Vijay Kumar Gupta1, Neeraj Basantani2, Ajay Sebastian Carvalho3, Ekata Gupta4.
Abstract
INTRODUCTION: In this double-blinded randomized controlled trial, we compare the long-term clinicoradiological effects of anterior cervical discectomy and fusion with polyether ether ketone implant (ACDF with PEEK) versus artificial cervical disc (cervical disc arthroplasty [CDA]) in patients with single and bilevel cervical degenerative disc disease (CDDD). AIMS: We aimed to compare the pain scores, cervical kinematics, and radiological outcomes in patients undergoing CDA and ACDF for single and bilevel CDDD over 5 years. SETTINGS ANDEntities:
Keywords: Anterior cervical discectomy and fusion; cervical arthrodesis; cervical arthroplasty; cervical degenerative disc disease; cervical disc arthroplasty; polyether ketone
Year: 2021 PMID: 35071069 PMCID: PMC8751538 DOI: 10.4103/ajns.AJNS_345_20
Source DB: PubMed Journal: Asian J Neurosurg
Baseline characteristics in the two groups
| CDA group | PEEK group |
| |
|---|---|---|---|
| Age, median (IQR) | 34 (29.25-39.50) | 41 (33-57) | 0.026 |
| Sex, | |||
| Female | 6 (24) | 4 (16) | 0.725 |
| Male | 19 (76) | 21 (84) | |
| Myelopathy, | 8 (32) | 7 (28) | 0.99 |
CDA - Cervical disc arthroplasty; PEEK - Polyether ether ketone; IQR - Interquartile range
Comparison of clinical findings in the cervical disc arthroplasty and polyether ether ketone group
| CDA group | Change median (95% CI), | PEEK group | Change median (95% CI), | ||
|---|---|---|---|---|---|
| VAS scores, median (IQR) | |||||
| Preoperative | 8.3 (7.7-8.9) | −4.65 (−4.95-−4.35), <0.001 | 8.0 (7.6-8.7) | 4.40 (−4.50-−4.30), <0.001 | 0.203 |
| Follow-up | 3.60 (3.20-4.10) | 3.90 (3.00-4.40) | |||
| NDI scores, median (IQR) | |||||
| Preoperative | 26.0 (24.0-27.0) | −18.50 (−19.00-−17.50), <0.001 | 26.0 (24.0-27.0) | −19.00 (−19.50-−18.00), <0.001 | 0.490 |
| Follow-up | 7.0 (6.0-8.0) | 7.0 (6.0-8.0) |
CDA - Cervical disc arthroplasty; PEEK - Polyether ether ketone; IQR - Interquartile range; CI - Confidence interval; VAS - Visual Analogue Scale; NDI - Neck disability index
Comparison of cervical range of motion in cervical disc arthroplasty and polyether ether ketone group
| CDA group | PEEK group | ||||
|---|---|---|---|---|---|
|
|
| ||||
| Values median (IQR) | Change median (95% CI), | Values median (IQR) | Change median (95% CI), | ||
| CROM | |||||
| Preoperative | 88.00 (84.00-96.00) | 8.50 (7.00-10.00), <0.001 | 83.00 (77-87) | −13.00 (−14.50-−12.00), <0.001 | <0.001 |
| Follow-up | 97.00 (96.00-102.00) | 68.00 (66.00-71.00) | |||
| Lateral flexion ROM | |||||
| Preoperative | 71.00 (66.00-74.00) | 9.50 (9.00-11.00), <0.001 | 71.00 (68.00-74.00) | −14.00 (−15.00-−12.00), <0.001 | <0.001 |
| Follow-up | 79.00 (77.00-83.00) | 57.00 (56.00-58.00) | |||
| Rotational ROM | |||||
| Preoperative | 112.00 (103.00-122.0) | 30.50 (27.00-33.50), <0.001 | 112.00 (102.00-120.0) | −19.00 (−22.00-−16.50), <0.001 | <0.001 |
| Follow-up | 142.00 (140.00-148.00) | 91.00 (84.00-100.00) | |||
CDA - Cervical disc arthroplasty; PEEK - Polyether ether ketone; ROM - Range of motion; CROM - Cervical ROM; CI - Confidence interval; IQR - Interquartile range
Figure 1Comparison of range of motion in the two groups
Comparison of radiological findings in cervical disc arthroplasty and polyether ether ketone groups
| CDA group | PEEK group | ||||
|---|---|---|---|---|---|
|
|
| ||||
| Values median (IQR) | Change median (95% CI), | Values median (IQR) | Change median (95% CI), | ||
| Sagittal ROM-upper level | |||||
| Preoperative | 9.10 (7.30-10.10) | 0.75 (0.65-0.80), <0.001 | 9.80 (8.3-10.7) | −0.25 (−0.35-−0.15), <0.001 | <0.001 |
| Follow-up | 9.90 (8.10-10.70) | 9.50 (7.9-10.70) | |||
| Sagittal ROM-lower level | |||||
| Preoperative | 8.80 (7.10-10.30) | 0.65 (0.55-0.75), <0.001 | 13.2 (10.3-14.0) | −2.00 (−2.55-−1.50), <0.001 | <0.001 |
| Follow-up | 9.30 (7.70-10.90) | 10.0 (8.6-11.80) | |||
| Sagittal ROM-implanted level | |||||
| Preoperative | 12.60 (11.40-13.90) | 0.30 (0.25-0.40), <0.001 | 13.2 (12.2-14.4) | −11.80 (−12.60-−11.15), <0.001 | <0.001 |
| Follow-up | 13.00 (11.70-14.30) | 1.60 (1.50-1.80) | |||
CDA - Cervical disc arthroplasty; PEEK - Polyether ether ketone; CI - Confidence interval; ROM - Range of motion; IQR - Interquartile range
Figure 2Comparison of sagittal range of motion in cervical disc arthroplasty and polyether ether ketone group
Comparison of functional spinal unit in cervical disc arthroplasty and polyether ether ketone groups
| AD group | PEEK group | ||||
|---|---|---|---|---|---|
|
|
| ||||
| Values median (IQR) | Change median (95% CI), | Values median (IQR) | Change median (95% CI), | ||
| FSU flexion | |||||
| Preoperative | 8.20 (6.90-8.90) | −1.20 (−1.60-−1.19), <0.001 | 7.90 (6.90-9.00) | −5.85 (−6.35-−5.45), <0.001 | <0.001 |
| Follow-up | 6.60 (5.60-7.70) | 2.10 (1.80-2.40) | |||
| Follow-up | −4.30 (−4.50-−3.80) | −2.00 (−2.40-−1.80) | |||
| FSU range of movement | |||||
| Preoperative | 14.10 (12.30-14.90) | −2.70 (−2.85-−2.65), <0.001 | 14.10 (13.00-15.10) | −10.05 (−10.65-−9.45), <0.001 | <0.001 |
| Follow-up | 11.20 (9.10-12.20) | 3.80 (3.60-4.70) | |||
PEEK - Polyether ether ketone; CI - Confidence interval; FSU - Functional spinal unit; IQR - Interquartile range; AD - Arthroplasty
Comparison of overall cervical alignment in cervical disc arthroplasty and polyether ether ketone group
| AD group | PEEK group | ||||
|---|---|---|---|---|---|
|
|
| ||||
| Values median (IQR) | Change median (95% CI), | Values median (IQR) | Change median (95% CI), | ||
| Overall cervical alignment-flexion | |||||
| Preoperative | 25.40 (23.90-27.40) | 1.00 (0.80-1.15), <0.001 | 22.80 (20.90-24.90) | −7.90 (−9.15-−6.90), <0.001 | <0.001 |
| Follow-up | 26.20 (24.30-28.40) | 14.70 (13.60-15.40) | |||
| Overall cervical alignment-ROM | |||||
| Preoperative | 50.70 (46.60-55.40) | 1.85 (1.50-2.20), <0.001 | 50.50 (49.10-52.40) | −16.80 (−18.70-−14.60), <0.001 | <0.001 |
| Follow-up | 52.30 (48.30-56.30) | 34.10 (31.20-35.40) | |||
IQR - Interquartile range; AD - Arthroplasty; PEEK - Polyether ether ketone; CI - Confidence interval; ROM - Range of motion
Comparison of disc height in cervical disc arthroplasty and polyether ether ketone group
| CDA group | PEEK group | ||||
|---|---|---|---|---|---|
|
|
| ||||
| Values median (IQR) | Change median (95% CI), | Values median (IQR) | Change median (95% CI), | ||
| Disc height-implanted level (posterior intervertebral height) | |||||
| Preoperative | 2.92 (2.52-4.59) | −0.01 (−0.02-0.02), 0.647 | 4.00 (3.78-4.84) | 0.83 (0.70-0.97), 0.001 | <0.001 |
| Follow-up | 2.90 (2.50-4.59) | 5.10 (4.18-5.58) | |||
| Disc height-lower level (posterior intervertebral height) | |||||
| Preoperative | 4.50 (3.22-5.08) | −0.01 (−0.03-0.00), 0.051 | 4.38 (3.82-5.02) | −0.39 (−0.46-−0.34), <0.001 | <0.001 |
| Follow-up | 4.49 (3.21-5.10) | −0.01 (−0.03-0.00), 0.051 | 3.94 (3.43-4.70) | −0.39 (−0.46-−0.34), <0.001 | <0.001 |
| Disc height-upper level (anterior intervertebral height) | |||||
| UL-AIH-pre | 4.90 (3.28-5.04) | 0.00 (−0.02-0.01), 0.855 | 4.32 (3.96-5.02) | −0.51 (−0.62-−0.39), <0.001 | <0.001 |
| UL-AIH-FU | 4.76 (3.28-5.01) | 4.01 (3.18-4.50) | |||
| Disc height-lower level (anterior intervertebral height) | |||||
| LL-AIH-pre | 4.76 (3.18-5.12) | −0.01 (−0.03-0.00), 0.066 | 4.36 (3.88-5.11) | −0.36 (−0.41-−0.31), <0.001 | <0.001 |
| LL-AIH-FU | 4.79 (3.16-5.06) | 4.01 (3.40-4.78) | |||
CDA - Cervical disc arthroplasty; PEEK - Polyether ether ketone; CI - Confidence interval; IQR - Interquartile range; AIH - Anterior intervertebral height; UL - Upper Level; LL - Lower Level; AIH - anterior intervertebral height