due to the convection caused by this force.[19,33] Thus, it was possible to maintain the quality of NMR measurements.
Figure 1
Evaluation
of interference in the 1H NMR spectra of
benzoquinone due to the position of the electrodes. Black line: electrodes
positioned 1 mm above the detection region. Red line: electrodes positioned
1 cm within the detection region. Signal attributed to (A) the standard
TSPd4, (B) water, (C) benzoquinone and hydroquinone. A 0.05 mol·L–1 benzoquinone solution, after a 10 min long reaction,
was used.
Evaluation
of interference in the 1H NMR spectra of
benzoquinone due to the position of the electrodes. Black line: electrodes
positioned 1 mm above the detection region. Red line: electrodes positioned
1 cm within the detection region. Signal attributed to (A) the standard
TSPd4, (B) water, (C) benzoquinone and hydroquinone. A 0.05 mol·L–1 benzoquinone solution, after a 10 min long reaction,
was used.The voltammetric profiles corresponding
to the redox processes
of benzoquinone in the presence and in the absence of the magnetic
field are shown in Figure . The changes observed in these results are mainly due to
the presence of the magnetic field, since the convection generated
by the magnetic force takes more molecules of benzoquinone to the
electrode, increasing the cathodic current, and the formed hydroquinone
molecules quickly diffuse to the bulk of the solution, reducing the
anodic current.[30] In the cathodic branch,
referring to the reduction of benzoquinone, the peak potential changes
from −0.38 V (ex situ) to somewhere around −0.80 V (in
situ), where the peak actually becomes less defined. The cathodic
peak currents were −0.11 mA (ex situ) and −0.44 mA (in
situ). Thus, it was found that the cathodic current was about 4 times
greater when the reaction was carried out in the presence of the magnetic
field.
Figure 2
Cyclic voltamograms acquired, using graphite–epoxy composite
electrodes, in situ (red line) and ex situ (black line).
Cyclic voltamograms acquired, using graphite–epoxy composite
electrodes, in situ (red line) and ex situ (black line).The chronoamperograms obtained during in situ and ex situ
measurements
with a fixed potential of −0.50 V (vs Ag/AgCl) are shown in Figure A). The analysis
of the chronoamperograms reveals differences in the current magnitudes
obtained in the different experimental conditions: during the ex situ
measurement the cathodic current remained close to −0.1 mA,
while in the in situ measurement, this parameter remained between
−0.3 and −0.4 mA, in agreement with the results obtained
by cyclic voltammetry. In addition to changes in the magnitude of
the current, it is possible to observe that the current never stabilized
during the in situ measurements, possibly due to the experimental
configuration, where a turbulent flow may be formed in between the
electrodes due to the relative orientations of the ion flow and the
magnetic field (Figure B).
Figure 3
(A) Chronoamperograms acquired during measurements, using the composite
graphite–epoxy electrodes, in situ (red line) and ex situ (black
line) electrodes. (B) Relative orientation of the magnetic field, B, current density, j, and resulting magnetic
force, FB.
(A) Chronoamperograms acquired during measurements, using the composite
graphite–epoxy electrodes, in situ (red line) and ex situ (black
line) electrodes. (B) Relative orientation of the magnetic field, B, current density, j, and resulting magnetic
force, FB.Figure shows the 1H NMR spectra obtained in situ during WE polarization at a
potential of −0.5 V vs Ag/AgCl, for the electro-reduction of
benzoquinone to hydroquinone. The peak corresponding to benzoquinone
diminishes in amplitude over time, as it is reduced while, conversely,
the peak of hydroquinone increases in amplitude over time, as it is
being formed. In addition, there is also a slight widening of the
signal referring to the 1H of benzoquinone during the in
situ monitoring. However, the separation between the peaks was not
compromised, which made it possible to quantify both species by their
respective peak areas (Figure ).
Figure 4
1H spectrum of in situ reduction of benzoquinone, performed
for 1 h.
Figure 5
Percentage of area over time, of the in situ
reduction of benzoquinone,
of each 1H peak of benzoquinone and of hydroquinone for
1 h.
1H spectrum of in situ reduction of benzoquinone, performed
for 1 h.Percentage of area over time, of the in situ
reduction of benzoquinone,
of each 1H peak of benzoquinone and of hydroquinone for
1 h.Figure shows the
ratio between the benzoquinone/hydroquinone peak area and the total
spectrum area, RS (eq ), during both in situ and ex situ experiments.
The points acquired in situ are represented by the filled squares
with benzoquinone in black and hydroquinone in blue, and those acquired
ex situ are represented by the white-filled squares. The data show
a consumption of 26.7% of the initial benzoquinone and the formation
of the same amount of hydroquinone when the reaction was carried out
inside the 14 T NMR spectrometer. However, when this reaction was
carried out outside the spectrometer, under the same concentration
and temperature conditions, the amount of benzoquinone consumed corresponded
to only 5.6% of the initial concentration. That is, the data obtained
by high resolution NMR spectroscopy confirm the electrochemical data,
showing that the reaction happens more quickly in the presence of
the magnetic field of the spectrometer (approximately 4 times faster).The set of results obtained from the in situ monitoring of the
reduction of benzoquinone using the composite graphite–epoxy
electrodes demonstrates their applicability during the electrochemistry–NMR
coupling. Despite the small electroactive area of these electrodes,
it was possible to convert around 25% of the benzoquinone into hydroquinone.
However, the quality of the NMR spectra was somewhat compromised during
the in situ monitoring of the electrochemical reaction, which was
observed by the enlargement and a small distortion of the peaks referring
to benzoquinone, at 6.85 ppm. This distortion, however, was not enough
to compromise the quantitative measurements, given that both NMR and
electrochemical results are in good agreement.
Isopropanol Oxidation
The electro-oxidation reaction
of isopropanol, ISP, was monitored in situ to demonstrate the versatility
of the graphite–epoxy composite electrodes. This reaction was
chosen to be monitored in real time as it has great importance in
the field of fuel cells,[42] in addition
to its main reaction product, acetone, being easily identifiable by 1H NMR, as it only displays a corresponding signal at 2.2 ppm.
For this study, however, it was necessary to modify the working electrode
to facilitate the reaction. The WE was modified through the electrodeposition
of Pt and the voltammetric profile confirming the presence of Pt on
the surface of the WE is shown by the gray line inFigure , where the Pt reduction peak
occurs at around +0.12 V (vs Ag/AgCl). This figure also shows the
cyclic voltammograms corresponding to the in situ and ex situ electrooxidation
of ISP, red and black lines, respectively. The anodic scan of this
voltamogram has two bands: the first one, whose maximum current is
obtained at +0.45 V (versus Ag/AgCl), appears due to the dehydrogenation
of the ISP, while the second band at +0.90 V corresponds to the bulk
oxidation of 2-propanol.[43] In addition,
the cathodic scan also has a band at +0.1 V vs Ag/AgCl due to the
reactivation current of the catalyst (Pt black).
Figure 6
Cyclic voltamograms acquired
using the platinized composite graphite–epoxy
electrode in situ (red) and ex situ (black) for the ISP electro-oxidation.
The cycles shown are number 10 out of 10 acquired.
Cyclic voltamograms acquired
using the platinized composite graphite–epoxy
electrode in situ (red) and ex situ (black) for the ISP electro-oxidation.
The cycles shown are number 10 out of 10 acquired.In the case of the ISP electro-oxidation reaction, the voltammetric
results obtained in situ and ex situ were very similar. A possible
explanation for this observation is that the maximum oxidation current
of alcohols is normally dependent on the formation of Pt oxides and
not on the alcohol diffusion.[44] Thus, even
with the effect of magnetoelectrolysis contributing to an increase
in mass transport, it was not possible to observe a significant change
in the voltammetric profile obtained in situ. In view of the similarity
between the cyclic voltammograms acquired in situ and ex situ, real-time 1H NMR measurements for the electro-oxidation of the ISP were
only performed in situ.Figure shows the
chronoamperogram corresponding to the oxidation of the ISP in the
presence of the magnetic field. Due to the adsorption of organic molecules
on the electrode surface, the catalytically active sites are blocked
causing a decrease in the oxidation current. For this reason, it was
necessary to periodically clean the electrode surface by applying
a potential of −0.5 V (vs Ag/AgCl) to reactivate the electrode
surface and prevent the equilibrium current from being achieved so
quickly. The applied potential was chosen because this experimental
condition favors the formation of acetone.
Figure 7
In situ chronoamperogram
of the ISP oxidation using the electrochemical
cell made with graphite and epoxy resin electrodes. The applied potential
was +0.45 V vs Ag/AgCl, while the cleaning potential was −0.5
V vs Ag/AgCl. Both were applied sequentially for 5 s.
In situ chronoamperogram
of the ISP oxidation using the electrochemical
cell made with graphite and epoxy resin electrodes. The applied potential
was +0.45 V vs Ag/AgCl, while the cleaning potential was −0.5
V vs Ag/AgCl. Both were applied sequentially for 5 s.Measured current density values (considering a geometric
area of
7.9 × 10–3 cm2, j+0.45V = 76 mA cm–2) are in the same
range as literature values obtained for the isopropanol electroxidation
reaction.[45] The electrochemically active
surface area (ECSA) was calculated using the hydrogen underpotential
region (Hupd, integrated from −0.4
to −0.15 V) and was found to be 110 cm2.The
influence of the platinum graphite–epoxy composite electrode’s
placement on the in situ monitoring of the ISP electro-oxidation was
verified (Figure ).
Just as in the case of benzoquinone reduction, the insertion of the
electrodes in the detection region also compromised the NMR spectral
quality. Thus, taking advantage of the agitation generated by the
presence of the electric current and the magnetic field, the electrodes
were kept just 1 mm above the detection region, since in these conditions
the NMR signals were very well-defined and the standard signal had
a FWHM of 0.66 Hz (Figure a).
Figure 8
Evaluation of interference in the 1H NMR spectra of
isopropanol due to the position of the electrodes. The electrodes
were positioned 1 mm above (black lines) and 1 cm within (red lines)
the detection region. Signal referring to (A) the standard TSPd4,
(B) ISP doublet, and (C) ISP multiplet.
Evaluation of interference in the 1H NMR spectra of
isopropanol due to the position of the electrodes. The electrodes
were positioned 1 mm above (black lines) and 1 cm within (red lines)
the detection region. Signal referring to (A) the standard TSPd4,
(B) ISP doublet, and (C) ISP multiplet.The resonance spectra obtained before and after the oxidation of
ISP as well as the correspondence of the observed 1H NMR
signals are shown in Figure . The peaks Ha and Hb are characteristic
of ISP. Ha are the hydrogen atoms of the methyl groups,
while Hb is the hydrogen bonded to the central carbon atom.
Ha produce a doublet signal in the NMR spectrum, with an
area 6 times larger than the multiplet signal of Hb, just
as expected by their ratio within the ISP molecule. Hc,
meanwhile, are the hydrogen atoms of acetone, which, due to their
equivalent chemical environment, produce a singlet in the NMR spectrum.
Figure 9
NMR spectra
obtained in situ using platinum–graphite epoxy
composite electrodes, before the reaction (black line) and after 120
min (red line) of ISP oxidation. The peaks for each 1H
nucleus are assigned in both spectra.
NMR spectra
obtained in situ using platinum–graphite epoxy
composite electrodes, before the reaction (black line) and after 120
min (red line) of ISP oxidation. The peaks for each 1H
nucleus are assigned in both spectra.Figure shows
the regions of the 1H NMR spectra corresponding to the
peaks Ha (Figure a) and Hc (Figure b) acquired during the in situ oxidation of ISP. It
is immediately clear from the figures that the spectral quality did
not deteriorate over time, given that the FWHM of the peaks remains
constant and no unfolding of the peaks was observed. As seen in Figure , the RS of Ha decreases over time, owing to the consumption
of ISP, and conversely, the RS of Hc increases over time, due to the formation of acetone. More
concretely, it was determined that only 1.6% of the ISP had been consumed
after 2 h while the concentration of acetone increased by the same
amount. The peak corresponding to Hb was not used for the
determination of ISP concentration due to its proximity to the solvent
signal (H2O + D2O) which, while not shown in Figure , is still close
enough to influence the area of Hb. In addition, the peaks
Ha and Hc correspond to the same number of hydrogen
atoms and thus have a 1:1 ratio, which facilitated the comparison.
Figure 10
1H spectra of in situ ISP oxidation over a 120 min long
reaction using the composite graphite–epoxy platinized electrodes.
ISP concentration was 0.5 mol·L–1 in HClO4 (0.1 mol·L–1). (A) ISP consumption
and (B) formation of acetone.
Figure 11
Graph
of the percentage of area with time for in situ ISP oxidation.
The areas correspond to the peaks referring to the hydrogens Ha (black line) and Hc (blue line). The former corresponding
to the ISP doublet and the latter to the acetone singlet, during 120
min reaction.
1H spectra of in situ ISP oxidation over a 120 min long
reaction using the composite graphite–epoxy platinized electrodes.
ISP concentration was 0.5 mol·L–1 in HClO4 (0.1 mol·L–1). (A) ISP consumption
and (B) formation of acetone.Graph
of the percentage of area with time for in situ ISP oxidation.
The areas correspond to the peaks referring to the hydrogens Ha (black line) and Hc (blue line). The former corresponding
to the ISP doublet and the latter to the acetone singlet, during 120
min reaction.The data shown demonstrate that
graphite–epoxy composite
electrodes are adequate for monitoring the electro-reduction of ISP
in real time if the electrodes are placed 1 mm above the NMR detection
region. However, due to the small surface area of the WE, the amount
of ISP consumed and, consequently, the amount of acetone formed was
very small, even after 2 h of reaction.
Comparison with Other Setups
Table shows a
summary of the main characteristics
of some setups used for in situ coupling between NMR and electrochemistry.
More information on current developments in the area of coupling between
electrochemistry and liquid NMR can be found in the recent review
published in 2021 by Pietrzak et al.[46]
Table 1
Comparison of Miniaturized Electrodes
for EC–NMR Coupling
carbon fiber electrode[2,19]
metallic wire electrode[5,19]
graphite–epoxi electrode
costs
low
low, metal
can be reused
low, disposable electrodes
modification of the electrode surface
It
is possible, but the attachment of other materials to the
fiber is difficult.
Different metals can be used (e.g.,
Pt, Pd, Au, etc.).
Different materials can be used (e.g.,
carbon supported catalysts,
etc.).
application for different analytes
Although electrochemically resistant, it can be applied to
a limited range of analytes when not modified.
broad
broad
NMR probe modifications required?
no
no
no
large surface area
yes, but difficult to quantify the
geometric area
yes, and it can be improved by increasing
the superficial roughness
of the material
It depends on the support material used
and the surface modification.
can be inserted
in detection region?
yes
no
no
Conclusions
The versatility of the developed cell was demonstrated with the
monitoring of the electro-reduction of benzoquinone and the electro-oxidation
of isopropanol both in situ and ex situ. It was possible to keep track
of the consumption of the reagents and their respective products in
real time. However, the percentage of ISP consumed in situ, 1.6%,
was significantly lower than the percentage of benzoquinone consumed
in situ, 26.7%, even with the platinum modification of the electrode
for the alcohol oxidation.The highlights of the developed graphite–epoxy-based
electrodes
are the simplicity of their fabrication and their low cost, with the
added benefit that both the support material and the catalyst can
be modified to suit the desired application. However, there are still
some limitations regarding the positioning of the electrode relative
to the detection region of the NMR spectrometer and the electrochemical
current generated and supported by this system. Regarding the restriction
of placing these electrodes inside the NMR detection region, the magnetoelectolysis
effect can be taken advantage of to increase the homogeneization rate
of the sample solution during the in situ reaction, thus reducing
the time between product formation and its detection. However, this
effect must also be kept in mind, evaluated, and considered during
the comparisons between in situ and ex situ experimental results.
As a final note, this type of electrode is recommended for researchers
who are starting their work with coupling electrochemistry and NMR
and who want to build their own electrodes.
Experimental Section
Chemicals
and Solutions
Benzoquine Electro-reduction
p-Benzoquinone
(98% purity) and D2O were acquired from Sigma-Aldrich,
and H2SO4 (95% purity) was acquired from Vetec.
Prior to electrochemical experiments, p-benzoquinone
was disolved in hot water (T = 70 °C) and stirred
for 30 min, after which the solution was allowed to cool down to 25
°C and filtered to obtain pure recrystallized p-benzoquinone (this was confirmed by 1NMR). The analyte
solution, containing 0.05 mol·L–1 of p-benzoquinone, was prepared in ultrapure water (18.2 MΩ·cm)
and the pH was adjusted to 1 using H2SO4, which
corresponds to a concentration of 0.1 mol·L–1 (supporting electrolyte).
Isopropanol Electro-oxidation
Isopropanol (98% purity),
perchloric acid (70% purity), chloroplatinic acid (37.5% purity),
lead acetate (95% purity),and deuterated water, D2O, were
all obtained from Sigma-Aldrich. The supporting electrolyte was a
0.1 mol·L–1 HClO4 solution . The
analyte solution used in the electrochemical experiments was prepared
in the supporting electrolyte solution and contained 0.5 mol·L–1 isopropanol.For both the benzoquinone and
isopropanol reactions, the total volume of the solution used was 660
μL, with 60 μL of D2O being diluted in 600
μL of the analyte solution.To platinize the working electrode,
a solution containing 0.02
mol·L–1 chloroplatinic acid and 3.2 ×
10–5 mol·L–1 lead acetate
was used.
Electrochemical Cell and Measurement Devices
Apparatus
Electrochemical measurements were performed
with an EmStat2 potentiostat (Utrecht, The Netherlands). 1H NMR experiments were carried out in a 600 MHz NMR spectrometer
(Ascend 600 Bruker).
Glass Capillaries
Commercial glass
capillaries with
a volume of 100 μL, of 5.9 and 12.5 cm in height for the reference
and for the WE and CE, respectively, from Blaubrand were used to make
the electrodes.
Electrochemical Cell
The cell developed
to monitor
the electro-reduction reaction of benzoquinone and the electro-oxidation
of the ISP consisted of a system of three electrodes in a 5 mm NMR
tube, as shown in Figure .
Figure 12
Electrochemical cell illustration. The cell was built in an NMR
tube with a 5 mm outer diameter. The cell is not to scale. WE, CE,
and RE refer to the working, counter, and reference electrodes, respectively.
Electrochemical cell illustration. The cell was built in an NMR
tube with a 5 mm outer diameter. The cell is not to scale. WE, CE,
and RE refer to the working, counter, and reference electrodes, respectively.The working electrode (WE) and the counter electrode
(CE) were
prepared with a mixture of graphite and epoxy resin (Caldofix) in
a 3:1 mass ratio, respectively, mixed under mechanical agitation until
the mixture was completely homogeneous (mixing lasted around 5 min).
This paste was compacted inside a commercial 100 μL glass capillary
with a height of 12.5 cm. The electrical contact with the graphite
paste was made before the resin was cured, which requires at least
24 h, with a 0.5 mm thick copper wire inserted at the opposite end
of the capillary (see Figure ). The reference electrode (RE) consisted of a miniaturized
Ag/AgCl electrode. This electrode was built in a 100 μL capillary
with a height of 5.9 cm. A Pt wire was fixed at one end of the glass
capillary by melting the glass and it was filled with a KCl-saturated
aqueous solution (between half and one centimeter of the Pt wire were
left both inside and outside the capillary). An AgCl film was electrodeposited
on a properly cleaned 0.1 mm thick Ag wire which was then inserted
in the solution-filled capillary. Finally, the Ag wire was fixed at
the other end of the capillary using a cianoacrylate resin (superglue).
No modification of the WE surface was needed to perfom the benzoquinone
reduction reaction; however, the electrode surface was platinized
to monitor the isopropanol oxidation reaction. To platinize the graphite–epoxy
electrode surface, a potential of −1 V vs Ag/AgCl was applied
for 10 min with the electrode being placed in a chloroplatinic acid
solution.
Electrical Connections
The electrochemical
cell was
connected to the potentiostat via a triple cable with a protective
copper mesh. Halfway between the cell and the potentiostat, chokes
(inductors with a 200 μH inductance) were attached on each cable
(i.e., there was one choke for each electrode). These were necessary
to filter the noise signals that are introduced during the electrochemical
measurements in the NMR spectra and vice versa.
Measurement
Parameters
Electrochemical Measurements
Electrochemical measurements
were performed using the cyclic voltammetry and chronoamperometry
techniques. The parameters used for both techniques are described
in Table . All electrodes
used were rinsed with ultrapure water and preconditioned using 30
cyclic voltammetry cycles, in the respective supporting electrolyte.
Table 2
Electrochemical Parameters Used for
the Benzoquinone Electro-reduction and Isopropanol Electro-oxidation
parameter
benzoquinone electro-reduction
isopropanol electro-oxidation
Cyclic Voltammetry
potential range (V vs Ag/AgCl)
–0.8 to
+0.8
–0.5 to +1.5
scan
rate (mV s–1)
10
100
potential step
(mV)
1.0
10
Chronoamperometry
potential (V vs Ag/AgCl)
–0.5
+0.45 for 5 s and −0.5
for 5 s
duration
(s)
3600
3600
The electrodes were positioned using the Bruker
sampler, as illustrated
in Figure . More
details on how to position the electrodes inside the NMR tube and
how the height in relation to the detection region was measured can
be found in the tutorial video of the Supporting Information of ref (19).
Figure 13
Adjustment of the height
of the electrodes in relation to the NMR
detection region using a standard Bruker accessory. (A) Electrode
positioned 1 mm above the detection region. (B) Electrode positioned
1 cm within the detection region.
Adjustment of the height
of the electrodes in relation to the NMR
detection region using a standard Bruker accessory. (A) Electrode
positioned 1 mm above the detection region. (B) Electrode positioned
1 cm within the detection region.
NMR Parameters and Sequences
Before the NMR measurements,
the “shimming” was performed, which consists of adjusting
the homogeneity of the magnetic field until the full width at half-maximum,
FWHM, of the chosen standard was close to 1 Hz. The standard, TSPd4
(3-(trimethylsilyl)propionic-2,2,3,3-d4 acid sodium salt with a purity of 98%, acquired from Sigma-Aldrich),
was mixed with the analyte solution, and the shimming was performed
with the electrochemical cell in the NMR spectrometer. In addition,
the 90° pulse was calibrated for each sample. All measurements
were performed at a temperature of 25 °C.
Benzoquinone
NMR Sequence Parameters
The zgpr sequence
(Bruker standard) was used to monitor the BQ reduction reaction. The
d1 recycling time was 5.00 s, the acquisition time (AQ) was 2.73 s,
the number of dummy scans (DS) was 4, and the number of scans (NS)
was 8. With these parameters the total acquisition time for each spectrum
was 93 s.
Isopropanol NMR Sequence Parameters
The zg sequence
(Bruker standard) was used to monitor the electro-oxidation of the
ISP. The d1 recycling time was 5.00 s, the acquisition time (AQ) was
3.00 s, the number of dummy scans (DS) was 4, and the number of scans
(NS) was 16. With these parameters the total acquisition time for
each spectrum was 160 s.The percentage of consumption of benzoquinone,
and that of hydroquinone production was calculated as the ratio between
the respective species’ peak and the total NMR spectrum area, RS, as shown in eq .where AS is the
area of the peak belonging to species S, and Areagent and Aproduct are the peak
areas of reagent and product, respectively. In the case of benzoquinone
electro-reduction, the reagent is benzoquinone and hydroquinone is
the product, while for the electro-oxidation of isopropanol, isopropanol
is the reagent and the product is acetone. All measurements were repeated
a minimum of three times each, with the electrodes maintaining a stable
response during the one and a half hours that the experiments lasted.
The electrodes were built to be disposable, so that a new and clean
surface could be used in each experiment, thus avoiding any cross
contamination.All the experiments, in situ and ex situ, were
carried out in the
same spectrometer. The term in situ refers to the electrochemical
reactions conducted inside the NMR spectrometer, while the ex situ
measurements (control) were conducted outside the spectrometer, and
the NMR spectra were acquired before and after the electrochemical
reactions.
Authors: Sergey A Krachkovskiy; J David Bazak; Peter Werhun; Bruce J Balcom; Ion C Halalay; Gillian R Goward Journal: J Am Chem Soc Date: 2016-06-16 Impact factor: 15.419