Does the geometry of scan bodies affect the alignment accuracy of computer-aided design in implant digital workflow: An in vitro study?

Yu Pan1, James K H Tsoi1, Walter Y H Lam1, Zhuofan Chen2, Edmond H N Pow1.   

Abstract

OBJECTIVES: To compare 2 implant scan bodies with different geometry on the accuracy of the virtual alignment process in the digital workflow.
MATERIALS AND METHODS: A master model of the edentulous maxilla with 6 implants and multiunit abutments (MUA) inserted was fabricated. Six dome-shaped and cuboidal scan bodies were mounted on the MUAs, respectively, and consecutively scanned by a laboratory scanner 10 times. The original scans were imported to a dental-specific CAD software and virtually aligned with the default CAD model in the implant library. Thus, 10 aligned models were created. Both the original scans and the aligned models were evaluated by an inspection software for deviation of the scan body surfaces, the centroids of scan body and MUA, the scan body center-axis, and the inter-MUA distances/angles. The two-sample T-test/Mann-Whitney U test were used to analyze the data with the level of significance set at 0.05.
RESULTS: The cuboidal group showed significant greater deviations of the model surface (13.9 µm vs. 10.7 µm) and the MUA centroids (31.7 µm vs. 22.8 µm) but smaller deviation of the inter-MUA angle (0.047° vs. 0.070°) than those of the dome-shaped group (p < .05). No significant differences in the deviation of scan body centroids, center-axis, and the inter-MUA distances between the 2 groups were found.
CONCLUSIONS: Virtual alignment of implant scan body affected the accuracy of the digital workflow for complete-arch implant-supported prostheses (up to ~30 µm/0.09°). Different geometries of the implant scan body could also influence the transfer accuracy in the CAD process.
© 2022 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  3-D imaging; CAD-CAM; dental implants; digital image processing; dimensional measurement accuracy; prosthesis design

Mesh:

Substances:

Year:  2022        PMID: 34971468     DOI: 10.1111/clr.13890

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  1 in total

1.  In Vitro Accuracy of Digital and Conventional Impressions for Full-Arch Implant-Supported Prostheses.

Authors:  Rani D'haese; Tom Vrombaut; Herman Roeykens; Stefan Vandeweghe
Journal:  J Clin Med       Date:  2022-01-25       Impact factor: 4.241

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.