| Literature DB >> 34981551 |
Clara Wai-Shan Lo1, Teresa Kam-Chi Tsui1, Ronald Ching-Wan Ma2,3, Michael Ho-Ming Chan1, Chung-Shun Ho1.
Abstract
Quantitation of plasma angiotensin (Ang) II, the active mediator of the renin-angiotensin system, is challenging owing to its low physiological concentration. We report a validated liquid chromatography-mass spectrometry (LCMS) method to overcome this challenge. Ang II was extracted from EDTA plasma by an offline solid-phase extraction procedure with a Waters MAX μElution plate. LCMS quantitation was performed on the Waters TQS system, monitoring the 3+ ions of the peptide. The analytical performance of the LCMS method was validated. The stability of Ang II was studied with or without the presence of a protease inhibitor. Local reference intervals were established from 143 healthy normotensive subjects (57% female, 21-60 years old). The Ang II LCMS method had a measurable range of 3.3-700 pmol/L. The between-batch precision coefficient of variation was <7% over Ang II concentrations of 8.6-110 pmol/L. No significant matrix interference and carryover were observed. There was no significant difference in Ang II concentration in EDTA blood and plasma for at least 2h and 1 h at room temperature, respectively. Ang II was stable for at least 1 year when stored at -80°C, with or without the protease inhibitor. Age-dependent Ang II reference intervals were established: 4.4-17.7 pmol/L (21-30 years) and 3.9-12.8 pmol/L (31-60 years). The present LCMS method is suitable for quantitation of plasma Ang II to study the renin-angiotensin system.Entities:
Keywords: Chinese population; LCMS; angiotensin II; reference intervals
Mesh:
Substances:
Year: 2022 PMID: 34981551 DOI: 10.1002/bmc.5318
Source DB: PubMed Journal: Biomed Chromatogr ISSN: 0269-3879 Impact factor: 1.902