| Literature DB >> 35004719 |
Hélène Legardeur1, Gessica Masiello-Fonjallaz1, Martine Jacot-Guillarmod1, Patrice Mathevet1.
Abstract
Introduction: Alignment of the uterine cervix with the vaginal canal is often required during insertion of an intrauterine contraceptive device (IUD). Currently available instruments are traumatic tenacula, which can cause pain and bleeding and represent an obstacle for certain patients to pursue their medical follow-up. A novel investigational cervical vacuum tenaculum, enables atraumatic traction of the cervix using a semi-circular suction pad, designed to conform to the anatomical shape of the external cervical os. Suction is generated by manually pulling out a sliding tube in a vacuum chamber.Entities:
Keywords: atraumatic tenaculum; bleeding; cervix; intrauterine contraceptive device; pain
Year: 2021 PMID: 35004719 PMCID: PMC8732360 DOI: 10.3389/fmed.2021.742182
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Figure 1The suction-based atraumatic tenaculum used in the study. 1. Main body; 2. Slider ring to apply and release suction; 3. Sliding tube to generate vacuum; 4. Semi-circular suction pad for affixation to the cervix.
Figure 2Study flow chart.
Participant-reported VAS pain scores at different stages of the procedure.
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|
|
|
|
|
| ||||
|
| 13 | 13 | 13 | 12 | 8 | 7 | 8 |
| Mean VAS ± SD | 0.0 ± 0.0 | 1.5 ± 4.3 | 7.7 ± 10.5 | 12.2 ± 11.3 | 27.8 ± 22.7 | 11.4 ± 20.4 | 9.8 ± 16.9 |
| Median VAS [IQR] | 0 [0–0] | 0 [0–0] | 4 [0–10] | 9 [2–23.5] | 35 [3.5–47.5] | 0 [0–30] | 2 [0–10] |
Complete data were only available for participants who experienced a successful procedure.
Satisfaction scores among operators, according to procedural success with the Carevix device.
|
|
|
|
| |||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
|
|
|
|
|
|
| |
| The insertion could be performed as planned | 5 | 0 | 0 | 1 | 2 | 5 |
| The device provides adequate visibility/access of the cervix during the procedure | 6 | 2 | 1 | 2 | 0 | 2 |
| I felt confident with the grasping quality of the cervix by suction | 5 | 2 | 2 | 1 | 0 | 3 |
| The device facilitates pulling and aligning the cervix with the uterine cavity | 7 | 0 | 0 | 1 | 0 | 5 |
| The overall handling of the device was satisfactory | 6 | 4 | 1 | 4 | 0 | 2 |