Prevention versus cure: Is BioGlue priming the optimal strategy against E-Vita NEO graft oozing?

Sven Z C P Tan1, Mohamad Bashir2.   

Abstract

BACKGROUND: Since the introduction of the E-Vita Open NEO aortic prosthesis in 2020, several incidences of post-anastomotic oozing from the polyester portion of the graft have emerged. The use of BioGlue to prime E-Vita Open NEO to prevent this has been suggested as a way to mitigate this worrying complication. We investigate the extent of graft oozing in E-Vita Open NEO and evaluate the use of BioGlue in preventing oozing, both experimentally and in terms of potential clinical complications.
MATERIALS AND METHODS: E-Vita Open NEO (in straight and branched configurations) was implanted in a perfused model. The distal stent graft and side branches were clamped, and the graft was pressurized with blood to 120 mmHg. The volume of blood (ml) oozing from the graft within 60 s was measured. Nonpressurized grafts were coated with BioGlue up to a thickness of 1, 2, and 3 mm, and the volume (mm3 ) of BioGlue required to do so was recorded.
RESULTS: Within 60 s, 250.0 ml of blood oozed from the grafts tested. 43.694, 87.389, and 174.778 mm3 of BioGlue were required to coat the device with 1, 2, and 3 mm of BioGlue.
CONCLUSION: Graft oozing from E-Vita Open NEO represents an omnipresent and worrying risk. The use of BioGlue herein is likely associated with several adverse consequences, which are an additional risk on top of that posed by graft oozing. These risks call into question the suitability of E-Vita Open NEO, especially when compared to alternative devices not affected by oozing.
© 2021 Wiley Periodicals LLC.

Entities:  

Keywords:  BioGlue; E-Vita Open NEO; frozen elephant trunk; oozing

Mesh:

Substances:

Year:  2021        PMID: 34907587     DOI: 10.1111/jocs.16179

Source DB:  PubMed          Journal:  J Card Surg        ISSN: 0886-0440            Impact factor:   1.620


  1 in total

1.  What Is the Long-Term Clinical Efficacy of the Thoraflex™ Hybrid Prosthesis for Aortic Arch Repair?

Authors:  Sven Z C P Tan; Matti Jubouri; Idhrees Mohammed; Mohamad Bashir
Journal:  Front Cardiovasc Med       Date:  2022-02-22
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.