Meseret N Teferra1, David A Hobbs1,2, Robyn A Clark3, Karen J Reynolds1.
Abstract
Background: With cardiovascular disease continuing to be the leading cause of death and the primary reason for hospitalization worldwide, there is an increased burden on healthcare facilities. Electronic-textile (e-textile)-based cardiac monitoring offers a viable option to allow cardiac rehabilitation programs to be conducted outside of the hospital.Entities:
Keywords: EASI ECG; Holter monitoring; ambulatory cardiac monitoring; electronic-textile electrodes; smart fabrics; wearable device; wearable sensors
Year: 2021 PMID: 34988133 PMCID: PMC8720778 DOI: 10.3389/fcvm.2021.806726
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Figure 1The e-textile ECG vest with textile electrodes and embedded wiring (right) the EASI electrodes attachment site on the smart ECG vest (RLD—the reference Right Leg Drive electrode) (left).
Figure 2The e-textile ECG electrodes; left—front view (user side), middle—back view with male snap fastner (attached to the ECG vest) and right—textile electrode attachment female snap fastner on the ECG vest.
Figure 3Lead-II representative ECG traces from different size textile electrodes acquired during climbing stairs (top–40 mm2; middle–60 mm2 and bottom–70 mm2 textile electrodes).
Results of SQI analysis based on lead-II ECGs from different size textile electrodes (40, 60, and 70 mm2).
|
|
|
| |||||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
| |
| Yawning | 0.049 ± 0.0049 | 0.043 ± 0.0030 | 0.9661 |
| 0.9946 | 5.2402 | 5.4238 | 6.2582 | |
| Deep Breathing | 0.046 ± 0.0067 | 0.041 ± 0.0040 | 0.9791 | 0.9780 | 0.9840 | 5.7759 | 8.0175 | 7.7144 | |
| Coughing | 0.036 ± 0.0067 | 0.044 ± 0.0039 | 0.9950 | 0.9920 | 0.9692 | 6.3124 | 7.3790 | 7.6144 | |
| Sideways | 0.055 ± 0.0080 | 0.055 ± 0.0055 | 0.9852 | 0.9880 | 0.9876 | 6.4403 | 7.8228 | 11.7176 | |
| Up | 0.121 ± 0.0182 | 0.098 ± 0.0155 |
| 0.9895 | 0.9937 | 5.6869 | 6.9605 | 7.2274 | |
| Sitting/Standing | 0.046 ± 0.0063 | 0.043 ± 0.0029 | 0.9812 | 0.9921 | 0.9943 | 7.8956 | 7.9628 | 9.1100 | |
| Stairs | 0.046 ± 0.0063 | 0.045 ± 0.0029 | 0.9828 | 0.9968 | 0.9921 | 7.8110 | 7.4803 | 8.7859 | |
Sideways, moving the hands sideways and moving them back to the midline horizontally; up, raise arms above the head and moving them back; sitting / standing, sitting / standing from a chair; stairs, climbing stairs; ApEn, Approximate entropy; basSQI, baseline power signal quality index; SNR, Signal to Noise Ratio; Bold italic - Red, moderate to intense lower frequency noise (basSQI <0.95); Bold, higher ApEn value.
Statistically significant (p < 0.05) compared to the ECG from the 70 mm.
Results of SQI analysis based on lead-II ECGs acquired from 3, 5, and 10 mm thick Textile electrodes.
|
|
|
|
| ||||||
|---|---|---|---|---|---|---|---|---|---|
|
| |||||||||
|
|
|
|
|
|
|
|
|
| |
| Yawning | 0.043 ± 0.0030 | 0.048 ± 0.0075 |
| 0.9680 | 0.9624 |
| 8.0865 | 5.7944 | 1.8421 |
| Deep breathing | 0.040 ± 0.0039 | 0.050 ± 0.0055 |
| 0.9863 | 0.9655 |
| 7.9564 | 5.7428 | 2.7611 |
| Coughing | 0.043 ± 0.0048 | 0.049 ± 0.0050 |
| 0.9709 | 0.9559 |
| 7.6567 | 5.8650 | 2.9173 |
| Sideways | 0.054 ± 0.0057 | 0.061 ± 0.0068 |
| 0.9921 | 0.9655 |
| 11.4961 | 6.9296 | 1.9336 |
| Up | 0.096 ± 0.0202 | 0.146 ± 0.0205 |
| 0.9942 | 0.9712 |
| 7.2397 | 6.2622 | 3.8354 |
| Sitting/standing | 0.042 ± 0.0031 | 0.053 ± 0.0043 |
| 0.9953 | 0.9743 |
| 8.4876 | 7.5886 | 4.2407 |
| Stairs | 0.046 ± 0.0025 | 0.056 ± 0.0043 |
| 0.9926 | 0.9692 |
| 8.7658 | 6.2901 | 3.8456 |
Sideways, moving the hands sideways and moving them back to the midline horizontally; up, raise arms above the head and moving them back; sitting / standing, sitting / standing from a chair; stairs, climbing stairs; ApEn, Approximate entropy; basSQI, baseline power signal quality index; SNR, Signal to Noise Ratio; Bold italic - Red, moderate to intense lower frequency noise (basSQI <0.95); Bold, higher ApEn value.
Statistically significant (p < 0.05) compared to the ECG from the 3 mm textile electrodes.
Figure 4Lead-II representative sideways ECG (top—Anterior; middle—medial and bottom—posterior AI electrodes placement).
Summary of the ECG characteristics based on “AI” electrodes placement.
|
|
|
|
| |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
| ||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Yawning | 0.990 | 0.986 |
| 3.08 | 1.89 | 1.13 | 269.1 | 105.3 | 41.3 | 0.039 ± 0.009 | 0.044 ± 0.011 | |
| Deep breathing | 0.988 | 0.987 |
| 3.15 | 2.01 | 1.23 | 258.1 | 105.4 | 40.6 | 0.039 ± 0.005 | 0.044 ± 0.007 | |
| Coughing | 0.995 | 0.998 | 0.992 | 3.27 | 2.12 | 1.35 | 304.6 | 140.5 | 59.1 | 0.032 ± 0.005 | 0.038 ± 0.005 | |
| Sideways | 0.992 | 0.990 | 0.965 | 3.04 | 2.07 | 1.26 | 221.1 | 110.2 | 43.2 | 0.039 ± 0.004 | 0.139 ± 0.031 | |
| Up | 0.993 | 0.992 | 0.964 | 2.89 | 1.93 | 1.24 | 228.8 | 119.6 | 49.4 | 0.074 ± 0.001 | 0.168 ± 0.027 | |
| Sitting/standing | 0.998 | 0.984 | 0.972 | 3.16 | 2.26 | 1.38 | 275.3 | 138.4 | 55.5 | 0.039 ± 0.005 | 0.065 ± 0.009 | |
| Stairs | 0.998 | 0.991 | 0.981 | 3.08 | 1.89 | 1.13 | 269.1 | 105.3 | 41.3 | 0.038 ± 0.004 | 0.074 ± 0.006 | |
Sideways, moving the hands sideways and moving them back to the midline horizontally; up, raise arms above the head and moving them back; sitting / standing, sitting / standing from a chair; stairs, climbing stairs; basSQI, baseline power signal quality index; ApEn, approximate entropy; Ant, AI electrodes placed at the left and right anterior axillary lines, respectively; Med, AI electrodes placed at the left and right medial axillary lines, respectively; Post, AI electrodes placed at the left and right posterior axillary lines, respectively; Bold italic - Red, moderate to intense lower frequency noise (basSQI <0.95); Bold, higher ApEn value.
Higher R-wave amplitude.
Statistically significant (p < 0.05) compared to the ECG from the Anterior AI textile electrodes placement.
SQI analysis results of lead-II ECG from textile electrodes and commercial wet-gel electrodes using textile ECG monitor.
|
|
|
| |||||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
| |
| Yawning | 0.038 ± 0.006 |
| 0.9822 | 0.9614 | 4.6507 | 7.6812 | 6.6747 | ||
| Deep breathing | 0.047 ± 0.004 | 0.039 ± 0.006 |
| 0.9900 | 0.9860 | 3.8205 | 7.1519 | 7.6097 | |
| Sideways | 0.045 ± 0.005 | 0.9858 | 0.9932 | 0.9701 | 6.7089 | 11.8897 | 7.7883 | ||
| Up | 0.094 ± 0.012 | 0.9633 | 0.9944 |
| 6.4031 | 8.3891 | 7.6526 | ||
| Sitting/standing | 0.043 ± 0.003 |
| 0.9953 | 0.9581 | 6.9461 | 9.1111 | 9.3364 | ||
| Stairs | 0.045 ± 0.003 | 0.059 ± 0.007 |
| 0.9926 | 0.9733 | 6.9189 | 8.9621 | 9.6773 | |
Sideways, moving the hands sideways and moving them back to the midline horizontally; up, raise arms above the head and moving them back; sitting / standing, sitting / standing from a chair; stairs, climbing stairs; ApEn, Approximate entropy; basSQI, baseline power signal quality index; SNR, Signal to Noise Ratio; Bold italic - Red, moderate to intense lower frequency noise (basSQI <0.95); Bold, higher ApEn value;
Statistically significant (p < 0.05) compared to the ECG from the wet textile electrodes.
Figure 5Lead-II representative ECG during up movement (top—wet textile electrodes and bottom—commercial wet-gel electrodes).
Figure 6V5 representative ECG during climbing stairs (top—Holter ECG and bottom—textile ECG monitor).
SQI analysis results of lead V5 ECG from the reference Holter monitor (using wet-gel commercial electrodes) and the textile ECG monitor (using textile electrodes).
|
|
|
| ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
| Deep breathing | 0.1268 ± 0.0126 | 0.9818 | 0.9898 | 20.1131 | 18.7119 | |
| Coughing | 0.1328 ± 0.0093 | 0.9639 | 0.9890 | 15.3232 | 16.8440 | |
| Sideways | 0.1459 ± 0.0329 | 0.9896 | 0.9904 | 13.6312 | 18.4488 | |
| Up | 0.1097 ± 0.0063 | 0.9862 | 0.9911 | 15.5721 | 16.0624 | |
| Sitting | 0.1143 ± 0.0102 | 0.9896 | 0.9867 | 16.0861 | 18.1898 | |
| A phone call | 0.1244 ± 0.0156 |
| 0.9920 | 14.6479 | 15.3835 | |
| Sitting / standing | 0.1400 ± 0.0184 |
|
| 9.1645 | 16.2594 | |
| Lying on a bed | 0.1126 ± 0.0052 |
| 0.9924 | 13.7291 | 15.0579 | |
| Stairs | 0.1268 ± 0.0126 |
| 0.9970 | 11.9310 | 17.0698 | |
Sideways, moving the hands sideways and moving them back to the midline horizontally; up, raise arms above the head and moving them back; sitting / standing, sitting / standing from a chair; stairs, climbing stairs; ApEn, Approximate entropy; basSQI, baseline power signal quality index; SNR, Signal to Noise Ratio; Bold italic - Red, moderate to intense lower frequency noise (basSQI <0.95); Bold, higher ApEn value.
Statistically significant (p < 0.05) compared to the ECG from the Textile ECG monitor.
Summary of the QRS duration and QTC intervals of ECG from the Holter monitor and textile-based ECG monitor during different body movements and activities of the daily living.
|
|
|
| ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
| |
| Deep breath | 333 | 0 | 332 | 1 | 0 | 333 | 0 | 333 | 0 | 0 |
| Coughing | 159 | 0 | 159 | 0 | 0 | 159 | 0 | 159 | 0 | 0 |
| Sideways | 147 | 0 | 146 | 1 | 0 | 147 | 0 | 146 | 0 | 1 |
| Up | 140 | 0 | 140 | 0 | 0 | 140 | 0 | 140 | 0 | 0 |
| Sitting | 261 | 1 | 202 | 1 | 58 | 262 | 0 | 261 | 0 | 0 |
| A phone call | 82 | 0 | 79 | 3 | 0 | 82 | 0 | 82 | 0 | 0 |
| Sitting / standing | 78 | 0 | 77 | 1 | 0 | 78 | 0 | 77 | 0 | 1 |
| Lying on a bed | 70 | 0 | 70 | 0 | 0 | 70 | 0 | 69 | 1 | 0 |
| Stairs | 160 | 0 | 158 | 0 | 2 | 160 | 0 | 160 | 0 | 0 |
|
| ||||||||||
|
|
| |||||||||
|
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
| |
| Sitting | 262 | 0 | 261 | 1 | 0 | 262 | 0 | 262 | 0 | 0 |
Sideways, moving the hands sideways and moving them back to the midline horizontally; up, raise arms above the head and moving them back; sitting / standing, sitting / standing from a chair; stairs, climbing stairs; QT.