Rutao Wang1,2,3, Hideyuki Kawashima2,4, Chao Gao1,2,3, Fangjun Mou1, Ping Li5, Junjie Zhang6, Jian Yang7, Jianfang Luo8, Darren Mylotte2, William Wijns2,9, Yoshinobu Onuma2, Osama Soliman2, Ling Tao1, Patrick W Serruys2,10.
Abstract
OBJECTIVES: To compare the quantitative angiographic aortic regurgitation (AR) of six self-expanding valves after transcatheter aortic valve replacement (TAVR).Entities:
Keywords: VitaFlow; aortic stenosis; paravalvular regurgitation; self-expanding valve; transcatheter aortic valve replacement; transcatheter heart valve
Year: 2022 PMID: 35146004 PMCID: PMC8821967 DOI: 10.3389/fcvm.2021.747174
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Figure 1The study flow chart.
Post-TAVR AR by videodensitometric assessment in valve in valve cases.
|
|
|
|---|---|
| 0.13 | 0.01 |
| 0.29 | 0.02 |
| 0.23 | 0.03 |
| 0.28 | 0.02 |
| 0.35 | 0.00 |
| 0.15 | 0.07 |
| Unanalyzable (Grade 3 by Sellers assessment) | 0.03 |
| Unanalyzable (Grade 3 by Sellers assessment) | 0.09 |
Figure 2Comparison of the mean LVOT-AR after TAVR among the six THVs. VitaFlow#: the mean LVOT-AR of the intermediate aortograms. VitaFlow*: the mean LVOT-AR of the final aortogram (after the second valve implantation). LVOT-AR, quantitative aortic regurgitation in the left ventricular outflow tract; TAVR, transcatheter aortic valve replacement; THV, transcatheter heart valve.
Figure 3Cumulative percentage of different degrees of post-TAVR AR by videodensitometric assessment. VitaFlow*: the analysis of LVOT-AR using the final aortograms (after the second valve implantation), VitaFlow#: the analysis of LVOT-AR using the intermediate aortograms. AR, aortic valve regurgitation; TAVR, transcatheter aortic valve replacement.
Figure 4Cumulative frequency curves of LVOT-AR after TAVR for VitaFlow (A) and the six THVs (B). The shaded background shows the area above 17% of AR, indicating moderate or severe regurgitation. LVOT-AR, quantitative aortic regurgitation in the left ventricular outflow tract; TAVR, transcatheter aortic valve replacement; THV, transcatheter heart valve. VitaFlow*: the analysis of LVOT-AR using the final aortogram (after the second valve implantation), VitaFlow#: the analysis of LVOT-AR using the intermediate aortograms. The red arrows show the LVOT-AR in the six valve-in-valve cases using the intermediate aortograms.
Figure 5The VitaFlow transcatheter aortic valve system.
Design features of the six THVs and their performance on post TAVR acute AR by videodensitometry analysis.
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|
| VitaFlow | MicroPort, Shanghai, China | 16/18 | 21, 24, 27, 30 | Bovine pericardium tissue valve. SE nitinol frame. | Yes | Not recapturable, not repositionable. Design with high radial force. | 7.3 ± 7.8% (n=107) | 8.6% ( |
| Evolut Pro | Medtronic Inc.,Dublin, Ireland | 16 | 23, 26, 29 | Porcine pericardium tissue valve. SE nitinol frame. | Yes | Recapturable, retrievable, repositionable. Does not require rapid pacing for deployment. 10% smaller in height than CoreValve. | 7.3 ± 6.5% | 5.3% ( |
| Evolut R | Medtronic Inc.,Dublin, Ireland | 14/16 | 23, 26, 29, 34 | Porcine pericardium tissue valve. SE nitinol frame. | No | Recapturable, retrievable, repositionable. Does not require rapid pacing for deployment. 10% smaller in height than CoreValve. | 7.9 ± 7.4% | 8.8% ( |
| Venus-A | Venus Medtech Inc., Hangzhou, China | 19 | 23, 26, 29, 32 | Porcine native aortic leaflets. SE nitinol frame. | No | Not retrievable, not repositionable. Designed with increased radial force at the initial 20 mm of the stent inflow segment. Has 3 positioning marker. | 8.9 ± 10.0% | 11.3% ( |
| Acurate Neo | Boston Scientific, Massachusetts, USA | 18 | Small (21–23), Medium (23–25), Large (25–27) | Porcine native aortic leaflets. SE nitinol frame. | No | Not retrievable, partially repositionable. Does not require fast pacing. Safe two-step, top-down deployment with stable hemodynamics. | 9.6 ± 9.2% | 14.2% ( |
| CoreValve | Medtronic Inc.,Dublin, Ireland | 18/20 | 26, 29, 31 | Porcine pericardium tissue valve. SE nitinol frame. | No | Not recapturable, not retrievable. Repositionable, does not require rapid pacing for deployment. | 13.7 ± 10.7% ( | 30.1% ( |