Graft Resistance Difference after HVAD to HeartMate 3 Left Ventricular Assist Device Exchange.

Kevin Bourque1, Matthew J Martin2, Daniel I Harjes2, David L Cassidy2, Francis D Pagani3, Robert L Kormos2.   

Abstract

BACKGROUND: A likely consequence of the discontinued distribution and sale of the HVAD System (Medtronic, Minneapolis, MN) will be an increase in replacement with the HeartMate 3 (Abbott, Chicago, IL) left ventricular assist device when device exchange is necessary. If part or all of the HVAD 10-mm-diameter outflow graft is retained during replacement, the HeartMate 3 will have to run at a higher speed than it would with its 14-mm-diameter graft.
METHODS: A steady-state, in vitro study was run with 250-mm-long samples of HVAD, HeartMate 3, and half-HVAD/half-HeartMate 3 grafts and additionally 125- and 375-mm-long samples of the HVAD graft. Flows of 3.0, 3.9, 4.3, 4.7, and 6.0 L/min were applied to encompass expected clinical conditions.
RESULTS: At typical and high flow rates of 4.3 and 6.0 L/min, HeartMate 3 rotor speeds with the full HVAD graft had to be increased relative to those with the HeartMate 3 graft from 5350 to 5700 and 6350 to 6900 rpm, respectively, with power consumption increases from 3.7 to 4.3 W (16%) and 5.5 to 6.8 W (24%), respectively.
CONCLUSIONS: The study did not elucidate a severe consequence of using a remnant HVAD graft during pump exchange, but the incremental risks of a higher rotor speed, disadvantage to the patient in battery runtime, and the general benefit of complete conversion to the HeartMate 3 graft should be balanced against other procedural considerations. Complete graft replacement during HVAD-to-HeartMate 3 conversion remains the preferred approach from an engineering point of view.
Copyright © 2022 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2022        PMID: 34990572     DOI: 10.1016/j.athoracsur.2021.11.065

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  1 in total

1.  Flow dynamics during the HeartWare HVAD to HeartMate 3 exchange: A computational study assessing differential graft lengths and surgical techniques.

Authors:  Jonathan Grinstein; Ryo Torii; Christopher Salerno; Pablo J Blanco; Hector Garcia-Garcia; Christos V Bourantas
Journal:  JTCVS Tech       Date:  2022-05-21
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.