Jugal Suthar1,2, Beatriz Prieto-Simon3,4, Gareth R Williams1, Stefan Guldin2. 1. UCL School of Pharmacy, University College London, 29-39 Brunswick Square, Bloomsbury, London, WC1N 1AX, United Kingdom. 2. Department of Chemical Engineering, University College London, Torrington Place, London, WC1E 7JE, United Kingdom. 3. Department of Electronic Engineering, Universitat Rovira i Virgili, 43007, Tarragona, Spain. 4. ICREA, Pg. Lluis Companys 23, 08010, Barcelona, Spain.
Abstract
The biomolecular contents of extracellular vesicles, such as exosomes, have been shown to be crucial in intercellular communication and disease propagation. As a result, there has been a recent surge in the exploration of novel biosensing platforms that can sensitively and specifically detect exosomal content such as proteins and nucleic acids, with a view toward application in diagnostic assays. Here, we demonstrate dual-mode and label-free detection of plasma exosomes using an electrochemical quartz crystal microbalance with dissipation monitoring (EQCM-D). The platform adopts a direct immunosensing approach to effectively capture exosomes via their surface protein expression of CD63. By combining QCM-D with a tandem in situ electrochemical impedance spectroscopy measurement, we are able to demonstrate relationships between mass, viscoelasticity and impedance inducing properties of each functional layer and analyte. In addition to lowering the limit of detection (by a factor of 2-4) to 6.71 × 107 exosome-sized particles (ESP) per mL in 25% v/v serum, the synergy between dissipation and impedance response introduces improved sensing specificity by offering further distinction between soft and rigid analytes, thereby promoting EQCM-D as an important technique for exosome analysis.
The biomolecular contents of extracellular vesicles, such as exosomes, have been shown to be crucial in intercellular communication and disease propagation. As a result, there has been a recent surge in the exploration of novel biosensing platforms that can sensitively and specifically detect exosomal content such as proteins and nucleic acids, with a view toward application in diagnostic assays. Here, we demonstrate dual-mode and label-free detection of plasma exosomes using an electrochemical quartz crystal microbalance with dissipation monitoring (EQCM-D). The platform adopts a direct immunosensing approach to effectively capture exosomes via their surface protein expression of CD63. By combining QCM-D with a tandem in situ electrochemical impedance spectroscopy measurement, we are able to demonstrate relationships between mass, viscoelasticity and impedance inducing properties of each functional layer and analyte. In addition to lowering the limit of detection (by a factor of 2-4) to 6.71 × 107 exosome-sized particles (ESP) per mL in 25% v/v serum, the synergy between dissipation and impedance response introduces improved sensing specificity by offering further distinction between soft and rigid analytes, thereby promoting EQCM-D as an important technique for exosome analysis.
The search
and application of
novel biosensing techniques for the detection of exosomal analytes
(such as proteins or microRNA) has grown rapidly in recent years.
This is fuelled by the discovery that these biomolecules play an essential
role in disease progression,[1,2] but also as potential
biomarkers for early disease diagnosis.[3,4] Minimally invasive
liquid biopsies will become a viable prospect if a suitable analytical
platform can be coupled with these vesicular sources of rich biomolecular
information. In this light, groups have explored a vast array of advanced
characterization techniques, including those principled on fluorescence,[5−7] absorbance,[8−10] electrochemistry,[11,12] plasmon resonance,[13,14] surface enhanced Raman spectroscopy,[15−17] and, more recently,
acoustic resonance in the form of a quartz crystal microbalance with
dissipation monitoring (QCM-D).[18−20]QCM-D is able to successfully
distinguish between exosomes and
non-exosomal particles based upon their viscoelastic properties as
well as their to their mass.[18] Not only
does this offer an additional discriminatory mode of measurement,
it exploits immuno-capture principles to specifically detect exosomes
based upon expression of the surface protein CD63. Nonetheless, the
platform’s limit of detection (LOD) was significantly poorer
than competing approaches. One method of improving platform sensitivity
is to sense exosomes through an immunosensor transduced via a complementary
mode of measurement in tandem with the QCM-D process. An alternative
method of detection could also present as a more holistic biosensor,
presenting richer insights on fundamental binding phenomena, while
providing a point of comparison against the frequency and dissipation
outputs from the acoustic approach. It is common to see QCM-D analysis
conducted alongside surface plasmon resonance (SPR) spectroscopy,
with the rationale being that QCM-D offers cosolvated mass of bound
adsorbates while plasmon resonance informs on dry mass, allowing for
solvation fraction determination.[21−25] Recently, a viable route toward dry mass determination
by QCM-D has been established via kinematic-viscosity matching.[27] Another notable synergistic technique is total-internal-reflection-fluorescence
microscopy.[26] Combining QCM-D with in situ ellipsometry measurements supports
similar conclusions in addition to greater understanding of the interfacial
optical properties.[28−31]An informative approach, particularly in the field of biosensing,
is the combination of electrochemical measurements in conjunction
with QCM-D, collectively termed EQCM-D.[32−34] The technique makes
use of a gold coated QCM sensor which also functions as a working
electrode (WE) within an electrochemical cell with a conventional
three electrode setup. This arrangement enables the conduct of routine
electrochemical measurements, such as electrochemical impedance spectroscopy
(EIS), cyclic, differential pulse, and square wave voltammetry, while
also capturing frequency and dissipation responses simultaneously,
in real-time.[35] The overall result should
be an improvement to EIS assays, due to unique information on bound
mass, interfacial structural changes, and binding kinetics offered
by QCM-D. Similarly, QCM-D is benefited by a greater understanding
of how distinct layers alter electron transfer processes at nanomolar
sensitivity via EIS.Traditionally, EQCM-D has been a popular
technique for the characterization
of electrochemical energy storage and conversion.[36,37] In a study with clinical relevance, an EQCM cytosensor detected
cell surface sialic acid, combining in situ cyclic voltammetry analysis
with QCM derived frequency response.[38] The
approach successfully differentiated sialic acid expressed between
normal and diabetic patient cells. EQCM-D was showcased as part of
a biosensor array for point-of-care detection of dengue fever by targeting
the NS1 antigen, with a LOD as low as 10 ng mL–1.[39] Gao et al. employed cyclic voltammetry
and acoustic frequency measurements in combination with an immunosensor
platform to successfully detect inflammation marker C-reactive protein
with a LOD of 0.02 μg mL–1. This represented
a significant amplification in response compared to an ELISA technique.[40] Srivastava et al. devised an EQCM platform with
an imprinted graphene-starch nanocomposite matrix, using differential
pulse voltammetry and frequency for epinephrine detection.[41] These studies demonstrated the lower LOD that
differential pulse voltammetry measurements offer compared to QCM,
thereby improving the sensitivity of the platform overall.In
spite of these successful applications, EQCM-D has yet to be
explored for the detection of exosomes as a target source of analytes.
Recently, however, electrochemical assessments (either on their own
or in conjunction with other techniques) have become leading modes
of measurement in the exosome field, with many reports exhibiting
superior sensitivity toward exosomes than optical based methods of
biosensing. Doldan et al. employed amperometry for detecting surface
located exosomal proteins using an immunosensor.[42] Amperometry was also adopted for the iMEX device, which
successfully profiled exosomal membrane proteins, including CD63,
from cancer patients.[43] More recently,
Moura et al. applied amperometry for cancer-specific exosome markers
(CD24, CD44, CD54, CD326).[44] All three
methods use direct or indirect labeling of the exosomal protein, or
response amplification through enzymatic digestion. Amperometry was
also used in tandem with colorimetric detection, offering a multimodal
measurement of placental alkaline phosphatase (PLAP)-positive exosomes.[12] Square wave voltammetry was implemented as part
of an aptasensor platform for exosomal CD63 detection.[45] Here, a linear potential sweep using a combination
of a square wave and staircase potential was applied to an electrode,
producing a differential current plot as opposed to more traditional
anodic and cathodic peaks. An et al. implemented a differential pulse
voltammetry based aptasensor to also detect exosomal CD63 by measuring
the reduction in current upon exosome binding.[46] EIS based detection of CD81-positive exosomes was recently
reported, where changes in impedance were found to be proportional
to the exosome concentration, as the physical barrier to electron
transfer across the WE was increased.[47,48] EIS was also
combined with voltammetry as part of an aptasensor platform to sensitively
detect CD63-positive exosomes in human serum.[10] While these reports did not assess EIS performance in more complex
matrices, they lay the foundation for the implementation of EIS as
the electrochemical method of choice for the investigation reported
herein.In this work, a method for EQCM-D detection of CD63-positive
exosomes
is presented. This combines principles of QCM-D and EIS in an attempt
to conduct label-free electrochemical transduction of binding events,
that can be compared with changes in viscoelasticity and adsorbed
mass at the working electrode surface, offering advancements to both
modes of assay. Initially, this work determines the most appropriate
electrochemical model to fit against the EIS data. Once established,
comparisons are made between impedance spectroscopy, frequency and
dissipation outputs from the EQCM-D platform. This methodology looks
to support measurements for each stage of immunofunctionalisation
and immunosensing process (from sensor fabrication to analyte detection).
Findings help to understand the influence of molecular rigidity and
softness of different layers on EQCM-D response. Furthermore, the
sensitivity and specificity of the EIS technique are explored against
a range of exosome concentrations in both buffer and complex media.
Experimental
Section
Materials
Materials were sourced as follows: antimouse
detection module for a Western blot WES machine and 12–230
kDa WES separation modules were acquired from Protein Simple (Bio-Techne,
Minneapolis, MN). For isolation and sample preparation, qEV original
size-exclusion chromatography (SEC) columns (Izon Science, UK), 0.45
μm filters (Merck Millipore, U.S.), HEPES buffered saline (HBS,
0.01 M HEPES, pH 7.4, 0.15 M NaCl) (GE Healthcare Life Sciences, Sweden),
Amicon Ultra-15 centrifugal filters (Merck Millipore, U.S.), 100 nm
polystyrene beads (Thermofisher Scientific, UK) and RIPA buffer (Sigma-Aldrich,
St. Louis, MO) were used. Mouse monoclonal anti-Alix (634 502, Biolegend
UK), mouse monoclonal anti-CD63 (353 013, Biolegend UK), mouse monoclonal
anti-CD9 (31 202, Biolegend UK), mouse monoclonal biotinylated anti-CD63
(353 017, Biolegend UK) and biotin-IgG isotype control antibody (400
103, Biolegend UK) were acquired for Western blot and immunosensing
experiments. Ferrocyanide (>98.5%, Honeywell) and ferricyanide
(99%,
ACROS Organics) were employed as the redox couple for EIS measurements.
Human plasma and streptavidin were purchased from Sigma-Aldrich, St.
Louis, MO. 5-MHz gold coated QCM sensors were purchased from QuartzPro,
Sweden.
Exosome Isolation and Characterization
Size-Exclusion Chromatography
SEC was implemented as
the exosome isolation technique from human plasma. The source plasma
was first filtered with a 0.45 μm filter (Merck Millipore, U.S.).
Thirty mL of clarified plasma was subsequently concentrated using
Amicon Ultra-15 centrifugal filters with a 10 kDa pore size cutoff
(Merck Millipore, U.S.). The filters were spun at 4000g for 30 min at 4 °C. Postspin, 0.5 mL of concentrated filtrate
was loaded onto a qEV 35 nm SEC column (Izon Science, UK). 0.2 μm
filtered HEPES buffered saline (HBS, 0.01 M HEPES, pH 7.4, 0.15 M
NaCl) (GE Healthcare Life Sciences, Sweden) was used as the eluting
buffer at a flow rate of 1 mL min–1. Twenty 1 mL
fractions were collected and stored at −80 °C.[18]
Western Blot Analysis of Final Isolate
Validation of
the SEC protocol was conducted by verifying exosome presence through
Western blot analysis using capillary gel electrophoresis. Based upon
previously reported SEC isolation protocols, SEC fractions 4, 5, 6,
7 were considered for onward protein characterization for exosome
presence.[49,50] Exosomal proteins Alix (97 kDa), tetraspanin
CD63 (57 kDa) and CD9 (24 kDa) were probed by chemiluminescent immunoassay,
using mouse monoclonal anti-Alix, mouse monoclonal anti-CD63 and mouse
monoclonal anti-CD9 as primary antibodies. The Western blot run was
conducted as per the manufacturer’s instruction (see Supporting Information (SI) for details).
NTA
Analysis of SEC Fractions
Based upon the protein
identification from the Western blot, the sixth isolation fraction
was chosen for concentration and hydrodynamic size characterization
of particulates nanoparticle tracking analysis (NTA) with the Nanosight
LM10 instrument (Malvern Instruments, UK). The machine was calibrated
with 100 nm polystyrene beads (Thermofisher Scientific, UK) prior
to fraction assessment. Measurement specifications were as follows:
532 nm green laser, five videos per fraction, 60 s video length, shutter
speed of 25–32 ms, camera gain of 400, camera level 15, lower
threshold of 910 and higher threshold of 11 180. Captured videos
were processed using the NTA software version 3.2, a detection threshold
of 5, auto settings for blur, minimum track length, and minimum particle
size. Measurements were carried out in static mode at room temperature.[18]
EQCM-D Measurements
General Methods for EQCM-D
Apparatus Setup and Sample Preparation
All EIS measurements
reported herein were conducted using a Q-Sense
Electrochemistry Module from Biolin Scientific (Sweden), in tandem
with a Q-Sense Analyzer instrument and a Gamry (UK) Reference 600
Plus potentiostat. The system used gold coated QCM sensors as the
WE, a platinum counter electrode (CE) and Ag/AgCl reference electrode
(RE) as part of a conventional three-electrode system. Data was acquired
with the Gamry Instrument Framework (v7.07) software and analyzed
using Gamry Echem Analyst (v7.07) software. When fitting the chosen
circuit model to the captured data, parameters were chosen based on
well-defined physical processes.EIS experiments were all carried
out with a frequency scan range of 10–1 Hz to 105 Hz at a 5 mV AC amplitude. The detection area was set at
0.79 cm2. An equimolar solution of 5 mM of K3[Fe(CN)6]/K4[Fe(CN)6] in 0.1 M KCl
was used for all measurements. Modified Randles cell circuit models
served to fit against EIS data and an optimal model was selected.
Impedance was determined after the formation of each layer, following
the addition of electrolyte into the chamber. The flow of electrolyte
through the chamber was paused during measurement acquisition. EIS
were captured in tandem with QCM-D response in all instances.All QCM-D measurements were performed on a Q-Sense E4 instrument
(Biolin Scientific, Sweden). Analysis of frequency and dissipation
response was conducted using the QTools software, version 3.0.17.560
(Biolin Scientific, Sweden). Changes in resonance frequency (Δf) were recorded from the third, fifth, seventh, ninth and
11th overtones. The presented data relates to the fifth overtone,
with variation of Δf between overtones being
10% or less. In all instances, samples were degassed prior to exchange
in the QCM flow module and AT-cut, 5-MHz gold coated quartz crystal
sensors with a 0.79 cm2 active area (Biolin, Sweden) were
used.To ensure reproducibility of each process, all analytes
were prepared
using the same degassed stock solutions to minimize impact of buffer
properties during sample exchange in observed responses. These were
prepared to identical volumes (0.25 mL per sensor). All reagents were
sourced from the same suppliers throughout the study to avoid influences
of differing characteristics or quality. In all cases, the analyte
was flowed at 10 μL min–1 and a sensor was
reserved to monitor for drift induced by buffer exchange. Drift in
frequency and dissipation exceeding 1.5 Hz/h and 2 × 10–7/h, respectively, was to be deemed excessive; however, this was not
ever encountered. All results presented are corrected for background
noise as determined by responses on an isotype control functionalized
sensor. A HBS buffer rinse was performed after the addition of every
functional layer or analyte, with frequency and dissipation responses
reported net or post-HBS rinse, to remove the response bias of weakly
bound analytes and any solvent specific effects.[18] For the purposes of this study, baseline is defined as
the response achieved immediately prior to the addition of an analyte.
Sensor Functionalization
An affinity based immunosensing
approach was employed as reported by Suthar et al.[18] A 1 mM ethanolic solution of SH-PEG (2 kDa)-Biotin and
spacer molecule SH-OEG (800 Da)-COOH at a 1:9 mol/mol ratio was flowed
across the sensor surface at 7.5 μL min–1 overnight
to form a self-assembled monolayer (SAM). Subsequently, a 100 μg
mL–1 solution of streptavidin (SAv) was flowed across
the sensor surface at 10 μL min–1, followed
by a rinse step of HBS buffer at 80 μL min–1. Twenty μg mL–1 of mouse monoclonal biotinylated
anti-CD63 was immobilized on the surface at 10 μL min–1, followed by another rinse step and response stabilization for 30
min prior to sample addition.
EQCM-D Detection of CD63-Positive
Exosomes
The QCM-D
immunosensor performance toward spiked CD63-positive exosomes was
assessed in tandem with electrochemical measurement (EQCM-D). Sensitivity
toward CD63-positive exosomes in HBS buffer and 25% v/v serum was
tested using titrated concentrations of ESPs verified by NTA. The
following concentrations were assessed: 5 × 107, 7.5
× 107, 1 × 108, 2.5 × 108, 5 × 108, 7.5 × 108, 1 ×
109, and 2.5 × 109 ESPs mL–1. To complete our sensitivity and specificity assessments, these
samples were applied to control sensors that employed a biotin-IgG
isotype control antibody in place of biotin-anti-CD63. This helped
to determine the signal-to-noise ratio (SNR), statistical significance,
solvent effects and noise in our setup.In this work, LOD is
defined as the concentration eliciting a SNR of 3 as per recommendation
by Shrivastava et al., with SNR being calculated as a response ratio
of the target sensor and control sensor.[51] For the purposes of this study, dynamic range is defined as the
range of concentrations between the limit of detection (LOD) and the
concentration where the saturation commences (nonlinearity appears).
Results and discussion
Advancing the detection of plasma
derived exosomes is imperative
to enable their full exploitation in diagnostic settings. We looked
to expand the exosome analytical toolkit by employing a dual-mode
method of analysis that combines in situ EIS measurements with QCM-D.
The result is an immunosensor which was transduced through both bulk
acoustic wave and electrochemical principles. A summary of the experimental
work flow is displayed in Figure .
Figure 1
Overview of the adopted experimental approach. CD63-positive
exosomes
were isolated from human plasma using SEC, prior to exosome verification
by nanoparticle tracking analysis, Western blot and subsequent biosensing
via the EQCM-D platform with the instrumental setup shown. CE: counter
electrode, RE: reference electrode, WE: working electrode, EIS: electrochemical
impedance spectroscopy, EQCM-D: electrochemical QCM-D, NTA: nanoparticle
tracking analysis.
Overview of the adopted experimental approach. CD63-positive
exosomes
were isolated from human plasma using SEC, prior to exosome verification
by nanoparticle tracking analysis, Western blot and subsequent biosensing
via the EQCM-D platform with the instrumental setup shown. CE: counter
electrode, RE: reference electrode, WE: working electrode, EIS: electrochemical
impedance spectroscopy, EQCM-D: electrochemical QCM-D, NTA: nanoparticle
tracking analysis.
SEC Isolation of Plasma
Exosomes
Exosome enriched proteins
such as cytosolic Alix, transmembrane CD63 and CD9 were probed using
Western blot in a capillary format (SI Figure S1A). Alix and CD9 were seen to be prominently expressed in
particles present in fractions 5 and 6, with smaller concentrations
of CD9 also seen in fractions 4 and 7. CD63 was solely detected in
fractions 6 and 7. As all three exosomal markers were present in fraction
6, this isolate was selected for downstream characterization and biosensing
due to favorable protein expression. These results underlined the
heterogeneity of exosome composition, which is further influenced
by the source media, where differing parent cells incite variable
marker expression in relation to their size.[52−54] NTA analysis
of fraction 6 found over 81% of particles falling within the exosomal
size range, with a mean size of 127 nm (SI Figure S1B). Collectively, this data confirms that plasma exosomes
expressing the target CD63 protein were successfully isolated with
the applied SEC protocol.[55−57] It also ensures the analyte possesses
the molecular composition required for the EQCM-D immunosensing approach,
which was based upon surface capture of CD63.
EIS Characterization of
Immunosensor Fabrication
To
reliably determine the resistance to charge transfer (Rct) from the EIS data, an appropriate circuit model was
required. SI Figure S2A displays the two
different circuit models that were evaluated. The bounded Warburg
model and constant phase element (CPE) with diffusion model both account
for key components of the electrochemical system. First, as a double-layer
capacitance is created at the interface between the conductive gold
WE and the adjacent liquid electrolyte, a CPE component was introduced
to account for the imperfect capacitor that is the dielectric layer.[58] Second, the impedance of electron transfer at
low frequencies is increased by the long distance of electron diffusion
through the bulk phase. This is especially relevant to the system
in question due to the large size of bound ESPs, thus it is addressed
by the inclusion of a Warburg element. The key difference between
the two models is that the bounded Warburg model assumes two time
constants, compared to the single time constant of the CPE with diffusion
model. The incorporation of two time constants assumes a greater complexity
in the dielectric features at the WE surface; that is, it accounts
for the varying degree of capacitance that is potentially introduced
by multiple sensing layers or layers with high heterogeneity in thickness
and coverage.EIS data captured after the formation of each
detection layer were fitted with both models (SI Figure S2B–E). The goodness of fits displayed in SI Table S2 were found to be largely similar
between the models for these sensing layers, considering an alpha-level
of 1%. However, some advantage was seen with the bounded Warburg model
when applied to impedance spectra acquired post-SAv addition (SI Figure S2E), as indicated by the inferior
goodness of fit value. Overall, this is expected, as layers comprised
of homogeneous small molecules (<15 nm) and that are known to incur
small changes in dissipation, such as the mixed-SAM, SAv and antibody,
do not induce significant changes in capacitance or increases in diffusion
related impedance, thus presenting similar impedance changes between
models (Figure ).
Displayed net differences refer to the determined Rct of the two models.
Figure 2
EIS characterization of immunosensor fabrication.
(A) Comparing
net change in Rct for each layer attained
from both cell models. Schematic of sensor functionalization approach
shown (not to scale). (B) EIS Nyquist plots for each sensing layer
fitted with the bounded Warburg model. (C) Bar graph showing mean
net change in Rct for each layer from
(B). Standard deviation determined from three independent experiments.
EIS characterization of immunosensor fabrication.
(A) Comparing
net change in Rct for each layer attained
from both cell models. Schematic of sensor functionalization approach
shown (not to scale). (B) EIS Nyquist plots for each sensing layer
fitted with the bounded Warburg model. (C) Bar graph showing mean
net change in Rct for each layer from
(B). Standard deviation determined from three independent experiments.Figure B illustrates
how impedance incrementally increased upon the addition of each layer,
demonstrated by the increasing diameter of the Nyquist plot semicircle.
As the biomolecular layering at the WE surface changes during the
sensing process, this increased the barrier and therefore resistance
to charge transfer across WE for current generation. This change in
charge transfer relates to the raised impedance of the overall system.
Similar impedance increases upon immunosensor fabrication have been
reported in literature.[59−61] It is interesting to note the
extent to which Rct increased for each
sensing layer (Figure C). Here, the net changes in resistance is the increase in Rct upon analyte addition from the previous (pre-analyte)
baseline. While the bare gold sensor displays minimal charge resistance,
the mixed-SAM increased the impedance of the system by ∼0.6
kOhm, followed by the anti-CD63 antibody (∼0.3 kOhm) and SAv
(∼0.1 kOhm).A possible explanation for the high impedance
induced by the PEG-SAM
is that it produced a homogeneous coverage across the entire WE surface,
while also assembling with high order and packing density, thereby
forming a complete barrier to electron transfer.[62,63] Furthermore, PEG at the WE surface has been shown to increase the
dielectric capacitance, likely contributing to the large increase
in impedance.[64,65] Another consideration is that
the mixed-SAM was formed directly on the WE gold surface. Therefore,
another reason for the large increase in Rct is that the impedance of a double-layer capacitor was more sensitive
to electrical potential when changes occur nearer to the WE surface,
as it disrupts the ion dense region but also reduces ion permeability.
Despite the SAv and antibody layer increasing impedance, they do so
to a lesser extent than has been in the case in other studies.[66,67] This could be a result of layers which are not homogeneously arranged
and instead present as a layer of discrete particles, with a suboptimal
surface coverage or an incomplete layer. In consequence, certain regions
may provide easier access to the underlying SAM for passing ions,
lowering the mean Rct, a notion also advocated
by Pali et al.[68]Crucially, however,
a divergence of ∼1 kOhm was seen between
the CPE with diffusion model compared to the bounded Warburg model
upon application of an ESP solution to the WE surface (SI Figures S2F and Figure A), resulting in a significant difference
in goodness of fit (SI Table S2). The introduction
of heterogeneous particles to the surface, up to 150 nm in size and
filled with cytosolic fluid, is likely to have disrupted the dielectric
capacitance at the WE, increased the diffusion length and raise the
barrier to electron transfer across the WE (also known as tunnelling
distance), thereby exaggerating the impedance (in excess of 8 kOhm)
at low frequencies compared to the functional layers formed prior
(Figure A). The overestimation
by the CPE with diffusion model could be attributed to the model not
sufficiently considering the complexity of capacitance at the surface
caused by highly heterogeneous exosome layer, both in terms of coverage
and individual size, thus the bounded Warburg model was selected as
the optimal approach moving forward for Rct determination. EIS data displayed herein shall showcase Rct1 as the more sensitive parameter.
Tandem
EQCM-D Detection of CD63-Positive Exosomes
The
EIS characterization of the sensor fabrication and exosome detection
process was acquired in tandem with QCM-D monitoring within a single
instrumental setup. Example EIS and QCM-D profiles captured on the
same WE are displayed in Figure . It is apparent that an increase in impedance coincides
with a decrease in frequency upon analyte addition to the sensor surface.
The addition of sensor mass upon adsorbate binding reduced the oscillatory
frequency of the sensor, while also providing a barrier to electron
transfer. Capture of exosomes to the WE resulted in a significant
increase in dissipation, due to the raised friction upon oscillation
in the adlayer. Moreover, the energy storage and loss of the viscoelastic
ESPs amplified the oscillatory decay, in line with results previously
reported.[18] This viscoelastic layer also
resulted in the largest increase in system impedance (∼5.5
kOhm). While the large ESP mass was likely a prominent contributor
to the EIS response, the size of the particles and their dissipation
display a stronger complementarity with the Rct.
Figure 3
Example of tandem EIS and QCM-D information acquisition. (A) EIS
response across entire sensing process (fabrication to ESP detection)
captured simultaneously alongside, (B) QCM-D response profile on the
same WE for SAM formation (from ethanolic solution) and the addition
of (C) SAv, biotin-anti CD63 and 7.5 × 108 ESPS mL–1 in HBS buffer. Vertical dashed lines represent the
start of respective sample uptake by the peristaltic pump.
Example of tandem EIS and QCM-D information acquisition. (A) EIS
response across entire sensing process (fabrication to ESP detection)
captured simultaneously alongside, (B) QCM-D response profile on the
same WE for SAM formation (from ethanolic solution) and the addition
of (C) SAv, biotin-anti CD63 and 7.5 × 108 ESPS mL–1 in HBS buffer. Vertical dashed lines represent the
start of respective sample uptake by the peristaltic pump.Response comparison during immunosensor fabrication between
the
three modes of measurement (EIS, frequency and dissipation) is displayed
as a proportional change in Figure . Responses were normalized to a baseline before the
addition of analyte. This is common practice in QCM-D and EIS analysis
to account for intersensor variability. Furthermore, it supports direct
comparison in responsiveness of the techniques. It is clear that of
the three modes of measurement, QCM-D derived dissipation exhibited
superior magnitude of response across all three sensing layers, with
significant increases for dissipative layers (SAMs and antibodies)
and decreases for non-dissipative layers (SAv). Layers that increased
dissipation also seemed to elicit higher proportional responses in
impedance rather than frequency. Conversely, SAv which was found to
decrease dissipation during oscillatory decay, induced a relatively
smaller impedance increase and far lower than QCM derived frequency.
These findings further support the proposal that dissipative layers
elicit higher impedance at the working electrode surface.
Figure 4
Normalized
response comparison upon sensing layer addition between
EIS, frequency and dissipation modes of analysis. Responses are taken
as a ratio of change from the baseline prior to analyte addition.
Standard deviation is determined from three independent experiments.
Normalized
response comparison upon sensing layer addition between
EIS, frequency and dissipation modes of analysis. Responses are taken
as a ratio of change from the baseline prior to analyte addition.
Standard deviation is determined from three independent experiments.
Sensitivity, LOD and Specificity of EQCM-D
Toward CD63-Positive
Exosomes
After characterizing EQCM-D responses for the immunosensor
fabrication process, the impact of titrated concentrations of ESPs
in HBS buffer and 25% v/v serum was explored (SI Figure S3 and Figure , respectively). SI Figure S3A and Figure A present
the corresponding Nyquist plots that show increasing half-circle
diameters as ESP concentration increases. The increase in Rct is expected, as higher concentrations of
ESPs resulted in a greater coverage of the WE surface with particles,
which are large and dissipative in nature, raising the tunnelling
distance and also the impedance to electron flow for eventual polarization.
This was similarly reported by Kilic et al. where increasing concentrations
of EVs resulted in Rct increments.[47] A second phase in the form of an upward inflection
was seen at lower frequencies for a few of the impedance profiles.
This was the effect of Warburg or diffusion related impedance induced
by the formation of an exosome layer at the surface, that forced reactants
to diffuse greater distances to the electrode surface.[69] This phenomenon was not apparent across all
concentrations, suggesting some heterogeneity of exosome layer coverage
between sensors.
Figure 5
EQCM-D performance against varying concentrations of ESPs
spiked
in 25% v/v serum. (A) Nyquist plots representing EIS response. (B)
Normalized response changes across EIS, frequency and dissipation.
(C) Change in system capacitance as exosome concentration increases.
EQCM-D performance against varying concentrations of ESPs
spiked
in 25% v/v serum. (A) Nyquist plots representing EIS response. (B)
Normalized response changes across EIS, frequency and dissipation.
(C) Change in system capacitance as exosome concentration increases.Response to these concentrations were compared
between EIS, frequency
and dissipation via normalized changes from the baseline prior to
ESP addition (SI Figures S3B and Figure B). A steady increase
in response is seen for all measurements along with ESP concentration.
At lower ESP concentrations, EIS was seen to be the most responsive
of the three techniques. Upon increasing the ESP content, dissipation
became more responsive than frequency prior to plateauing along with
EIS at the highest concentrations as the detection reached saturation.
Once again, we see a correlation between the highly dissipative ESP
analyte and a pronounced response in electrochemical impedance, in
contrast to a comparatively small frequency change. Furthermore, EIS
was seen to be more responsive than dissipation for the highly viscoelastic
ESPs, compared to the smaller antibody and SAM adsorbates in Figure , where dissipation
exhibited the stronger relative responses. It can be argued that the
increased electron diffusion path length of the larger sized ESPs
has a greater influence on impedance than their acoustic dissipation,
compared to the SAM and antibody molecules.Nonetheless, frequency
response was seen to increase slightly along
with ESP concentration, which could be accounted for by the fact that
frequency is a measure of the cosolvated mass. The solvent coupling
includes the liquid in the interstitial spaces between adsorbed ESPs.[70] The amount of liquid that is trapped is influenced
by a few factors. Generally, if the higher coverage of ESPs was creating
a homogeneous layer at the sensor surface, it would result in a displacement
of entrapped liquid, thus reducing solvent contribution to frequency
response.[71] However, it is likely that
even at high ESP concentrations, the surface coverage is incomplete,
making it important to consider the hydrodynamically trapped liquid
coats surrounding each immobilized particle. It is also known that
higher liquid content of the particle results in higher solvent entrapment
within the particle layer, which is clearly applicable to the cytosol
filled exosomes.[72] Furthermore, solvent
association is seen to be higher for layers comprised of particles
with heterogeneous height-to-width ratios. Exosomes already possess
variable sizes and are assumed to deform at the WE surface giving
low height-to-width ratios, thereby supporting the feasibility of
this theory.[13] Lastly, entrapment is shown
to be prevalent where particles have a disparate distribution/lateral
organization.[71] The combination of liquid
filled particles, particle deformation and heterogeneity in distribution
may all contribute to stronger hydrodynamic effects at higher ESP
concentrations, leading to the more pronounced frequency response.
The increased trapping of ions within this layer may also be the underlying
reason behind the increase in system dielectric capacitance as exosome
concentration increases (Figure C). Alternatively, the relative plateauing in dissipation
response at the highest concentrations could be due to a crowding
effect of ESPs at the sensor surface, which reduces the hydrodynamic
interaction between particles and inhibits the rocking and translational
movements of the bound ESPs where dissipation normally occurs.[72]Moreover, no significant matrix effects
were observed when comparing
the performance of the EQCM-D platform to titrated ESP concentrations
between HBS buffer and more complex media (25% v/v serum), with the
differences in linear response gradient of <10%. This is despite
the likely presence of nonspecific particles (proteins, lipoproteins
and lipids) that could alter the detection and affect the relative
response between the techniques. Figure B shows that the introduction of serum into
the buffer caused an overall increase in the system dissipation compared
to buffer alone (SI Figure S3B), likely
due to a small degree of nonspecific binding and the sensitivity to
an increase in running buffer viscosity. Crucially, however, the relative
outputs from both EIS and QCM-D remain unaffected by non-exosomal
artifacts. As the overall response relationships between techniques
are unchanged, it points to a high platform specificity, likely attributed
to the nonfouling properties offered by the mixed-PEG SAM.[18]This notion was further enforced when
comparing the target platform
response to an ESP sample with a control surface. The control surface
had anti-CD63 antibodies replaced with nonspecific IgG antibodies. Figure A confirmed a small
difference in impedance response between the control surface (1105
Ohms) and a baseline reading (858 Ohms) conducted in just running
buffer. The increase in impedance could be attributed to the binding
of some nonspecific proteins to the WE surface from the 25% v/v serum
matrix. Crucially, an identical sample (1 × 108 ESPs
mL–1) elicited an impedance reading of 4512 Ohms
on the target surface, equating to a SNR of 4.1. This phenomena is
also reflected in the tandem QCM-D results in Figure B. Minimal increases in dissipation, or decrease
in frequency, was seen upon addition of ESPs, demonstrating a low
propensity for nonspecific binding.
Figure 6
Specificity assessment of the EQCM-D platform.
(A) EIS response
to 1 × 108 ESPs mL–1 on a control
surface compared to a target surface. (B) Tandem QCM-D frequency and
dissipation response toward 1 × 108 ESPs mL–1 captured on a control functionalized sensor.
Specificity assessment of the EQCM-D platform.
(A) EIS response
to 1 × 108 ESPs mL–1 on a control
surface compared to a target surface. (B) Tandem QCM-D frequency and
dissipation response toward 1 × 108 ESPs mL–1 captured on a control functionalized sensor.A full set of isotype control data for the entire range of ESP
concentrations in both HBS buffer and 25% v/v serum is displayed in SI Figure S4. The small degree of nonspecific
binding introduced by the complex serum is seen across all EPS concentrations
and across all modes of measurements. The responsiveness to this nonspecific
binding is exacerbated at ESP concentrations >1 × 109 ESPs mL–1 for frequency measurement and EIS read-out
in particular. This could indicate that the artifactual binding on
the control sensor has minimal impact on layer viscoelasticity, hence
marginally impacting dissipation response compared to HBS buffer
alone. This information supported the generation of SNR across all
ESP concentrations (SI Figure S5) in both
HBS buffer and 25% serum. Again, it is reasonable to suggest that
frequency is most significantly affected by the introduction of more
complex media, resulting in a significant reduction in SNR compared
to EIS and frequency. While dissipation seemingly displays lower SNR
overall, it is the least affected by the increase in matrix complexity.
A plateauing of SNR at the highest ESP concentration could indicate
signal saturation relative to noise.The working concentrations
of the three analytical principles employed
by the EQCM-D platform toward CD63-positive exosomes in 25% buffer
were derived from normalized response curves, with the LOD displayed
in Table . The statistical
significance of the target sensor response compared to the control
sensor at the respective LOD concentrations is displayed in SI Figure S6. Although frequency response was
seen to possess a larger dynamic range of measurement; dissipation
and EIS exhibited a superior LOD. Furthermore, as EIS was more effective
at minimizing background signal, it delivered a favorable SNR across
the range of concentrations, resulting in a lower LOD than dissipation
response by approximately half an order of magnitude (Table ). This aligns with other reports
of the more sensitive EIS lowering the LOD of the overall detection
system.[39,41,73] These detection
limits are 1–2 orders of magnitude better than some recent
biosensing attempts of exosomes, all while being conducted in complex
media.[74−78]
Table 1
EQCM-D LOD Values.
LOD
(ESPs mL–1)
Sample media
QCMfreq
QCMdiss
EIS
HBS buffer
1.71 × 108
1.08 × 108
5.34 × 107
25% v/v serum
2.15 × 108
1.25 × 108
6.71 × 107
Overall, these results
underline the capability of the EQCM-D based
immunosensor to effectively discern CD63-positive exosomes in a sensitive
and specific manner. The clinical significance of these findings includes
being able to selectively detect CD63-positive exosomes in minimal
sample volumes and biological fluids where average exosome concentrations
are lower than native levels.
Conclusion
In
conclusion, we report a significant extension to a previously
established QCM-D immunosensing method for CD63-positive exosomes,
by incorporating an additional mode of in situ measurement in the
form of EIS. Using a circuit model which incorporates two time constants
to account for dielectric complexity, we successfully showcase a multimodal
approach that offers label-free analysis with high specificity from
complex matrices. Uniquely, all this was achieved while using the
same sensing surface to detect exosomes based upon their mass, viscoelasticity,
and impedance inducing properties. Such an approach allowed for analytical
comparison between techniques, which suggested a synergy between EIS
and QCM-D, as determined by the magnitude of dissipation by an analyte
and resulting increase in impedance at the working electrode. This
allowed further discrimination between rigid and viscoelastic particles,
which is of particular utility for fluid filled exosome analytes.
Besides the complementarity between the techniques, impact was delivered
through a lowering of the LOD to 6.71 × 107 ESPs mL–1 in 25% v/v serum, compared to 2.15 × 108 and 1.25 × 108 ESPs mL–1 for QCM-D frequency and dissipation techniques, respectively.Ultimately, the strengths of both QCM-D and EIS are enhanced when
applied in combination due to the unique insight offered by each technique.
This work exhibits a more sophisticated biosensor, with QCM-D offering
information on binding kinetics, mass, and interfacial structuring
during layer formation, while EIS improves the sensitivity and detection
limit of the platform overall. This methodology can now be used to
detect a library of exosomal markers with diagnostic potential or
for quality control. The authors encourage the research community
to incorporate EIS data into the ongoing efforts to derive quantitative
exosome information using acoustic-based detection platforms.
Authors: Deborah L M Rupert; Ganesh V Shelke; Gustav Emilsson; Virginia Claudio; Stephan Block; Cecilia Lässer; Andreas Dahlin; Jan O Lötvall; Marta Bally; Vladimir P Zhdanov; Fredrik Höök Journal: Anal Chem Date: 2016-10-04 Impact factor: 6.986
Authors: Stuart D Ibsen; Jennifer Wright; Jean M Lewis; Sejung Kim; Seo-Yeon Ko; Jiye Ong; Sareh Manouchehri; Ankit Vyas; Johnny Akers; Clark C Chen; Bob S Carter; Sadik C Esener; Michael J Heller Journal: ACS Nano Date: 2017-07-03 Impact factor: 15.881
Authors: Sara Cavallaro; Josef Horak; Petra Hååg; Dhanu Gupta; Christiane Stiller; Siddharth S Sahu; André Görgens; Hithesh K Gatty; Kristina Viktorsson; Samir El Andaloussi; Rolf Lewensohn; Amelie E Karlström; Jan Linnros; Apurba Dev Journal: ACS Sens Date: 2019-05-07 Impact factor: 7.711
Authors: Reka A Haraszti; Marie-Cecile Didiot; Ellen Sapp; John Leszyk; Scott A Shaffer; Hannah E Rockwell; Fei Gao; Niven R Narain; Marian DiFiglia; Michael A Kiebish; Neil Aronin; Anastasia Khvorova Journal: J Extracell Vesicles Date: 2016-11-17
Authors: Jugal Suthar; Alberto Alvarez-Fernandez; Alaric Taylor; Maximiliano J Fornerod; Gareth R Williams; Stefan Guldin Journal: ACS Appl Nano Mater Date: 2022-08-19