Literature DB >> 3512123

Efficacy of an automated external defibrillator in the management of out-of-hospital cardiac arrest: validation of the diagnostic algorithm and initial clinical experience in a rural environment.

K R Stults, D D Brown, R E Kerber.   

Abstract

Automatic external defibrillators (AEDs) may have advantages over manual defibrillation in managing prehospital cardiac arrest, particularly in rural communities. We conducted a two-part evaluation of a commercially available AED. We first established the diagnostic accuracy of the AED's rhythm recognition algorithm by challenging it with 205 cardiac arrest rhythms previously recorded from actual patients in the field. The AED demonstrated 100% specificity and 92% sensitivity for ventricular fibrillation (VF) in this nonclinical setting. We then compared the clinical efficacy of AEDs in 18 small communities (study group) with that of manual defibrillation in 18 additional communities (control group) of similar size. Ambulance technicians using manual defibrillators correctly diagnosed VF more frequently than the AEDs (98% vs 83%; p less than .025). Specificity for VF was similar in the two groups (100% for AEDs vs 94% for technicians; p greater than .10). AEDs were able to deliver shocks more quickly than was possible with the manual defibrillators (1.56 vs 2.77 min; p less than .001). The ability of the AEDs to terminate VF was excellent, converting VF in 28 of 29 (97%) patients to some other rhythm compared with only 37 of 53 (70%) patients in the control group (p less than .01). Hospital admission and discharge rates were similar for the two groups. Ten of the 35 (29%) patients managed with AEDs achieved admission and six (17%) were ultimately discharged. In the control group 17 of 53 (33%) patients with VF were admitted and seven (13%) were discharged (p less than .75). AEDs are an effective alternative to manual defibrillation in small communities.

Entities:  

Mesh:

Year:  1986        PMID: 3512123     DOI: 10.1161/01.cir.73.4.701

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  13 in total

1.  Cost-effectiveness of automated external defibrillator deployment in selected public locations.

Authors:  Peter Cram; Sandeep Vijan; A Mark Fendrick
Journal:  J Gen Intern Med       Date:  2003-09       Impact factor: 5.128

2.  "Heartstart Scotland"--initial experience of a national scheme for out of hospital defibrillation.

Authors:  S M Cobbe; M J Redmond; J M Watson; J Hollingworth; D J Carrington
Journal:  BMJ       Date:  1991-06-22

3.  Emergency medicine: expanding the use of automatic external defibrillators to home and community.

Authors:  R O Cummins
Journal:  West J Med       Date:  1987-05

Review 4.  Ventricular defibrillation: the Belfast experience.

Authors:  G W Dalzell; S R Cunningham; C M Wilson; J D Allen; J Anderson; A A Adgey
Journal:  Br Heart J       Date:  1987-11

5.  Treatment of cardiac arrest by ambulance staff.

Authors:  G Walters; E Glucksman
Journal:  Br Med J (Clin Res Ed)       Date:  1988-01-23

6.  Automated external defibrillation.

Authors:  M C Colquhoun
Journal:  Br J Gen Pract       Date:  1993-03       Impact factor: 5.386

7.  Effect of reducing ambulance response times on deaths from out of hospital cardiac arrest: cohort study.

Authors:  J P Pell; J M Sirel; A K Marsden; I Ford; S M Cobbe
Journal:  BMJ       Date:  2001-06-09

8.  Cardiac arrest in Ontario: circumstances, community response, role of prehospital defibrillation and predictors of survival.

Authors:  R J Brison; J R Davidson; J F Dreyer; G Jones; J Maloney; D P Munkley; H M O'Connor; B H Rowe
Journal:  CMAJ       Date:  1992-07-15       Impact factor: 8.262

9.  Part 12: Education, implementation, and teams: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations.

Authors:  Jasmeet Soar; Mary E Mancini; Farhan Bhanji; John E Billi; Jennifer Dennett; Judith Finn; Matthew Huei-Ming Ma; Gavin D Perkins; David L Rodgers; Mary Fran Hazinski; Ian Jacobs; Peter T Morley
Journal:  Resuscitation       Date:  2010-10       Impact factor: 5.262

10.  Mortality within hospital after resuscitation from ventricular fibrillation outside hospital.

Authors:  W Dickey; A A Adgey
Journal:  Br Heart J       Date:  1992-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.