Literature DB >> 35000470

No Mortality Signal With Stellarex Low-Dose Paclitaxel DCB: ILLUMENATE Pivotal 4-Year Outcomes.

Sean P Lyden1, Peter L Faries2, Khusrow A K Niazi3, Ravish Sachar4, Ash Jain5, Marianne Brodmann6, Martin Werner7, Ami Sood8, Prakash Krishnan9.   

Abstract

BACKGROUND: Paclitaxel-coated balloons have shown safety and efficacy in the short- to intermediate-term; however, long-term data remain limited.
OBJECTIVES: To report late safety and efficacy outcomes for a low-dose paclitaxel drug-coated balloon (DCB) compared with percutaneous transluminal angioplasty (PTA) in femoropopliteal lesions from a large randomized controlled trial (RCT).
METHODS: ILLUMENATE Pivotal is a multicenter, single-blind RCT conducted across 43 US and EU centers to examine the safety and efficacy of the Stellarex DCB for the treatment of femoropopliteal disease. Assessments were recorded for all active patients at 36 and 48 months. Vital status of patients formally exited from the study was also collected.
RESULTS: Primary patency through 36 months for patients treated with DCB was significantly higher compared with PTA (p=0.016). The primary safety endpoint through 36 months was 77.4% and 72.4%, respectively (p=0.377). Kaplan-Meier analysis indicated that a higher proportion of DCB subjects were event-free compared with PTA at all study visits. The rate of major adverse event (MAE) through 48 months was 32.9% in the DCB group and 37.9% in the PTA group (p=0.428). No differences in the rate of mortality were evident through 48 months of follow-up with 15.6% in the DCB group and 15.2% in the PTA group (p=0.929).
CONCLUSIONS: Stellarex DCB was associated with significantly higher patency compared with PTA through 3 years with no mortality difference detected through 4 years. The data from the ILLUMENATE Pivotal RCT support the long-term safety and efficacy of the low-dose Stellarex DCB.

Entities:  

Keywords:  CD-TLR; ILLUMENATE Pivotal; Stellarex; drug-coated balloon; femoropopliteal artery; mortality; paclitaxel; peripheral artery disease

Year:  2022        PMID: 35000470     DOI: 10.1177/15266028211068769

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  2 in total

1.  Long-Term Outcomes of the 150 mm Drug-Coated Balloon Cohort from the IN.PACT Global Study.

Authors:  Marianne Brodmann; Wouter Lansink; Katharina Guetl; Antonio Micari; Jeremiah Menk; Thomas Zeller
Journal:  Cardiovasc Intervent Radiol       Date:  2022-07-21       Impact factor: 2.797

2.  Safety of paclitaxel-coated devices in the femoropopliteal arteries: A systematic review and meta-analysis.

Authors:  Chenyang Zhang; Guosheng Yin
Journal:  PLoS One       Date:  2022-10-13       Impact factor: 3.752