Hirosumi Miyakawa1, Huai-Hsun Hsu1, Mizuki Ogawa1, Yuichi Miyagawa1, Naoyuki Takemura1.
Abstract
This study compared canine and feline fibroblast growth factor (FGF)-23 concentration measurements between automated chemiluminescence assay (CLEIA) and enzyme-linked immunosorbent assay (ELISA). Seventy serum samples each from dogs and cats were evaluated. FGF-23 measurements by CLEIA significantly correlated with those of ELISA in both dogs and cats. The Bland-Altman test showed that FGF-23 between CLEIA and ELISA had fixed and proportional biases, respectively, in both dogs and cats. Measurements by CLEIA were lower than those of ELISA, especially in higher serum FGF-23 concentrations. This study showed that FGF-23 concentrations in dogs and cats can be evaluated by automated CLEIA. However, FGF-23 cannot be directly compared between CLEIA and ELISA.Entities:
Keywords: biomarker; cat; dog; mineral metabolism
Mesh:
Substances:
Year: 2021 PMID: 34980757 PMCID: PMC8920724 DOI: 10.1292/jvms.21-0566
Source DB: PubMed Journal: J Vet Med Sci ISSN: 0916-7250 Impact factor: 1.267
Fig. 1.Dilution linearity in canine serum fibroblast growth factor-23 concentration of 47,310 pg/ml (A) and feline serum fibroblast growth factor-23 concentration of 22,546 pg/ml (B) measured by automated chemiluminescence immunoassay.
Storage stability of serum fibroblast growth factor (FGF)-23 concentrations using chemiluminescence assay in dogs and cats
| Storage period (day) | Canine FGF-23 (pg/ml) recovery percentage | Feline FGF-23 (pg/ml) recovery percentage | ||||||
|---|---|---|---|---|---|---|---|---|
| Room temperature | 4°C | −20°C | −80°C | Room temperature | 4°C | −20°C | −80°C | |
| 1 | 166 (52%) | 263 (82%) | 355 (112%) | 315 (99%) | 241 (81%) | 284 (95%) | 282 (94%) | 300 (101%) |
| 3 | 135 (43%) | 248 (78%) | 329 (104%) | 322 (102%) | 182 (61%) | 288 (97%) | 290 (97%) | 303 (102%) |
| 7 | 104 (33%) | 210 (66%) | 339 (107%) | 335 (106%) | 83 (28%) | 278 (93%) | 299 (100%) | 305 (102%) |
| 28 | No data | No data | 302 (95%) | 302 (95%) | No data | No data | 270 (90%) | 263 (88%) |
The storage stability was represented by recovery percentage, dividing the measurement by the baseline value.
Fig. 2.Scatter plot of serum fibroblast growth factor-23 concentrations with sandwich enzyme-linked immunosorbent assay (ELISA) and automated chemiluminescence immunoassay (CLEIA) in dogs (A) and cats (B). Dashed lines represent a regression line; r: correlation coefficient.
Fig. 3.Bland–Altman plot of serum fibroblast growth factor-23 concentrations with sandwich enzyme-linked immunosorbent assay (ELISA) and automated chemiluminescence immunoassay (CLEIA) in dogs (A) and cats (B). Dashed lines show limit of agreement for difference.