| Literature DB >> 35156817 |
Xin Huang1, Hanna Ahola1, Matthew Daly2, Chiara Nitride2,3, En Clare Mills2, Tuula Sontag-Strohm1.
Abstract
Pure oats are generally accepted to be safe for most celiac patients, and consumption of oats provides advantageous dietary fibers. However, oats can be contaminated by gluten proteins from wheat, barley, and/or rye. The analytical challenge lies in the reliability of the quantification method and how to maintain the contamination level under a gluten-free food threshold of 20 mg/kg. In this study, we investigated barley-spiked oat flour samples at four levels using four gluten ELISA kits. The largest recovery variance was with the R5 kit that gave 5-6 times overestimation; the G12 kit cross-reacted with oat proteins and gave 4-5 times overestimation at all spiked levels. The Total Gluten and Morinaga kits gave satisfactory recoveries. Total barley hordeins were isolated and characterized to be used as a common calibrator in all four kits aiming at harmonizing the results and to test the kits' performance. Immunoblotting of total hordein isolate revealed that Total Gluten and Morinaga antibodies provided an overall detection, while R5 and G12 antibodies recognized specific hordein groups leading to a larger difference when wheat and barley were used as the calibrant. Calibration with total hordein isolate corrected the overestimation problem and decreased the variability between the four gluten kits.Entities:
Keywords: G12; R5; calibrator; celiac disease; codex; overestimation; reference material
Mesh:
Substances:
Year: 2022 PMID: 35156817 PMCID: PMC8880376 DOI: 10.1021/acs.jafc.1c07715
Source DB: PubMed Journal: J Agric Food Chem ISSN: 0021-8561 Impact factor: 5.279
Characteristics of Four Gluten ELISA Kits
| R5 | G12 | Total Gluten | Morinaga | |
|---|---|---|---|---|
| validation | AACC 38-50.01 | AACC 38-52.01 | AOAC SMPR | AOAC PTM 011804[ |
| AOAC OMA 2012.01 | AOAC OMA 2014.03 | 2017.021 | FDA 2013 | |
| AOAC PTM 120601[ | ||||
| sample size | 1 g | 1 g | 1 g | 1 g |
| (I) sample to extraction buffer ratio (w/v) | (I) 1:40 | (I) 1:40 | (I) 1:40 | (I) 1:19 |
| (II) further dilution (v/v) | (II) 1:12.5 | (II) 1:12.5 | (II) 1:12.5 | (II) 1:20 |
| extraction conditions | (a) 50
°C, 40 min (patented
cocktail solution) | (a) 50 °C, 40 min (extraction buffer) | (a) 50 °C,
40 min (patented
cocktail solution) | extraction buffer including, 2-ME, SDS AT, overnight |
| (b) AT | (b) AT, 60 min (60% v/v ethanol) | (b) AT, 60 min (60% v/v ethanol) | ||
| antibodies | mAb R5 | mAb G12 | mAb R5; mAb HMW GS; mAb LMW GSs | pAb wheat proteins |
| quantification range | LOD: 0.5 mg/kg gliadin | LOD: 2 mg/kg gluten | LOQ: 5 mg/kg gluten | LOD: 0.31 mg/kg wheat proteins |
| LOQ: 2.5 mg/kg gliadin | LOQ: 4 mg/kg gluten | LOQ: 0.78 mg/kg wheat proteins | ||
| ELISA antibody binding steps: (1) first reaction; (2) washing; (3) second reaction; (4) washing; (5) color reaction | 1:30 min | 1:20 min | 1:20 min | 1:60 min |
| 2 & 4:3 times | 2 & 4:5 times | 2 & 4:3 times | 2 & 4:6 times | |
| 3:30 min | 3:20 min | 3:20 min | 3:30 min | |
| 5:30 min | 5:20 min | 5:10 min | 5:20 min | |
| calibrant | PWG gliadins | vital gluten | total gluten | wheat proteins |
| calibration function | cubic spline | dose–response curve, provided excel sheet | 4-parameter function | 4-parameter curve fit (cubic regression) |
| gluten calculation | gliadin × 2 | as it is | as it is | wheat proteins × 0.85 |
G12 kit instruction sample size 0.25 g, in actual test sample size was 1 g for all tests.
Patented cocktail solution including 2-mercaptoethanol (2-ME), guanidine hydrochloride, phosphate buffered saline, WO 02/092633.
AT, ambient temperature, 20–25 °C
In this study, all ELISA calculations were conducted by agonist response-variable slope (four parameters) by GraphPad Prism 8.0.2.
A mixture of 14 wheat cultivars and extracted based on the Japanese official guideline.[50]
Figure 1Characterization of the total hordein isolate by reverse-phase liquid chromatography (panel A) and SDS-PAGE (panel B). Lane 1, Novex Sharp Prestained protein standard. Lane 2, albumin + globulin extract of barley flour. Lane 3 and dotted line 3, total hordein extract from barley flour with albumin + globulin removed. Lane 4 and black line 4, total hordein isolate. Lane 5, Mark 12 unstained protein standard. The letters indicated the classification of hordeins, D, D-hordein; C, C-hordein; B/γ, B/γ-hordein.
Figure 2Immunoblot of total hordein isolate by an antibody-conjugate from four ELISA kits. Protein load to the gel lane A 2.5 μg and lane B 0.5 μg.
Figure 3Comparison of each kit’s own calibration curve and total hordein calibration curve in four ELISA kits: (A) R5 gliadin vs total hordein ng/mL, (B) G12 gluten vs total hordein ng/mL, (C) Total Gluten total gluten vs total hordein, (D) Morinaga wheat proteins vs total hordein ng/mL. The total hordein was hordein isolate × protein content ng/mL. The curves were matched on a gravimetric basis. The curve was produced using agonist response-variable slope (four parameters) nonlinear fit, and the EC50 value was calculated from the best-fit curve values by GraphPad Prism 8.0.2 (San Diego, CA).
Calibration Results mg/kg Gluten Proteins from Barley-Spiked Oat Samples from Kit Standard Calibration and Total Hordein Calibration
| kit calibration ± SD (mg/kg) | kit calibration recovery (%) | total hordein calibration ± SD (mg/kg) | total hordein calibration recovery (%) | |
|---|---|---|---|---|
| 0 mg/kg | ||||
| R5 | 5.3 ± 0.3 | <LOQ | ||
| G12 | 9.8 ± 0.3 | <LOQ | ||
| Total Gluten | <LOQ | <LOQ | ||
| Morinaga | 3.1 ± 0.1 | 3.9 ± 0.1 | ||
| 4 mg/kg | ||||
| R5 | 24.5 ± 3.8 | 613 | <LOQ | 95 |
| G12 | 17.2 ± 1.0 | 429 | <LOQ | 54 |
| Total Gluten | 6.9 ± 0.7 | 173 | 3.7 ± 0.8 | 92 |
| Morinaga | 3.9 ± 1.0 | 98 | 4.9 ± 1.2 | 123 |
| 16 mg/kg | ||||
| R5 | 80.1 ± 6.8 | 501 | 11.6 ± 0.9 | 72 |
| G12 | 84.9 ± 10.5 | 531 | 10.2 ± 1.0 | 64 |
| Total Gluten | 17.4 ± 1.0 | 109 | 10.2 ± 1.3 | 64 |
| Morinaga | 13.0 ± 1.6 | 82 | 15.7 ± 1.7 | 98 |
| 40 mg/kg | ||||
| R5 | 218.2 ± 25.0 | 545 | 32.8 ± 3.6 | 82 |
| G12 | 210.0 ± 6.9 | 525 | 25.4 ± 0.7 | 63 |
| Total Gluten | 44.3 ± 4.0 | 111 | 30.5 ± 4.7 | 76 |
| Morinaga | 31.1 ± 4.0 | 78 | 37.7 ± 4.4 | 94 |
| 80 mg/kg | ||||
| R5 | 398.2 ± 19.0 | 498 | 57.5 ± 2.5 | 72 |
| G12 | 298.6 ± 26.5 | 373 | 38.1 ± 2.9 | 48 |
| Total Gluten | 85.6 ± 11.7 | 107 | 51.8 ± 6.9 | 65 |
| Morinaga | 58.7 ± 1.7 | 73 | 71.5 ± 1.9 | 89 |
One extraction replicate was
The LOQ of hordein calibration was calculated from the lowest hordein standard concentration.
Figure 4Youden plots of the z-score of two levels (16 mg/kg vs 80 mg/kg) of barley-spiked oats measured with four ELISA kits using the (A) kits’ own calibration and (B) total hordein isolate calibration. The black cross set at z = 0 was the theoretical spiking content, and the black rectangle was set to|z| ≤ 2, which was considered satisfactory in the proficiency assessment (ISO 13528). The blue cross was set at the mean value of all z-scores, and the blue rectangle was set at 2 times standard deviation of all z-scores. Three points from each ELISA kit indicated three individual extraction replicates.