Elmira Almukhambetova1, Murat Almukhambetov1, Abdugani Musayev1, Ainur Yeshmanova1, Vildan Indershiyev1, Zhadira Kalhodzhaeva1.
Abstract
INTRODUCTION: One of the trends in the development of medical technologies is considered to be telemedicine. This study aimed to evaluate the accuracy of a remote electrocardiogram (ECG) analysis and transmission system in prehospital setting.Entities:
Keywords: Cardiovascular system; cardiovascular diseases; diagnosis; health services administration; quality of health care
Year: 2022 PMID: 35072094 PMCID: PMC8771155 DOI: 10.22037/aaem.v10i1.1399
Source DB: PubMed Journal: Arch Acad Emerg Med ISSN: 2645-4904
Figure 1Example of a 73 years old patient's ECG recorded by a paramedic in 12 leads and sent to a specialist. Interpretation of ECG by "Poly-Spectrum" recording device, which was confirmed by a specialist:
Sensitivity and specificity of the electrocardiogram (ECG) analysis device in diagnosis of different ECG abnormalities compared with cardiologist’s interpretation
| Abnormality | N (%) | TP | FP | FN | TN | Sensitivity | Specificity | Accuracy |
|---|---|---|---|---|---|---|---|---|
| Atrial fibrillation | 1387 (7.2) | 1311 | 607 | 76 | 17878 | 94,5 | 96,7 | 96,6 |
| Atrial flutter | 139 (0.7) | 108 | 56 | 31 | 19126 | 77,7 | 99,7 | 99,6 |
| LVH | 4527 (23.5) | 3678 | 1353 | 849 | 14738 | 81,3 | 91,6 | 89,3 |
| AV block | 559 (2.9) | 473 | 195 | 86 | 18706 | 84,6 | 98,9 | 89,3 |
| Extra-systole | 1888 (9.8) | 1659 | 935 | 229 | 17377 | 87,9 | 94,9 | 94,2 |
| RBBB | 1522 (7.9) | 1397 | 848 | 125 | 17743 | 91,8 | 95,4 | 95,2 |
| LBBB | 2639 (13.7) | 2411 | 1258 | 228 | 16626 | 91,4 | 92,9 | 92,8 |
| Ischemia | 1291 (6.7) | 1168 | 489 | 123 | 17974 | 90,5 | 97,4 | 96,9 |
| Lower MI | 167 (6.8) | 148 | 84 | 19 | 19098 | 88,6 | 99,6 | 99,5 |
| Anterior-lateral MI | 153 (0.9) | 138 | 78 | 15 | 19112 | 90,2 | 99,6 | 99,5 |
| Sinus rhythm | 16992 (88.2) | 17817 | 283 | 782 | 1448 | 95,8 | 83,7 | 94,8 |
| Normal | 12179 (63.2) | 7439 | 2920 | 4740 | 11826 | 61,1 | 80,2 | 71,6 |
All measures are presented with 95% confidence interval. N: number; FP: false positive; FN: false negative; TP: true positive; TN: true negative; LVH: left ventricular hypertrophy; AV: atrioventricular; RBBB: right bundle brunch block; LBBB: left bundle brunch block; MI: myocardial infarction.