Fig. 1( A ) Example of a normal, physiological readout from the GTT from our group. The time between consecutive drops (d) is measured at 5 at the start of measurement (arrow 1), with gradual reduction in flow shown by increasing (d) ( arrow 2 ) and eventual occlusion time ( arrow 3 and also shown as readout on top left-hand corner of graph as OT 370). ( B ) Reproduction of readout of “normal” curve from Clavería et al showing at the start of the measurement, (d) is already 10, with rapid increase in (d) at 10 seconds (on x-axis) with resultant OT 40 seconds, which represents an erroneous reading—in contrast to data from a number of prior studies all showing average OT in normal volunteers between 350–490 seconds. 28 29 34 35