Liliana Tribuna1, Pedro Belo Oliveira1,2, Alba Iruela3, João Marques1, Paulo Santos1, Tiago Teixeira1,4.
Abstract
This study aimed at establishing native T1 reference values for a Canon Vantage Galan 3T system and comparing them with previously published values from different vendors. A total of 20 healthy volunteers (55% Women; 33.9 ± 11.1 years) underwent left ventricular T1 mapping at 3T MR. A MOLLI 5(3)3 sequence was used, acquiring three short-axis slices. Native T1 values are shown as means (±standard deviation) and Student's independent samples t-test was used to test gender differences in T1 values. Pearson's correlation coefficient analysis was used to compare two processes of T1 analysis. The results show a global native T1 mean value of 1124.9 ± 55.2 ms (exponential analysis), that of women being statistically higher than men (1163 ± 30.5 vs. 1077.9 ± 39.5 ms, respectively; p < 0.001). There were no specific tendencies for T1 times in different ventricular slices. We found a strong correlation (0.977, p < 0.001) with T1 times derived from parametric maps (1136.4 ± 60.2 ms). Native T1 reference values for a Canon 3T scanner were provided, and they are on par with those already reported from other vendors for a similar sequence. We also found a correlation between native T1 and gender, with higher values for women.Entities:
Keywords: MOLLI; cardiac MR; native T1 mapping
Year: 2021 PMID: 34943571 PMCID: PMC8699831 DOI: 10.3390/diagnostics11122334
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Typical sequence parameters of the CMR protocol performed on all participants of the study.
| Base Sequence | TR (ms) | TE (ms) | FA (°) | FOV (mm) | Matrix | Slice Thickness (mm) | Gap (mm) | Number of Slices | BW (Hz) | |
|---|---|---|---|---|---|---|---|---|---|---|
| T2 AX | FASE | 9100 | 80 | 90 | 450 × 340 | 224 × 256 | 8 | 2 | 16–18 (Aortic arch to liver) | 651 |
| Cine 2ch | SSFP | 2 | 1.5 | 55 | 400–450 × 400–420 | 224 × 192 | 8 | 2 | 1 | 1302 |
| Cine 4ch | 1 | |||||||||
| Cine SA | 9–11 (depending on heart size) | |||||||||
| T2 FS AX | FSE | 2036 | 60 | 90 | 400–450 × 300 | 160 × 256 | 8 | 6–8 | 3 (base, midventricular and apex) | 651 |
| T1 mapping | MOLLI 5(3)3 | 3.9 | 1.4 | 13 | 400 × 450 | 128 × 208 | 8 | 2 | 3 (base, midventricular and apex) | 408 |
TR: repetition time; TE: echo time; FA: flip angle; FOV: field of view; BW: bandwidth; FASE: fast advanced spin echo sequence; SSFP: steady-state free precession sequence; FSE; fast spin echo sequence.
Figure 1Exponential determination of T1 relaxation times.
Figure 2Parametric T1 maps from midventricular slice, without (left) and with (right) T1 determination.
Demographic and CMR gender characteristics of study population.
| Overall Population | Women | Men | ||
|---|---|---|---|---|
| Age (years) | 33.9 ± 11.2 | 32.1 ± 10.0 | 36.1 ± 12.7 | 0.440 |
| LVEDV (mL/m2) | 75.6 ± 17.7 | 66.7 ± 8.4 | 75.6 ± 17.7 | 0.160 |
| LVESV (mL/m2) | 30.6 ± 10.6 | 23.1 ± 3.7 | 30.6 ± 10.6 | 0.040 |
| LVMM (g/m2) | 58.4 ± 7.1 | 44.0 ± 10.1 | 58.4 ± 7.1 | 0.002 |
| LVEF (%) | 60.6 ± 6.1 | 65.1 ± 5.3 | 60.6 ± 6.1 | 0.090 |
| T2 | 1.3 ± 0.3 | 1.2 ± 0.2 | 1.3 ± 0.3 | 0.550 |
LVEDV: Left Ventricular End Diastolic Volume Index; LVESV: Left Ventricular End Systolic Volume Index; LVMM: Left ventricular myocardial mass; LVEF: Left ventricular ejection fraction; * p value for women and men comparison.
Native T1 values estimates.
| T1 Estimates | Overall Population | Women | Men | |
|---|---|---|---|---|
| Global (ms) | 1124.9 | 1163.4 | 1077.9 | <0.001 |
| Base (ms) | 1133.0 | 1161.9 | 1097.6 | 0.001 |
| Midventricular (ms) | 1119.5 | 1160.0 | 1069.9 | <0.001 |
| Apex (ms) | 1125.9 | 1168.3 | 1067.8 | <0.001 |
* p value for Women and Men.
Figure 3Correlation between T1 values acquired by map analysis and values acquired by exponential analysis.
Tolerance intervals in the studies using a MOLLI 5(3)3 sequence.
| Study |
| Native T1 | Tolerance Interval |
|---|---|---|---|
| Teixeira et al. [ | 9 | 1208 ± 18 ms | 1170.8–1245.0 |
| Dong et al. [ | 69 | 1202 ± 45 ms | 1150.4–1253.6 |
| Granitz et al. [ | 58 | 1184 ± 38 ms | 1139.4–1228.6 |
| Tribuna et al. Canon | 20 | 1125 ± 55 ms | 1042.4–1207.4 |
n = sample size.