Elena N Iomdina1, Nina Yu Kushnarevich2.
Abstract
PURPOSE: To compare the effectiveness of transpalpebral scleral tonometry (TPST) and corneal pneumotonometry in children, and assess the discomfort level when measuring intraocular pressure (IOP) by these methods.Entities:
Keywords: Children; Discomfort level score; Intraocular pressure; Noncontact pneumotonometry; Transpalpebral scleral tonometry
Mesh:
Year: 2022 PMID: 35088357 PMCID: PMC9122879 DOI: 10.1007/s10792-021-02158-5
Source DB: PubMed Journal: Int Ophthalmol ISSN: 0165-5701 Impact factor: 2.029
Fig. 1Measurement of the IOP in a child using an EASYTON transpalpebral tonometer
The IOP values (mm Hg) for the whole group (42 children, 84 eyes), as well as for the right (OD) and left (OS) eyes individually, obtained using an EASYTON and a Reichert 7 AutoTonometer
| EASYTON 18.3 ± 2.3 (± 0.3)* | Reichert 7 AutoTonometer 17.1 ± 3.9 (± 0.4) | ||
|---|---|---|---|
| OD | OS | OD | OS |
| 18.1 ± 2.2 (± 0.3)† | 18.2 ± 2.3 (± 0.3) | 17.0 ± 3.9 (± 0.6)†† | 17.2 ± 3.8 (± 0.6) |
*Difference with the corresponding indicator of pneumotonometry is significant, p = 0.02
†Difference with the corresponding indicator of the fellow eyes is not significant, p = 0.83
††Difference with the corresponding indicator of the fellow eyes is not significant, p = 0.81
IOP values (mm Hg) in myopic children (33 children, 64 eyes) across the whole myopic group, as well as for the right (OD) and left (OS) eyes individually, obtained using an EASYTON and a Reichert 7 AutoTonometer
| EASYTON 18.2 ± 2.3 (± 0.3)* | Reichert 7 AutoTonometer 16.9 ± 3.8 (± 0.5) | ||
|---|---|---|---|
| OD | OS | OD | OS |
| 18.1 ± 2.3 (± 0.4)† | 18.3 ± 2.2 (± 0.3) | 16.7 ± 3.9 (± 0.7) †† | 17.1 ± 3.8 (± 0.7) |
*Difference with the corresponding indicator of pneumotonometry is significant, p = 0.04
†Difference with the corresponding indicator of the fellow eyes is not significant, p = 0.69
††Difference with the corresponding indicator of the fellow eyes is not significant, p = 0.68
IOP values (mm Hg) in children with hyperopia (9 children, 18 eyes) across the whole hyperopic group, as well as for the right (OD) and left (OS) eyes individually, obtained using an EASYTON and a Reichert 7 AutoTonometer
| EASYTON 18.5 ± 2.3 (± 0.8)† | Reichert 7 AutoTonometer 18.2 ± 4.0 (± 1.0) | ||
|---|---|---|---|
| OD | OS | OD | OS |
| 18.5 ± 2.1 (± 0.7)†† | 18.6 ± 2.4 (± 0.9) | 18.4 ± 4.0 (± 1.4)††† | 18.0 ± 4.3 (± 1.5) |
†Difference with the corresponding indicator of pneumotonometry is not significant, p = 0.82
††Difference with the corresponding indicator of the fellow eyes is not significant p = 0.92
†††Difference with the corresponding indicator of the fellow eyes is not significant p = 0.84