Mingzheng Peng1, Lingming Yu2, Yang Zhou3, Yunhai Yang1, Qingquan Luo1, Xinghua Cheng1.
Abstract
BACKGROUND: Accurate localization of early lung cancer, manifested as solitary pulmonary nodules (SPNs) on computed tomography (CT), is critical in sublobar lung resection. The AR-assisted localization of SPNs was evaluated using a pig animal model.Entities:
Keywords: Solitary pulmonary nodules (SPNs); animal model; augmented reality; localization; sublobar lung resection
Year: 2021 PMID: 35004248 PMCID: PMC8674605 DOI: 10.21037/tlcr-21-554
Source DB: PubMed Journal: Transl Lung Cancer Res ISSN: 2218-6751
Figure 1A pilot study of HoloLens AR-assisted localization of point with plastic thoracic model. (A) A 3D-printed plastic thoracic model; (B) the localization under the assistance of AR was performing; (C) a blue point (in white circle) draw with marker pen was presumed as SPN, and consequently, was precisely hit without seeing it directly; (D) the view from the HoloLens during the process, and the red dot (in while circle) was in the exact position of the blue point; the green line was a plan path for puncture. This image is published with the participant’s consent.
The accuracy of Augmented Reality-assisted localization of targets were evaluated in plastic model
| Cases | Difference between actual depth and expected depth | Mean ± standard deviation (mm) | Hit rate in 1cm |
|---|---|---|---|
| Case1 | 2.50 mm, 1.40 mm, 3.00 mm, 0.50 mm, 1.10 mm | 1.70±1.03 | 100% |
| Case2 | 0.00 mm, 0.70 mm, 1.60 mm, 3.00 mm, 2.20 mm | 1.50±1.19 | 100% |
| Case3 | 0.30 mm, 1.80 mm, 2.80 mm, 0.60 mm, 0.00 mm | 1.10±1.17 | 100% |
| Total | – | 1.43±1.08 | 100% |
Figure 2The reconstruction of SPNs and radiological analysis. (A-C) Coronal, transversal, and sagittal images of SPN (Red dot); (D) simulation and 3D reconstruction of the SPN and plan puncturing path (Red line).
Figure 3The procedure of localization of SPNs in pig animal. (A,B) The researcher adjusted the interpupillary distance in the HoloLens system. After the HoloLens was worn and adjusted to a comfortable state, the AR surgical navigation software was opened; (C) the preset SPNs were processed for 3D-reconstruction using Materialise Mimics V17.0, and the puncturing path and depth were planned; (D-F) SPNs were punctured using hook-wire under the aid of AR planning position and path. This image is published with the participant’s consent.
Figure 4Representative radiological images of localization of SPN. (A) a SPN was selected and marked as red dot in red circle before puncture; (B) this SPN was accurately hit using hook-wire.
The accuracy of Augmented Reality-assisted localization of solitary pulmonary nodules were evaluated in pig animal study.
| Cases | The angle between puncturing line and expected path (°) | Distance between actual entry point and expected point (mm) | Difference between actual depth and expected depth (mm) | Hit rate in | Hit rate in 2 cm |
|---|---|---|---|---|---|
| Case 1 | |||||
| Right | 16.12±7.32 | 8.74±7.65 | 14.56±9.23 | 80% | 100% |
| Left | 16.95±5.02 | 10.26±2.57 | 5.95±1.79 | 100% | 100% |
| Case 2 | |||||
| Right | 8.8±3.74 | 6±2.11 | 4.94±4.47 | 100% | 100% |
| Left | 7.53±6.28 | 3.75±3.95 | 9.05±6.97 | 80% | 100% |
| Case 3 | |||||
| Right | 13.08±6.70 | 7.97±7.92 | 15.85±11.2 | 50% | 100% |
| Left | 14.74±5.23 | 5.76±4.34 | 18.62±9.80 | 50% | 100% |
| Total | 14.52±6.04 | 8.74±5.07 | 9.42±7.95 | 76.67% | 100% |
Advantages and disadvantages of most common used localization methods of solitary pulmonary nodules
| Drawbacks | Application in VATS | Error rate | Complications | Pain | Short maintenance time | Blind locating area | High requirement of apparatus | Expenditure | CT Radiation | Transferring risk |
|---|---|---|---|---|---|---|---|---|---|---|
| Mid-operation | Almost been abandoned | |||||||||
| Palpation | ● | |||||||||
| Ultrasound | ● | ● | ||||||||
| Pre-operation | Currently used | |||||||||
| Hook-wire localization under CT | ● | ● | ● | ● | ● | ● | ● | |||
| Radioactive iodine oil | ● | ● | ● | ● | ● | ● | ● | ● | ||
| Methylene blue mapping | ● | ● | ● | ● | ● | ● | ● | ● | ● | |
| Magnetic navigation guidance | ● | ● | ● | ● | ● | |||||
| AR-assisted localization | ● | |||||||||
Black dot indicates conform to the situation.
The comparison of localization of solitary pulmonary nodules under the CT guidance and augmented reality assistance
| Localization under CT guidance | Localization under augmented reality | |
|---|---|---|
| Time | 20–30 min | 5–10 min |
| Radiation | Repeated radiation | No radiation |
| Waiting time | 0–2 hours | 0 hours |
| Pain | Local anesthesia, intense pain | General anesthesia before surgery |
| Requirement on apparatus | CT, immovable | HoloLens, portable |
| Requirement on circumstance | CT room | Non, operation room |
| Expenditure | CT maintenance cost, localization cost for 2–3 staffs | Model reconstruction cost |