| Literature DB >> 35024761 |
Zeinab Ghassabi1, Roman V Kuranov2,3, Joel S Schuman1,4,5,6,7,8, Ronald Zambrano1, Mengfei Wu1, Mengling Liu1, Behnam Tayebi1,5, Yuanbo Wang3, Ian Rubinoff2, Xiaorong Liu9, Gadi Wollstein1,4, Hao F Zhang2, Hiroshi Ishikawa1,4,10.
Abstract
Purpose: Growing evidence suggests that dendrite retraction or degeneration in a subpopulation of the retinal ganglion cells (RGCs) may precede detectable soma abnormalities and RGC death in glaucoma. Visualization of the lamellar structure of the inner plexiform layer (IPL) could advance clinical management and fundamental understanding of glaucoma. We investigated whether visible-light optical coherence tomography (vis-OCT) could detect the difference in the IPL sublayer thicknesses between small cohorts of healthy and glaucomatous subjects. Method: We imaged nine healthy and five glaucomatous subjects with vis-OCT. Four of the healthy subjects were scanned three times each in two separate visits, and five healthy and five glaucoma subjects were scanned three times during a single visit. IPL sublayers were manually segmented using averaged A-line profiles.Entities:
Mesh:
Year: 2022 PMID: 35024761 PMCID: PMC8762683 DOI: 10.1167/iovs.63.1.18
Source DB: PubMed Journal: Invest Ophthalmol Vis Sci ISSN: 0146-0404 Impact factor: 4.799
Figure 1.(A) Illustration of the overall speckle-reduction raster scanning protocol in vis-OCT. (B) Illustration of srA-line acquisition as highlighted by the dashed box in panel A. The red dots indicate the spatial locations of each regular A-line.
Figure 2.(A) A speckle-reduced vis-OCT image from a healthy eye. Horizontal bar: 500 µm; vertical bar: 50 µm. (B) Magnified view of the region highlighted by the dashed box in panel A (15 srA-lines segments). (C) Depth-resolved OCT amplitude profile of the IPL sublayers. We averaged 15 srA-lines, corresponding to approximately 88 µm along the lateral direction, within the highlighted region in panel A. (D) Illustration of the lamination of ganglion cells from RNFL to the IPL. The “red” ganglion cells (ON center) are laminating dendrites to the “b” sublamella of the IPL whereas “blue” cells (OFF center) laminate to the “a” sublamella. The “green” ganglion cell is bi-laminating. (E) A speckle-reduced vis-OCT image from a glaucoma eye. (F) A magnified view of the region highlighted by the dashed box in panel E. (G) Depth-resolved line profile of the glaucoma eye IPL sublayers.
Subject Demographics
| Subjects | Number | Age (y), Median (Q1, Q3) | MD Range (dB), Median (Q1, Q3) | Global RNFL Thickness, Median (Q1, Q3) |
|---|---|---|---|---|
| Healthy | 9 | 47.0 (33.0, 56.0) | 0.23 (−0.1,0.7) | 94.0 (89.0,105.0) |
| Glaucoma | 5 | 64.0 (49.5,67.5) | −19.4 (−23.2, −11.0) | 55.0 (52.5,69.5) |
|
| — | 0.02 | 0.002 | 0.004 |
Intrasession Repeatability for Healthy and Glaucoma Subjects
| Entire IPL CV (%) | Sublayer L1 CV (%) | Sublayer L2 CV (%) | Sublayer L3 CV (%) | |||||
|---|---|---|---|---|---|---|---|---|
| Median | Range | Median | Range | Median | Range | Median | Range | |
| Healthy | 3.1 | 0.0–4.3 | 5.6 | 5.1–6.2 | 6.9 | 3.5–8.8 | 5.6 | 5.4–5.6 |
| Glaucoma | 1.8 | 1.7–3.0 | 6.0 | 5.1–6.0 | 7.7 | 3.7–8.3 | 6.2 | 5.6–10.0 |
|
| 0.44 | 0.95 | 0.69 | 0.007 | ||||
Intersession Repeatability for Healthy Subjects
| Entire IPL CV (%) | Sub layer L1 CV (%) | Sublayer L2 CV (%) | Sublayer L3 CV (%) | |||||
|---|---|---|---|---|---|---|---|---|
| Median | Range | Median | Range | Median | Range | Median | Range | |
| Healthy | 2.9 | 1.6–4.3 | 5.4 | 5.2–5.5 | 6.8 | 5.1–7.8 | 5.6 | 5.1–6.2 |
Measured IPL Thickness and Calculated P Values Based on Mixed-Effects Models Comparing the Difference Between Glaucoma and Healthy Subjects
| Entire IPL Thickness (mm), Mean ± SD | Sublayer L1 Thickness (mm), Mean ± SD | Sublayer L2 Thickness (mm), Mean ± SD | Sublayer L3 Thickness (mm), Mean ± SD | |
|---|---|---|---|---|
| Healthy | 40.1 ± 1.7 | 11.4 ± 0.9 | 17.5 ± 1.4 | 11.3 ± 0.8 |
| Glaucoma | 36.2 ± 1.5 | 11.0 ± 0.9 | 14.2 ± 1.8 | 10.5 ± 0.8 |
|
| 0.003 | 0.9 | 0.003 | 0.05 |
Comparison of the Local RNFL Thickness at IPL Sampling Locations Between Commercial NIR-OCT and Experimental vis-OCT
| RNFL Thickness vis-OCT (µm), Mean ± SD | RNFL Thickness NIR-OCT (µm), Mean ± SD | Total Retina Thickness vis-OCT (µm), Mean ± SD | Total Retina Thickness NIR-OCT (µm), Mean ± SD | |
|---|---|---|---|---|
| Healthy | 24.5 ± 6.0 | 24.5 ± 6.2 | 323.6 ± 10 | 323.5 ± 9.7 |
| Glaucoma | 23.5 ± 3.9 | 23.4 ± 3.7 | 290.1 ± 17.7 | 290.6 ± 17.3 |
|
| 0.94 | 0.99 | ||