A K M S Inam1,2, Martina A Costa Angeli1, Bajramshahe Shkodra1, Ali Douaki1, Enrico Avancini1, Luca Magagnin3, Luisa Petti1, Paolo Lugli1. 1. Faculty of Science and Technology, Free University of Bozen-Bolzano, Bolzano 39100, Italy. 2. Department of Nutrition and Food Engineering, Daffodil International University, Dhaka 1207, Bangladesh. 3. Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Milano 20133, Italy.
Abstract
Nitrate (NO3 -) contamination is becoming a major concern due to the negative effects of an excessive NO3 - presence in water which can have detrimental effects on human health. Sensitive, real-time, low-cost, and portable measurement systems able to detect extremely low concentrations of NO3 - in water are thus becoming extremely important. In this work, we present a novel method to realize a low-cost and easy to fabricate amperometric sensor capable of detecting small concentrations of NO3 - in real water samples. The novel fabrication technique combines printing of a silver (Ag) working electrode with subsequent modification of the electrode with electrodeposited copper (Cu) nanoclusters. The process was tuned in order to reach optimized sensor response, with a high catalytic activity toward electroreduction of NO3 - (sensitivity: 19.578 μA/mM), as well as a low limit of detection (LOD: 0.207 nM or 0.012 μg/L) and a good dynamic linear concentration range (0.05 to 5 mM or 31 to 310 mg/L). The sensors were tested against possible interference analytes (NO2 -, Cl-, SO4 2-, HCO3 -, CH3COO-, Fe2+, Fe3+, Mn2+, Na+, and Cu2+) yielding only negligible effects [maximum standard deviation (SD) was 3.9 μA]. The proposed sensors were also used to detect NO3 - in real samples, including tap and river water, through the standard addition method, and the results were compared with the outcomes of high-performance liquid chromatography (HPLC). Temperature stability (maximum SD 3.09 μA), stability over time (maximum SD 3.69 μA), reproducibility (maximum SD 3.20 μA), and repeatability (maximum two-time useable) of this sensor were also investigated.
Nitrate (NO3 -) contamination is becoming a major concern due to the negative effects of an excessive NO3 - presence in water which can have detrimental effects on human health. Sensitive, real-time, low-cost, and portable measurement systems able to detect extremely low concentrations of NO3 - in water are thus becoming extremely important. In this work, we present a novel method to realize a low-cost and easy to fabricate amperometric sensor capable of detecting small concentrations of NO3 - in real water samples. The novel fabrication technique combines printing of a silver (Ag) working electrode with subsequent modification of the electrode with electrodeposited copper (Cu) nanoclusters. The process was tuned in order to reach optimized sensor response, with a high catalytic activity toward electroreduction of NO3 - (sensitivity: 19.578 μA/mM), as well as a low limit of detection (LOD: 0.207 nM or 0.012 μg/L) and a good dynamic linear concentration range (0.05 to 5 mM or 31 to 310 mg/L). The sensors were tested against possible interference analytes (NO2 -, Cl-, SO4 2-, HCO3 -, CH3COO-, Fe2+, Fe3+, Mn2+, Na+, and Cu2+) yielding only negligible effects [maximum standard deviation (SD) was 3.9 μA]. The proposed sensors were also used to detect NO3 - in real samples, including tap and river water, through the standard addition method, and the results were compared with the outcomes of high-performance liquid chromatography (HPLC). Temperature stability (maximum SD 3.09 μA), stability over time (maximum SD 3.69 μA), reproducibility (maximum SD 3.20 μA), and repeatability (maximum two-time useable) of this sensor were also investigated.
In the past decades,
contamination of water sources by industrial
and agricultural activities has become a major concern all over the
world. Among the various water and soil contaminants, the substances
most incriminated are surely nitrate ions (NO3–).[1] Indeed, NO3– ions are widely used not only in fertilizers but also as an additive
to enhance color and flavor or as an agent to prevent food poisoning
from Clostridium botulinum in food
industries.[2] However, a high level of NO3– has several detrimental effects on human
health since NO3– can be converted into
different harmful nitrogen compounds such as nitrite (NO2–), nitric oxide, and N-nitrosamines,
which can cause liver disease and gastric cancer.[3] An excessive NO3– intake is
also responsible for infant methemoglobinemia, commonly known as blue
baby syndrome.[4] Because of its toxic influence
on human health, the World Health Organization (WHO) and European
Directives have set the maximum contaminant level (MCL) of NO3– in public drinking water to be 50 mg/L
(ca. 0.8 mM).[5] It is therefore of uttermost
importance to determine the correct levels of NO3–, especially in drinking water.To date, different methodologies
have been developed to identify
the level of NO3– in water, such as flow
injection analysis, capillary electrophoresis, ion chromatography,
liquid chromatography-tandem mass spectrometry, spectrophotometry,
chemiluminescence, high-performance liquid chromatography (HPLC),
and gas chromatography–mass spectrometry.[6,7] Although
these techniques are sensitive and specific and allow a wide range
of detection, several drawbacks are connected to their use, especially
the need to utilize highly expensive, time-consuming, sophisticated
instruments and the requirement of trained personnel. Alternative
methods able to detect NO3– in a proper,
quick, and cost-effective way without sacrificing sensitivity and
selectivity are therefore of great interest. In this context, electrochemical
sensors are a promising class of analytical devices as alternatives
to the above-mentioned screening techniques. A wide range of electrochemical
sensors have been proven to be promising in comparison to traditional
methods because of their simplicity, selectivity, portability, and
miniaturization.[8,9]Up to now, various types
of electrochemical sensors have been used
to detect NO3– in water, such as potentiometric,
amperometric, and conductometric with or without the incorporation
of enzymes.[10−13] The nonenzymatic electrochemical sensors for NO3– detection can be realized using different types of
sensitive materials. For example, different metals like copper (Cu),
platinum (Pt), silver (Ag), and gold (Au) have been used as catalysts
for electroreduction of NO3– in amperometric
electrochemical sensors.[14−17] Among these electrocatalytic metals, Cu has been
proven to be one of the most effective metals to catalyze the electroreduction
of NO3–,[18,19] mainly because
of its high conductivity (5.8 × 107 S/m) that improves
the charge transfer; in addition, compared to other metals, Cu is
also less expensive.[20,21] Recently, researchers have demonstrated
the ability to lower the detection limit [limit of detection (LOD)]
of electrochemical NO3– sensors at a
concentration as low as 10 μM (620 μg/L) by increasing
the electroactive surface area using nanostructured Cu.[10,22] For example, Essousi et al.[23] were able
to improve the LOD using a working electrode (WE) made of ion-imprinted
polymer coated with Cu nanoparticles, whereas Wu et al.[24] realized the NO3– sensor with a LOD low as 12.2 μM (7.44 μg/L) with a
linear range of 50–600 μM (3.1–37.2 mg/L) using
Cu nanoparticles by thermal oxidation. Therefore, taking advantage
of the above cited methods in a more cost-effective way, we decided
to deposit a rough Cu nanostructured layer on top of our WE using
electrodeposition. Electrodeposition is a well-established, easy,
cost-effective, and large-area scalable deposition technique of pure
metal, metal alloy, or oxide, where the growth process can be easily
kinetically controlled by changing the deposition time and current
density in cyclic voltammetry (CV) or chronoamperometry.[25−27] In comparison with other deposition methods, such as e-beam evaporation,
sputtering, and pulsed laser deposition, electrodeposition does not
need expensive equipment for ultrahigh vacuum and high temperature.[28] Moreover, Cu can be easily electrodeposited
in the form of nanoclusters or nanoparticles on top of different bulky
electrodes, such as glassy carbon, graphite felt, and disk electrodes,
as reported, for example, by Bagheri et al.,[10] Lu et al.,[29] and Stortini et al.,[30] respectively. Additionally, we decided to use
screen printing, a fabrication technique that enables the development
of flexible, disposable, and cheap electrochemical sensors.[31] To the best of our knowledge, there is no work
reporting modification of amperometric electrochemical printed silver
WEs with Cu electrodeposition to detect NO3– in water.In this work, we developed a flexible screen-printed
amperometric
electrochemical NO3– sensor functionalized
by electrodeposited Cu metal nanoclusters. The uniqueness of the proposed
sensor consists in the possibility of combining two low-cost and scalable
techniques such as screen printing and electrodeposition, which allows
to realize cost-effective, disposable, easy to use sensors on a polymeric
substrate with a sensing performance (LOD and linear detection range)
comparable to that reported by other authors who employed complex
multilayer structures.[15,32] The proposed sensor showed a
high capability of detecting NO3– in
water with a low calculated LOD (0.207 nM or 0.012 μg/L) and
a wide dynamic concentration range (50 to 5000 μM or 31 to 310
mg/L) by using linear sweep voltammetry (LSV). The analysis of the
effect of most common interfering analytes, for example, Cl–, NO2–, SO42–, HCO3–, Fe2+, Fe3+, CH3COO–, Mn2+, Na+, and Cu2+, yielded negligible
effects on NO3– detection. Moreover,
sensor stability over time and temperature, reproducibility, and repeatability
were also investigated. Finally, the proposed sensors were employed
to detect NO3− in
tap and river water and were in good agreement with HPLC results with
a relative recovery (RR) of 102.3 and 107.5% and a coefficient of
variance of 3.65 and 2.86%, respectively.
Results and Discussion
Figure shows the
fabrication process and detection mechanism of the proposed cost-effective,
easy to fabricate NO3– sensor. Briefly,
Ag and AgCl were screen-printed on a flexible polyethylene (PET) film,
and the WE was functionalized with electrodeposited nanoclusterized
Cu. More information on the sensor fabrication can be found in the
experimental section.
Figure 1
Schematic illustration of the fabrication process and
the detection
mechanism of the proposed nitrate (NO3–) sensor: (A) Screen-printed silver (Ag) WE (B) modified with electrodeposited
copper (Cu), leading to (C) NO3– reduction.
(D) Micrograph of the cost-effective, flexible, screen-printed electrochemical
NO3– sensor.
Schematic illustration of the fabrication process and
the detection
mechanism of the proposed nitrate (NO3–) sensor: (A) Screen-printed silver (Ag) WE (B) modified with electrodeposited
copper (Cu), leading to (C) NO3– reduction.
(D) Micrograph of the cost-effective, flexible, screen-printed electrochemical
NO3– sensor.
Morphological
and Compositional Characterization
Figure A shows the WE morphology
after the Cu electrodeposition. As shown in the figure, the deposited
Cu is characterized by a globular nanocluster shape, where the diameter
of each globule is in the range of 0.5–1 μm. The obtained
morphology and size of the Cu nanocluster were uniform all over the
WE and consistent with the results previously obtained by Li et al.[33] The Cu coverage and uniformity are clearly visible
if Figure A is compared
with the scanning electron microscopy (SEM) image of the bare Ag electrode
(Figure S2A). Cu deposited on the surface
of the WE was characterized by X-ray diffraction (XRD) to evaluate
the crystallographic structure. The XRD pattern, as presented in Figure B, shows peaks at
positions (2θ) of 50.6, 59.1, and 88.6°corresponding to
the Bragg reflections of crystalline Cu(111), (200), and (220), respectively.
Additionally, crystalline Ag was also indicated by the presence of
peaks at positions of 44.5, 51.5, 76.1, and 92.8°. This pattern
confirms the successful coverage of crystalline Cu on top of the Ag
WE. Moreover, the observation was quite similar to that of Chen et
al.[34] who showed that the morphology of
the Cu deposited by electrocrystallization is controlled by the different
surface energies of the crystallographic plane.[35] The composition of the surface of the WE estimated by performing
energy-dispersive X-ray spectroscopy (EDS) analysis is presented in Figure S3. The analysis of the considered EDS
peaks reveals concentrations of 80% of Cu, 15% of Ag, 4% of C, and 0.5% of O, which are in the expected range, considering
the type of deposition and processes used to fabricate the sample,
and also reveals that most of the electrode is well covered with electrodeposited
Cu.
Figure 2
Surface morphology and composition of Cu deposited on the screen-printed
Ag WE: (A) SEM micrograph showing the surface characterized by a uniform
deposition of Cu nanoclusters. (B) XRD pattern of the Cu electrodeposited
on top of the Ag WE.
Surface morphology and composition of Cu deposited on the screen-printed
Ag WE: (A) SEM micrograph showing the surface characterized by a uniform
deposition of Cu nanoclusters. (B) XRD pattern of the Cu electrodeposited
on top of the Ag WE.The results from a 3D
optical profilometer showed that the thickness
and roughness of the WE electrode increased after the Cu deposition
from 10.1 ± 0.7 to 14.2 ± 1.5 μm and from 1.4 ±
0.4 to 5.7 ± 0.1 μm, respectively. This observation was
consistent with the SEM micrographs (Figures A and S2), proving
that the electrodeposition process increased the electrode surface
area by the formation of Cu nanoclusters.
Electrochemical Characterization
Figure A shows
the CV obtained at a scan rate of
100 mV for the bare Ag electrode and the Cu/Ag electrode in a blank
solution (0 mM NO3– in 0.1 M KCl). In
the case of the bare Ag electrode, the curve shows a reduction peak
at −1.15 V due to the reduction of Ag, which disappears in
the case of the Ag/Cu electrode, proving the full Cu coverage. Instead,
in the latter electrode, the CV shows one cathodic peak at −0.2
V (E1) and another at −0.6 V (E2). Peaks
E1 and E2 are ascribed to the reduction of Cu(I)
and Cu(II), as described by eqs and 2
Figure 3
(A) CV analysis using
Ag electrode (black) and Ag electrode with
electrodeposited Cu modification (Ag/Cu, red) in a blank solution
(0 mM of NO3–) of 0.1 M KCl electrolyte.
(B) CV analysis of Ag (black) and Ag/Cu (red) with the presence of
3 mM NO3– in 0.1 M KCl. Screen-printed
AgCl was used as the RE.
(A) CV analysis using
Ag electrode (black) and Ag electrode with
electrodeposited Cu modification (Ag/Cu, red) in a blank solution
(0 mM of NO3–) of 0.1 M KCl electrolyte.
(B) CV analysis of Ag (black) and Ag/Cu (red) with the presence of
3 mM NO3– in 0.1 M KCl. Screen-printed
AgCl was used as the RE.After the addition of
3 mM NO3– in
the electrolyte solution, the CV of the bare Ag electrode remains
unchanged; instead, two additional reduction peaks (E3 and
E4) appeared for the Ag/Cu electrode (Figure B) which correspond to the
reduction of NO3– (E3) and
NO2– (E4) ions that happen
consecutively.[36] Indeed, as described by eqs and 4, the primary product of NO3– reduction
is NO2–, which is in turn reduced to
form NH3 during the potential scan[37]The cathodic
peaks for NO3– and NO2– reduction occurred at −0.86 and
−1.15 V, respectively, as also reported by Lotfi Zadeh Zhad
and Lai.[38]The NO3– reduction peak amplitude
can be influenced by the processing parameters of the Cu electrodeposition,
such as the deposition potential and the time or the cycle number
of deposition performed using CV as shown by Li et al.[33] To optimize the Cu deposition process, six different
numbers of CV cycles (2, 5, 7, 10, 15, and 20) were performed on similar
WEs. After the electrodeposition, each electrode, characterized by
different CV cycles of Cu electrodeposition, was examined with four
different NO3– concentrations (0, 0.1,
0.8, and 1.6 mM) as plotted in Figure S1, where the reduction peak current versus the NO3– concentration is shown. The sensitivity of each electrode
was calculated and plotted with the current reduction peak at 1.6
mM of NO3– versus the number of CV cycles
in Figure . The sensitivity
as well as the reduction peak current reached the highest value for
10 cycles of CV of Cu electrodeposition, and then on increasing the
CV cycles, they stabilized at a plateau value. From this result, we
can infer that 10 CV cycles of Cu deposition correspond to the highest
value of the electroactive surface area that provides more electrochemical
accessibility for NO3–[39] and hence regarded as the optimized CV cycle for Cu deposition.
More than 10 cycles of CV may generate excess Cu deposition which
reduces the electrode porosity and so the electroactive surface area.
Indeed, SEM images taken from samples modified with 2, 5, 10, and
15 CV cycles revealed that with 5 CV cycles, the Cu nanoclusters were
nonuniform all over the WE and the underneath Ag was clearly visible
(Figure S2B). On the other hand, 10 cycles
showed uniform distribution of the Cu nanoclusters all over the Ag
electrode surface (Figure S2C), and from
15 CV cycles, the surface showed an increase of Cu but only clusterized
in the area at a higher Cu concentration (Figure S2D).
Figure 4
Sensitivity and current reduction peak at 1.6 mM of the
NO3– sensor prepared through different
CV cycles
of Cu electrodeposition.
Sensitivity and current reduction peak at 1.6 mM of the
NO3– sensor prepared through different
CV cycles
of Cu electrodeposition.To evaluate the nature
of the NO3– electrochemical reaction
of screen-printed Cu/Ag electrodes, the
effect of the scan rate (50 to 500 mV s–1) on the
reduction peak current was investigated, as shown in Figure A. The cathodic peak current
increased linearly with the increase of the scan rate (Figure B), suggesting a diffusion-controlled
reduction process as described by the Randles-Sevcik[40]eq where ip is the
peak current, n is the number of electron transfers
(here it is 2 for NO3–), α is the
cathodic electron transfer coefficient, A is the
active surface area (cm2), Do is the diffusion coefficient (2.0 × 10–6 cm2 s–1 for NO3–),[41] v is the scan rate (V s–1), and C is the NO3– concentration (mol
cm–3). On the contrary, the potential (Ep) at which the NO3– reduction
occurs shifted negatively with the increment of scan rate as shown
in Figure S4A. This characteristic behavior
is associated with a diffusion-controlled irreversible electron transfer
process.[42]
Figure 5
(A) Cyclic voltammogram at 3.0 mM NO3– at various scan rates. (B) Linear relationship
of peak current versus
square root of the scan rate of the Cu/Ag electrode.
(A) Cyclic voltammogram at 3.0 mM NO3– at various scan rates. (B) Linear relationship
of peak current versus
square root of the scan rate of the Cu/Ag electrode.To investigate the kinetics of an electrode, the half-peak
potential
(Ep/2) is often examined.[43] The magnitude of ΔEp/2 (= Ep – Ep/2) was calculated and plotted against the scan rate as shown
in Figure S4B. It is noticeable that ΔEp/2 is constant at various scan rates (from
50 to 500 mV s–1) with an average value of 80 ±
2 mV, proving that the transfer coefficient of NO3– reduction reaction was independent of the scan rate.By employing eq ,
the effective electrochemical surface areas of both Ag and Cu/Ag WEs
were calculated and found to be 0.062 and 0.111 cm2, respectively,
showing that the Cu electrodeposition induced an increase in the effective
surface of around 77%.
Sensor Performance for NO3– Detection
To evaluate the analytical performance
of the sensor, LSV was employed
with different concentrations of NO3– in 0.1 M KCl solution (Figure A). The calibration curve, shown in Figure B, was realized by averaging
the NO3– reduction peak current of three
samples at each concentration (0.05 to 5 mM of NaNO3) with
the standard deviation (SD). The curve showed a linear detection range
from 0.05 to 5 mM with a sensitivity of 19.58 μA/mM and a coefficient
of determination (R2) of 99.06% indicating
an excellent linear fit.
Figure 6
(A) LSV at different concentrations of NO3– in 0.1 M KCl. (B) Calibration curve of
the Cu/Ag sensor for NO3– detection (NO3– reduction peak current vs NO3– concentration).
Each point represents the average peak current performed by three
sensors, where the SD with error bars is shown.
(A) LSV at different concentrations of NO3– in 0.1 M KCl. (B) Calibration curve of
the Cu/Ag sensor for NO3– detection (NO3– reduction peak current vs NO3– concentration).
Each point represents the average peak current performed by three
sensors, where the SD with error bars is shown.LOD was calculated from the following formulawhere I0 is the
generated peak current at 0 mM NO3– and m is the slope of the linear response curve calculated by
the following formulawhere I1 is the
generated current for concentration C1 and I0 is the generated current for C0 or blank measurement.[47]The calculated LOD was 0.207 nM, which is significantly lower
compared
to other reported nitrate sensors realized using screen-printed carbon
electrode as shown in Table .[11,20,44−46] Also the linear detection range (0.05–5 mM) and sensitivity
(19.08 μA/mM) of the proposed sensor revealed good performance
when compared with previously reported screen-printed NO3– sensors (Table ). Furthermore, NO3– reduction
reactions were investigated using KCl solution as a neutral medium.
Few research works have been done for NO3– reduction reaction within neutral pH;[36,37] indeed, Na2SO4 pH 2.0 has been used in most cases to obtain
better sensitivity.[10,30] In this work, the KCl electrolyte
solution has proved to be an excellent medium and provided the opportunity
to directly measure NO3– content in real
water samples without the need to change the electrolyte pH and without
interfering with the sensor’s performance.
Table 1
Comparison between the Performance
of Different Screen-Printed NO3– Sensors
electrode
method
linear range (mM)
sensitivity (μA/mM)
LOD (nM)
R2
refs
screen-printed graphite
amperometry
0.1–10
0.12
100
0.999
(11)
screen-printed carbon
amperometry
0.015–0.250
0.005
5500
0.996
(44)
screen-printed carbon
potentiometry
0.1–100
100
(45)
screen-printed carbon
amperometry
0.01–0.25
3.13
0.97
(46)
screen-printed carbon
potentiometry
0.001–10
1000
(20)
screen-printed silver
amperometry
0.05–5
19.08
0.21
0.987
this work
We also found that the amplitude of the reduction peak of Cu (II)
(peak E2 in Figure B) increased linearly with the increasing
NO3– concentration. This observation
can be explained by the catalytic effect of Cu. Filimonov and Shcherbakov[48] showed that cuprous ion exhibits the catalytic
effect with any nitrogen-containing compound which is electrochemically
active, and thus, Cu(II) peak (Figure S5) can also be used to detect NO3–.
Effect of Temperature
The effect of temperature on
the sensor performance is another important aspect to evaluate since
it can directly impact the kinetics of the sensor’s electrochemical
reaction. Thus, the NO3– reduction peak
current of the sensor in the presence of 0.8 mM of NO3– was evaluated in a range of temperature from 10 to
40 °C. Figure shows that the overall reduction peak current increased with increasing
temperature, although it was stable between 25 and 35 °C. This
is an expected behavior since as reported by Cho et al.,[49] the reaction rate increases with temperature
leading to higher current. Despite this phenomenon that can be accounted,
for example, by integrating a temperature sensor on the back of the
polymeric substrate, the sensors showed good performance at different
temperatures with a maximum SD of 3.09 μA.
Figure 7
Effect of temperature
(10, 15, 20, 25, 30, 35, and 40 °C)
on sensor response (reduction peak current of 0.8 mM NO3–).
Effect of temperature
(10, 15, 20, 25, 30, 35, and 40 °C)
on sensor response (reduction peak current of 0.8 mM NO3–).
Interference Study
There are potential anions and cations
(Cl–, NO2–, SO42–, HCO3–, Fe2+, Fe3+,
CH3COO–, Mn2+, Na+, and Cu2+) commonly found in water that can interfere
with NO3– detection,[10,39] and hence a comprehensive interference study was performed in this
work. At first, the influence of Cl– was investigated
as it is considered as one of the most potential interfering agents[14,50] and also the main component of the electrolyte solution employed.
Since 0.1 M KCl was used as the electrolyte in this work, two other
concentrations of KCl (0.5 and 1 M) have been prepared and mixed with
0.8 mM NO3– to investigate the possible
interference on the cathodic reduction peak, as shown in Figure S6A. The currents for 0.1, 0.5, and 1
M KCl were 33.23 ±2.28, 34.01 ± 1.83, and 32.75 ± 0.52
μA, respectively, showing that the KCl concentration did not
interfere with NO3– detection.The selectivity of the Ag/Cu sensor was investigated by evaluating
the change of the reduction peak of NO3– (at −0.85 V) in the presence of 0.8 mM of different possible
interfering ions (NO2–, SO42–, HCO3–, Fe2+, Fe3+,
CH3COO–, Mn2+, Na+, and Cu2+). As shown in Figure S6B, the amplitudes of the current peak at −0.85 V for all these
interferents were similar to the value of the blank solution (0 mM
NaNO3 in 0.1 M KCl). In addition, these interferents were
prepared in the presence of 0.8 mM NO3– in 0.1 M KCl solution in order to evaluate the possible cointerference.
It was found that the current amplitude of the NO3– reduction peak of the coexistence of different interferents
were within the same range of pure NO3–. Each experiment was done in triplicate, and the results are shown
in Figure with error
bars from SD. The maximum variation of the reduction current peak
was found for the SO42– solution as 3.9
μA. From the data, it is observable that there was very little
or no interference because of the presence of these ions.
Figure 8
Interference
study using a reduction peak current of only 0.8 mM
NO3– (orange color) and 0.8 mM other
interferents (NO2–, SO42–, HCO3–, Fe2+, Fe3+, CH3COO–, Mn2+, Na+, and Cu2+) (blue color) in the presence of 0.8 mM NO3–.
Interference
study using a reduction peak current of only 0.8 mM
NO3– (orange color) and 0.8 mM other
interferents (NO2–, SO42–, HCO3–, Fe2+, Fe3+, CH3COO–, Mn2+, Na+, and Cu2+) (blue color) in the presence of 0.8 mM NO3–.
Reproducibility, Repeatability,
and Stability of the Sensor
Important parameters for the
sensor evaluation in practical applications
are repeatability, reproducibility, and stability over time. The repeatability
and reproducibility of the proposed sensors were assessed via repeated
measurements performed on the same electrode and on three electrodes
at the same conditions, respectively. Specifically, the repeatability
test was performed using the same sensor 10 consecutive times and
comparing the amplitude of the reduction current peak at 0.8 mM of
NO3– with the first measurement to calculate
the relative SD (RSD). Each time after examining the sensor with NO3–, the sensor was rinsed with DI water and
dried with compressed air and kept ready for the next test. The sensor
has stable behavior only up to the second measurement, which is characterized
by a reduction of 3% of the NO3– cathodic
peak from the first one (Figure S7). Instead,
from the third measurement, the sensor showed a very high reduction
of the peak current which was 24% less compared to the first test.
This was an expected result since a screen printable sensor is meant
to be disposable after one-time use (or maximum two times) for the
gradual decaying of the sensor, and indeed, the continuous applying
of the potential can degrade the performance of the pseudo-reference
electrode (RE).[51]For the reproducibility
test, the samples were tested at five different concentrations of
NO3– (0.05, 0.1, 0.8, 1.6, and 3.0 mM),
while the measurement on each concentration was performed three times
and RSD was found to be 0.82, 1.50, 2.30, 1.73, and 2.16 μA,
respectively, showing good reproducibility. This result is ascribed
to the good quality of the overall fabrication process: both screen
printing and electrodeposition guarantee an excellent homogeneity
and reproducibility of the deposition of the sensitive material.[52]To investigate the shelf life of the proposed
sensor, NO3– reduction current peak stability
versus time
was tested for 1 month. In this test, all sensors were fabricated
at the same time and kept in a cool dry place at room temperature
(RT). Tests were performed using the same NO3– concentration (0.8 mM) every week. The average results of the reduction
current peaks from five samples are shown in Figure with error bars. The error bars from the
SD values increased gradually, remaining below 2.00 μA after
the third week, and reached a maximum of 3.69 μA after the fourth
week. From the calibration curve, the SD value of 0.8 mM concentration
was 2.3 μA. The performance of the sensor degrades and fluctuates
over time, likely as a consequence of the formation of the oxidative
layer on top of both Ag and Cu. This can be solved by preparing and
storing the sensor in a nitrogen atmosphere and avoiding contamination.
Figure 9
Stability
test for the Cu/Ag sensor: the reduction peak (average
of five samples) of 0.8 mM NO3– repeated
each week until 1 month.
Stability
test for the Cu/Ag sensor: the reduction peak (average
of five samples) of 0.8 mM NO3– repeated
each week until 1 month.
NO3– Detection in Water
To investigate the applicability of
the proposed sensor, it is important
to test it in real water samples. Thus, the sensor was tested with
tap and river water by using the standard addition method.[53] 0.8 mM of NO3– was
added to the real sample without any extra preparation or purification.
Each experiment was done in triplicate under identical conditions.
The results are shown in Table with RSD and RR. The results of Ag/Cu sensors were in good
agreement with those of HPLC for both tap and river water with a RR
of 102.3 and 107.5% with a coefficient of variance of 3.65 and 2.86%,
respectively. HPLC was calibrated (R2 =
0.9988) before the real water sample test using double distilled water
with a wide range of NO3– concentration
from 0.1 to 6 mM. These results indicated that the matrix effect was
almost negligible with respect to the performance of the sensor in
real sample analysis, showing promising feasibility of the employment
of the proposed sensor for the determination of NO3– in water samples.
Table 2
NO3– Detection
in Different Water Samples
sample
added (mM)
detected by sensor (mM)
RR (%)
detected by HPLC (mM)
tap water
0.8
0.838 ± 0.045
102.3
0.858 ± 0.040
river water
0.8
0.840 ± 0.038
107.5
0.903 ± 0.028
Conclusions
In
this work, a flexible, low-cost, easy to fabricate screen-printed
electrochemical sensor was presented for NO3– detection in water. Initially, the sensor fabrication was optimized
in terms of the number of cycles of CV for Cu electrodeposition. The
resulting sensor was characterized by nanocluster formed Cu with a
77.4% increase of the electroactive surface area if compared with
the bare silver electrode. Under optimal conditions, NO3– could be quantitatively determined in the range
extending from 0.05 to 5.0 mM with a calculated LOD of 0.218 nM in
neutral media. The sensor showed a high reproducibility with a maximum
RSD of 2.60 μA. From the repeatability test, it was confirmed
that the same sensor can be used only twice, which is acceptable for
a low-cost disposable sensor. Furthermore, the stability test proved
that the sensor can be kept in a normal environment for 3 weeks with
minimal change in the reduction peak current. The effects of different
interfering ions to the sensors were negligible proving the selectivity
of the sensor in real water measurements. Additionally, the sensor
showed good performance at different temperatures (maximum SD was
3.09 μA). Finally, this electrochemical sensor was investigated
with real samples, tap and river water, and validated against HPLC,
proving the ability to the sensor to be used in a real application.
The proposed sensor can be easily implemented in the industrial sector
as a low-cost, fast, specific, and sensitive sensor for NO3– detection. In the future, the sensor can be further
improved in terms of sensitivity by using a different transducing
platform, such as an electrolyte-gated field-effect transistor, and
in terms of portability by realizing a custom-made portable read-out
system.
Experimental Section
Reagents and Apparatus
Chemicals
in this work (all
were of analytical grade) were used without further purification.
Double-distilled water (resistivity below 18.2 Ω cm) was used
for the preparation of all solutions. Sodium nitrate (NaNO3), copper sulfate (CuSO4·5H2O), sulfuric
acid (H2SO4), potassium chloride (KCl), sodium
bicarbonate (NaHCO3), sodium sulfate (Na2SO4), sodium chloride (NaCl), ferrous sulfate (FeSO4), ferric chloride hexa hydrate (FeCl3·6H2O), sodium acetate (CH3COONa), manganese sulfate (MnSO4), and sodium nitrite (NaNO2) were purchased from
Merck KGaA (Germany). A flexible 125 μm thick PET was used as
a substrate (Rauch GmbH, Germany). Silver chloride (AgCl) (ECI 6038E)
and Ag (ECI 1011) screen printable inks were purchased from LOCTITE
E&C (CA, USA) and used for the electrode fabrication. 0.1 M CuSO4·5H2O (pH adjusted to 2.0 by 0.1 M H2SO4) was used for the electrodeposition of Cu, and 0.1
M KCl was used as electrolyte in all measurements. Tap water was collected
from the lab (city water supply network in Bolzano, Italy) and river
water was collected from the Adige river in Bolzano, Italy. Cu electrodeposition
and all electrochemical measurements were performed by using VersaSTAT
4 electrochemical workstation (Princeton Applied Research, USA) at
RT. To determine the NO3– concentration
of tap and river water samples, a 1525 Waters HPLC system (Waters
Corporations, MA, USA) was used. The HPLC was equipped with a binary
pump, an auto-sampler injection system, a Symmetry C18 Column (2.1
× 50 mm, 3.5 μm), and a photodiode array detector (PDA
2998) set at 286 nm.
Electrode Fabrication
A semiautomatic
screen-printing
machine (C290, Aurel automation S.P.A., Italy) was used to print the
typical three-electrode amperometric electrochemical sensor structure
on the PET flexible substrate. The screen-printed flexible electrodes
consist of an Ag WE of 4 mm diameter working area, an Ag counter electrode,
and an Ag/AgCl pseudo-RE, as shown in Figure . The fabrication of the screen-printed three-electrode
structure involved two steps as described by Shkodra et al.[54] Initially, the Ag ink was screen-printed on
cleaned PET and annealed at 120 °C for 15 min; subsequently,
the upper half of the RE was coated with the AgCl ink, and the electrodes
were annealed again at 120 °C for 15 min (see Figure A). Afterward, the electrodes
were ultrasonically cleaned in isopropyl alcohol and double distilled
water for 5 min in a bath sonicator (CP 104, Vetrotecnica, Italy)
at RT.
Copper Nanocluster Deposition
For the Cu nanocluster
deposition, 0.1 M CuSO4·5H2O solution was
prepared in acidic condition (pH = 2.0) by using 0.1 M H2SO4. Electrodeposition was performed on the WE electrode
by CV using different cycles (from 2 to 20) at RT while scanning potentials
between −1.0 and 0 V with a scan rate of 0.1V s–1, following the procedure from Mumtarin et al.[36] (see Figure B).CV in the potential
range from −0.1 to −1.4 V at different scan rates (50
to 500 mV s–1) was used to evaluate the electrochemical
nature of the NO3– reaction at the modified
WE electrode surface and to calculate its electroactive surface area.
To perform sensitivity tests, LSV was applied in the potential range
of −0.1 to −1.4 V at 0.01 V s–1 scan rate using different NO3– concentrations
(0.05 to 5.0 mM) in the KCl electrolyte solution. LSV over CV was
chosen because LSV can clearly show each peak of the consecutive reactions
of NO3– reduction.[55] To examine the effect of temperature, the proposed sensor
was tested by changing the electrolyte temperature from 10 to 40 °C.
The other tests, namely, the interference study, repeatability, reproducibility,
stability, and real sample analysis were examined using the MCL of
NO3– in water, 0.8 mM, set by WHO and
European Directives.[30]SEM
(Quanta 600F, FEI, USA), XRD, and EDS measurements were performed
to reveal the surface morphology of Cu nanocluster deposition and
also the crystal structure of Cu on Ag electrodes and the elemental
analysis on the electrode surface. To acquire XRD measurements, an
Italstructures IPD3000 diffractometer equipped with a Co anode source
(line focus), a multilayer monochromator to suppress k–β radiation, and fixed 100 μm slits was used.
Samples were positioned in reflection geometry with a fixed 5°angle
with respect to the incident beam to maximize the signal from the
sample surface with respect to the substrate; powder patterns were
acquired by means of an Inel CPS120 detector over 5–120°
2-theta range (0.03° per channel). The total acquisition time
for each sample was 1800 s. EDS (Bruker, Quantax 200 6/30) with the
no-standard P/B ZAF package was used to record EDS spectra with an
acquisition time of 30 s and an acceleration voltage of 20 kV. The
thickness and roughness of Ag and Cu-deposited Ag (Cu/Ag) were evaluated
by a noncontact 3D optical profilometer (ProFilm3D from Filmetrics,
Unterhaching, Germany).
Authors: A K M Sarwar Inam; Martina Aurora Costa Angeli; Ali Douaki; Bajramshahe Shkodra; Paolo Lugli; Luisa Petti Journal: Sensors (Basel) Date: 2022-04-02 Impact factor: 3.576