Validation of an equine fitness tracker: ECG quality and arrhythmia detection.

Fe Ter Woort1, Guillaume Dubois2, Grace Tansley3, Marie Didier2, Lidwien Verdegaal3, Samantha Franklin3, Emmanuelle Van Erck-Westergren1.   

Abstract

BACKGROUND: Cardiac arrhythmias in exercising horses are the focus of much interest, both in terms of what is considered normal and potential associations with poor performance and sudden cardiac death. One barrier to performing large-scale studies is the lack of an easily applicable device, to allow recording of large numbers of high-quality exercising electrocardiograms (ECGs). The EquimetreTM is a new wearable device which records a single lead ECG, amongst other parameters. Validation of such wearable devices is essential before further studies are undertaken.
OBJECTIVES: To evaluate the quality of ECG using the EquimetreTM and compare arrhythmia detection during exercise with the reference TelevetTM system. STUDY
DESIGN: Prospective blinded clinical study.
METHODS: Simultaneous ECGs were recorded with both systems in 49 healthy horses during exercise. High intensity exercise (>40km/h) was performed by 29 racehorses, and lower intensity exercise for the remainder of the racehorses and show jumpers. Tracings were excluded if >10% artefact was present (duration of artefact relative to duration of exercise). For included ECGs, the duration of artefact was recorded and compared. ECGs were evaluated using Kubios premium software. Arrhythmia detection (yes/no) and arrhythmia classification (sinus arrhythmia, narrow complex of similar morphology to the sinus complexes or wide complex with a different morphology) were compared using Cohen's Kappa coefficient.
RESULTS: Nine TelevetTM ECGs and 3 EquimetreTM ECGs were excluded due to artefact >10%. TelevetTM ECGs included significantly more artefact during exercise than EquimetreTM ECGs (5% vs. 0.25% p<0.001). Arrhythmia analysis was performed on 38 horses' paired ECGs. The Kappa coefficient was excellent for arrhythmia detection (K=0.97) and arrhythmia classification (K=0.93). MAIN LIMITATIONS: Relatively low numbers of horses with arrhythmias (n=21) were included. The ECG recordings only provided one lead, making arrhythmia classification challenging in some cases.
CONCLUSIONS: The Equimetre device provides a reliable ECG for arrhythmia detection during exercise. This system may be useful clinically and for future large-scale investigations into the occurrence and significance of exercising arrhythmias. This article is protected by copyright. All rights reserved.

Entities:  

Keywords:  arrhythmia; exercise; horse

Year:  2022        PMID: 35138653     DOI: 10.1111/evj.13565

Source DB:  PubMed          Journal:  Equine Vet J        ISSN: 0425-1644            Impact factor:   2.888