| Literature DB >> 34950572 |
Mattia Squarcia1,2, Mireia Mora2,3,4, Gloria Aranda2,3, Enrique Carrero5, Daniel Martínez6, Ramona Jerez5, Ricard Valero5, Joan Berenguer1, Irene Halperin2,3,4, Felicia A Hanzu2,3,4.
Abstract
AIM: The short-term and long-term efficacy of different thermal percutaneous ablation techniques remains a topical issue. Our group implemented percutaneous laser ablation (LA), a moving-shot technique to increase efficiency and reduce costs and variability of LA by applying multiple lower-intensity energy illuminations (MLIEI) covering the nodular volume (V) through changes in position of a single laser fiber within the thyroid nodule. The aim of the present study was to evaluate the efficacy of the single-fiber LA-MLIEI during a 5-year follow-up and to identify possible predictors of the final outcome.Entities:
Keywords: long-term response; short-term response; single fiber; thyroid nodules; unique session
Year: 2021 PMID: 34950572 PMCID: PMC8691264 DOI: 10.3389/fonc.2021.584265
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 6.244
Figure 1Single-fiber laser ablation multiple lower-intensity energy illuminations (LA-MLIEI) technique. (A, A1) Positions of the operator and patient. The patient is placed in a supine position with mild neck extension, and the operator stands close to the patient’s head. The operator’s left hand holds the ultrasound (US) probe and the right hand the electrode. (A2) US image of the LA-MLIEI with trans-isthmic approach. This transverse US image shows the needle shaft (red arrow), in the thyroid nodule, and the hyperechoic area (green arrow) from tissue heating and vaporization during illuminations. (B, B1) B-mode and drawing of the LA-MLIEI technique. The optic fiber (dotted green arrow) is inserted through the isthmus to visualize the entire length. Illuminations are sequentially performed from the deepest portion of the nodule (red circles Nr. 1 and 4) to the superficial area (red circles Nr. 3 and 6) by pulling back the tip of the optic fiber. (B2) Contrast-enhanced US imaging and drawing of the LA-MLIEI technique. The procedure is repeated with different optic fiber angulations (dotted green arrows) to cover all the volume (V) of the target nodule. Illuminations are performed with lower energies in the proximities of the peripheral danger triangle (in red) and higher energies in the central area. (C) Sequential US images of the LA-MLIEI technique. (C1) Transverse US image showing the tip of the optic fiber. (C2) Hyperechoic area in the thyroid nodule appearing during optic fiber firing. (C3) Overlapping areas of gas formation during optic fiber moving shot. (C4) At the end of the procedure, the complete volume of the nodule appears diffusely hypoechogenic with multiple hyperechogenic gas bubbles dispersed in the treated area. (D) Postprocedural axial (D1) and sagital (D2) contrast-enhanced US images showing complete ablation of the target nodule with the presence of enhancing tissue in the location of the danger triangle, indicating preserved safety margins.
Clinical, demographic, and laboratory features of all patients at baseline and overall treatment characteristics and side effects.
| Variable | |
|---|---|
| Sex, male/female (n) | 7/23 |
| Age (years) | 62.3 ± 15.6 |
| BMI (kg/m2) | 28.2 ± 5.9 |
| Uninodular/multinodular Goiter (%) | 72/36 |
| Nodule type (solid/microcystic) | 21/7 |
| Thyroid nodule volume (ml) | 18.9 ± 21.2 |
| TSH (mU/ml) | 2.8 ± 2.3 |
| FT4 (ng/dl) | 1.1 ± 0.1 |
| FT3 (ng/dl) | 1.1 ± 0.2 |
| TPOAb (mUI/ml) (n/%) | 13/43.3 |
| Treatment duration (min) | 45 ± 6 |
| Total administered energy (J) | 2,954.7 ± 1,689.4 |
| Energy/V (J/ml) | 231.9 ± 179.0 |
| Number of illuminations | 5.1 ± 7.7 |
| Intraprocedural events (n/%) | 1/3.3 |
| Side effects (n/%) | 3/9.9 |
Values are expressed as mean ± SD unless otherwise specified.
BMI, body mass index; TSH, thyrotropin; FT4, free T4; FT3, free T3; TPOAb, anti-thyroid peroxidase antibodies; V, volume; n, number.
Normal range TSH 0.4–4.0 mUI/L.
Normal range FT4: 0.8–2.0 ng/dl.
Normal range FT3: 0.7–1.9 ng/ml.
TPOAb < 60 mUI/ml.
Figure 2Nodule reduction during follow-up after single-fiber laser ablation multiple lower-intensity energy illumination (LA-MLIEI) technique. (A) Percent (%) of nodule volume (V) reduction in the responder and non-responder groups during the 60 months’ follow-up. Values are expressed as mean of % V reduction ± SD. (B) Maximum percent (%) of nodule V reduction in the responder and non-responder groups during the 60 months’ follow-up. Values are expressed as mean of % V reduction ± SD. (C) Axial and sagittal US images of the target nodule obtained before the treatment and 12 months after the treatment showing a significant volume (V) decrease.
Clinical and treatment characteristics in responders and non-responders to the single-fiber laser ablation multiple lower-intensity energy illuminations (LA-MLIEI) treatment.
| Clinical features | Responders | Non-responders | p-Value |
|---|---|---|---|
| Sex, female/male (n) | 16/4 (20) | 7/3 (10) | ns |
| Age (years) | 60.5 ± 15.1 | 65.9 ± 16.8 | ns |
| Nodule type (solid/microcystic) | 16/4 | 5/5 | 0.05 |
| Thyroid nodule volume (ml) | 14.2 ± 8.6 | 30.4 ± 31.6 | 0.03 |
| Total administered energy (J) | 2,816.1 ± 1,840.9 | 3,232.0 ± 1,383.3 | ns |
| Energy/V (J/ml) | 268.3 ± 203.8 | 159.3 ± 82.3 | ns |
| Number of illuminations | 5.5 ± 2.6 | 6.2 ± 3.8 | ns |
| Intraprocedural events (n/%) | 1/3.3 | – | ns |
| Side effects (n/%) | 2/6.6 | 1/3.3 | ns |
Values are expressed as mean ± SD unless otherwise specified. Statistical significance was set at p < 0.05.
V, volume; n, number.
Progression in volume (V) decrease after single-fiber laser ablation multiple lower-intensity energy illuminations (LA-MLIEI) treatment in responders and non-responders during follow-up.
| V 0m (ml) | V 1m (ml) | V 3m (ml) | V 6m (ml) | V 12m (ml) | V 24m (ml) | V 36m (ml) | V 48m (ml) | V 60m (ml) | |
|---|---|---|---|---|---|---|---|---|---|
|
| 14.2 ± 8.7 | 9.4 ± 7.5 | 7.7 ± 6.3 | 6.6 ± 5.0 | 5.0 ± 2.7 | 4.4 ± 2.6 | 3.7 ± 1.3 | 3.9 ± 2.0 | 4.1 ± 2.1 |
|
| 30.4 ± 31.6 | 22.3 ± 21.9 | 20.5 ± 20.3 | 21.2 ± 26.1 | 19.8 ± 21.0 | 18.8 ± 22.8 | 16.3 ± 4.6 | 16.7 ± 19.0 | 17.2 ± 13.6 |
|
| 0.03 | 0.02 | 0.01 | 0.02 | <0.01 | 0.04 | 0.03 | <0.01 | <0.01 |
Values are expressed as mean ± SD unless otherwise specified.
V, volume; m, months; LA, laser ablation.