| Literature DB >> 35121143 |
Sanjeev S Mukherjee1, Debabrata Bera2, Ayan Kar2.
Abstract
Implantable cardioverter defibrillator (ICD) shocks are always a matter of anxiety and concern for both patient and clinician. We report a case of ischemic cardiomyopathy who reported to us after receiving two shocks within 48 hours. The therapy was inappropriate as noted from device logs. EMI was identified as the incriminating cause. We discuss the reasons why he received the therapy despite active 'Noise reversion' and 'SecureSense' algorithms to avoid noise /EMI related oversensing.Entities:
Keywords: Defibrillator; ICD; Inappropriate shock; Physiotherapy machine
Year: 2022 PMID: 35121143 PMCID: PMC8981156 DOI: 10.1016/j.ipej.2022.01.006
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Fig. 112 lead ECG showing ventricular paced rhythm with wide fragmented QRS fitting with ischemic cardiomyopathy.
Fig. 2AStored electrograms showing noise in both channels. The discrimination (FF) channel shows more distinct EGM’S.
Fig. 2BStored EGM shows treated episode in the VF zone. It shows oversensing of noise which stops after 15J shock and return of sensed rhythm.
Fig. 3After the paced/sensed R wave ‘threshold starts’ at programmed decay delay. 1) threshold start (sensed 50% R wave/1 mV if paced) 2) If decay delay programmed 0 ms (hypothetical) 3) decay delay at 95 ms (this case).