Abstract

OBJECTIVE: To compare the retentions of different designs of cobalt-chromium (Co-Cr), pure titanium (CP Ti), and titanium alloy (Ti-6Al-4V) removable partial denture (RPD) circumferential clasps manufactured by selective laser melting (SLM) and to analyze the stress distribution of these clasps during the removal from abutment teeth.
METHODS: Clasps with clasp arm size A (1.9 mm width/1.1 mm thickness at the body and 0.8-taper) or B (1.2 times A) and 0.25 mm or 0.50 mm undercut engagement were modeled on a prepared first premolar die, named as designs A1, A2, A3, and A4, respectively. The density and elastic modulus of SLM-built Co-Cr, CP Ti, and Ti-6Al-4V were measured and given to different groups of clasps. The density, elastic modulus, and Poisson ' s ratio of enamel were given to the die. The control group was the cast Co-Cr clasp with design A1, to which the density and elastic modulus of cast Co-Cr alloy were given. The Poisson's ratio of all metals was 0.33. The initial 5 N dislodging force was applied, and the maximum displacement of the clasp along the insertion path was computed. The load was reapplied with an increment of 5 N than in the last simulation until the clasp was completely dislodged. The retentive force range of different groups of clasps was obtained. The retentive forces of the SLM-built Co-Cr, CP Ti, and Ti-6Al-4V clasps with equivalent computed retentive force range to the control group were validated through the insertion/removal experiment. The von Mises stress distributions of these three groups of SLM-built clasps under 15 N loads were analyzed.
RESULTS: SLM-built Co-Cr, CP Ti, and Ti-6Al-4V clasps with designs B1 or B2, and Co-Cr clasps with design A2 had higher retentive forces than those of the control group. SLM-built CP Ti and Ti-6Al-4V clasps with design A1 had lower retentive forces than those of the control group. SLM-built Co-Cr clasp with design A1 and SLM-built CP Ti and Ti-6Al-4V clasps with design A2 had equivalent retentive forces to those of the control group. The insertion/removal experiment showed that the measured retentive forces of these three groups of SLM-built clasps were (21.57±5.41) N, (19.75±4.47) N, and (19.32±2.04) N, respectively. No statistically significant measured retentive force difference was found among these three groups of SLM-built clasps (P>0.05). The maximum von Mises stress of these three groups of SLM-built clasps exceeded their responding yield strength except for the Ti-6Al-4V one.
CONCLUSION: SLM-built Co-Cr circumferential clasps had higher retention than CP Ti and Ti-6Al-4V ones with the same clasp arm size and undercut engagement. The retention of SLM-built circumferential clasps could be adjusted by changing the undercut engagement and clasp arm size. If SLM-built circumferential clasps are used in clinical practice, the Ti-6Al-4V clasp with clasp arm size A and 0.50 mm undercut engagement is recommended considering the long-term use of RPD in the patient's mouth.

Entities:  

Keywords:  Circumferential clasp; Design parameters; Finite element analyses; Retention, dental prosthesis; Selective laser melting

Mesh:

Substances:

Year:  2022        PMID: 35165476      PMCID: PMC8860646     

Source DB:  PubMed          Journal:  Beijing Da Xue Xue Bao Yi Xue Ban        ISSN: 1671-167X


  14 in total

1.  Prediction of permanent deformation in cast clasps for denture prostheses using a validated nonlinear finite element model.

Authors:  Ahmad Abdel Aziz Mahmoud; Noriyuki Wakabayashi; Hidekazu Takahashi
Journal:  Dent Mater       Date:  2006-03-20       Impact factor: 5.304

2.  Pre-overloading to extend fatigue life of cast clasps.

Authors:  A Mahmoud
Journal:  J Dent Res       Date:  2007-09       Impact factor: 6.116

3.  Accuracy of removable partial denture framework fabricated by casting with a 3D printed pattern and selective laser sintering.

Authors:  Akinori Tasaka; Takahiro Shimizu; Yoshimitsu Kato; Haruna Okano; Yuki Ida; Shizuo Higuchi; Shuichiro Yamashita
Journal:  J Prosthodont Res       Date:  2019-08-26       Impact factor: 4.642

4.  Fatigue strength of Co-Cr-Mo alloy clasps prepared by selective laser melting.

Authors:  Yuka Kajima; Atsushi Takaichi; Takayuki Nakamoto; Takahiro Kimura; Yoshiaki Yogo; Maki Ashida; Hisashi Doi; Naoyuki Nomura; Hidekazu Takahashi; Takao Hanawa; Noriyuki Wakabayashi
Journal:  J Mech Behav Biomed Mater       Date:  2016-03-04

5.  The effect of build orientation on the microstructure and properties of selective laser melting Ti-6Al-4V for removable partial denture clasps.

Authors:  Wenqiang Xie; Meihua Zheng; Jieqi Wang; Xiaoyu Li
Journal:  J Prosthet Dent       Date:  2019-04-11       Impact factor: 3.426

6.  Fitness accuracy and retentive forces of additive manufactured titanium clasp.

Authors:  Kazuya Takahashi; Mana Torii; Toyoki Nakata; Noboru Kawamura; Hidemasa Shimpo; Chikahiro Ohkubo
Journal:  J Prosthodont Res       Date:  2020-02-13       Impact factor: 4.642

7.  Effect of different materials and undercut on the removal force and stress distribution in circumferential clasps during direct retainer action in removable partial dentures.

Authors:  João Paulo Mendes Tribst; Amanda Maria de Oliveira Dal Piva; Alexandre Luiz Souto Borges; Rodrigo Máximo Araújo; João Maurício Ferraz da Silva; Marco Antonio Bottino; Cornelis Johannes Kleverlaan; Niek de Jager
Journal:  Dent Mater       Date:  2019-11-30       Impact factor: 5.304

8.  An analytical model to design circumferential clasps for laser-sintered removable partial dentures.

Authors:  Ammar A Alsheghri; Omar Alageel; Eric Caron; Ovidiu Ciobanu; Faleh Tamimi; Jun Song
Journal:  Dent Mater       Date:  2018-06-21       Impact factor: 5.304

9.  The effect of environment (Dry and Natural Saliva) on clasp retention: In vitro study - Part I.

Authors:  Bilal Mourshed; Nashwan Mohammed Qaed; Hashem Motahir Al-Shamiri; Nader Alaizari; Saleh Sulaiman Alhamdah; Amal Alfaqih
Journal:  Eur J Dent       Date:  2017 Jul-Sep

10.  A three-dimensional finite element analysis of mechanical function for 4 removable partial denture designs with 3 framework materials: CoCr, Ti-6Al-4V alloy and PEEK.

Authors:  Xin Chen; Bochun Mao; Zhuoli Zhu; Jiayi Yu; Yuqing Lu; Qianqian Zhang; Li Yue; Haiyang Yu
Journal:  Sci Rep       Date:  2019-09-27       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.