Dimensional changes of commercial and novel polyvinyl siloxane impression materials following sodium hypochlorite disinfection.

Shahab Ud Din1, Muhammad Sajid2, Asfia Saeed2, Farooq Ahmad Chaudhary1, Mohammad Khursheed Alam3, Juneda Sarfraz1, Bilal Ahmed1, Mangala Patel4.   

Abstract

BACKGROUND: Dental impressions are used to record anatomy of teeth and surrounding oral structures. Impression materials become contaminated with saliva and blood requiring disinfection, which may have negative impact on dimensional stability of materials.
OBJECTIVE: Comparatively evaluate linear dimensional changes of synthesized Tetra-functional (dimethylsilyl) orthosilicate (TFDMOS) containing Polyvinylsiloxane (PVS) impressions following sodium hypochlorite disinfection.
METHODS: Percentage dimensional changes of three commercial PVS (Elite HD Monophase, Extrude and Aquasil Ultra Monophase) and five experimental PVS impression materials were measured. Experimental material contained novel cross-linking agent (TFDMOS) and a non-ionic surfactant (Rhodasurf CET-2) that is Exp-A (without TFDMOS), Exp-B (with TFDMOS), Exp-C (TFDMOS+ 2% Rhodasurf CET-2), Exp-D (TFDMOS+ 2.5% Rhodasurf CET-2) Exp-E (TFDMOS+ 3% Rhodasurf CET-2). Samples were made using rectangular stainless-steel molds (40 × 10 × 3 mm3) and linear dimensional changes were measured using a calibrated travelling microscope at 10× magnification after immersion in distilled water (D.W) and 1% Sodium Hypochlorite solution at two different time intervals i.e., 30 min and 24 h.
RESULTS: Samples immersed in 1% NaOCl showed significant (p < 0.05) dimensional changes after 30 min of immersion. Exp-E showed significantly greater dimensional changes than their control (Exp-A and Exp-B). In distilled water, there were no significant difference among the tested materials. Aquasil exhibited highest expansion (0.06%) in both solutions. At 24 h, among the commercial materials, Extrude had the greatest expansion followed by Aquasil and Elite in DW while Aquasil showed the greatest expansion followed by Extrude and Elite in NaOCl.
CONCLUSION: Experimental PVS had linear dimensional changes within the ISO 4823; 2015 recommended range. However, extended immersion can negatively affect the linear dimensions.
© 2022 Ud Din et al.

Entities:  

Keywords:  Dimensional changes; Disinfection; Impression materials; Sodium hypochlorite; Vinyl polysiloxane

Year:  2022        PMID: 35174015      PMCID: PMC8802710          DOI: 10.7717/peerj.12812

Source DB:  PubMed          Journal:  PeerJ        ISSN: 2167-8359            Impact factor:   2.984


Introduction

Dental impressions are used to accurately record and reproduce the shape, relationship of teeth with surrounding oral structures (Kumari & Nandeeshwar, 2015). However, during impression making, the material becomes contaminated with saliva and often with blood which may pose health hazard to the dental team (Ud Din et al., 2018). Different methods of disinfection, including spraying and immersion of impression material are commonly employed (Khan & Mushtaq, 2018). However, the American Dental Association (ADA) recommends the immersion technique because it allows direct contact of disinfectant solutions with all surfaces of the impression (Samra & Bhide, 2018). Disinfection through immersion reduces the risk of cross infection, yet alteration in dimensions of impression materials and negative impact on quality of the resulting cast has often been reported (Guiraldo et al., 2012; Martin, Martin & Jedynakiewicz, 2007). This might result in dimensional changes in dental cast prosthesis and ultimately affect the fit of the final restoration (Asopa et al., 2020). Iwasaki et al. (2016) reported marked changes in the dimensions of alginate impressions due to water sorption after prolonged immersion in a disinfectant solution (Iwasaki et al., 2016). Similarly, Babiker, Khalifa & Alhajj (2018) reported significant dimensional changes of gypsum casts obtained from irreversible hydrocolloids after disinfection by NaOCl immersion (Babiker et al., 2018). The recommended products for disinfection of the impression materials are sodium hypochlorite, chlorhexidine, glutaraldehyde and iodine agents (Correia-Sousa et al., 2013). Among numerous disinfectants available, sodium hypochlorite has been recommended by the ADA and Environmental Protection Agency (EPA) for the disinfection of impressions due to its low cost and broad spectrum antimicrobial activity (Correia-Sousa et al., 2013; Khan & Mushtaq, 2018). A study by Al-Enazi and Naik on the efficacy of 1% sodium hypochlorite and 2% glutaraldehyde disinfectant sprays on impression materials (alginate and addition silicone rubber-based impression material) established that 1% sodium hypochlorite yielded better results (Al Enazi & Naik, 2016). Polyvinylsiloxane is popular due to its ability to precisely record tissue details, excellent physical properties and easy manipulation, and is preferred for making final impressions of fixed dental prosthesis (Azevedo et al., 2019). Nevertheless, these materials have lower tear strength and percentage elongation compared to polysulphide. Additionally, the hydrophobic nature of the material demands a dry field of operation to accurately record fine tissue details (Ud Din et al., 2018). Different researchers have incorporated surfactant to the material composition to synthesize hydrophilic addition silicones, but limited literature is available on methods to improve the tear strength of the material (Lee et al., 2004). Formerly, we have carried out synthesis of novel TFDMOS containing polyvinyl siloxane with a non-ionic surfactant (Rhodasurf CET-2) to improve the tear strength and hydrophilicity of conventional addition silicones. It was observed that experimental formulation had significantly higher tear strength compared to commercial material and that tear strength was directly proportional to the concentration of non-ionic surfactant (Din et al., 2018). Considering the variation in material properties corresponding to the composition of the material, the effect of non-ionic surfactant on dimensional stability of materials especially after disinfection needed further exploration. It was hypothesized that the experimental formulations have higher water sorption due to the presence of non-ionic surfactant, resulting in noticeable dimensional changes. Extensive use of polyvinylsiloxane in dentistry and varieties of formulation available necessitated a comprehensive comparison of material particularly in terms of dimensional changes following long term disinfection. Therefore, the objective of this study was two-fold including formulation of experimental polyvinylsiloxane containing TFDMOS and evaluation of the influence of storage media that is distilled water and 1% sodium hypochlorite on the linear dimensional changes over the period of 24 h, in commercial and synthesized PVS impression materials.

Materials and Methods

In this in-vitro experimental study, three commercially available medium bodied PVS materials, Aquasil (Ultra, Dentsply, PA, USA), Elite HD Monophase (Zhermack, Badia Polesine, Italy) and Extrude (Kerr, CA, USA) were selected. In addition, five experimental PVS materials were prepared as base and catalyst paste by weighing all components using a four beam balance (Mettler, Toledo Ltd, Model AG204, UK) (Ud Din et al., 2018). The composition of experimental materials is illustrated in Table 1. To prevent premature polymerisation, equal amount of the prepared material was weighed with a four beam balance and was packed into separate compartments of an auto-mixing cartridge and stored at 4 °C. Exp-A served as a control for Exp-B, while Exp-B was used as a control for Exp-C, D and E. Rhodasurf CET-2 (non-ionic surfactant) was incorporated in the base paste of Exp-B at 2, 2.5 and 3% to formulate Exp C, D and E respectively.
Table 1

Composition of Experimental Polyvinyl siloxane impression materials.

ComponentsBase paste (Wt %)
Exp-AExp-BExp-CExp-DExp-E
Vinyl-terminated dimethylpolysiloxane (Mw 62700)39.9039.9037.9537.4636.98
Polymethylhydrosiloxane (Mw 2270)1.100.770.740.730.72
Tetra-functional (dimethylsilyl) orthosilicate (Mw 329)0.330.320.310.31
Components Catalyst paste (Wt %)
Exp-A Exp-B Exp-C Exp-D Exp-E
Vinyl-terminated dimethylpolysiloxane (Mw 62700)40.7240.7239.5139.5139.51
Platinum (0.05 M)0.060.061.271.271.27
Palladium (<1 µm)0.230.230.220.220.22
Rhodasurf CET-22.002.503.00
A total of 80 samples (n = 10 for each material) were prepared by introducing the material from automixing syringe into the preformed stainless steel mold (40 × 10 × 2 mm3) sandwiched between two metal plates lined by acetate paper. The assembly was placed under a hydraulic press (MESTRA MOD-030350; Talleres Mestraitua, Bizkaia, Spain). Commercial material was allowed to polymerize for the time specified by the manufacturers, while experimental formulation were polymerized for 10 min (Ud Din et al., 2018; Din et al., 2018). After polymerization, the material was removed from the mold and the specimens were divided into two test groups, each consisting of five samples from each material. Group 1 was immersed in 1% NaOCl and Group 2 in distilled water solution for 24 h at 23 ± 1 °C in an oven (Qualicool incubators; LTE Scientific Ltd, Oldham, UK). At predetermined time periods, which were 10 and 30 min (replicate at-office disinfection), and 60 min and 24 h (simulating the transportation time for the impression to reach the dental laboratory) each sample was removed from the respective liquid and blot dried. Linear dimensional changes were recorded using a calibrated travelling microscope (Chesterman, Sheffield, England) by measuring the distance between the fixed edge and free end of the sample (movable pin) at 10× magnification. Data were presented as mean and standard deviation using SPSS Version 22 (Armonk NY IBM Corp, Armonk, NY, USA). Analysis of variance with post hoc Tukey’s test was performed to statistically compare all PVS materials at different time points and p value of 0.05 was considered as significant.

Results

Figures 1 and 2 show percentage linear dimensional changes of impression materials in DW and 1% NaOCl after 24 h immersion. Aquasil exhibited the highest expansion (0.06%) in both solutions while Exp-A exhibited the least expansion (0.01%) in DW. For Group 1, a significant difference (p ≤ 0.05) in dimensions were observed after 30 min of immersion. Inter-group analysis revealed statistically similar dimensional changes among all tested materials, except for Exp-E (Table 2). The surfactant modified Exp-E showed significantly (p ≤ 0.05) greater dimensional changes than their control groups (Exp A & Exp B). Among commercial materials, although no significant changes in dimensions were observed, Aquasil showed the highest expansion followed by Extrude and Elite in 1% NaOCl solution.
Figure 1

Mean linear dimensional changes with Standard Deviation of commercilal and experimental PVS impression materials over the period of 24 h following immersion in Distal water.

Figure 2

Mean linear dimensional changes with Standard Deviation of commercilal and experimental PVS impression materials over the period of 24 h following immersion in 1% NaOCl.

Table 2

Comparison of linear dimensional changes of impression materials stored in 1% NaOCl disinfectant solution at 30 min.

MaterailsMeanS.DF (p-value)Post –hoc Tukey testp-value
Aquasil0.0580.0293.657(0.005)Exp E vs Aquasil0.634
Extrude0.0460.027Exp E vs Extrude0.252
Elite0.0210.028Exp E vs Elite0.012
Exp A0.0100.023Exp E vs Exp A0.003
Exp B0.0310.028Exp E vs Exp B0.047
Exp C0.0410.023Exp E vs Exp C0.159
Exp D0.0320.029Exp E vs Exp D0.053
Exp E0.0400.023
There were no significant differences among the commercial and experimental materials for Group 2 immersed in distilled water. Extrude had the highest expansion followed by Aquasil and Elite at 24 h. Exp B showed slightly higher, but statistically similar expansion to Exp-A, while Exp-C, D and E exhibited higher expansion in each immersion solution compared to Exp-B (control) at each interval. Also, expansion of experimental materials was directly related to the concentration of the surfactant.

Discussion

Impression making is an important aspect of prostheses fabrication. In the oral cavity, these materials come in contact with saliva and blood (Khan & Mushtaq, 2018). This demands disinfection of materials to prevent cross-infection in dental clinics and hospital. However, it is important that the impressions remain dimensionally stable after disinfection. The results of the present study revealed that immersion of addition silicones impression materials in distilled water (Group 2) had no significant impact on the dimensional stability compared to disinfection of impression with 1% NaOCl (Group 1) for 24 h which resulted in expansion of the tested materials. However, the dimensional changes observed were not clinically relevant as the values were within the permitted range of ≤0.5% as recommended by ISO 4823 (2000); 2015 (ISO) and ADA specification 19 (Association ANSiAD, 2007); thus, indicating high precision of the experimental material. The limited amount of linear expansion could be attributed to the isotropic expansion of the samples as they adhere to the PTFE trough of the travelling microscope. These findings were supported by Carvalhal et al. (2011) who investigated the dimensional changes of polysulphide, polyether, addition and condensation silicone after immersion disinfection with 0.5% NaOCl and 2% glutaraldehyde. They observed insignificant changes in the linear dimensions of the materials and recommended that all synthetic elastomers can be safely disinfected by immersion in 0.5% NaOCl and 2% glutaraldehyde (Carvalhal et al., 2011). In a similar study, Samra & Bhide (2018) observed clinically acceptable dimensional changes for alginate and PVS after disinfection with ultraviolet rays and immersion disinfection with sodium hypochlorite, but changes in cross-arch space and inter-abutment distance on gypsum casts were noted for materials disinfected by glutaraldehyde immersion (Samra & Bhide, 2018). Furthermore, Exp-E had significantly greater dimensional changes after immersion disinfection with NaOCl. This was in line with a number of studies that indicated hydrophilic silicones have greater tendency to absorb water and expand (Silva & Salvador, 2004). Results also indicated that longer immersion periods can affect the materials. These findings were similar to those noted by Johnson (1986), who reported long-term dimensional stability of PVS to be unsatisfactory (Thouati et al., 1996). Similarly, Kumari & Nandeeshwar (2015) observed significant difference between non-disinfected and disinfected specimens of PVS and polyether after immersion of 16 h in 0.525% sodium hypochlorite (Kumari & Nandeeshwar, 2015). On contrary, Nassar et al. (2017) found no statistically significant difference between dimensions of vinyl polyether silicone at the time of fabrication and after storage for 1 and 2 weeks in 2.5% buffered solution of glutaraldehyde and reported that dimensional changes within the material obeyed ANSI/ADA standards. The limitation of this study was that the accuracy of the impression material was determined by studying the stability of the impression itself and that the setting expansion of the gypsum cast was not taken into account. However, the results confirmed the high precision of the experimental materials, suggesting routine use of these materials for impression of fixed and removable partial dentures, complete dentures, precision attachments and implants. However, it is suggested to evaluate the effects of other disinfectant systems.

Conclusions

All three commercial and five experimental polyvinylsiloxane impressions materials exhibited linear expansion within the recommended range outlined by ANSI and ISO 4823 (2000); 2015 following 24 h of immersion, though extended immersion time amplified the dimensional changes within all the groups. Therefore, following disinfection, the silicone impression materials must be poured within 24 h.

Raw data.

The raw data collected during the experimental work in the laboratory for all commercial and experimental materials used in this study including individual files for all materials separately and a combined file after analysis the data along with the graphic representation.arrahcs/EvoloPy-FS. Click here for additional data file.
  14 in total

1.  Dimensional change of elastomeric materials after immersion in disinfectant solutions for different times.

Authors:  Cintia Iara Oda Carvalhal; Jose Antonio Nunes de Mello; Lourenco Correr Sobrinho; Americo Bertolazzo Correr; Mario Alexandre Coelho Sinhoreti
Journal:  J Contemp Dent Pract       Date:  2011-07-01

2.  The dimensional stability of dental impression materials following immersion in disinfecting solutions.

Authors:  N Martin; M V Martin; N M Jedynakiewicz
Journal:  Dent Mater       Date:  2007-03-13       Impact factor: 5.304

3.  Effect of the disinfection technique on the linear dimensional stability of dental impression materials.

Authors:  Sávio Marcelo Leite Moreira da Silva; Milton Carlos Gonçalves Salvador
Journal:  J Appl Oral Sci       Date:  2004-09       Impact factor: 2.698

4.  Surface detail reproduction and dimensional accuracy of stone models: influence of disinfectant solutions and alginate impression materials.

Authors:  Ricardo Danil Guiraldo; Thaís Teixeira Borsato; Sandrine Bittencourt Berger; Murilo Baena Lopes; Alcides Gonini; Mário Alexandre Coelho Sinhoreti
Journal:  Braz Dent J       Date:  2012

5.  Accuracy of addition silicones as a function of technique.

Authors:  G H Johnson; R G Craig
Journal:  J Prosthet Dent       Date:  1986-02       Impact factor: 3.426

6.  Effects of immersion disinfection of agar-alginate combined impressions on the surface properties of stone casts.

Authors:  Yukiko Iwasaki; Hisako Hiraguchi; Eriko Iwasaki; Takayuki Yoneyama
Journal:  Dent Mater J       Date:  2016       Impact factor: 2.102

7.  The dimensional accuracy of polyvinyl siloxane impression materials using two different impression techniques: An in vitro study.

Authors:  Nirmala Kumari; D B Nandeeshwar
Journal:  J Indian Prosthodont Soc       Date:  2015 Jul-Sep

8.  The effect of prolonged storage and disinfection on the dimensional stability of 5 vinyl polyether silicone impression materials.

Authors:  Usama Nassar; Carlos Flores-Mir; Giseon Heo; Ysidora Torrealba
Journal:  J Adv Prosthodont       Date:  2017-06-19       Impact factor: 1.904

9.  A simple and effective method for addition silicone impression disinfection.

Authors:  Maria João Azevedo; Inês Correia; Ana Portela; Benedita Sampaio-Maia
Journal:  J Adv Prosthodont       Date:  2019-06-26       Impact factor: 1.904

10.  Effect of heat sterilization and chemical method of sterilization on the polyvinyl siloxane impression material. A comparative study.

Authors:  Swati Joshi Asopa; U Narendra Padiyar; Sumit Verma; Prerna Suri; Nagaveni S Somayaji; Indu Cherangapadath Radhakrishnan
Journal:  J Family Med Prim Care       Date:  2020-03-26
View more
  1 in total

1.  Physico-Mechanical Properties of Commercially Available Tissue Conditioner Modified with Synthesized Chitosan Oligosaccharide.

Authors:  Asfia Saeed; Shahreen Zahid; Muhammad Sajid; Shahab Ud Din; Mohammad Khursheed Alam; Farooq Ahmad Chaudhary; Muhammad Kaleem; Haytham Jamil Alswairki; Huda Abutayyem
Journal:  Polymers (Basel)       Date:  2022-03-18       Impact factor: 4.329

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.