| Literature DB >> 35059558 |
Natee Sirinvaravong1, Mark Heimann1, Steve Liskov1, Gan-Xin Yan1.
Abstract
BACKGROUND: Atrial dissociation (AD) is described as the existence of two simultaneous electrically isolated atrial rhythms. Theoretically, detection of dual atrial rhythms with a sufficiently high rate by pacemaker can lead to automatic mode switching and associated pacemaker syndrome. Such a clinical observation has not been reported before in the literature. CASEEntities:
Keywords: Atrial dissociation; Atrial tachycardia; Case report; Dual atrial rhythms; Pacemaker syndrome
Year: 2022 PMID: 35059558 PMCID: PMC8765788 DOI: 10.1093/ehjcr/ytab531
Source DB: PubMed Journal: Eur Heart J Case Rep ISSN: 2514-2119
Figure 1The presenting electrocardiogram showed ventricular paced rhythm at 70 b.p.m. with retrograde P waves most clearly visible in the inferior leads.
Figure 2(A) Presenting electrocardiogram. (B) Atrial and ventricular electrograms from pacemaker interrogation revealing two independent simultaneous atrial rhythms and ventricular paced rhythm. Note retrograde P waves (double-ended arrows) and atrial tachycardia (small arrows). Atrial electrogram: 5 mm/mV; ventricular electrogram: 1 mm/mV.
Figure 3Atrial and ventricular electrograms showed sinus bradycardia with atrial premature depolarization (marked as ‘AS’) with anterograde atrioventricular conduction. Atrial tachycardia is electrically isolated and not recognized by the device. Atrial electrogram: 20 mm/mV; ventricular electrogram: 1 mm/mV.
Figure 4(A) Surface electrocardiogram after atrial lead sensitivity adjustment after the pacemaker was temporarily programmed to AAI at 30 b.p.m. (B) Magnified lead V3 from A showing AT activity.
Figure 5(A) Electrocardiogram after atrial lead sensitivity adjustment. AAIR mode, backup with DDDR mode, was restored. The rhythm was atrial-paced and ventricular-sensed. (B) Electrograms at outpatient follow-up no longer showed atrial tachycardia.
| 4 years prior | Single-chamber atrial pacemaker implant for sinus node dysfunction. |
| 3 years prior | Tricuspid valve replacement. Addition of a right ventricular epicardial lead. Decreased atrial lead sensing and increased pacing threshold were noted during subsequent outpatient follow-up. |
| 5 weeks prior | Complete loss of sensing and capture of right atrial lead were noted. |
| 2 weeks prior | Patient underwent atrial lead revision. |
| 1 week prior | Increased dyspnoea and fatigue started. |
| Index hospital visit | Dual atrial rhythms leading to ‘inappropriate’ automatic mode switching were identified. After atrial lead sensitivity was adjusted, atrial-paced rhythm was restored. |
| Day 1 | Heart failure symptoms resolved. |
| One month after discharge | The patient remained asymptomatic with atrial-paced rhythm. |