Marta Saccher1, Shinnosuke Kawasaki2, Martina Proietti Onori3, Geeske M van Woerden3,4, Vasiliki Giagka2,5, Ronald Dekker2,6.
Abstract
BACKGROUND: Microelectrode arrays (MEA) enable the measurement and stimulation of the electrical activity of cultured cells. The integration of other neuromodulation methods will significantly enhance the application range of MEAs to study their effects on neurons. A neuromodulation method that is recently gaining more attention is focused ultrasound neuromodulation (FUS), which has the potential to treat neurological disorders reversibly and precisely.Entities:
Keywords: Bioelectronics; Focused ultrasound; Multielectrode arrays; Neuromodulation; Ultrasound
Year: 2022 PMID: 35081966 PMCID: PMC8793260 DOI: 10.1186/s42234-021-00083-7
Source DB: PubMed Journal: Bioelectron Med ISSN: 2332-8886
Fig. 1Experiment features. A) Microscope view of MEA electrodes. In the centre the 12 recording electrodes and at the sides, the 4 reference electrodes. B) Sonication protocol. C) Spike train features
Fig. 2Experimental setup. A) 3D model of the experimental setup. The black box at the bottom is the MEA headstage, at the top of which the custom-made lid is placed. The circular holes in the lid are aligned to the underlying MEA plate. The ultrasound transducer is inserted in the transducer holder structure which can then be positioned in the engraved alignment holes for precise centering with respect to the MEA well. B) Cross-section view of the experimental setup showing the relative position of the transducer with respect to the neuronal culture. C) Detail picture of the experimental setup showing the transducer in the support structure on the left and on the right, the wells with the coloured cap are used to record the trigger signal. D) Schematic view of the complete experimental setup showing the connections between the various elements
COMSOL parameters used for the hippocampal neurons
| Parameter | Value |
|---|---|
| Speed of sound | 1500 [m/s] |
| Density | 1045 [kg/m 3] |
| Thermal conductivity | 0.55 [W/m/K] |
| Heat capacity | 3696 [J/kg/K] |
| Shear modulus | 2 [kPa] |
Fig. 3Schlieren setup. A) View of the complete setup. B) Result of ultrasound field simulation in COMSOL. C) Interference pattern generated by the reflection of the ultrasound field on a hard surface. The distance between the ultrasound transducer and the reflecting surface is the same as the distance between the transducer and the bottom of the MEA well (6 mm). D) Heat profile generated by the ultrasound transducer at intensity 4 times higher than the experimental conditions. The shape of the thermocouple is also visible
Fig. 4Ultrasound neuromodulation results. A) Example of raster plot of one of the wells used in the experiment, for a treatment duration of 15 seconds. Dark green shaded area indicates the US-ON period, while stripes-patterned area indicates the US-OFF period, namely the post period. B) Same as A) but for a treatment duration of 60 seconds. Dark green shaded area indicates the first 15 seconds of US-ON period and light green area indicates the remaining 45 seconds. Stripes-patterned area indicates the US-OFF period. C) Spike rate trend corresponding to the raster plot in A), calculated for 15 seconds bins over the duration of the experiment. Dark green shaded area indicates the US-ON period and stripe-patterned area indicates the US-OFF period. D) Spike rate trend corresponding to the raster plot in B), calculated for 15 seconds bins over the duration of the experiment. Dark green shaded area indicates the first 15 seconds of US-ON and light green area the remaining 45 seconds. Stripe-patterned area indicates the period of US-OFF