FIGURE 1Mean imprecision profile of high-sensitive cardiac troponin I (hs-cTnI) determined on Architect analyser (Abbott, Chicago, IL, USA), using analytical coefficient of variation (CVa) and mean hs-cTnI concentrations provided in supplementary data from [2]. CVa1 and CVa2 are the analytical coefficients of variation corresponding to the hs-cTnI concentrations before (C1) and after (C2) an 8-week period with (red) or without (green) continuous positive airway pressure (CPAP) treatment, and calculated according to the equation of the nonlinear regression curve. The adjusted analytical change limits (adjACL) were calculated according to the formula ±1.96×(CVa12+CVa22)1/2 [4].