Literature DB >> 35171140

Intraoperative adjustment of implantable collamer lens vault by lens rotation aided by intraoperative OCT.

Roberto Zaldivar1, Roger Zaldivar, Pablo Adamek, Alejandro Cerviño.   

Abstract

PURPOSE: To assess the use of intraoperative optical coherence tomography (OCT) to detect high vault during implantation surgery and guide intraoperative vault adjustment by implantable collamer lens (ICL) rotation.
SETTING: Instituto Zaldivar SA. Mendoza, Argentina.
DESIGN: Prospective single-arm observational study.
METHODS: 25 eyes of 25 consecutive patients showing high intraoperative vault who had undergone ICL rotation from horizontal to vertical or oblique position were evaluated. Central ICL vault was measured both intraoperatively using the microscope-integrated iOCT mounted on a standard surgical microscope (HS Hi-R NEO 900A NIR) and postoperatively using the Casia 2 swept-source OCT at 4 hours, 1 day, and 1 month.
RESULTS: 25 eyes of 25 consecutive patients showing high intraoperative vault who had undergone ICL rotation from horizontal to vertical or oblique position were evaluated. The mean vault value obtained intraoperatively was 1147.88 ± 188.36 μm and changed significantly to 739.76 ± 194.97 μm after lens rotation to either vertical (n = 19, 76% of cases) or oblique (n = 6, 24% of cases) positions (mean difference 408.12 ± 213.57 μm, P < .001). The amount of change due to lens rotation was significantly correlated with white-to-white distance ( r = -0.480, P = .015) and vault before rotation ( r = -0.564, P = .003). The mean vault values were 758.40 ± 187.10 μm, 729.73 ± 227.86 μm, and 661.88 ± 275.17 μm at 4 hours, 24 hours, and 1 month postsurgery, respectively.
CONCLUSIONS: Intraoperative adjustment of ICL vault by lens rotation using intraoperative OCT was an effective procedure to obtain an optimal vault.
Copyright © 2022 Published by Wolters Kluwer on behalf of ASCRS and ESCRS.

Entities:  

Mesh:

Year:  2022        PMID: 35171140     DOI: 10.1097/j.jcrs.0000000000000915

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.528


  2 in total

1.  Comparison of Vault Measurements Using a Swept-Source OCT-Based Optical Biometer and Anterior Segment OCT.

Authors:  Yali Du; Chuang Jin; Shengjie Yin; Geng Wang; Qian Ma; Yuancun Li; Binyao Chen; Hongxi Wang; Kunliang Qiu; Mingzhi Zhang
Journal:  Front Med (Lausanne)       Date:  2022-06-23

2.  Clinical Prediction of Inadequate Vault in Eyes With Thick Lens After Implantable Collamer Lens Implantation Using Iris Morphology.

Authors:  Zhikun Yang; Lihui Meng; Xinyu Zhao; Youxin Chen; Yan Luo
Journal:  Front Med (Lausanne)       Date:  2022-06-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.