Literature DB >> 34973652

Technical description of a novel device for external ventricular drainage in neonatal and pediatric patients: Results from a single referral center experience.

Alessandro Consales1, Giuseppe Di Perna2, Laura C De Angelis3, Mattia Pacetti1, Alberto Balestrino4, Marcello Ravegnani1, Marco Pavanello1, Francesca Secci1, Luca A Ramenghi5, Gianluca Piatelli1, Armando Cama1.   

Abstract

BACKGROUND: Since external ventricular drainage (EVD) related infections are usually due to skin flora, an extradural intra/extra-cranial accessory device, developed for pediatric patients under three years of age undergoing EVD positioning, is described. The aim of this paper is to provide technical description of this device, underlining the possibility to reduce infective risk and to prevent EVD dislocation.
METHODS: Patients undergoing A-D device EVD placement between 1990 and 2017 at authors' institution were retrospectively considered. The device was made of a fully MRI-compatible inert material (Ketron-Peek-1000), composed of two pieces securely fixable to the skull, bridging the catheter directly from the epidural space to the extracranial space without letting it come in contact with the skin.
RESULTS: A total number of 350 patients were considered. The mean age was 1.4 years, being the youngest patient a newborn of 25 weeks of gestational age. Mean time of EVD maintenance was 45 days, ranging from 21 to 81 days. 2 cases (0.6%) of EVD related infections were reported, while, pull-out of the ventricular catheter occurred in 3 cases (0.9%). No cases of bone fractures related to the clamp effect provided by A-D device were reported in the series
CONCLUSIONS: This device could represent a safe and feasible option to reduce EVD related infections and catheter pull-out in pediatric patients. The encouraging results could strength the aim of the device to allow safer and longer length of CSF drainage. Moreover, the fully MRI-compatible nature and its non-magnetic properties allow to use it with neuronavigation systems.
Copyright © 2021 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Anti-displacement device; Catheter pull-out; Cerebrospinal fluid infection; External ventricular drain; Hydrocephalus

Mesh:

Year:  2021        PMID: 34973652     DOI: 10.1016/j.clineuro.2021.107100

Source DB:  PubMed          Journal:  Clin Neurol Neurosurg        ISSN: 0303-8467            Impact factor:   1.876


  1 in total

1.  Surgical Antimicrobial Prophylaxis in Neonates and Children Undergoing Neurosurgery: A RAND/UCLA Appropriateness Method Consensus Study.

Authors:  Susanna Esposito; Mino Zucchelli; Sonia Bianchini; Laura Nicoletti; Sara Monaco; Erika Rigotti; Laura Venditto; Cinzia Auriti; Caterina Caminiti; Elio Castagnola; Giorgio Conti; Maia De Luca; Daniele Donà; Luisa Galli; Silvia Garazzino; Stefania La Grutta; Laura Lancella; Mario Lima; Giuseppe Maglietta; Gloria Pelizzo; Nicola Petrosillo; Giorgio Piacentini; Simone Pizzi; Alessandro Simonini; Simonetta Tesoro; Elisabetta Venturini; Fabio Mosca; Annamaria Staiano; Nicola Principi
Journal:  Antibiotics (Basel)       Date:  2022-06-26