Darío Sánchez Martín1, Reinier Oropesa-Nuñez2, Teresa Zardán Gómez de la Torre1. 1. Department of Material Sciences and Engineering, Division of Nanotechnology and Functional Materials, Ångström Laboratory, Uppsala University, 751 21 Uppsala, Sweden. 2. Department of Material Sciences and Engineering, Division of Solid-State Physics, Ångström Laboratory, Uppsala University, 751 21 Uppsala, Sweden.
Abstract
Visual detection of rolling circle amplification products (RCPs) has been achieved by specific aggregation with magnetic nanoparticles. The method presented here reliably generates aggregates in 1.5 h; these are visible to the naked eye in samples containing at least 0.4 fmol of RCPs. In addition, alternate current susceptometry and absorbance spectroscopy have also been used to quantify the amplified products. The specificity of the detection method was tested, and no non-specific aggregation was detected in samples containing up to 20 fmol of non-complementary amplified DNA. This method is a versatile tool for detecting pathogenic DNA in point-of-care diagnostics, with no readout equipment required. However, chips and automated assays can be used in conjugation with the developed method since detection and quantification can be achieved by commercially available readout instruments.
Visual detection of rolling circle amplification products (RCPs) has been achieved by specific aggregation with magnetic nanoparticles. The method presented here reliably generates aggregates in 1.5 h; these are visible to the naked eye in samples containing at least 0.4 fmol of RCPs. In addition, alternate current susceptometry and absorbance spectroscopy have also been used to quantify the amplified products. The specificity of the detection method was tested, and no non-specific aggregation was detected in samples containing up to 20 fmol of non-complementary amplified DNA. This method is a versatile tool for detecting pathogenic DNA in point-of-care diagnostics, with no readout equipment required. However, chips and automated assays can be used in conjugation with the developed method since detection and quantification can be achieved by commercially available readout instruments.
There is an increasing
need for point-of-care (POC) pathogen diagnostics
in low-income countries and at small healthcare centers. Fast methods
that do not need expensive equipment, complicated machinery, or trained
personnel are valuable assets in the fight against disease across
the globe.[1] These methods rely on detection
of biological molecules.[2,3]Methods that detect
specific DNA sequences enable the development
of a wide range of biomedical applications since all living organisms
have DNA. Furthermore, these methods can usually be readily adapted
to detect specific sequences of interest.[4,5] The
most common method of determining the presence of a specific DNA sequence
in a sample is polymerase chain reaction (PCR) amplification, followed
by confirmation of DNA presence through gel electrophoresis or fluorescence-based
methods.[6,7] However, PCR requires trained personnel
and expensive materials,[6] thus making it
unsuitable for POC diagnostics. A number of alternative DNA amplification
methods that are simple and efficient have been developed in recent
years. One of them is rolling circle amplification (RCA), an isothermal
amplification method where a DNA polymerase creates a long single-stranded
DNA (ssDNA) molecule out of a circular DNA molecule.[8,9] The circular template can be obtained using a so-called padlock
probe, a linear DNA oligonucleotide phosphorylated on the 5′
end which hybridizes through both ends to a target DNA sequence. When
this hybridization occurs, both ends are brought together, and a ligase
can use the 5′ phosphate to link the ends, creating a circular
template.[10,11] Ligation is followed by an RCA reaction,
which creates long ssDNA products of hundreds of repeats complementary
to the padlock probe.[9] The RCA products
(RCPs) then collapse in random-coiled, micrometer-sized structures.[12] RCA is suitable for POC diagnostics since it
is conducted at a single temperature and does not require highly trained
personnel or expensive equipment.RCPs can be detected and analyzed
using a variety of readout techniques
including gel electrophoresis,[13,14] electrochemical techniques,[15−17] and by colorimetric[18−20] and spectroscopic visualization.[21,22] One of the most common techniques to detect RCPs is optically, using
fluorescence analysis.[23] This readout method
is based on the binding of agents to the amplified DNA, giving a fluorescent
signal, or through fluorescence-labeled detection oligonucleotides
(DOs) that are complementary to the repeating sequence in the RCPs.
Because fluorophores are light sensitive and need to be handled and
stored properly, they can be unsuitable for POC diagnostics. As an
alternative, magnetic nanoparticles (MNPs) can be used as labels where
their signal can be measured by magnetometers. MNPs are physically
stable, easy to handle, and not sensitive to light.[24] To date, many different types of sensors that can measure
the magnetic field from magnetic particles have been developed.[25,26] Among magnetic field sensors commonly used for biosensing are superconducting
quantum interference devices (SQUIDs),[27−29] inductive methods,[30−32] fluxgates,[33−35] and magnetoresistive sensors.[36,37] The SQUIDs are very sensitive, but they are large in size and require
cryogenics, which makes them expensive and not suited for POC. Sensors
based on induction, fluxgates, and magnetoresistivity are smaller
in size and more flexible in their design. Due to their smaller size,
they often require smaller sample volumes and can potentially be integrated
with a microfluidic sample preparation device.The Brownian
relaxation detection principle is an inductive method
where MNPs are coupled to DOs that have the ability to bind specifically
to complementary RCPs. The Brownian relaxation frequency of the MNPs
changes to lower frequencies when they are bound to RCPs as these
molecules are much bigger in size than the MNPs. The change in the
relaxation frequency can be measured by alternate current (AC) susceptometry.
This is the underlying working principle of the volume-amplified magnetic
nanobead detection assay (VAM-NDA).[38,39] Tabletop AC
susceptometers are commercially available and have been used for VAM-NDA.
An earlier study has shown that RCPs can be detected in around 2 min
in a portable AC susceptometer.[40] Recent
research has also demonstrated that different types of nanoparticles
can form aggregates in the presence of the analyte of interest, leading
to visual identification of the reaction products.[22,41,42] This has been carried out in combination
with PCR,[41] or with RCA, using gold[22] or MNPs.[42] However,
the studies using RCA indicated practical limitations such as long
assay times (>3.5 h), the need for in-lab synthesis of nanoparticles,
and non-specific binding of the particles, which can easily lead to
false results.In this work, we explored the possibility of
combining the VAM-NDA
with MNP aggregation to develop a biosensing method for POC diagnostics
based on visual detection. Synthetic Vibrio cholerae DNA was used as a model target sequence for this study.
Results and Discussion
We have developed a detection method based on the formation of
visible aggregates resulting from the interaction of MNPs and RCPs.
The colloidal particles in the MNP solution used in the study gave
the solution a dark brown color. The hue of the solution became lighter
when the MNPs and RCPs aggregated and precipitated. This allowed a
multi-readout assay using naked eye detection of precipitated MNP-RCP
aggregates, absorbance/turbidity measurement of the solution color,
and AC susceptibility readout.This approach yields results
in around 1.5 h with samples containing
as low as 0.4 fmol of RCPs (in theory each target molecule yields
a maximum of an RCP). We optimized the incubation step of the VAM-NDA
so that the MNPs could bind to the RCPs to form millimeter-sized aggregates
that easily precipitated to the bottom of the sample tubes. The differences
between the two protocols are outlined in Scheme . Figure A shows how the formed aggregates are visible to the
naked eye, facilitating detection of specific DNA targets, as no readout
system is required. As can be seen in Figure B, the aggregation protocol resulted in compact,
individual aggregates even with the lowest RCP amounts (0.4 fmol).
In contrast, the VAM-NDA protocol generated more disperse and smaller
clusters and only in samples between 4 and 20 fmol of RCPs.
Scheme 1
Comparison of the Two Methods; Schematic Showing the
Differences
between the VAM-NDA and the Aggregation Protocol
DO
= detection oligonucleotides;
MNP = magnetic nanoparticles; RCA = rolling circle amplification;
and VAM-NDA = volume-amplified magnetic nanobead detection assay.
Figure 1
Aggregates
can be seen with the naked eye and under a microscope.
(A) 96-well plate containing aggregates formed in samples with different
RCP amounts using the aggregation protocol. (B) Comparison of samples
using the aggregate and VAM-NDA protocol under an optical microscope
(4× objective). Photographs were taken by T. Zardán Gómez
de la Torre and micrographs were taken by D. Sańchez Martín.
Aggregates
can be seen with the naked eye and under a microscope.
(A) 96-well plate containing aggregates formed in samples with different
RCP amounts using the aggregation protocol. (B) Comparison of samples
using the aggregate and VAM-NDA protocol under an optical microscope
(4× objective). Photographs were taken by T. Zardán Gómez
de la Torre and micrographs were taken by D. Sańchez Martín.
Comparison of the Two Methods; Schematic Showing the
Differences
between the VAM-NDA and the Aggregation Protocol
DO
= detection oligonucleotides;
MNP = magnetic nanoparticles; RCA = rolling circle amplification;
and VAM-NDA = volume-amplified magnetic nanobead detection assay.MNPs and RCPs cannot bind and aggregate at room
temperature (Figure S1). Incubation at
a high temperature
is required to open the collapsed RCPs and allow MNP binding. Aggregation
was also enhanced by high NaCl concentration. NaCl concentrations
of 1 M or less raises the melting temperature of double-stranded DNA,[44] allowing stronger binding of the DOs, as it
balances the negative charges of DNA. Several NaCl concentrations
and temperatures were analyzed for the incubation step to form the
aggregates (Figure S2). It was found that
using 500 mM of NaCl and incubation at 80 °C were the optimal
conditions to form compact aggregates. We presume that the high temperature
and high NaCl concentration provide a good environment for the RCPs
to open and the MNPs to bind. This results in RCPs and MNPs forming
loose viscous aggregates in less than 2 min. Figure S3 contains images showing the process of aggregation when
the MNPs were mixed with the RCPs. Once the tubes were spun down or
the samples left to cool, the aggregates collapsed and precipitated,
becoming tight and stable, as shown in Figure A. Heating or vortexing did not separate
the aggregates, although vortexing caused them to break into smaller
pieces. Others have reported that non-functionalized gold NPs aggregate
at room temperature in the presence of salt.[45] This was not the case for our MNPs as can be seen by the lack of
aggregation in the negative control (NC) samples. We used atomic force
microscopy, specifically magnetic force microscopy, to ensure that
the aggregates contained both DNA and MNPs. Images are shown in Figure S4. We observed MNPs of the expected size,
along with what seemed to be ssDNA.Visual identification of
the aggregates does not provide a quantitative
estimate of DNA in the sample. In the VAM-NDA approach, immobilized
MNPs have a lower Brownian relaxation frequency than free MNPs, as
detected in a commercially available tabletop AC susceptometer. We
assumed that the same was true for the aggregate approach. The magnetization
responses for both detection methods are presented in Figure A,B, and the magnetization
spectra are presented in Figure S5. At
0.4 fmol of RCPs, both approaches resulted in a significant drop in
the values of the imaginary part of the frequency peaks, which represent
unbound MNPs with a nominal size of 100 nm. The peaks decreased with
higher amounts of RCPs, becoming nearly flat at 20 fmol. There was
no change in the limit of detection (LOD) (as defined in Materials and Methods) for the new method compared
to the VAM-NDA, meaning that accurate quantification of RCPs is still
possible with the aggregate approach.
Figure 2
Number of RCPs can be quantified using
an AC susceptometer and
an absorbance reader. (A,B) Average peak maximum of the imaginary
part of the complex magnetization spectra (at 128 Hz) for the aggregate
(A) and VAM-NDA (B) protocol samples. (C,D) Absorbance values at 350
nm of samples using the aggregate (C) and VAM-NDA (D) protocols. Error
bars represent standard deviation (sd) based on N = 3. The dashed line indicates the values for the NCs minus 3*sd,
representing the LOD.
Number of RCPs can be quantified using
an AC susceptometer and
an absorbance reader. (A,B) Average peak maximum of the imaginary
part of the complex magnetization spectra (at 128 Hz) for the aggregate
(A) and VAM-NDA (B) protocol samples. (C,D) Absorbance values at 350
nm of samples using the aggregate (C) and VAM-NDA (D) protocols. Error
bars represent standard deviation (sd) based on N = 3. The dashed line indicates the values for the NCs minus 3*sd,
representing the LOD.The formation of aggregates
makes the sample solutions clearer
than the solutions in the NC samples (see Figure A). This is due to the large aggregates precipitating
out of the solution. We theorized that an absorbance measurement at
a short wavelength, outside the absorption range of DNA molecules,
could be used to quantify unprecipitated MNPs in the sample. The results
are shown in Figure C,D. With the aggregate protocol, there was a significant difference
from the NCs in samples containing 1, 4, and 20 fmol of RCPs but not
in samples containing 0.4 fmol. This was not the case with the VAM-NDA
protocol, where there was a substantial difference from the NCs only
in samples containing 20 fmol of RCPs. There was only a small difference
from the NCs for the 4 fmol samples and no difference for the samples
containing 0.4 and 1 fmol. Thus, the aggregation protocol allows better
quantification through absorbance than the VAM-NDA method. This is
an advantage of the new approach since absorbance spectroscopy is
a simple readout technique. In addition, several samples can be placed
on a 96-well plate and measured at the same time in a matter of minutes,
making the protocol faster and allowing automatization. The possibility
of using an absorbance reader makes the method much more useful as
many laboratories and health centers around the world have absorbance
readers nowadays. Lastly, we addressed the specificity of the new
approach. The padlock probe-based RCA protocol has good specificity
because the padlock probe is ligated only when a complementary sequence
is present.[10] However, it is possible that
DNA contaminants are present in a sample or even that unwanted circularization
and amplification of the DNA target might occur. To test the specificity
of the aggregate formation, we changed the DO bound to the MNPs to
a DO with a sequence that was not complementary to the RCPs. No aggregation
was observed, and the AC susceptometry results confirmed the absence
of aggregates. The results are shown in Figure ; there were no significant differences between
samples ranging from 0.4 to 20 fmol of RCPs and the NCs. These results
indicate that the aggregation protocol is specific. The protocol is
thus reliable as it depends on the specificity of both the padlock
probe ligation and the DO recognition of RCPs. Other research has
been carried out where MNPs show to aggregate with RCPs independently
of their sequence,[42] but that is not the
case here.
Figure 3
Aggregation of MNPs is specific to DNA that is complementary to
the DOs. Average peak maximum of the imaginary part of the complex
magnetization spectra (128 Hz) for different RCP amounts and MNPs
with non-complementary DOs. Error bars represent sd based on N = 3. The dashed line indicates the values for NCs minus
3*sd, representing the LOD.
Aggregation of MNPs is specific to DNA that is complementary to
the DOs. Average peak maximum of the imaginary part of the complex
magnetization spectra (128 Hz) for different RCP amounts and MNPs
with non-complementary DOs. Error bars represent sd based on N = 3. The dashed line indicates the values for NCs minus
3*sd, representing the LOD.Table presents
several DNA detection methods based on isothermal amplification and
optical detection. The sensitivity of our method varies from 0.4 to
1 fmol (2–5 pM in 200 μL) depending on the readout technique.
Our detection method is more sensitive than the detection techniques
based on nucleic acid sequenced-based amplification (NASBA) and hybridization
chain reaction (HCR). Our method has also the advantage to be more
rapid compared to these detection methods. The assays based on RCA
have as good as or better sensitivity compared to our method but lack
in detection speed. These assays have also showed other practical
limitations, such as in-lab synthesis of the nanoparticles, and non-specific
binding of the particles to the RCPs, which has not been observed
in the current study.
Table 1
Comparison of Different
DNA Detection
Methods Based on Isothermal Amplification and Optical Detection
amplification
method
readout technique
LOD
assay time
reference
NASBA
fluorescens
0.1 μM
about 2 h
(46)
HCR
colorimetric
(AuNP)
500 pM
about 4 h
(47)
HRCA
colorimetric (phenol red)
3 pM
about 3 h
(18)
RCA
colorimetric (AuNPs)
70 fM
less than 4 h
(22)
RCA
visual
(magnetic particles)
0.62 fmol
6 h
(42)
RCA
visual, spectroscopy, and magnetic
0.4–1 fmol (2–5 pM) depending on the readout
less than 2 h
this work
Overall, using RCA with functionalized MNPs provides same or in
some cases an increase in sensitivity but offers a more rapid detection
speed compared to most of the presented methods. This makes it a robust
method for POC diagnostics in healthcare centers lacking experienced
laboratory personnel or laboratories with sterile conditions.
Conclusions
We have demonstrated that this detection approach creates visible
aggregates from a DNA target sequence amplified through RCA and commercially
available MNPs. The method provides high sensitivity and specificity
coupled with visual and instrumental readouts, and the entire assay
used here takes less than 2 h. It not only keeps the sensitivity of
the VAM-NDA in a tabletop AC susceptometer, such as the DynoMag, but
also allows visual identification and optical quantification through
absorbance measurements. The most important factor is the potential
for running samples in a 96-well plate, which would allow fast quantification
of many samples at the same time. Visual identification without readout
equipment is also possible if a qualitative response (positive/negative
result) is all that is needed. In addition, the protocol is shown
to be very specific.The proposed method requires long ssDNA
products with complementary
sequences to a DO to form visible aggregates, and while we based our
method in the VAM-NDA, aggregation of MNPs to RCPs is likely to happen
if we were to use other RCA-based techniques producing ssDNA (such
as circle-to-circle amplification) or perhaps even loop-mediated isothermal
amplification.The purpose of the study was to showcase the
aggregation step and
how that opens avenues into naked-eye detection or absorbance-based
quantification for POC diagnostics. Nevertheless, many questions remain
about the aggregation itself. It has shown to be highly specific,
and also to require high temperature and salt concentration, and we
intend to investigate this phenomenon further to improve the aggregation
of RCPs and MNPs when it comes to sensitivity, time, and ease of use.
We hope this novel method will aid in the design of robust, specific,
and sensitive diagnostic tests for POC diagnostics.
Materials and
Methods
MNPs and Reagents
The core–shell MNPs used in
this study were purchased from Micromod Partikeltechnologie, Germany.
The MNPs had a core of 75–80 wt % magnetite, which was encapsulated
in cross-linked hydroxyethyl starch, with surface streptavidin (product
code 10-19-102). These particles had a diameter of around 100 nm and
were suspended in phosphate-buffered saline (PBS) at a concentration
of 10 mg/mL. The theory behind the dynamic magnetic properties of
the NPs is outlined in supporting theory in the Supporting Information.All oligonucleotides were purchased
from Biomers, Ulm, Germany, and all reagents were purchased from Thermo
Fisher Scientific, Waltham, MA, USA, unless otherwise specified. The
oligonucleotides used in this work are listed in Table .
A
ligation reaction to circularize 20 nM of padlock probes was performed
(1× phi29 buffer, 1 mM ATP, 20 nM padlock probe, 60 nM V. cholerae synthetic target DNA, 0.1 U/μL
T4 ligase) for 15 min at 37 °C. The circularized padlock probes
were then amplified in an RCA reaction (6.66 nM circularized padlock
probe, 1× phi29 buffer, 0.2 mg/mL bovine serum albumin, 0.25
mM dNTPs, and 0.33 U/μL phi29 polymerase) for 1 h at 37 °C
followed by a 5 min enzymatic inactivation at 65 °C.
Functionalization
of MNPs with DO and Hybridization between
Functionalized MNPs and RCPs
Aggregation Protocol
Non-functionalized
MNPs were washed
three times with washing buffer [6.6 mM Tris-HCL pH 8, 3.3 mM ethylenediaminetetraacetic
acid (EDTA), 66 mM NaCl, 0.06% Tween-20] using a permanent magnet.
The MNPs were then resuspended in washing buffer with 240 nM of DO V. cholerae and an MNP concentration of 4 mg/mL,
and the mixture was incubated for 1 min at room temperature.The formation of visible aggregates was achieved as follows: the
RCPs were diluted to the desired concentration with a 500 mM NaCl
solution. 20 μL of the RCA solution was incubated in a thermoblock
at 80 °C for 20 min, and 5 μL of the functionalized MNPs
(4 mg/mL) was added immediately after. The tube was lightly vortexed.
The aggregates were formed at room temperature over about 2 min, and
the solution was then spun down to collect condensed water at the
bottom of the tube.For the optimization of the aggregate protocol,
the following NaCl
concentrations and incubation temperatures were analyzed: 250 mM,
500 mM, 750 mM, and 1 M NaCl and 60, 70, 80, and 90 °C, respectively.
VAM-NDA Protocol
This procedure closely followed that
of Oropesa-Nuñez et al.[43] Briefly, after the MNPs were washed and functionalized,
as described previously, they were resuspended in PBS to a concentration
of 1 mg/mL MNPs and 60 nM DO V. cholerae. The RCPs were serially diluted with hybridization buffer (0.1 M
Tris-HCl pH 8, 0.1 M EDTA, 0.5% Tween 20, 2.5 M NaCl) to achieve the
desired RCP concentrations. 20 μL of RCPs was mixed with 20
μL of functionalized MNPs and incubated for 20 min at 55 °C.
Measurements were performed immediately after.
AC Susceptibility
Measurements
Samples were pipetted
into vials, and a final volume of 200 μL was obtained by adding
PBS. The frequency-dependent magnetic susceptibility measurements
were conducted in an AC susceptometer (DynoMag, RISE Acreo, Göteborg,
Sweden) at room temperature from 5 to 250,000 Hz at 21 logarithmically
equidistant frequency points. Triplicates were run for each sample type.
Absorbance
Measurements
Samples with DNA, NCs (no DNA),
and buffer blanks (no MNPs) were placed in a 96-well plate (transparent
and flat wells), and PBS was added to bring the volume to 200 μL/well.
Individual absorbance measurements were carried out at 350 nm with
a 10 ms settle time and 4 × 4 filled circle multi-well reads
(1500 μm away from well edges) in an Infinite 200 plate reader
(Tecan, Sweden). Blank values were subtracted from all samples. Triplicates
were run for each sample type.
Imaging of DNA Samples
Optical microscopy was used
to evaluate the morphology of the aggregates. The samples were photographed
on a 96-well plate with an Olympus BX60 optical microscope and an
OMAX A3580U3 camera (Figure B).The samples were also photographed with a Fujifilm
X-T3 camera and a Fujinon XF 90/2 R LM WR lens (Figure A).
Specificity Measurements
To test the specificity of
the aggregation step, RCPs at different concentrations were incubated,
as described previously. After the incubation, MNPs functionalized
with DO C2CA V. cholerae, a sequence
not complementary to the RCPs, were added to the RCPs. The samples
were then assessed in an AC susceptometer, as described previously.
Triplicates were run for each sample type.
Limit of Detection
The LOD was defined as the average
value of the NCs minus three times the sd. A sample was considered
detectable if its value plus sd was below the LOD. The LOD was calculated
for both absorbance and AC susceptibility measurements.