Pan Luo1, Jonathan Harrist2, Giansergio Menduni3, Rabah Mesdour4, Nathan StMichel2, Angelo Sampaolo3. 1. EXPEC Advanced Research Center, Saudi Aramco, Dhahran 31311, Saudi Arabia. 2. Houston Research Center, Aramco Americas, Houston, Texas 77084, United States. 3. Polysense Lab, University and Politecnico of Bari, Bari 70126, Italy. 4. Unconventional Reservoir Engineering Department, Saudi Aramco, Dhahran 31311, Saudi Arabia.
Abstract
Natural gas is sampled and produced throughout the lifespan of a petroleum field. Gas composition and isotope data are critical inputs in the exploration and field development, such as gas show identification, petroleum system analysis, fluid characterization, and production monitoring. On-site gas analysis is usually conducted within a mud gas unit, which is operationally unavailable after drilling. Gas samples need to be taken from the field and shipped back to the laboratory for gas chromatography and isotope-ratio mass spectrometry analyses. Results are usually without sufficient resolution to fully characterize the heterogeneity and dynamics of fluids within the reservoir and the production system. In addition, it often takes a considerable time to obtain the results using the traditional method. A novel QEPAS (quartz-enhanced photoacoustic spectroscopy) sensor system was developed to move gas composition analyses to field for quasi-real-time characterization and monitoring. With respect to previously reported QEPAS prototypes for trace gas detection, the new system realized measuring concentrations of methane (C1), ethane (C2), and propane (C3) in gas phase within the percentage range that is typically encountered in natural gas samples from oil and gas fields. A gas mixing enclosure was used to dilute the natural gas-like mixtures in nitrogen gas (N2) to avoid the saturation of QEPAS signals. An iterative analysis based on multilinear regression of QEPAS spectra was developed to filter out the influence of gas matrix variation from multiple hydrocarbon components. The advance in simultaneous measuring hydrocarbon gases and expanded linearity range of QEPAS, with previously reported detection of H2S, CO2, and gas isotopes (12CO2/13CO2, 13CH4/12CH4), opens a way to use the advanced sensing technology for in situ and real-time gas detection and chemical analysis in the oil industry.
Natural gas is sampled and produced throughout the lifespan of a petroleum field. Gas composition and isotope data are critical inputs in the exploration and field development, such as gas show identification, petroleum system analysis, fluid characterization, and production monitoring. On-site gas analysis is usually conducted within a mud gas unit, which is operationally unavailable after drilling. Gas samples need to be taken from the field and shipped back to the laboratory for gas chromatography and isotope-ratio mass spectrometry analyses. Results are usually without sufficient resolution to fully characterize the heterogeneity and dynamics of fluids within the reservoir and the production system. In addition, it often takes a considerable time to obtain the results using the traditional method. A novel QEPAS (quartz-enhanced photoacoustic spectroscopy) sensor system was developed to move gas composition analyses to field for quasi-real-time characterization and monitoring. With respect to previously reported QEPAS prototypes for trace gas detection, the new system realized measuring concentrations of methane (C1), ethane (C2), and propane (C3) in gas phase within the percentage range that is typically encountered in natural gas samples from oil and gas fields. A gas mixing enclosure was used to dilute the natural gas-like mixtures in nitrogen gas (N2) to avoid the saturation of QEPAS signals. An iterative analysis based on multilinear regression of QEPAS spectra was developed to filter out the influence of gas matrix variation from multiple hydrocarbon components. The advance in simultaneous measuring hydrocarbon gases and expanded linearity range of QEPAS, with previously reported detection of H2S, CO2, and gas isotopes (12CO2/13CO2, 13CH4/12CH4), opens a way to use the advanced sensing technology for in situ and real-time gas detection and chemical analysis in the oil industry.
Gas is ubiquitous in the
subsurface, as free gas in porous space,
adsorbed gas on kerogen/mineral surface, or dissolved gas in water
and oil. Gas is sampled or produced throughout the lifespan of a field,
including mud gas logging, formation and well testing, and production.
Detecting and measuring gas is a basic analysis in the oil industry
that provides gas composition and properties for petroleum system
analysis, formation evaluation, PVT study, reservoir simulation, production
monitoring, safety, and economic evaluation.[1−3]QEPAS
(quartz-enhanced photoacoustic spectroscopy) is a laser-based
optical absorption spectroscopy for gas sensing, relying on quartz
tuning forks (QTFs) as sound-to-current transducers. This technique
was invented at Rice University and reported in 2002.[4] These first prototypes employed standard 32 kHz tuning
forks, normally used for time keeping. The Jet Propulsion laboratory,
NASA, was involved very quickly in the development mainly for early
fire warming.[5,6] NASA still have active R&D
and application of QEPAS sensors in trace gas sensing (e.g., CH4, HCl, NO2, H2CO, SO2, and
CO2) relevant to spacecraft environmental monitoring and
advanced life support (https://isda.jsc.nasa.gov/experiment/exper/9272). QEPAS was proposed to the oil industry in the 2008 Offshore Technology
Conference[7,8] for monitoring hydrogen sulfide (H2S), carbon dioxide (CO2), and methane (C1) inside the
annulus of flexible risers because the presence of the sour gases
can dramatically influence corrosion fatigue levels and may cause
safety issues in operation. Based on our best knowledge and investigation
in the market, the proposal has not been applied in the field and
commercialized in the market. Starting from 2013, the Polysense Laboratory
in the Polytechnic University of Bari developed custom resonators
on purpose for gas sensing applications.[9]In a collaboration between the Polysense Laboratory, Rice
University,
and Saudi Aramco,[10−13] a new generation of a QEPAS spectrometer has been designed and prototyped
in a cake box size as a portable gas analyzer for multiple hydrocarbon
components. The sensor box demonstrated the capability of the QEPAS
technique in measuring concentrations of C1 and ethane (C2) in ppb
sensitivity level and propane (C3) in ppm level by employing a single
interband cascade laser (ICL) emitting at 3.345 μm, and the
potential to detect 12CH4 and 13CH4 isotopologues at ppb level by using a quantum cascade laser
(QCL) operating around 7.730 μm. This paper aims to introduce
QEPAS technology to the oil industry, illustrate the realization of
a QEPAS-based gas analysis and calibration system, and summarize benching
testing results and performance. This paper discusses the advantages
and limits of the new technique and some potential applications in
the laboratory and oil and gas fields.
Gas Sensing
and Chemical Analysis
There are many types of gas sensors
(detectors/monitors/analyzers),
which are mainly categorized into four groups based on the detection
technology: semiconductor, electrochemical, analytical (gas chromatography
(GC), spectrometry), and laser optical absorption sensors.[9,14] In the oil industry, for operation safety, infrared point and catalytic
bead (pellistor) as two major sensors are used for monitoring combustible
(hydrocarbon) gas, and electrochemical sensors are used for H2S detecting and warning. These sensors usually detect single
gas component/species in ambient air, suffering from drift, cross-response
to other gases, and changing humidity levels.[15] Therefore, semiconductor and electrochemical sensors are not suitable
for quantitative and chemical analysis in downhole, flow line, and
laboratory analytics environments.During drilling, exploration,
reservoir characterization, and production
monitoring, fluid (gas) samples are taken from drilling mud, reservoir
formation, wellhead, or separator. The samples are usually injected
into GC, connecting with alternative detectors for chemical composition
analysis. The most common detectors are flame ionization detectors
(FIDs) for hydrocarbons, and a thermal conductivity detector (TCD)
for non-hydrocarbon gases (e.g., N2, CO2, H2S, O2, H2, He, and Ar).[1,16] Recently, a quadrupole mass spectrometer (QMS) was developed for
rapid and direct analysis of C1–C10 hydrocarbons and common
inorganic volatiles in mud gas logging.[17] Isotope ratios (e.g., 13C/12C, D/H) of each
gas component, as geochemical fingerprints, are commonly determined
by GC connected with isotope ratio mass spectroscopy (GC-IRMS).[18] These are quite big and delicate instruments,
requiring stable and strict laboratory conditions and sophisticated
experts for operation and maintenance. Consequently, such high precision
laboratory analyzers (GC/MS/IRMS) are incompatible with a tough and
dynamic downhole environment and well site conditions. Transporting
field fluid samples to a laboratory provides delayed, sparse, and
sometimes unrepresentative data, which do not help in near real-time
and high economic value decisions for drilling, formation/well testing,
and on-site troubleshooting.[20,21] In addition, the sampling,
transporting, sample preparation, and routine laboratory analysis
are expensive operations, which are not a pragmatic workflow to generate
high resolution/time-lapse data for fluid heterogeneity and dynamics
studies.Laser-based optical absorption technology offers non-contact,
fast
response, minimal drift, high specificity, low maintenance requirements,
and continuous monitoring, for gas detection and chemical analysis.[15,22,23] Several techniques, based on
the Lambert–Beer law, measuring the optical absorption at specific
wavelength are developed, including non-dispersive infra-red, spectrophotometry,
tunable laser absorption spectroscopy(TLAS)/tunable diode laser absorption
spectroscopy (TDLAS), cavity ring down spectroscopy (CRDS)/cavity-enhanced
absorption spectroscopy (CEAS)/integrated cavity output spectroscopy
(ICOS), and photoacoustic spectroscopy (PAS).PAS, based on
the photoacoustic effect discovered by Alexander
Graham Bell in 1880, is an indirect optical absorption technique that
does not require an optical detector and has a laser wavelength-independent
responsivity. As shown in Figure , when a modulated laser output is absorbed by a target
gas, the absorbed laser energy at characteristic wavelengths induces
heating and expansion to create a vibration of gas molecules at the
resonant frequency, subsequently resulting in the generation of an
acoustic wave. A microphone transduces the pressure wave into an electrical
signal proportional to the concentration (mol %, vol %) of the target
gas. QEPAS is an improved approach to photoacoustic detection by replacing
the microphone with a piezoelectric QTF as a sharply resonant acoustic
transducer to detect weak photoacoustic excitation, allowing the sensor
to be made in an extremely small size.[9,11]
Figure 1
Schematic diagram
shows the principal of PAS for gas detection
and the use of QTF to enhance the photoacoustic spectroscopy (QEPAS).
Schematic diagram
shows the principal of PAS for gas detection
and the use of QTF to enhance the photoacoustic spectroscopy (QEPAS).Among the main optical sensors, QEPAS has been
proven to be a leading-edge
technique for out-of-laboratory detection for trace gas because of
extremely high sensitivity (down to ppb/ppt), level of compactness,
immunity to environmental noise and its proven reliability, ruggedness,
and in-situ operation.[9,11,14,24,25] In a similar
way to PAS, QEPAS does not require an optical detector; it reaches
high detection sensitivity within the short optical pathlength, and
the sound wave detection by the tuning fork is insensitive to the
laser wavelength used for exciting the optical transition. These factors,
together with the modularity of the whole system, represent the main
advantages with respect to other laser-based techniques, such as TDLAS,
CRDS/CEAS, and other multi-pass based spectroscopy.[9,11,15,26] The core part
of a QEPAS sensor is the tuning fork (QTF), whose dimensions are pretty
comparable with a coin, and the whole sensor could be made very compact,
for example, fitting into a 2-inch internal diameter pipe for downhole
operations.[10,11] The resonance of QTF presented
high frequency stability (with frequency shift of ∼0.035ppm/C[2] in the temperature from −40 to 90 °C),[13] suggesting that QEPAS sensors may operate in
a large temperature range. In addition, the QEPAS technique has proved
that it can detect multiple gas species and their isotopes.[13,27] The performances of the QEPAS spectrometers are also continuously
improving in terms of precision, accuracy, detection limits, and response
time. Furthermore, the spectral range of operation has reached the
THz region, where other concerned non-hydrocarbon components such
as H2S can be more easily discriminated by the absorption
bands of the alkane molecules.[28] Therefore,
QEPAS may overcome the abovementioned disadvantages for conventional
gas sensors and laboratory-based instruments and be applied in oil
and gas fields for in-situ and real-time gas detection and chemical
analysis.
QEPAS Gas Analyzer System
The QEPAS
gas analyzer system consists of five major components
(Figure ): sample
introduction enclosure, sample mixing enclosure, sample drying enclosure,
QEPAS sensor and laser controller enclosure, and PC with QEPAS specific
software.
Figure 2
Enclosures and sample line for the QEPAS gas analyzer system.
Enclosures and sample line for the QEPAS gas analyzer system.The QEPAS spectrometer fits in a 30 cm × 10
cm × 20 cm
box (Figure a), containing
the optical components, an ICL operating in the spectral range 3342–3349
nm, the acoustic detection module (ADM), the gas line, the pressure
meter and a power meter for alignment purposes. The ADM is composed
of a standard 32 kHz QTF equipped with a dual tube resonator system
for sound wave amplification. A PCIe-6363 NI-DAQ provides the modulation
signal to the ICL and acquires the QTF signal. A LabVIEW-based software
was implemented to manage the current driver/temperature controller
(Thorlabs ITC4002QCL) and monitor the values of both the pressure
meter and the power meter. A dedicated sub-routine of the software
acts as a lock-in detector for the QEPAS signal demodulation at different
integration times.
Figure 3
Realization of a QEPAS-based gas analysis system. (a)
QEPAS sensor
box; (b) QEPAS, sample preparation enclosures and PC are assembled
in a rack.
Realization of a QEPAS-based gas analysis system. (a)
QEPAS sensor
box; (b) QEPAS, sample preparation enclosures and PC are assembled
in a rack.The sensing system was complemented
with a gas sample control system,
composed of components for sample drying, quantitative mixing (dilution)
and handling the common gas samplers (cylinder and isotube) used in
the industry. All components were mounted in a rack (Figure b) as a prototype of the gas
analyzer for benching tests.The system was designed to meet
the following design parameters:measuring major hydrocarbon gas components
in typical concentrations in the natural gas field (C1: 70–100%,
C2: 1–10%, and C3: <2%);[1,3]handling common gas cylinders (including
isotube);preparing
gas samples that need to
be diluted with dry nitrogen (N2) and dehumidified to the
level of water vapor less than 300 ppm;calculating and reporting gas concentrations,
ratios of hydrocarbons and sample pressure.
Sample Introduction
The QEPAS prototype
system can be connected to a gas cylinder or an isotube through the
sample introduction enclosure. Any common type of a gas cylinder can
be connected to the QEPAS with a pressure regulator and appropriate
adapters. In order to provide a driving force to fill the sensor chamber,
it is necessary to pressurize the gas sample to 80 psi.
Gas Mixing Enclosure
The gas sample
flows through the gas mixing enclosure that dilutes 1-part of the
sample with 9-parts of N2. Considering the high hydrocarbon
concentration range under investigation, the gas sample needs to be
diluted to avoid gas direct absorption, thus lowering the optical
power available to stimulate the photoacoustic effect, and the saturation
of electronic components. The dilution also helps the detection such
that they do not to exceed the linearity range set by the gain of
the transimpedance amplifier. This enclosure uses a commercially available
gas blender to mix gas from the sample port with N2 from
an external cylinder.
Gas Drying Enclosure
The gas sample
needs to be dried before entering the QEPAS chamber to minimize the
effect of moisture. In fact, CH4 may rapidly relax the
energy through collisions with water vapor molecules, thus the photoacoustic
generation efficiency strongly depends on H2O concentration
variations in the 3 μm wavelength range.[11] The drying enclosure uses a monotube dryer to dehumidify
gas samples down to 300 ppm of water vapor. The monotube dryer uses
a Nafion tube to transfer moisture from the sample line to a purging
line surrounding the Nafion tube.
QEPAS
Sensor with a Laser Controller
QEPAS measures the absorption
of a gas sample via photoacoustic excitation,
by focusing a tunable laser beam between the prongs of a tuning fork.
The laser wavelength is modulated across the absorption peak of the
target gas. The gas modulated absorption generates a pressure wave.
The modulation frequency matches the vibrational fundament mode of
the QTF. The quartz piezoelectric property allows the pressure wave
to be converted into an electric signal. The amplitude of the signal
from the QTF is proportional to the gas concentration.[14] The ADM houses a standard QTF resonant at f0 = 32 767 Hz and the cylindrical acoustic
resonators. A mid-infrared ICL emitting at a central wavelength of
3345 nm was employed. An optical power meter is used to align the
beam through the acoustic resonator and between the QTF prongs. The
alignment is improved, minimizing the noise acquired in the QEPAS
scan.
QEPAS Software
Two programs are developed
for operating the QEPAS gas analyzer. The first program is developed
by Aramco Houston Research Center for monitoring the sample preparation
system. PolySense Lab developed a LabVIEW-based software to both control
the laser source and acquire and process the QTF data, employing a
PCIe-6363 NI-DAQ. The modulation frequency provided to the laser is f0/2. A dedicated subroutine is implemented to
acquire the QTF signal and retrieve its f0 component. This 2f-detection allow for the acquisition
of a background-free signal.[14]
Laboratory Testing, Calibration, and Performance
In
a previous study, a QEPAS sensor capable of detecting C1, C2,
and C3 in a nitrogen-based mixture was reported.[12,13] The hydrocarbon concentrations ranged from 1 to 1000 ppm and the
measurements demonstrated a good linearity between the QEPAS signal
and the concentration of each individual component in the range (Table ). The Allan deviation
analysis showed that for 1 s integration time the detection limit
for C1, C2, and C3 is ∼90, ∼7 ppb, and ∼3 ppm,
respectively. The detection limit achieved for C2 made a record for
the QEPAS technique, and measuring C3 added a new capability to the
technique. These limits are well below the sensitivity needed for
a sensor aiming at hydrocarbon detection in petroleum exploration
and engineering, where concentrations of hydrocarbon gases are expected
to be generally much above the ppm scale (usually %). C1–C2
and C2–C3 mixtures were also tested in the previous studies
to demonstrate the capability of detecting multiple hydrocarbons by
QEPAS.
Table 1
Performance Comparison of QEPAS with
Representative Optical Sensors for Hydrocarbon Gas Detection
CRDS
TILDAS
QEPAS
C1
C2
C1
C2
C1
C2
C3
precision
30 ppb
10 ppb
300 ppt
50 ppb
90 ppb
7 ppb
3 ppm
response time
<1 s
<1 s
<1 s
<1 s
<1 s
<1 s
<1 s
Sampaolo et al., (2020)[13]
linearity range
1–5000 ppm
0–500 ppm
0–100 ppm
0–50 ppm
0–1000 ppm
0–1000 ppm
10–1000 ppm
linearity range
4–1000 ppm
2–100 ppm
200–1000 ppm
Sampaolo et al., (2019)[12]
linearity range
100 000–1000 000 ppm (10–100%)
10 000–100 000 ppm (1–10%)
2000–20 000 ppm (0.2–2%)
this
study
In this
study, the QEPAS gas analyzer system including the sensor
box and new components of sampling preparation (Figure ) with enlarged concentration ranges for
individual C1, C2, and C3 that are typically encountered in natural
gas samples from oil and gas fields (Table , Figures –6) for linearity assessment is presented. A series of C1–C2
and C1–C2–C3 mixtures were tested, confirming that QEPAS
can measure C1, C2, and C3 in the three-hydrocarbon-components gas
mixture and extending this capability also at concentrations in the
percentage scale.
Figure 4
(a) QEPAS spectra for methane testing, (b) relation between
peak
amplitude and concentration for calibration peaks at 55 mA, and (c)
relation between peak amplitude and concentration for methane peaks
at 62 mA.
Figure 6
(a) QEPAS spectra for propane testing and (b) relation
between
peak amplitude and concentration for propane peaks at 22 mA.
(a) QEPAS spectra for methane testing, (b) relation between
peak
amplitude and concentration for calibration peaks at 55 mA, and (c)
relation between peak amplitude and concentration for methane peaks
at 62 mA.(a) QEPAS spectra and for ethane testing and
(b) relation between
peak amplitude and concentration for ethane peaks at 44 mA.(a) QEPAS spectra for propane testing and (b) relation
between
peak amplitude and concentration for propane peaks at 22 mA.
Single Component (C1, C2, and C3) Testing
Standard C1, C2, and C3 gases with certified concentrations and
high purity (99.9999%) N2 were used to prepare artificial
gas samples. Each gas was diluted using the N2 into different
concentrations to generate a hydrocarbon mixture that may be encountered
in natural gas samples from oil and gas fields. C1 samples were diluted
in a very large concentration range from 10% to 100% for the variation
of C1 concentration in oil-associated gas (least C1), condensate,
wet gas, dry gas,[3] or biogenetic gas (predominated
C1).[2] C2 and C3 samples were diluted in
the range of 1 to 10% and 0.2 to 2%, respectively.Measurements
were performed across the range of concentrations for each single
gas component. Two samples taken at each target concentration were
analyzed to check the reproducibility. QEPAS spectra and calibration
for the C1, C2, and C3 QEPAS spectra, respectively, are shown in Figures –6. The higher concentration scales considered in
this work required the detection strategy to consider the sensible
variations of the gas matrix and the non-linearity arising from the
high absorption coefficients. Indeed, the photoacoustic generation
is sensitive to the target molecule relaxation dynamics, which depends
on the energy transfer efficiency through collisions with other molecules
in the gas mixture. Furthermore, the percentage concentrations reached
in this investigation, combined with the intense line-strengths characterizing
the C1, C2 absorption features exploited by Sampaolo et al. (2019,
2020),[12,13] determined a non-linear relation between
the QEPAS signal recorded at those wavelengths and the concentration
of the target molecule. This is due to the Lambert–Beer law,
which rules the light absorption through a gas sample as:where I0 is the
incident light, l is the interaction pathlength,
and α is the absorption coefficient.This relation can
be linearized asfor small absorptions, which is the case trace gas detection via
QEPAS. This is not the case for the C1 two-fold structure with the
strongest peak at 55 mA, C2, C3 calibrations.In Figure , it
can be easily noticed that the C1 QEPAS peak signal at 55 mA is saturating
while approaching the 100% concentration (10% after dilution in pure
N2) and the shape of the QEPAS spectrum also suffers from
a progressive distortion due to the increased absorption. On the contrary,
the QEPAS signal at 62 mA shows a good linearity as a function of
the concentration because of the much lower line-strength associated
with the excited optical transition.In Figure , the
calibration of the 44 mA C2 peak signal is plotted. The QEPAS signal
amplitude shows a sub-linear trend versus concentration, demonstrating
how direct absorption impacts on a C2 calibration curve even in a
concentration range one order of magnitude lower with respect to C1.
This is due to an overall C2 cross-section at 44 mA approximately
twice as large as C1 at atmospheric pressure. However, this calibration
curve is a monotonically strictly increasing function of the concentration,
and it can be fitted by a polynomial.
Figure 5
(a) QEPAS spectra and for ethane testing and
(b) relation between
peak amplitude and concentration for ethane peaks at 44 mA.
The portion of the QEPAS
spectrum used as a diagnostic feature
for evaluating the broadband absorption of C3 was identified in the
20–23 mA window, in which a linear behavior (Figure ) has been verified as well.
C1–C2 and C1–C2–C3 Gas
Mixtures Testing
While C1 and C2 detection is straightforward,
thanks to the well-defined absorption peaks in the ICL operating range,
C3 detection requires the extraction of its characteristic broadband
absorption profiles which merge with C1 and C2 background signals
in the ICL tuning range. Previous studies showed no cross-talk between
C1 and C2 in the QEPAS spectra of C1–C2 gas mixtures and developed
a fitting procedure with a linear combination of reference spectra
to retrieve C2 and C3 concentrations with high accuracy in the C2–C3
mixtures under test.[12,13] Because of the abovementioned
reasons, the detection scale extension to per cents presented in this
work required the implementation of a multilinear regression, based
on the iterative analysis of different windows of the QEPAS spectra.
The algorithm exploits the saturated absorption of C1 to evaluate
the effect of the gas sample matrix on methane and the calibration
curves related to each diagnostic peak to extract C1, C2, C3 concentrations.
In terms of energy relaxation dynamics within gas samples with fluctuating
backgrounds, the cross-correlation among the first three alkanes will
be illustrated in detail in a further article, together with a systematic
description of the multivariate approaches that can be successfully
implemented.Two- (C1–C2) and three-components (C1–C2–C3)
gas mixtures were used in the tests to evaluate the performances of
the QEPAS systems, in terms of accuracy, precision, and linearity
between actual and measured concentration in multicomponent gas mixtures.
Certified 10% C1/N2, 10% C2/N2, and 10% C3/N2 cylinders were diluted by pure N2 in the gas mixing
enclosure to make gas samples in the representative range as observed
in the field (C1: 70–100%, C2: 2–10%, C3: 0.4–2%).
Their actual concentrations and QEPAS-measured values are compared
and displayed in Table and Figure for
C1–C2 mixture, and Table and Figure for C1–C2–C3 mixture.
Table 2
Gas Concentrations (%) and Ratios
of C1–C2 Mixture in the Bench Testing
C1 (%)
C2 (%)
C1/C2
mixture#
actual
measured
actual
measured
actual
measured
1
98.0
98.0
2.0
1.8
49.0
54.4
2
95.0
94.7
5.0
4.6
19.0
20.6
3
90.0
88.2
10.0
10.0
9.0
8.8
4
90.0
90.4
5.0
4.6
18.0
19.7
5
80.0
81.5
10.0
10.0
8.0
8.2
6
80.0
81.9
5.0
5.1
16.0
16.1
7
70.0
70.4
10.0
10.0
7.0
7.0
8
70.0
70.6
5.0
5.2
14.0
13.6
Figure 7
Comparison of actual
and measured gas concentrations and ratios
in the bench testing on C1–C2 mixture.
Table 3
Gas Concentrations (%) and Ratios
of C1–C2–C3 in the Bench Testing
C1 (%)
C2 (%)
C3 (%)
C1/C2
C1/(C2+C3)
C1/(C1+C2+C3)
mixture#
actual
measured
actual
measured
actual
measured
actual
Measured
actual
measured
actual
measured
1
93.0
96.0
2.0
2.0
1.0
0.7
46.5
48.0
31.0
35.6
0.97
0.97
2
90.0
91.6
8.0
8.1
0.4
0.2
11.3
11.3
10.7
11.0
0.91
0.92
3
90.0
92.5
5.0
4.5
1.0
0.6
18.0
20.6
15.0
18.1
0.94
0.95
4
85.0
84.3
10.0
10.0
1.0
0.7
8.5
8.4
7.7
7.9
0.89
0.89
5
80.0
80.6
10.0
10.0
2.0
1.6
8.0
8.1
6.7
6.9
0.87
0.87
Figure 8
Comparison
of actual and measured gas concentrations and ratios
in the bench testing on C1–C2–C3 mixture.
Comparison of actual
and measured gas concentrations and ratios
in the bench testing on C1–C2 mixture.Comparison
of actual and measured gas concentrations and ratios
in the bench testing on C1–C2–C3 mixture.For C1–C2 mixture, the measured
C1 and C2 concentrations
are very close to their actual values, showing good linearity with
R2 > 0.99 (Figure a,b). The C1/C2 ratio, which is commonly used as a
proxy of
gas dryness and a geochemical fingerprint,[2] also shows a good linearity (R2 >
0.99, Figure c), but
with a deviation
when C1 presented the highest concentration (98%) and C2 presented
the lowest concentration (2%) in the samples under test. This is mainly
due to the efficiency of the analysis method in discriminating relatively
low C2 concentrations from the C1 background absorption signal.For the first time, the detection of C1, C2, and C3 in a gas mixture
at the percentage scale using QEPAS technology is reported. As shown
in Figure for the
C1–C2–C3 mixture, comparisons between actual and measured
data demonstrate a fairly good linearity (R2 > 0.977). C1 concentration values are the most immediate evidence
of the margin for improvement that the current analysis method has
in quantifying the effect of matrix variations on the detection of
individual components. On the other hand, C3 underestimation is an
indication that the present spectral range is not the optimum choice
for C3 detection in natural gas such as mixtures. Indeed, the identification
of a more characterizing fingerprint for C3, clearly distinguishable
with respect to the absorption background of the other hydrocarbons
and relying on a higher absorption cross-section, would definitely
help the analysis method in accurately retrieving propane concentration.
Obtaining accurate concentrations for components in a gas mixture
consisting of homologous compounds (e.g., hydrocarbons) is much more
challenging than dealing with single or two components.[13,27] However, the good correlation and linearity between all measured
gas concentrations and ratios and their actual values suggest that
QEPAS sensors are capable of detecting C1, C2, and C3 in oil and gas
applications and improved algorithms may generate more accurate results.
Several aspects will be pursued to enhance the performance of the
QEPAS system:quantitative control in the gas mixing
enclosure;selective
features for better defining
C3 in the background with C1 and C2;testing in a larger concentration range
for C2 and C3 to have a more dynamic calibration curve;trying a partial least square multivariate
analysis and machine learning method to optimize the procedure to
calculate gas concentrations in the mixture.Currently, we suggest using the ratios of C1/(C1 + C2 + C3)
as
a dryness parameter/geochemical fingerprint in the application of
the QEPAS system; because the gas ratio as a procedure of normalization
eliminates the error and demonstrates better results than direct use
of the concentrations (Figure d,f).
Discussion
Gas Species, Multicomponent Mixtures, and
Natural Gas Samples
The QEPAS system reported in this study
represents an advancement of QEPAS sensor prototypes demonstrated
by Sampaolo et al. (2019; 2020)[12,13] by expanding the detection
range to per cents scale and setting the basis for C3 detection in
multicomponent mixtures mainly based on hydrocarbons. More than 25
gas species now can be detected and quantified by QEPAS,[9,11,14,24,25,29−31] including the most common and business-impacted gas species (C1,
C2, C3, CO2, and H2S) in the upstream of oil
industry (Figure ).
The best noise equivalent concentration, as a parameter of minimum
detection limit, were reported as a function of the employed laser
wavelength.[9,30] Although GC-FID/TCD and QMS used
in laboratory and mud gas unit can measure the full range of hydrocarbon
gas species (C1–C10) and more inorganic gas components, QEPAS
demonstrates extremely high sensitivity as ppb level of some gases
can be detected (Figure ) over GC-FID/TCD and QMS (usually ppm level).[16,17]
Figure 9
Minimum
detection limits and operating spectral region for QEPAS-based
gas sensors. The gas species with interest from the oil industry are
marked with red labels.
Minimum
detection limits and operating spectral region for QEPAS-based
gas sensors. The gas species with interest from the oil industry are
marked with red labels.QEPAS can reach high
detection sensitivity at the short optical
pathlength and have immunity to environmental noise, representing
the main distinct advantages among optical absorption spectroscopies.[9,30] In contrast to those well-developed optical absorption techniques
(e.g., CRDS and TILDAS), QEPAS detection displays some cons and pros
(Table ). The three
sensors listed in the table are portable devices sharing the advantage
of a fairly low response time and high precision and sensitivity,
thus allowing for measurements in real-time and/or in a continuous
monitoring mode. As discussed by Sampaolo et al., (2020),[13] the outstanding precision in the ppt range provided
by the TILDAS sensor limits on the other hand the detection range,
leading to the deployment of this device mainly in environmental monitoring
of C1 and C2 (as stray gas) and scientific studies (e.g., space exploration).
Another distinct advantage of QEPAS over TDLAS and other absorption
spectroscopy is that its performance can be improved when the output
of the excitation laser source is increased.[24] The QEPAS sensor tested shows comparable performances in terms of
precision compared to the CRDS sensor, and additionally offers C3
concentration with a precision in the ppm range. In the study by quantitative
dilution performed in the gas mixing enclosure of our QEPAS system,
the detection ranges have been significantly enlarged for natural
gas samples from oil and gas fields (Table ).QEPAS has demonstrated the capability
of detecting multiple gas
species,[9,23,27,30] which is a main advantage over “unconventional”
semiconductor and electrochemical gas sensors. Individual H2S, CO2, and C1 gas (diluted in N2) were successfully
tested in laboratory with QEPAS, and the technology was introduced
to oil industry for continuous monitoring the chemical environment
inside flexible risers by Weppenaar et al., (2009).[7] However, application of QEPAS to chemical composition analysis
on multicomponent gas mixtures is difficult because of complex processes
linked to the speed of sound, molecular relaxation dynamics, and overlapping
of absorption lines. A QEPAS using near-infrared, fiber-coupled diode
lasers as an excitation source and data analysis algorithm were developed
to determine chemical composition of gas mixtures with H2S, CO2, and C1 at concentrations 0–100%.[27] The pilot study provided a guideline and suggested
that it was possible to utilize multivariate regression, nonlinear
least square fitting, or machine learning for deriving gas concentrations
regardless of the complexity of the involved phenomena.Simultaneous
multicomponent gas detection using a standard QEPAS
approach is not feasible because the standard configuration is based
on a sequential excitation of separated absorption features, or a
continuous wavelength scan over a broad absorption to discriminate
the different contributions to the spectrum. The force sensing-based
QTF cannot recognize the gas component inducing the acoustic wave.
In both cases, there is a time delay in measuring two or multiple
gas target concentrations.[30] The fundamental
resonance frequency was recently reduced via custom QTFs, allowing
the simultaneous excitation of the QTF on both the fundamental and
first overtone modes employing the same or different laser sources.
A QTF frequency division multiplexing technique was developed for
the first demonstration of simultaneous dual-gas (C2H2/H2O) sensing by QEPAS.[32] A number of dual-gas (two components) can be simultaneously detected
by QEPAS, however, the triple-gas (three components) still require
a wavelength scan. We demonstrate in the paper a QEPAS gas analyzer
that is capable of a fast response detection of C1, C2, and C3 using
a single ICL source operating around 3.3 μm, representing a
breakthrough of QEPAS technology for hydrocarbon sensing applications.Expanding the detection for C3+ gas species and including H2S and CO2 in a QEPAS sensor is a very challenging
task, but very important for applying the QEPAS sensor in analyzing
natural gas samples from drilling, downhole sampling, well testing,
and production. In the natural gas sample, the background absorption
will rise up from C2, C3 and the heavier molecules (C4, C5, etc.),[13] and there might be cross-talk between the partial
overlapping lines of CO2 and H2S.[27] In this case, each molecule must be spectrally
characterized to generate a reference spectrum as a function of the
laser wavenumber λ at a fixed concentration. Then, a fitting
procedure based on linear combination of reference spectra can be
constructed for correlating the QEPAS response with gas concentration.[13,30]
Isotope Analysis
Gas isotopes, in
particular the isotopic composition or isotope ratio of a gas component
(e.g., δ13C of CH4), are very useful in
determining the origin of gas, estimating thermal maturity, and representing
a fingerprint of reservoir compartmentalization study and production
allocation.[2,18]An early study using the
modulation cancellation method separated H218O and H216O in water vapor from QEPAS signal.
The sensitivity in measuring the deviation from a standard sample
was 1.4‰ for δ18O, in 200 s of integration
time.[33] A QEPAS sensor with a continuous-wave
ICL, emitting at 4.35 μm, was capable of distinguishing 12CO2 and 13CO2 and measuring
δ13C of CO2 with an average precision
<1‰, which fulfills the precision requirement of most medical
and field applications.[34] A mid-IR QCL
emitting at 7.73 μm, was used in a QEPAS sensor to selectively
detect 12CH4 and 13CH4 isotopologues at ppb sensitivity level.[13] A standard methane sample was tested, and the standard abundances
of ∼98.82% 12CH4 and 1.11% 13CH4 were verified. The QEPAS sensor is expected to detect
the variation of δ13C below 1‰ with an integration
time in the order of magnitude of tens of seconds for methane with
percents concentration in a gas mixture, typical of natural gas samples.
The two prototypes for C1 isotope and hydrocarbon gas (C1–C3)
concentrations determination can be potentially integrated into a
QEPAS sensor box.[13]Isotope ratios
are critical geochemical parameters or fingerprints
for petroleum system analysis, fluid characterization, and injection
and production surveillance. The advances in isotope analysis with
QEPAS sensors add a strength of the optical sensing technology in
geochemical analysis, which would promote the application of QEPAS
gas sensing in on-site fluid characterization and real-time/time-lapse
monitoring.
Conclusions and the Way Forward
QEPAS is a novel laser-based optical absorption technology for
gas sensing and chemical analysis. The technology is versatile in
detecting trace gas and measuring concentrations and isotopic compositions
of multiple gas components. Recent advances have demonstrated the
features of QEPAS sensor, including extremely high sensitivity (down
to ppb/ppt), high level of compactness, immunity to environmental
noise, insensitivity of resonator to the excitation wavelength, potential
for continuous monitoring, reliability, and robustness for in-situ
operation.QEPAS takes the main advantages in multicomponent
gas detection
and quantitative analysis over semiconductors and electrochemical
sensors. QEPAS currently cannot measure the extensive gas species
for their concentrations and isotopic compositions in natural gas
samples as the laboratory-based analytic instruments (e.g., GC-FID/TCD,
QMS, IRMS). However, the optical sensor is much smaller and relatively
cheaper, requiring less maintenance and offering quicker analysis
with higher sensitivity than the laboratory-based instruments. As
discussed, the two advantages of the QEPAS over other optical sensing
techniques (e.g., CRDS, ICOS, TILDAS) are (1) the capability to measure
C3 concentration and (2) extremely high sensitivity in short optical
pathlengths. Comparing technical specifications and performances is
beyond the scope of the paper.Being beneficial from the features
and advantages, QEPAS has been
proposed for field applications in monitoring sour gas (H2S, CO2) in an offshore riser,[7] downhole gas analysis,[10] early fire,[6] leak detection, and health-safety-environment
monitoring.[30] A typical QEPAS sensor system
can be compacted in a small size with light weight, supporting the
deployment of the sensor in a portable analyzer, a downhole tool,
and an unmanned aerial vehicle.[24,35] Several real-world
applications of QEPAS have already been reported,[36,37] including carbon monoxide (CO) detection in urban area, monitoring
of methane (CH4) in landfills, and leak detection of sulfur
hexafluoride (SF6) in a vacuum-seal test station for diesel
engine injector.In this study, a QEPAS-based hydrocarbon gas
analysis system was
developed and tested. The QEPAS sensor is highly sensitive for hydrocarbon
gas detection (ppb to 10 ppm level). The sensor is not suitable for
direct analyzing natural gas samples with hydrocarbon components in
ppm to % level. We integrated multiple enclosures for gas introducing,
mixing, and drying with a QEPAS sensor box in a rack and developed
the associated software to monitor sample preparation and retrieve
the hydrocarbon concentrations. We realized a new QEPAS prototype
for detecting hydrocarbon gas (C1, C2, and C3) and measuring their
concentrations for the oil industry sample by quantitative dilution
and optical sensing. The implementation of quantitative dilution in
our system expands the linearity range of the QEPAS detection from
ppm to % range.The prototype system was tested with C1, C2,
and C3 single components
and C1–C2 and C1–C2–C3 mixtures, with a series
of concentration gradients that are typically encountered in natural
gas samples from oil and gas fields. Good linearity relations between
QEPAS response and concentration at % level are shown for the single,
binary and trinary components, confirming that measuring concentrations
in % level and multicomponent gas detection can be achieved by the
system. Especially, it is the first time to report the trinary C1–C2–C3
mixtures at % level.Although there are some deviations of measured
C1%, C3%, and C1/(C2
+ C3) ratio compared to actual values, all measured concentrations
and gas ratios illustrate good linear correlation with actual values
(Figure ), suggesting
that it is highly possible to generate more accurate results by improving
calibration curves and the multivariate approach implemented.Future developments and applications of QEPAS technology in the
oil industry include (1) testing natural gas samples and developing
the detection strategy to determine gas concentrations and isotopic
compositions in the real samples (complex mixture); (2) integrating
CO2 and H2S sensors with the hydrocarbon gas
sensor in a compact device; (3) reducing the size and increasing robustness
for field deployment; and (4) developing the applications of on-site
gas analysis while drilling, testing, and production, by the advanced
sensing, to support real-time decision making and time-lapse operation.
Authors: Marilena Giglio; Andrea Zifarelli; Angelo Sampaolo; Giansergio Menduni; Arianna Elefante; Romain Blanchard; Christian Pfluegl; Mark F Witinski; Daryoosh Vakhshoori; Hongpeng Wu; Vittorio M N Passaro; Pietro Patimisco; Frank K Tittel; Lei Dong; Vincenzo Spagnolo Journal: Photoacoustics Date: 2019-12-26