| Literature DB >> 35090007 |
Pier D Lambiase1, Dominic A Theuns2, Francis Murgatroyd3, Craig Barr4, Lars Eckardt5, Petr Neuzil6, Marcoen Scholten7, Margaret Hood8, Jȕrgen Kuschyk9, Amy J Brisben10, Nathan Carter10, Timothy M Stivland10, Reinoud Knops11, Lucas V A Boersma11,12.
Abstract
AIMS: To report 5-year outcomes of EFFORTLESS registry patients with early generation subcutaneous implantable cardioverter-defibrillator (S-ICD) devices. METHODS ANDEntities:
Keywords: Implantable cardioverter-defibrillator; Primary prevention; Secondary prevention; Subcutaneous ICD; Sudden death
Mesh:
Year: 2022 PMID: 35090007 PMCID: PMC9156377 DOI: 10.1093/eurheartj/ehab921
Source DB: PubMed Journal: Eur Heart J ISSN: 0195-668X Impact factor: 35.855
Complications, overall, year 1, and years 2–5
| Complication term | Overall ( | Year 1 ( | Years 2–5, all ( | Years 2–5, complication in year 1 ( | Years 2–5, no complication in year 1 ( | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Events | Subjects | Events | Subjects | Events | Subjects | Events | Subjects | Events | Subjects | |
|
|
|
|
|
| ||||||
| Infection requiring device removal | 34 | 31 (3.2) | 26 | 25 (2.5) | 8 | 8 (0.9) | 2 | 2 (3.6) | 6 | 6 (0.7) |
| Erosion | 23 | 23 (2.3) | 10 | 10 (1.0) | 13 | 13 (1.4) | 4 | 4 (7.1) | 9 | 9 (1.1) |
| Inappropriate shock requiring intervention: cardiac oversensing | 17 | 16 (1.6) | 5 | 5 (0.5) | 12 | 11 (1.2) | 0 | 0 (0.0) | 12 | 11 (1.3) |
| Other procedural complications | 12 | 12 (1.2) | 7 | 7 (0.7) | 5 | 5 (0.5) | 0 | 0 (0.0) | 5 | 5 (0.6) |
| Discomfort | 11 | 11 (1.1) | 3 | 3 (0.3) | 8 | 8 (0.9) | 2 | 2 (3.6) | 6 | 6 (0.7) |
| Haematoma | 9 | 9 (0.9) | 8 | 8 (0.8) | 1 | 1 (0.1) | 0 | 0 (0.0) | 1 | 1 (0.1) |
| PG movement | 8 | 6 (0.6) | 5 | 3 (0.3) | 3 | 3 (0.3) | 0 | 0 (0.0) | 3 | 3 (0.4) |
| Premature cell battery depletion | 8 | 8 (0.8) | 2 | 2 (0.2) | 6 | 6 (0.7) | 0 | 0 (0.0) | 6 | 6 (0.7) |
| Sub-optimal electrode position | 8 | 8 (0.8) | 8 | 8 (0.8) | 0 | 0 (0.0) | 0 | 0 (0.0) | 0 | 0 (0.0) |
| Electrode movement | 7 | 7 (0.7) | 6 | 6 (0.6) | 1 | 1 (0.1) | 0 | 0 (0.0) | 1 | 1 (0.1) |
| Incision/superficial infection | 6 | 6 (0.6) | 5 | 5 (0.5) | 1 | 1 (0.1) | 0 | 0 (0.0) | 1 | 1 (0.1) |
| Unable to convert: during procedure | 5 | 5 (0.5) | 4 | 4 (0.4) | 1 | 1 (0.1) | 0 | 0 (0.0) | 1 | 1 (0.1) |
| Inappropriate shock requiring intervention: non-cardiac oversensing | 4 | 4 (0.4) | 2 | 2 (0.2) | 2 | 2 (0.2) | 0 | 0 (0.0) | 2 | 2 (0.2) |
| Inappropriate shock requiring intervention: SVT above discrimination zone (normal device function) | 4 | 4 (0.4) | 1 | 1 (0.1) | 3 | 3 (0.3) | 0 | 0 (0.0) | 3 | 3 (0.4) |
| Other technical complications | 4 | 4 (0.4) | 1 | 1 (0.1) | 3 | 3 (0.3) | 0 | 0 (0.0) | 3 | 3 (0.4) |
| Inability to communicate with the device | 3 | 3 (0.3) | 0 | 0 (0.0) | 3 | 3 (0.3) | 0 | 0 (0.0) | 3 | 3 (0.4) |
| Sub-optimal PG and electrode position | 3 | 3 (0.3) | 3 | 3 (0.3) | 0 | 0 (0.0) | 0 | 0 (0.0) | 0 | 0 (0.0) |
| Sub-optimal PG position | 1 | 1 (0.1) | 1 | 1 (0.1) | 0 | 0 (0.0) | 0 | 0 (0.0) | 0 | 0 (0.0) |
| Total | 167 | 141 (14.3) | 97 | 87 (8.8) | 70 | 60 (6.6) | 8 | 6 (10.7) | 62 | 54 (6.3) |
PG, pulse generator; SVT, supraventricular tachycardia.