Yufei Ma1, Yinqiu Hu1, Shunda Qiao1, Ziting Lang1, Xiaonan Liu1, Ying He1, Vincenzo Spagnolo2. 1. National Key Laboratory of Science and Technology on Tunable Laser, Harbin Institute of Technology, Harbin 150001, China. 2. PolySense Lab, Dipartimento Interateneo di Fisica, University and Politecnico of Bari, Via Amendola 173, Bari, Italy.
Abstract
In this paper, we report on the performance of quartz tuning fork (QTF) based laser spectroscopy sensing employing multiple QTFs. To avoid that resonance frequency mismatching of the QTFs degrades the sensor performance, two types of resonance frequency matching method are here proposed. A system based on the coupling of two sensing modules, one based on quartz-enhanced photoacoustic spectroscopy (QEPAS) and one on light-induced thermoelastic spectroscopy (LITES) technique, was realized to validate the proposed methods. Each module employed a different QTF (QTF1 and QTF2, respectively). Operating temperature or pressure of QTF2 were regulated to match the resonance frequency of QTF1, which operated at 25.0 °C and atmospheric pressure. Without regulation, the difference between QTF1 and QTF2 resonance frequencies was 2.42 Hz and the superposition coefficient η was only 54.7%. When the temperature regulation was carried out, at a QTF2 operating temperature of 67.5 °C, an optimal η value of 95.0% was obtained. For the pressure regulation approach, if operating QTF2 at pressure of 500 Torr, η reached a value of 97.2%. The obtained results show that the proposed two methods are effective in resonance frequency matching of QTFs for gas sensing systems.
In this paper, we report on the performance of quartz tuning fork (QTF) based laser spectroscopy sensing employing multiple QTFs. To avoid that resonance frequency mismatching of the QTFs degrades the sensor performance, two types of resonance frequency matching method are here proposed. A system based on the coupling of two sensing modules, one based on quartz-enhanced photoacoustic spectroscopy (QEPAS) and one on light-induced thermoelastic spectroscopy (LITES) technique, was realized to validate the proposed methods. Each module employed a different QTF (QTF1 and QTF2, respectively). Operating temperature or pressure of QTF2 were regulated to match the resonance frequency of QTF1, which operated at 25.0 °C and atmospheric pressure. Without regulation, the difference between QTF1 and QTF2 resonance frequencies was 2.42 Hz and the superposition coefficient η was only 54.7%. When the temperature regulation was carried out, at a QTF2 operating temperature of 67.5 °C, an optimal η value of 95.0% was obtained. For the pressure regulation approach, if operating QTF2 at pressure of 500 Torr, η reached a value of 97.2%. The obtained results show that the proposed two methods are effective in resonance frequency matching of QTFs for gas sensing systems.
Trace gases are ubiquitous, and their detection found applications in various fields such as fire alarm [1], combustion diagnosis [2], life science [3], electrical safety monitoring [4] and warning of toxic and flammable gases [5]. Therefore, the development of highly sensitive trace gas detection technique is of great significance. Owing to the advantages of high selectivity and sensitivity, non-invasive and real-time detection [6], [7], [8], optical trace gas sensing techniques are widely adopted.In 2002, quartz-enhanced photoacoustic spectroscopy (QEPAS) was firstly proposed [9]. It consists in an indirect absorption spectroscopy technique exploiting the photoacoustic effect discovered by A.G. Bell in 1880 [10]. When a modulated laser passed through a gas sample and is selectively absorbed by the trace gas specie to be detected, acoustic waves are generated. If the laser source is focused between the prong of quartz tuning fork (QTF) the pressure waves put the QTF prongs in oscillation. These vibrations generate a charge distribution due to the piezoelectric effect [11], [12], which is collected by the metal pattern deposited on the QTF surface [13], [14]. Compared with traditional optical trace gas sensing techniques such as tunable diode laser absorption spectroscopy and microphone based photoacoustic spectroscopy, QEPAS provides additional figures of merit, like compactness, laser wavelength independence and immunity to the ambient noise due to the sharp frequency response of the QTF [15], [16], [17], [18], [19], [20], [21], [22]. A novel QTF-based gas sensing method is light-induced thermoelastic spectroscopy (LITES), which was firstly reported in 2018 [23]. After absorbed by target gas, the modulated beam laser hits the QTF surface generating a modulated localized heating. The temperature changes in QTF induced by photothermal conversion results in thermoelastic expansion and contraction [24], [25], [26], [27], [28]. These light-induced deformations, once again, generate a charge distribution that can be collected by the metal pattern, generating an electrical signal which results proportional to the portion of absorbed light from the gas sample. In LITES, QTFs can operate as a narrow-bandwidth (1 Hz), fast-response (tens of kHz), broadband, high-responsivity infrared photodetector, suitable for tunable laser-based absorption spectroscopy [29], [30], [31], [32], [33] for the remote and standoff trace gas detection and can be used in some harsh conditions such as combustion field.The performance of QTF based gas sensors can be improved by exploiting multiple QTF approaches and superimposing the signals generated from each QTF. Previously, three examples of multiple QTF-based trace gas detection methods have been proposed: 1) a multiple-quartz-enhanced photoacoustic spectroscopy (M-QEPAS) sensor which employed two QTFs to detect acoustic wave signals [34]; 2) a multiple-light-induced thermoelastic spectroscopy sensor (M-LITES) which employed two QTFs act as photodetectors [35]; 3) a coupled quartz-enhanced photoacoustic spectroscopy & light-induced thermoelastic spectroscopy (QEPAS-LITES) sensor which employed two QTFs collecting the photoacoustic wave signal and thermoelastic signal, respectively [36]. Although they demonstrated better performance with respect to single QTF setups, their improvements were limited by resonance frequency mismatching between the two QTFs employed, impeding their perfect signal superposition. For example, in M-QEPAS, only a 1.7 times signal enhancement was obtained [34], instead of the 2 times improvement expected for perfect resonance frequency matching. Resonance frequency mismatching is typical also between standard 32 KHz QTFs due to the relative uncertainty of ±10 µm in the exact dimensions of the prongs and the deposited gold pattern [13].The QTF in plane flexural resonance frequency f0 depends on the prongs geometry, the quartz elastic properties and the surrounding gas pressure and temperature. With respect to standard operating conditions (25.0 °C under vacuum), frequency shift of f0 (Δf) can be estimated using the following equation [37], [38]:Where δ is the prong thickness of QTF, η is the dynamic viscosity of the surrounding gas, ρ is the density of quartz, W is the prong width of QTF, ρg is the gas density, c1 and c2 are geometry dependent parameters. In particular, if the surrounding pressure increases, f0 will decrease due to gas viscosity effects and this viscosity can be in turn modified by varying the operating temperature. Thereby, f0 can be tuned by varying the operating thermodynamic conditions (pressure and temperature), allowing to recover possible QTF resonance frequency mismatches.In this manuscript, two types of resonance frequency matching methods are proposed. A QEPAS-LITES setup, implementing two QTFs, were adopted to validate the proposed methods. Resonance frequency matching was realized by adjusting the temperature and pressure of the QTF used for LITES. Acetylene (C2H2) was selected as test gas to evaluate the performance of these two techniques.
Experimental setup
The schematic of the experimental setup used for QTF resonance frequency matching based on temperature and pressure regulation is depicted in Fig. 1. QTF1 and QTF2 were used to detect QEPAS and LITES signal, respectively. C2H2 with a concentration of 2% was selected as test gas to assess the performance of these two techniques. To target the absorption line of C2H2 at 6534.37 cm−1 (1530.37 nm), a fiber-coupled, distributed feedback (DFB), continuous wave (CW) diode laser with a center wavelength of 1.53 µm and output power of 20 mW was as employed. Wavelength modulation spectroscopy (WMS) and second harmonics (2f) detection strategy were applied. In detail, a bias voltage and a sinusoidal voltage generated by a lock-in amplifier were fed to the tunable diode laser to vary the emission wavelength. By slowly varying the bias voltage the output wavelength of the laser was scanned across the absorption line of C2H2. In our experiment, the injection current of the laser was increased from 89 mA to 129 mA and the central current was set to 109 mA at 28 °C. Sinusoidal voltage was added to modulate the laser at a frequency f, so to excite the acoustic wave signal and thermoelastic signals for QEPAS and LITES, respectively. QTFs with a standard resonance frequency of 30.7 kHz were used, which have dimensions of 6 mm, 0.6 mm and 0.36 mm in height, prong width and thickness, respectively. QTF1 was enclosed in a 26 mm-long gas cell C1, filled with 2% C2H2:N2, and equipped with two input and output windows made of calcium fluoride (CaF2). The flow rate of C2H2 was 100 ml/min. The laser beam was collimated by a fiber collimator (FC) and pass through the QTF1 prongs without illuminating it. The laser vertical position was set 0.7 mm below the prongs top to maximize the QEPAS signal [39]. The divergent laser transmitted from C1 was focused on QTF2 by lens L with focal length of 30 mm. The focal position corresponds to the base of one of the prongs of QTF2, where the maximum strain field occurs [25], [30]. The overall optical pathlength from FC to QTF2 surface was 73 mm. However, due to the design that C2H2 only existed in gas cell C1, an effective optical path of 20 mm was obtained for LITES approach.
Fig. 1
Schematic diagrams of resonance frequency matching in QTFs for QEPAS-LITES based on temperature regulation and pressure regulation. Σ: adder. C1: 26 mm-long gas cell. H: steel heating cell. PTC: ceramic heating plate. PT100: temperature sensor. C2: 26 mm-long gas cell.
Schematic diagrams of resonance frequency matching in QTFs for QEPAS-LITES based on temperature regulation and pressure regulation. Σ: adder. C1: 26 mm-long gas cell. H: steel heating cell. PTC: ceramic heating plate. PT100: temperature sensor. C2: 26 mm-long gas cell.For resonance frequency matching based on temperature regulation, QTF2 was installed in steel heating cell H. It had a hole at the front, allowing the laser light to hit the QTF2 and was heated by a ceramic heating plate PTC. A temperature sensor PT100 was placed close to QTF2 to monitor the temperature T2, which was fed back to a thermostat. T2 was varied from 25.0 °C to 70.0 °C. The control accuracy for temperature regulation was 0.1 °C. For resonance frequency matching based on pressure regulation, QTF2 was installed in gas cell C2, similar to C1 but filled with pure nitrogen (N2). N2 pressure P2 was set by a pressure controller and a pump and was varied from 750 Torr to 475 Torr. The control accuracy for pressure regulation was 2 Torr. The flow rate of N2 was 20 ml/min. C2H2 was only present in C1, where QTF1 was located, and C1 was kept at normal temperature (25.0 °C) and atmospheric pressure. The regulation of temperature and pressure worked only for QTF2, to avoid any influence on physical properties of the C2H2:N2 gas mixture. The overall optical pathlength and effective optical path remained unchanged in temperature regulation and pressure regulation. Superposition of QEPAS signal and LITES signal generated from QTF1 and QTF2, respectively, was realized by an adder. The added signal was sent to the lock-in amplifier for demodulation. Integration time of 60 ms was adopted for 2f demodulation with the bandwidth of 1.118 Hz. In the process of demodulation, a reference sinusoidal signal with specific frequency and phase was used to extract the signal by correlation operation. The reference frequency was an integral multiple of the modulation frequency (nf), and the phase was the one maximizing the X component. The X component for the second harmonic component (2f) of the system was analyzed.
Experimental results and discussions
QEPAS-LITES without resonance frequency matching
For realizing QEPAS-LITES setup, two QTFs were randomly selected from a set of QTF operating around 30.7 kHz to simulate a general situation. Compared with the commonly used QTF around 32 kHz, QTF with lower resonance frequency 30.7 kHz had longer accumulation time, which was beneficial to the improvement of signal amplitude. The optical excitation method was used to test the selected QTFs. The injection current of laser was kept at 109 mA to match output wavelength with the absorption peak of C2H2. Laser was then modulated by a sinusoidal voltage with a modulation frequency f. f was varied to retrieve the frequency response of QTF. The QTF as the detector of QEPAS module was denoted by QTF1 and another QTF as the detector of LITES module was denoted by QTF2. The characteristic parameters (resonance frequency f0 and quality factor Q) for QTF1 (f1 and Q1) and QTF2 (f2 and Q2) were retrieved by a Lorentzian fit of the measured data, as shown in Fig. 2. For two QTFs resulted: f1 = 30,707.75 Hz, f2 = 30,710.17 Hz with related bandwidth Δf1 = 2.42 Hz, Δf2 = 2.23 Hz. Q1 = 12,689 and Q2 = 13,771 were calculated based on the relation Q = f/Δf. A difference of ~2.42 Hz between f1 and f2 was measured.
Fig. 2
Peak-normalized Lorentzian fit of frequency responses for QTFs.
Peak-normalized Lorentzian fit of frequency responses for QTFs.Since the 2f QEPAS C2H2 signal S1 is affected by the modulation amplitude, this value has to be optimized. The 2f signal peak value as a function of laser modulation depth was measured by keeping fixed the laser emission wavelength at 1530.37 nm, modulating its current at f1/2, while demodulating the QTF electrical signal at f1. The S1 peak signal as a function of the modulation depth is shown in Fig. 3. The maximum value was obtained for a modulation depth of 13.8 mA, which was the value used in all the following experiments.
Fig. 3
S1 peak signal as a function of the modulation current.
S1 peak signal as a function of the modulation current.The 2f signals measured when only the QEPAS module (S1) or the LITES module (S2) were considered was retrieved using modulation frequencies of f1/2 and f2/2, respectively. Constant f1/2 as the modulation frequency, the signal for system (S3) would increase with the decrease of |f2 - f1| and achieve the maximum when |f2−f1| = 0. Therefore, f1/2 was adopted as the system modulation frequency to show the superposition effect for system. As shown in Fig. 4, peak signals of 28.18 μV, 69.76 μV and 53.54 μV were measured for S1, S2 and S3, respectively, when operating at a temperature of 25.0 °C and a pressure of 750 Torr. The superposition effect evaluated by a coefficient η was measured considering the S1, S2 and S3 peak values and η is defined in percent as:
Fig. 4
S1, S2 and S3 spectral scans of the C2H2 absorption at 25.0 °C and atmospheric pressure.
S1, S2 and S3 spectral scans of the C2H2 absorption at 25.0 °C and atmospheric pressure.In the ideal case of perfect frequency matching η = 100%, while from the data of Fig. 4, η results only ~54.7%, due to resonance frequency mismatching.To explain the above result of poor η, a theoretical model was presented. Due to the resonance characteristics of QTF, when the modulation frequency f deviates from half of the resonance frequency (f0/2), the attenuation in Lorentzian form for QEPAS or LITES signal generated by QTF occurs. The maximum signal would be obtained when resonance frequency f0 = 2f. This attenuation can be expressed as:where S(f, f0, Δf0) is the peak value of 2f QEPAS or LITES signal generated by QTF with a resonance frequency of f0 and a bandwidth of Δf0 in modulation frequency f, S is the peak value of the maximum 2f signal generated by QTF when f = f0/2. Therefore, the peak value of system signal (S3) can be expressed as:where S1 is the peak value of the maximum 2f QEPAS signal generated by QTF1 when f = f1/2, S2 is the peak value of the maximum 2f LITES signal generated by QTF2 when f = f2/2, S3(f, f1, f2) is the peak value of system signal in modulation frequency f. According to formula 4, for QTFs without frequency matching, the generated signals can hardly reach the maximum value S1 and S2 in modulation frequency f, the system signal will be far from the ideal value as a result. A simple simulation based on this formula was carried out through the substitution of S1, S2, f1, f2, Δf1, Δf2. When f = f1/2, S3 peak of 60.21 μV and η of 61.5% were calculated, which were close to the experimental result.
Resonance frequency matching based on temperature regulation
In this section, QTF1 and QTF2 were the QTFs which had been used in Section 3.1. The QTF with resonance frequency of 30,707.75 Hz was QTF1 for QEPAS module and another QTF with resonance frequency of 30,710.17 Hz was QTF2 for LITES module. The performances of QTFs remained unchanged that peak 2f QEPAS signal for QTF1 (S1) was 28.18 μV and peak 2f LITES signal for QTF2 (S2) was 69.76 μV.The operating temperature T2 of QTF2 was regulated to tune its resonance frequency. The relationships between f2, Q2 with T2 were measured by optical excitation method and the obtained results are shown in Fig. 5. When T2 increased from 25.0 °C to 35.0 °C, f2 also increased from 30,710.17 Hz to 30,710.49 Hz. While, at larger T2, f2 decreased, down to a value of 30,706.77 Hz at T2 = 70.0 °C. The frequency matching f2 = f1 was achieved at T2 = 67.5 °C. Q2 generally showed a downward trend from 13771 to 11373 with the increase of T2 due to the increased energy dissipation. At the temperature of 67.5 °C, Q2 = 12137.
Fig. 5
Resonance frequency f2 and quality factor Q2 of QTF2 as a function of T2. The frequency matching between the two QTFs f2 = f1 = 30,707.75 Hz was achieved when T2 = 67.5 °C.
Resonance frequency f2 and quality factor Q2 of QTF2 as a function of T2. The frequency matching between the two QTFs f2 = f1 = 30,707.75 Hz was achieved when T2 = 67.5 °C.To test the efficacy of the frequency matching based on temperature regulation, we performed a series of C2H2 detection measurements while varying T. The laser modulation frequency was set at f1/2 when operating only with the QEPAS module or for the QEPAS+LITES measurements, while if only the LITES module was operated a modulation frequencies f2/2 was used. Increasing T2, due to the minor change of Q, the S2 peak value remained almost unchanged. The relationship between the S3 peak value and T2 is shown in Fig. 6. Increasing T2, a minimum S3 value of 42.31 μV was measured at 35.0 °C, while the maximum S3 value of 93.04 μV was achieved at 67.5 °C. Indeed, the frequency difference between the QTFs reached its highest value at T2 = 35.0 °C, resulting in a poor superposition effect. When T2 increased from 35.0 °C to 67.5 °C, the frequency difference between QTFs was gradually reduced, reaching a good matching. Consequently, S3 increased achieving a maximum value of 93.04 μV. When T2 further rose to 70.0 °C, the frequency difference increased and S3 dropped to 85.87 μV.
Fig. 6
S3 peak signal as a function of T2. The maximum value of 93.04 µV was measured for T2 = 67.5 °C.
S3 peak signal as a function of T2. The maximum value of 93.04 µV was measured for T2 = 67.5 °C.At the optimal T2 = 67.5 °C value and modulation frequency f1/2 = f2/2 = 15353.87 Hz, the measured 2f spectral scan for the QEPAS-LITES system (S3) is shown in Fig. 7 and compared with the corresponding S1 and S2 spectra. Being the signals peak values S1 = 28.18 μV, S2 = 69.76 μV and S3 = 93.04 μV, the superposition coefficient η for QEPAS-LITES system resulted ~95.0%, significantly improved if compared with the 54.7% value measured without frequency matching. However, for a perfect matching S3 should be 97.94 μV, i.e., S1 + S2, a value higher than the measured one. This slight difference is correlated to two main effects: a) loss of signals when superimposed by the adder; b) phase difference between QEPAS and LITES signals.
Fig. 7
2f spectral scans measured at T2 = 67.5 °C and a modulation frequency f1/2 = f2/2 = 15,353.87 Hz.
2f spectral scans measured at T2 = 67.5 °C and a modulation frequency f1/2 = f2/2 = 15,353.87 Hz.
Resonance frequency matching based on pressure regulation
In this section, QTF1 and QTF2 were the QTFs which had been used in Section 3.1. However, due to the characteristic that QTF resonance frequency increases with the decrease of pressure, QTFs were interchanged compared to Section 3.1. The QTF with higher resonance frequency of 30,710.17 Hz was QTF1 for QEPAS module and another QTF with lower resonance frequency of 30,707.75 Hz was QTF2 for LITES module. The same as Section 3.1, peak 2f QEPAS signal for QTF1 (S1) was 28.18 μV and peak 2f LITES signal for QTF2 (S2) was 69.76 μV.For these experiments, the temperature of the system was kept at 25.0 °C, while the pressure P2 for QTF2 was regulated to tune its resonance frequency and hence to achieve a frequency matching with QTF1. The relationship between f2, Q2 with P2 were measured by optical excitation method and the results are shown in Fig. 8. When P2 decreased from 750 Torr to 475 Torr, f2 increased from 30,707.75 Hz to 30,710.49 Hz. At 500 Torr, a frequency matched condition was achieved (f2 = f1 = 30,710.17 Hz). As expected, Q2 showed an upward trend from 12,689 to 21,327 with the decrease of P2. At pressure of 500 Torr, Q2 = 19,686.
Fig. 8
Resonance frequency f2 and quality factor Q2 of QTF2 as a function of P2.
Resonance frequency f2 and quality factor Q2 of QTF2 as a function of P2.Once again, we performed a series of C2H2 sensing while varying P and modulated the laser at f1/2 when operating only with the QEPAS module or for the QEPAS+LITES measurements, while at f2/2 if only the LITES module is operated, as already described in Section 3.2. The relationship between S2 and S3 peak values with P2 is shown in Fig. 9. When P2 decreased from 750 Torr to 475 Torr, S2 peak increased from 69.76 μV to 87.91 μV due to the related increase of Q2. At P2 = 500 Torr, QTF1 and QTF2 were in frequency matching, and S3 peak reached its maximum value of 110.73 μV.
Fig. 9
S2 and S3 peak signal as a function of P2. The maximum value of 110.73 µV for S3 was measured at P2 = 500 Torr.
S2 and S3 peak signal as a function of P2. The maximum value of 110.73 µV for S3 was measured at P2 = 500 Torr.At the optimal P2 of 500 Torr value and modulation frequency f1/2 = f2/2 = 15,355.08 Hz, the measured S3 spectral scan is shown in Fig. 10 and compared with the corresponding S1 and S2 spectra.
Fig. 10
2f spectral scans measured at P2 = 500 Torr and modulation frequency f1/2 = f2/2 = 15,355.08 Hz.
2f spectral scans measured at P2 = 500 Torr and modulation frequency f1/2 = f2/2 = 15,355.08 Hz.The peak values resulted S1 = 28.18 μV, S2 = 85.79 μV and S3 = 110.73 μV. Also in this case, S3 resulted lower than the (S1 + S2) = 113.97 μV, for the same motivations reported in Section 3.2. However, the superposition coefficient η for QEPAS-LITES system resulted ~97.2%, even higher than the value measured for resonance frequency matching based on temperature regulation (~95.0%).
Conclusion
One way to improve the performance of QTF-based gas sensing systems is the employment of multiple QTFs and superimposition of their signals. However, resonance frequency mismatching between the QTFs leads to a reduction in performance. Here, we proposed two different methods to achieve QTFs frequency matching in a QEPAS-LITES sensing system, based on a temperature or a pressure regulation of one of the QTF. C2H2 with a concentration of 2% in N2 was selected as the target gas and two QTFs with resonance frequency f0 around 30.7 kHz (at atmospheric pressure) where employed in a QEPAS-LITES setup to validate the proposed methods. Without resonance frequency regulation, a superposition coefficient η of only 54.7% was determined, due to a resonance frequency mismatch of ~2.42 Hz. For the temperature regulation methods, the resonance frequency matching condition was achieved for QTF2 operating at a temperature of 67.5 °C and η increased to 95.0%. In case of resonance frequency matching based on pressure regulation, QTF2 operated at a pressure of 500 Torr and η increased up to 97.2%. The obtained results demonstrated that the proposed two methods are effective in nearly fully recover the theoretical enhancement factor (η = 100%). This is not only beneficial for multiple QTFs based laser spectroscopy sensing but can also be implemented for any QTF-based optical microscope and imaging sensing technique [40], [41], [42], [43].
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.