Nicola Blake1, Luciano Pereira1, David J Eaton2, Deirdre Dobson1.
Abstract
Surface-guided radiotherapy (SGRT) can assist with patient setup by providing a real-time feedback mechanism over the whole patient treatment surface. It also has the potential to reduce the number of close contacts between staff and the patient, which is advocated for infection control during the COVID-19 pandemic. Residual translations and rotations (post-CBCT) were acquired following a conventional setup protocol (using permanent marks and lasers) and an SGRT setup protocol. The SGRT protocol resulted in one of the two therapeutic radiographers not having any close contact (<2m) with a patient during setup. Data from 702 imaging sessions showed similar setup accuracy with either protocol, fewer large translations and fewer repeat setup occurrences using the SGRT protocol. The potential of SGRT for infection control should be recognised alongside other benefits.Entities:
Keywords: COVID; Infection control; Lung cancer; SGRT
Year: 2021 PMID: 34988300 PMCID: PMC8710980 DOI: 10.1016/j.tipsro.2021.11.005
Source DB: PubMed Journal: Tech Innov Patient Support Radiat Oncol ISSN: 2405-6324
Fig 1Example region of interest for SGRT localisation.
Summary of translations and rotations comparing the different setup protocols.
| Pitch (°) | Roll (°) | Yaw (°) | Vertical (cm) | Longitudinal (cm) | Lateral (cm) | |
|---|---|---|---|---|---|---|
| 0.4 ± 1.2 | −0.4 ± 1.3 | 0.2 ± 1.4 | −0.1 ± 0.2 | 0.0 ± 0.5 | 0.0 ± 0.3 | |
| 0.3 ± 1.1 | 0.0 ± 1.3 | 0.3 ± 1.5 | 0.2 ± 0.3 | 0.0 ± 0.3 | 0.0 ± 0.3 | |
| 0.20 | <0.01 | 0.05 | <0.01 | 0.12 | 0.86 | |
Results of staff survey regarding experience of SGRTP.
| Yes | No | No difference | |
|---|---|---|---|
| Do you think setting patients up using AlignRT is more accurate than a conventional set up with PMs? | 76% | 12% | 12% |
| Do you find setting patients up using AlignRT easier than a conventional set up with PMs? | 84% | 8% | 8% |