| Literature DB >> 34940646 |
Raphaela Fritsche-Guenther1, Yoann Gloaguen1,2, Anna Bauer1, Tobias Opialla1, Stefan Kempa3, Christina A Fleming4, Henry Paul Redmond4, Jennifer A Kirwan1.
Abstract
Using manual derivatization in gas chromatography-mass spectrometry samples have varying equilibration times before analysis which increases technical variability and limits the number of potential samples analyzed. By contrast, automated derivatization methods can derivatize and inject each sample in an identical manner. We present a fully automated (on-line) derivatization method used for targeted analysis of different matrices. We describe method optimization and compare results from using off-line and on-line derivatization protocols, including the robustness and reproducibility of the methods. Our final parameters for the derivatization process were 20 µL of methoxyamine (MeOx) in pyridine for 60 min at 30 °C followed by 80 µL N-Methyl-N-trimethylsilyltrifluoracetamide (MSTFA) for 30 min at 30 °C combined with 4 h of equilibration time. The repeatability test in plasma and liver revealed a median relative standard deviation (RSD) of 16% and 10%, respectively. Serum samples showed a consistent intra-batch median RSD of 20% with an inter-batch variability of 27% across three batches. The direct comparison of on-line versus off-line demonstrated that on-line was fit for purpose and improves repeatability with a measured median RSD of 11% compared to 17% using the same method off-line. In summary, we recommend that optimized on-line methods may improve results for metabolomics and should be used where available.Entities:
Keywords: automated derivatization; gas-chromatography mass spectrometry; metabolomics; on-line derivatization; optimization; quality assurance (QA); quality control (QC); validation
Year: 2021 PMID: 34940646 PMCID: PMC8703763 DOI: 10.3390/metabo11120888
Source DB: PubMed Journal: Metabolites ISSN: 2218-1989
Figure 1Concentration range of 30/42 detected compounds in mouse liver. A calibration mixture at high (1:1), middle (1:10) and low (1:100) concentration was used for comparison (one for calibration mixture and n = 18 for liver). A minimum/maximum normalization was performed on the calibration mixture for the single compounds.
Figure 2Workflow for off-line and on-line parameter optimization.
Combined results for the optimization of the on-line parameters for derivatization. The 1:10 calibration mix dilution was used.
| Parameter- | MeOx Volume | Time | Temperature | Equilibration |
|---|---|---|---|---|
| Analyzed replicates | 20 µL: 5 | 30/30 min: 4 | 30 °C: 4 | 0 h: 4 |
| Detected compounds | 20 µL: 32 | 30/30 min: 33 | 30 °C: 38 | 0 h: 34 |
| Median RSD (%) | 20 µL: 17 | 30/30 min: 23 | 30 °C: 10 | 0 h: 11 |
Repeatability and reproducibility test. Results are from plasma and liver (single batch) and pooled serum quality control (QC) samples (analyzed over three large batches of around 115 samples per batch). RSD: relative standard deviation.
|
| Plasma | Liver | Batch 1 | Batch 2 | Batch 3 |
|---|---|---|---|---|---|
| Number of metabolites | |||||
| Number of missing values | 11 (0.8%) | 0 (0%) | 24 (11%) | 25 (12%) | 25 (12%) |
| Median RSD | 16% | 10% | 21% | 20% | 19% |
| RSD range | 11–28% | 2–56% | 3–42% | 12–69% | 13–39% |
Comparison of on-line and off-line settings. RSD: relative standard deviation.
| Parameter | On-Line | Off-Line (Original) | Off-Line with On-Line Settings (OLOLP) |
|---|---|---|---|
| Replicates | 9 | 9 | 8 |
| Number metabolites | |||
| Number of missing values | 14 (6%) | 7 (3%) | 5 (3%) |
| Median RSD | 11% | 21% | 17% |
| RSD range | 4–38% | 4–48% | 2–50% |
Figure 3Distribution of individual metabolite relative standard deviation (RSD in %) for the different protocols used. The orange and green lines indicate RSD thresholds of 30% and 15%, respectively.
Figure 4Schematic of a typical automatic derivatization sequence. The multicolored diagram in the foreground displays the order of events for a single typical sample. The background diagram shows how these events are overlapped with different samples to maximize GC-MS instrument time. The GC-MS runs that are not attached to a derivatization sequence (salmon colored) represent MSTFA washes which are injected throughout the run to assess for any carryover.