Treatment of C1.1 (AO-41) tibial plateau fracture: A finite element analysis of single medial, lateral and dual plating.

Norin Forna1, Florin Munteanu2, Sînziana Anca Butnaru Moldoveanu2, Liliana Savin1, Paul Dan Sîrbu1, Bogdan Puha1.   

Abstract

Bicondylar tibial plateau fractures pose many challenges in surgical treatment. The aim of the present study was to analyze three methods of reduction, single medial, single lateral, and dual plating, for the treatment of a bicondylar tibial plateau fracture, through finite element analysis (FEA). A simple metaphyseal fracture, type C1.1 (AO-41) was modeled on a CT-derived 3D model of the knee. Lateral and medial proximal tibial polyaxial plates with screws were modeled and placed accordingly for the three methods of reduction. Simulation of physiological type loading corresponding to the maximal weight acceptance phase during a slow walking gait cycle was performed using FEA. Values of stress and strain were recorded near the fracture lines. Dual plating provided a decrease of stress and strain in the tibial plateau area. However, the differences in the values among the three cases were small. The stress concentration areas were located in the vicinity of the fracture, predominantly in the area of the tibial plateau. Considering the limitations of the present study, the results revealed that dual plating leads to smaller stress and strain values near the fracture lines in the tibial plateau area. However, values obtained for single lateral plating are close in range. Considering the complexity of the surgical approach for dual plating, single lateral plating may be a solution for good reduction with fewer surgical risks and complications. Further studies on the C1.1 fracture (AO-41) are needed to analyze the complex issue of reducing and stabilizing such a fracture and to characterize the postoperative state while providing predictable parameters for an optimal result. Copyright: © Forna et al.

Entities:  

Keywords:  dual plating; finite element analysis; fracture reduction; lateral plating; stabilization; tibial plateau

Year:  2022        PMID: 35126701      PMCID: PMC8794547          DOI: 10.3892/etm.2022.11121

Source DB:  PubMed          Journal:  Exp Ther Med        ISSN: 1792-0981            Impact factor:   2.447


Introduction

Bicondylar tibial plateau fractures are common injuries of the lower limb, leading to the destabilization of both the medial and lateral condyles (1-3). This type of fracture accounts for up to 39% of all tibial plateau fractures (4). The bicondylar fracture occurs as a result of significant trauma involving a high-energy mechanism in which the varus or valgus is combined with axial loading (3). The C1.1 (AO-41) fracture, from the AO/OTA classification, is a complex variation of bicondylar fractures where the fracture line is described as having an inverted ‘Y’ trajectory (5). In the case of a bicondylar fracture, the reduction and fixation techniques usually involve an open reduction and an internal fixation using a plate and screws that can be applied laterally, medially, or combined (4). Dual plating is considered the most stable type of fixation for bicondylar tibial plateau fractures. However, the surgical approach for dual plating is complex, involving two incisions and soft-tissue manipulation. Moreover, for this type of fracture, soft-tissue complications are an important concern associated with dual plating (4). Plates with polyaxial stability (PAS) offer the possibility to adapt the position of the screws to the fracture trajectory. Due to this, appropriate fixation and stabilization can be achieved through a single plate with lateral fixation (6). However, each case of bicondylar fracture must be analyzed in the context of specific clinical features, considering the trajectory of the fracture and the affected soft tissues. This is necessary because no single method can guarantee optimal results for the full range of possible clinical cases (4). The selection of an inappropriate treatment for the bicondylar fracture can lead to a reduction in the range of motion of the joint and joint instability. There are, of course, several other issues that contribute to poor results in the treatment of bicondylar plateau fractures. The advanced age of >50 years of a patient has been associated with poor postoperative results (7-9). Smoking, the existence of an open fracture, and compartment syndrome can increase the risk of infection (10,11). In assessing the outcome of a fracture at the knee joint, the Rasmussen and Iowa scores are the only tools specifically designed for this (12,13). However, these tools have not been validated with a rigorous methodology to assess the results of patients with bicondylar tibial plateau fractures (4). The finite element analysis (FEA) method is particularly advantageous as it allows for the virtual reconstruction of the anatomical and physiological conditions of this type of fracture. Bone shape and fracture trajectory, as well as biomechanical conditions, such as the force generated by muscle deformation, which in turn is transmitted through tendons to the bone, can be simulated (14). It is quite difficult to make an experimental device, in vitro, that respects the anatomical and physiological conditions as can be done in FEA. This method has been widely used in trauma research in orthopedics, to determine whether there are stress concentrators in the implants or osteosynthesis used, that would lead to their deformation or rupture, but also to analyze the effect of these devices on the bone tissue (14-16). For the C1.1 type bicondylar fracture (AO-41), the scientific literature presents a lack of FEA studies to distinguish between different simulated scenarios and to analyze the results obtained. The aim of the present study was to analyze three methods of reduction, single medial, single lateral and dual plating, for the treatment of the bicondylar tibial plateau fracture, C1.1 (AO-41), using FEA to determine the stress and strains at the level of the reduced and stabilized fracture.

Materials and methods

Knee model

For this study, a 3D model of the knee was obtained using CT images in Slicer3D (http://www.slicer.org) and 3D modeling with further processing was carried out using Autodesk ReCap Photo version 2019 and Autodesk Fusion360 (Autodesk, Inc.). Following the bone reconstruction, the menisci and cartilage tissue were modeled in Autodesk Fusion360 (Autodesk, Inc.) (Fig. 1).
Figure 1

A 3D model of the knee. (A) Frontal view of the knee model. (B) Exploded view of the knee model. (C) Modeled C1.1 (AO-41) fracture.

Method selection and load simulation

A simple metaphyseal fracture, type C1.1 (AO-41), according to the AO/OTA classification, was modeled, with an inverted ‘Y’ trajectory. For this type of fracture, three methods of fracture reduction were selected, single medial, single lateral and dual plating. Proximal polyaxial tibial plates and screws of 3 mm diameter, with no threads were modeled and placed accordingly for the three methods of reduction. To ensure contact areas between the plate and tibia, very small volumes of material were cut from the plates, following the contour of the contact surfaces. These volumes were very small and negligible in terms of plate thickness, having only the role of providing small contact areas. The positioning of the plates and screws on the fractured 3D model is presented in Fig. 2.
Figure 2

Positioning of the plates and screws on the modeled fracture for the three methods of reduction selected. Frontal plane view of (A) single lateral plating, (B) single medial plating, (C) dual plating. Medial descriptive view of (D) single lateral plating, (E) single medial plating, (F) dual plating.

Load simulation was performed in the Simulation Mechanical 2017 software (Autodesk, Inc.). Materials assigned to all parts were considered materials with linear isotropic and elastic mechanical properties and were characterized by Young's modulus and the Poisson's ratio, the values of which are presented in Table I (17-19).
Table I

Material properties.

Material (Refs.)Young's Modulus (MPa)Poisson's ratio
Cortical bone (17)17,0000.33
Trabecular bone (17)5,0000.33
Cartilage (18)50.46
Menisci (18)590.49
Titanium (19)110,0000.3

Determining the magnitude of the forces

To obtain the appropriate magnitude of the forces applied, the peak value in the maximal weight acceptance phase that occurs during a slow walking gait cycle was considered. In a complete gait cycle, the reaction force has two peaks occurring at weight acceptance and push-off, as well as a minimum value in mid-stance. The first peak was considered for the calculation of the applied forces. The approximation of the position, as a percentage of the gait cycle, of this first peak was produced based on the data reported by Arnold and his collaborators (20). Thus, the first peak of the ground reaction force occurs at 16% of the gait cycle at a walking speed of 1 m/sec. The forces were calculated for an individual with a body mass of 70 kg. The applied loads were represented by the forces of the quadriceps muscle (FQ), semitendinosus (FSt) and semimembranosus (FSm), medial and lateral gastrocnemius (FGm, FGl), biceps femoris (FBf), as well as the ground reaction force (GRF), as revealed in Fig. 3.
Figure 3

Descriptive image of forces and constraints applied to the 3D model. (A) Anterior and (B) posterior view. FQ, quadriceps muscle; FSt, semitendinosus; FSm, semimembranosus; FGm, medial gastrocnemius; FGl, lateral gastrocnemius; FBf, biceps femoris; GRF, ground reaction force.

The magnitude of these forces was calculated using the formula: where Fm is muscle force, Ts is specific tension, PCSA is physiological cross-sectional area, and VEMG is normalized value of muscle activity during maximal weight acceptance. For the specific tension of the muscle, an average value of 22.5 N/cm2 was used, and the physiological cross-sectional area, specific to each muscle, was selected from available data from the literature (21,22). The normalized value of the muscle activity, during maximal weight acceptance, was approximated from the data reported by Arnold and his collaborators (20). The directions of the muscle forces were established according to data from the literature (22-26). In the frontal plane, an angle of 3˚ between the mechanical and the vertical axis was considered. In the simulation interface, the directions of the forces were set to the vertical axis. In the frontal plane, FQ was oriented in the direction of the Q angle (14˚), calculated to the vertical axis, FBf 11.8˚ medial, FSm 7˚ lateral, FSt 5.1˚ lateral, FGl 4.8˚ lateral, and FGm 5.3˚ medial to the mechanical axis of the tibia (23-27). For the sagittal plane, the knee was considered to be in extension, and the anatomical axis of the tibia, which corresponded to the mechanical axis, was vertical. In the sagittal plane, FQ, FGl, and FGm were considered parallel to the tibial axis and FBf 7.3˚ anterior to the tibial axis, FSm oriented at 16.1˚ and FSt at 19.6˚ posterior to the tibial axis (24-26). Regarding the direction of the GRF, the step size was extremely small postoperatively. Thus, GRF acted normally on the tibial surface, the angle between the tibia and the reaction force being negligible. The magnitude of the GRF was approximated based on data reported by Arnold and his collaborators, at maximum weight acceptance (20). Table II reveals the rounded values of the muscular forces as well as the GRF. The fibula bone was not included in this study. However, the contact area between the fibula and the tibia was considered as the area of application of the force developed by the biceps femoris muscle, which has its insertion in the fibula. Constraints were applied to the diaphyseal surface of the femur, completely restricting the movement of this surface. The type of contact selected between the simulation elements was bonded. A descriptive image of the forces and constraints applied to the 3D model is shown in Fig. 3.
Table II

Rounded values of applied forces (N).

FQFSmFStFBfFGlFGmGRF
2135013422462720

FQ, quadriceps muscle; FSt, semitendinosus; FSm, semimembranosus; FGm, medial gastrocnemius; FGl, lateral gastrocnemius; FBf, biceps femoris; GRF, ground reaction force.

Results

Analysis of maximum principal stress and strain values

In this study, the values of maximum principal stress and strain were analyzed (Figs. 4 and 5). Values were recorded at the cortical level near the fracture lines, in two areas, the central frontal area (where the fracture lines split) and the tibial plateau area. To eliminate artifacts represented by high local values, an average of 3 values recorded in the vicinity of the area of interest was made. The highest stress and strain values were recorded near the fracture lines, in the tibial plateau area. Dual plating led to a decrease of stress and strain values in this region. In the center frontal area, values were similar between the three methods of reduction analyzed.
Figure 4

Maximum principal stress recorded at cortical bone level, in the vicinity of the fracture lines, in the center frontal and tibial plateau region.

Figure 5

Maximum principal strain recorded at cortical bone level, in the vicinity of the fracture lines, in the center frontal and tibial plateau region.

Stress distribution

The distribution of the stress values at the level of the tibia was also analyzed. For an optimal view of the stress distribution, the color legend was changed to highlight the areas of stress concentration. Thus, the upper-stress limit for this stress distribution was selected as 10 MPa. This value does not represent the maximum stress obtained but is only a value selected for a stronger exposure of the high-stress concentration areas. The stress concentration areas were located in the region of the fracture, predominantly in the area of the tibial plateau, as is evident in Fig. 6, Fig. 7 and Fig. 8. Small areas of stress concentration were evident in regions where the forces were applied and the plates came in direct contact with the bone. These areas appear in all three cases due to the same loading applied.
Figure 6

Distribution of the maximum principal stress for single medial plating. Descriptive (A) anterior, (B) antero-lateral, and (C) postero-medial view.

Figure 7

Distribution of the maximum principal stress for single lateral plating. Descriptive (A) anterior, (B) antero-lateral, and (C) postero-medial view.

Figure 8

Distribution of the maximum principal stress for dual plating. Descriptive (A) anterior, (B) antero-lateral, and (C) postero-medial view.

Discussion

The aim of the present study was to analyze the complex reduction and fixation of the tibial plateau C1.1 (AO-41) type fracture, considering three methods of reduction, single medial, single lateral and dual, by using proximal polyaxial plates and screws. The results revealed that both stress and strain increased in the region of the tibial plateau. The stresses and strains increased due to the forces applied but also the fixture of the plates. The bone discontinuity caused by the fracture path caused a large amount of stress on the plate when loading was applied, which was then transmitted to the bone tissue near the screws and further along to the fracture lines. Strain values can be analyzed in relation to the physiological intervals defined by Frost through the mechanostat theory (28). The theory defines the mechanism of adaptation of the bone tissue to the action of a mechanical stimulus that is maintained in a specific area by the variation of bone mass (29,30). In other words, the action of a mechanical stimulus leads to a remodeling of the bone tissue, for the bone to respond favorably to the applied loads. The response of bone tissue can be quantified by analyzing the values of the strain generated (28,31). There are four intervals of bone strain, <1,000 µɛ (absence of an applied load, potential for tissue atrophy), 1,000-1,500 µɛ (tissue homeostasis), 1,500-3,000 µɛ (physiological load that promotes bone remodeling), and >3,000 µɛ (overload associated with fractures and bone resorption) (28,31). The values of strain obtained in both areas were found in the minimum strain range. Although it is considered that in the present study the strain was not enough to achieve bone remodeling, it is noteworthy that the simulation was performed taking into account forces applied during a smaller walking speed. However, the average speed for normal walking is reported at approximately 1.33 m/sec, while slow walking speed is approximately 0.66 m/sec (32). By decreasing the walking speed, the muscle forces also decrease, resulting in lower values of stress and strain. At lower speeds, acceleration is also smaller. The same acceleration can be transposed to the system of forces in the human body, i.e., the lower the acceleration, the lower the forces exerted by the muscles, and thus, the stress transmitted to the bone tissue becomes smaller. In addition, the variation of the GRF in a complete walking cycle may change and in the context of a fracture, the variation can also extend to the walking phases. As the speed of movement or the angle of the GRF changes by increasing the step size, the load on the tibia may increase accordingly. The variation obtained shows that dual plating leads to smaller stress and strain values near the fracture lines in the tibial plateau area. Dual plating is considered to be the most stable type of fixation for bicondylar tibial plateau fractures; however, the surgical approach involved is complex and soft-tissue complications are an important concern associated with dual plating (4). In the present study, values obtained for single lateral plating were close in range. Considering the complexity of the surgical approach for dual plating, single lateral plating may be a solution for good reduction with fewer surgical risks and complications. Moreover, plates with PAS provide the means to adapt the position of the screws to the fracture trajectory. Thus, appropriate fixation and stabilization can be achieved through a single plate with lateral fixation (6). We observed that the stress concentration areas are located in the fracture area, more specifically in the tibial plateau region, in the vicinity of the fracture line. The location of the stress concentration areas may derive from the loading generated by the muscular forces and also the type of contact between the surfaces. The type of contact between the elements is a limitation of FEA studies as it is difficult to characterize the contact between the fracture fragments, postoperatively. For this type of fracture, there is a lack of FEA studies to help distinguish between different simulated scenarios and to analyze the results obtained. As aforementioned, the existing studies describe only from a clinical point of view, the impact of different methods of reduction and osteosynthesis of simple metaphyseal fracture of the tibial plateau. The FEA study by Samsami et al addresses the bicondylar tibial plateau fracture; however, the fracture trajectory is extremely different from that of the present study (33). Additionally, the loading mode differs, forces of 350 N and 250 N were applied to the tibial plateau region in that study (33). Being a type of fracture characterized by increased heterogeneity from patient to patient, it also has a high risk of complications of treatment (34). Among the reported complications, there are associated lesions of the menisci and cruciate ligaments, vascular or nerve lesions, vicious consolidation, knee stiffness, and osteoarthritis (4). The heterogeneity of the fracture is also highlighted by the variety of technical solutions for reduction and stabilization. Normal means are open reduction and internal fixation in the bicondylar fractures of the tibial plateau (4,35). In the context of a lack of FEA studies addressing the C1.1 fracture type (AO-41) and the limitations of the FEA method, the results obtained in the present study need to be analyzed together with several aspects. The materials assigned to the 3D models were considered elastic linear. The values of the muscle PCSA used to calculate the force were obtained from cadaveric specimens and may be smaller than in vivo measured values in patients (22). Furthermore, the normalized EMG activity values were taken from a gait cycle of 1 m/sec speed. Although smaller than the average walking speed, slow walking speed is approximately 0.66 m/sec (32). A smaller walking speed may lead to smaller applied forces and thus to decreased stress and strain. Nevertheless, studies are required to determine the variation of muscle forces in a slow walking cycle, in the context of a reduced and stabilized bicondylar tibial plateau fracture. Concurrently, it is necessary to determine the variation of the GRF that occurs during a postoperative slow walking cycle, to obtain appropriate reference values. It is imperative that such studies be conducted on postoperative patients for the accuracy of data. Considering the limitations of this study, the results of the present study have shown that dual plating leads to smaller stress and strain values near the fracture lines in the tibial plateau area. However, values obtained for single lateral plating were close in range. Considering the complexity of the surgical approach for dual plating, single lateral plating may be a solution for good reduction with fewer surgical risks and complications. Plates with polyaxial stability allow for adaptation of the position of the screws to the fracture trajectory. This provides the means for appropriate fixation and stabilization through a single plate with lateral fixation. The values of strain obtained in both areas are found in the minimum strain range. Although in the present study, it is considered that the bone is not loaded enough to achieve bone remodeling or homeostasis, it is notable that the stresses transmitted by the plates and screws should not cause associated trauma but should only stabilize and reduce the fracture to promote healing. The present study can be further developed to explore forces that occur in other moments of a slow walking cycle to describe the complete loading of the fractured tibia. However, studies are required to determine the variation of muscle forces that occur in a slow walking cycle, in the context of a reduced and stabilized bicondylar fracture at the tibial plateau. Further studies on the C1.1 fracture (AO-41) are required to analyze the complex issue of reducing and stabilizing such a fracture and to characterize the postoperative state while providing predictable parameters for an optimal result.
  30 in total

1.  Orientation of tendons in vivo with active and passive knee muscles.

Authors:  Sietske Aalbersberg; Idsart Kingma; Janet L Ronsky; Richard Frayne; Jaap H van Dieën
Journal:  J Biomech       Date:  2005-09       Impact factor: 2.712

Review 2.  How accurately can we predict the fracture load of the proximal femur using finite element models?

Authors:  Sven van den Munckhof; Amir Abbas Zadpoor
Journal:  Clin Biomech (Bristol, Avon)       Date:  2014-01-10       Impact factor: 2.063

3.  Long-term follow-up after fractures of the tibial and fibular shafts.

Authors:  T C Merchant; F R Dietz
Journal:  J Bone Joint Surg Am       Date:  1989-04       Impact factor: 5.284

4.  Adaptive bone-remodeling theory applied to prosthetic-design analysis.

Authors:  R Huiskes; H Weinans; H J Grootenboer; M Dalstra; B Fudala; T J Slooff
Journal:  J Biomech       Date:  1987       Impact factor: 2.712

5.  Bone "mass" and the "mechanostat": a proposal.

Authors:  H M Frost
Journal:  Anat Rec       Date:  1987-09

6.  Tibial condylar fractures. Impairment of knee joint stability as an indication for surgical treatment.

Authors:  P S Rasmussen
Journal:  J Bone Joint Surg Am       Date:  1973-10       Impact factor: 5.284

Review 7.  Bicondylar Tibial Plateau Fractures: A Critical Analysis Review.

Authors:  Adam K Lee; Seth A Cooper; Cory Collinge
Journal:  JBJS Rev       Date:  2018-02

8.  Functional outcomes of severe bicondylar tibial plateau fractures treated with dual incisions and medial and lateral plates.

Authors:  David P Barei; Sean E Nork; William J Mills; Chad P Coles; M Bradford Henley; Stephen K Benirschke
Journal:  J Bone Joint Surg Am       Date:  2006-08       Impact factor: 5.284

9.  Prolonged operative time increases infection rate in tibial plateau fractures.

Authors:  Matthew Colman; Adam Wright; Gary Gruen; Peter Siska; Hans-Christoph Pape; Ivan Tarkin
Journal:  Injury       Date:  2012-11-28       Impact factor: 2.586

10.  Risk factors of infection after ORIF of bicondylar tibial plateau fractures.

Authors:  Brent J Morris; R Zackary Unger; Kristin R Archer; Shannon L Mathis; Aaron M Perdue; William T Obremskey
Journal:  J Orthop Trauma       Date:  2013-09       Impact factor: 2.512

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.